Difference between revisions of "NVGate Specification"

From OROS Wiki
Jump to navigation Jump to search
(Correction de fautes d'orthographe et de balisage (typos, wiki markup); aucun parametre technique modifie)
(Mise en forme: tableaux en wikitable + colonne categorie en en-tetes, gras superflu retire, hierarchie/coherence des titres, ponctuation et unites (2.048 S/s, kHz/s, espaces))
 
Line 3: Line 3:
The following official specifications concern NVGate<sup>®</sup> software for OR10, OR34, OR35, OR36, OR38 and Mobi-Pack<sup>®</sup> multi-analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the instrument connected to the PC. This specification applies for NVGate version 12 or upper and Teamwork instruments.
The following official specifications concern NVGate<sup>®</sup> software for OR10, OR34, OR35, OR36, OR38 and Mobi-Pack<sup>®</sup> multi-analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the instrument connected to the PC. This specification applies for NVGate version 12 or upper and Teamwork instruments.


====Modules====
==Modules==
NVGate<sup>®</sup> software is based on a general-purpose platform where optional analysis modules called plug-in analyzers are added.
NVGate<sup>®</sup> software is based on a general-purpose platform where optional analysis modules called plug-in analyzers are added.


Line 19: Line 19:
* '''''General-purpose analysis''''' for monitoring and result tracking.
* '''''General-purpose analysis''''' for monitoring and result tracking.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Signal sources'''
! rowspan="3" | Signal sources
|
! Front-end
'''Front-end '''
|Controls dynamic inputs, generators, external synchronization/tach. input and parametric (DC) inputs.
|Controls dynamic inputs, generators, external synchronization/tach. input and parametric (DC) inputs.


|-
|-
|
! Recorder
'''Recorder'''
|Record and store time domain signal on analyzer or PC hard disc. Records dynamic, parametric (DC) and ext. synch/tach inputs with multiple frequencies.
|Record and store time domain signal on analyzer or PC hard disc. Records dynamic, parametric (DC) and ext. synch/tach inputs with multiple frequencies.


|-
|-
|
! Player
'''Player'''
|Visualize, listen, playback on output or generate signals for post-analysis purpose. Based on recorded or imported time domain signal file.
|Visualize, listen, playback on output or generate signals for post-analysis purpose. Based on recorded or imported time domain signal file.


|-
|-
|align = "center" rowspan = "4"|'''Shared resources'''
! rowspan="4" | Shared resources
|
! Filters
'''Filters'''
|High/low pass, Band/Stop pass, single and double integrator, differentiator, A and C acoustic laws
|High/low pass, Band/Stop pass, single and double integrator, differentiator, A and C acoustic laws


|-
|-
|
! Events
'''Events'''
|Edge detection, DC levels, Delta DC levels, RPM, Delta RPM, Manual, Time period, combinations.
|Edge detection, DC levels, Delta DC levels, RPM, Delta RPM, Manual, Time period, combinations.


|-
|-
|
! Weighting windows
'''Weighting windows'''
|Programmable Force and response, Hanning, Hamming, Kaiser Bessel, flat-top, uniform.  
|Programmable Force and response, Hanning, Hamming, Kaiser Bessel, flat-top, uniform.  


|-
|-
|
! Tachometers
'''Tachometers'''
|Compute RPM from ext. sync and input pulses, DC, CAN. Fractional, simulated and combined tach.
|Compute RPM from ext. sync and input pulses, DC, CAN. Fractional, simulated and combined tach.


|-
|-
|align = "center" rowspan = "4"|'''Operational tools'''
! rowspan="4" | Operational tools
|
! Transducers management
'''Transducers management '''
|Automatic calibration, database management, calibration history, automatic setup of inputs, TEDS management.
|Automatic calibration, database management, calibration history, automatic setup of inputs, TEDS management.


|-
|-
|
! Masks
'''Masks'''
|Mask editor for spectra, profiles, 1/n octave, order spectra
|Mask editor for spectra, profiles, 1/n octave, order spectra


|-
|-
|
! Report
'''Report'''
|On-line report editor, automated report generation, word or Excel files.
|On-line report editor, automated report generation, word or Excel files.


|-
|-
|'''Automation''' '''tools'''
|Automation tools
|Macros, sequence from Excel<sup>®</sup> sheet, alarms based on mask comparison, customizable control panel, restricted user profiles, remote controller interface.
|Macros, sequence from Excel<sup>®</sup> sheet, alarms based on mask comparison, customizable control panel, restricted user profiles, remote controller interface.


|-
|-
|align = "center" rowspan = "2"|'''General purpose analysis'''
! rowspan="2" | General purpose analysis
|
! Monitor
'''Monitor'''
|4 channels FFT analyzer with input hot swap. Statistical extraction (RMS, Max, Min, Kurtosis) from time domain with programmable band-pass filter.  
|4 channels FFT analyzer with input hot swap. Statistical extraction (RMS, Max, Min, Kurtosis) from time domain with programmable band-pass filter.  


|-
|-
|
! Waterfall
'''Waterfall'''
|Synchronized stack for 2D (spectra) and scalar (order, overall). 3D, color map and profiles displays. Extraction capabilities.
|Synchronized stack for 2D (spectra) and scalar (order, overall). 3D, color map and profiles displays. Extraction capabilities.


|}
|}


'''NVGate<sup>®</sup> Plug-ins analyzers'''
===NVGate<sup>®</sup> Plug-ins analyzers===


In addition to the standard functions featured in the software platform, NVGate<sup>®</sup> can receive additional plug-in analyzers that offer flexible configuration of independent analysis modes.  
In addition to the standard functions featured in the software platform, NVGate<sup>®</sup> can receive additional plug-in analyzers that offer flexible configuration of independent analysis modes.  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center"|'''FFT'''
! FFT
|'''Fast Fourier Transform''' - Time to Frequency conversion with multiple average domains, combinable trigger, zooms and cross functions.  
|Fast Fourier Transform - Time to Frequency conversion with multiple average domains, combinable trigger, zooms and cross functions.  
|General-purpose frequency analysis, machinery signature, modal analysis acquisition (hammer or shaker), machinery diagnostic.
|General-purpose frequency analysis, machinery signature, modal analysis acquisition (hammer or shaker), machinery diagnostic.


|-
|-
|align = "center"|'''FFT Add-ons'''
! FFT Add-ons
|&nbsp;
|&nbsp;
|&nbsp;
|&nbsp;


|-
|-
|align = "center"|'''CBT'''
! CBT
|'''Constant Band Tracking''' - Order extraction at constant bandwidth
|Constant Band Tracking - Order extraction at constant bandwidth
|Gears mesh analysis, gearboxes noise and vibration tracking.
|Gears mesh analysis, gearboxes noise and vibration tracking.


|-
|-
|align = "center"|'''FFTDiag'''
! FFTDiag
|'''Cepstrum, Auto '''and cross '''correlation''', '''DC''', '''Min/Max, Pk, pk/pk '''and '''crest factor'''.
|Cepstrum, Auto and cross correlation, DC, Min/Max, Pk, pk/pk and crest factor.
|Set of functions dedicated to the diagnostics of industrial machineries.
|Set of functions dedicated to the diagnostics of industrial machineries.


|-
|-
|align = "center"|'''TDA'''
! TDA
|'''Time Domain Analysis''' –Time view (oscilloscope). True statistical extraction, averaging.
|Time Domain Analysis –Time view (oscilloscope). True statistical extraction, averaging.
|Machinery diagnostic, long term acquisition monitoring, critical acquisition control, large structure damping control, shocks determination, machinery health control, trend analysis.
|Machinery diagnostic, long term acquisition monitoring, critical acquisition control, large structure damping control, shocks determination, machinery health control, trend analysis.


|-
|-
|align = "center"|'''SOA'''
! SOA
|'''Synchronous Order Analysis''' - Time to Order re-sampling with angle or order averaging, acoustic weightings, multiple tach management.
|Synchronous Order Analysis - Time to Order re-sampling with angle or order averaging, acoustic weightings, multiple tach management.
|Rotating machinery balancing. In vehicle order to noise correlation R&D. Pump, compressors, turbines and speed reducer/multiplier test.  
|Rotating machinery balancing. In vehicle order to noise correlation R&D. Pump, compressors, turbines and speed reducer/multiplier test.  


|-
|-
|align = "center"|'''SOADiag <br>(SOA Add-on)'''
! SOADiag <br>(SOA Add-on)
|'''Cepstrum, '''Auto''' '''and cross '''correlation''', Revolution '''synchronous statistics''', Order transmission Function ('''ORF™'''), '''X''' functions
|Cepstrum, Auto and cross correlation, Revolution synchronous statistics, Order transmission Function (ORF™), X functions
|Set of diagnostic functions based on the synchronous order analysis for rotating part transmission and forced vibration extraction.
|Set of diagnostic functions based on the synchronous order analysis for rotating part transmission and forced vibration extraction.


|-
|-
|align = "center"|'''OCT'''
! OCT
|'''1/n Octave Constant Percentage Band filters''' - 1/n octave detectors set with multiple averaging modes.
|1/n Octave Constant Percentage Band filters - 1/n octave detectors set with multiple averaging modes.
|General-purpose industrial noise analysis. Vehicle acoustics R&D. Noise test.
|General-purpose industrial noise analysis. Vehicle acoustics R&D. Noise test.


|-
|-
|align = "center"|'''OVA'''
! OVA
|'''Overall Acoustics Detectors''' - 3 detectors <nowiki>+</nowiki> 1 peak detection per channel with parallel averaging and acoustic weightings.
|Overall Acoustics Detectors - 3 detectors <nowiki>+</nowiki> 1 peak detection per channel with parallel averaging and acoustic weightings.
|Industrial acoustic, sound power measurements. Machinery with cycles, R&D and test.
|Industrial acoustic, sound power measurements. Machinery with cycles, R&D and test.


Line 144: Line 132:
Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software also features options:  
Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software also features options:  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" bgcolor = "#D9D9D9"|'''IVC'''
! IVC
|bgcolor = "#D9D9D9"|Instantaneous angular velocity converter  
! Instantaneous angular velocity converter
|bgcolor = "#D9D9D9"|Torsional measurements from Ext. Synch inputs
! Torsional measurements from Ext. Synch inputs


|-
|-
|align = "center"|'''CTE'''
! CTE
|Combined Tachometer editor
|Combined Tachometer editor
|Tachometer math editor - 2 inputs, 1 output
|Tachometer math editor - 2 inputs, 1 output


|-
|-
|align = "center" bgcolor = "#D9D9D9"|'''VIN'''
! VIN
|bgcolor = "#D9D9D9"|Virtual Inputs
! Virtual Inputs
|bgcolor = "#D9D9D9"|Real-time math combination of dynamic inputs
! Real-time math combination of dynamic inputs


|-
|-
|align = "center"|'''VDC'''
! VDC
|Virtual DC
|Virtual DC
|Real-time math combination of parametric inputs
|Real-time math combination of parametric inputs


|-
|-
|align = "center"|'''A-Samp'''
! A-Samp
|Angular sampling for SOA
|Angular sampling for SOA
|Real-time Angular sampling of inputs
|Real-time Angular sampling of inputs
Line 177: Line 165:
The front-end module gathers the different input and output settings available in the instrument unit. In addition, the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to the PC) which is useful for set-up.
The front-end module gathers the different input and output settings available in the instrument unit. In addition, the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to the PC) which is useful for set-up.


'''Front-end settings'''
'''Front-end Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|rowspan = "10"|'''Dynamic inputs'''
! rowspan="10" | Dynamic inputs
|Sampling clock
|Sampling clock
|'''102.4 kS/s to 2,048 S/s''' or '''65.536 kS/s to 3,200 S/s'''
|102.4 kS/s to 2.048 S/s or 65.536 kS/s to 3,200 S/s


|-
|-
|LEMO<sup>2</sup> connectors management
|LEMO<sup>2</sup> connectors management
|'''200 V '''polarization''' on/off''' per block of 8 input
|200 V polarization on/off per block of 8 input


|-
|-
|Coupling<sup>3,4</sup>
|Coupling<sup>3,4</sup>
|'''AC''' (0.35 Hz) - '''DC''' - '''ICP''' - '''ICP <nowiki>+</nowiki> TEDS''' - '''AC float -''' '''DC float''' - '''Grounded'''
|AC (0.35 Hz) - DC - ICP - ICP <nowiki>+</nowiki> TEDS - AC float - DC float - Grounded


|-
|-
|Node information<sup>3</sup>
|Node information<sup>3</sup>
|'''Label''' - '''component''' - '''node''' - '''direction''' (± X ,Y,Z) - '''type''' (translation, rotation)
|Label - component - node - direction (± X ,Y,Z) - type (translation, rotation)


|-
|-
Line 202: Line 190:
|-
|-
|Sensitivity<sup>3,4</sup>
|Sensitivity<sup>3,4</sup>
|User defined in '''V/unit.'''
|User defined in V/unit.


|-
|-
|Range<sup>2</sup>
|Range<sup>2</sup>
|'''Pk to pk''' expressed in the associated unit, (taking in account sensitivity) - '''linear '''or '''dB'''
|Pk to pk expressed in the associated unit, (taking in account sensitivity) - linear or dB


|-
|-
|External conditioner compensation<sup>3</sup>
|External conditioner compensation<sup>3</sup>
|'''Gain''' - '''polarity''' - '''offset'''.
|Gain - polarity - offset.


|-
|-


||Filter<sup>3</sup>
||Filter<sup>3</sup>
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)
|Any NVGate filter can independently be applied to each dynamic input (See Filters §)


|-
|-
Line 223: Line 211:


|-
|-
|rowspan = "13"|'''Ext. sync inputs'''
! rowspan="13" | Ext. sync inputs
|Sampling
|Sampling
|over-sampled 64 time the dynamic inputs sampling
|over-sampled 64 times the dynamic inputs sampling


|-
|-
|Coupling<sup>3</sup>
|Coupling<sup>3</sup>
|'''AC''' (0.35 Hz) - '''DC'''
|AC (0.35 Hz) - DC


|-
|-
Line 241: Line 229:
|-
|-
|Sensitivity<sup>3,4</sup>
|Sensitivity<sup>3,4</sup>
|User defined in '''V/unit'''.
|User defined in V/unit.


|-
|-
|Range<sup>3</sup>
|Range<sup>3</sup>
|'''Pk to pk''' expressed in the associated unit, (taking in account sensitivity) -  '''linear or dB'''
|Pk to pk expressed in the associated unit, (taking in account sensitivity) -  linear or dB


|-
|-
|External conditioner compensation<sup>3</sup>
|External conditioner compensation<sup>3</sup>
|'''Gain''', '''offset'''.
|Gain, offset.


|-
|-
|Edge detection<sup>3</sup>
|Edge detection<sup>3</sup>
|'''Threshold''' (in associated unit) - '''Edge''' (rise, fall) - '''Hysteresis''' - '''hold off''' (sec)
|Threshold (in associated unit) - Edge (rise, fall) - Hysteresis - hold off (sec)


|-
|-
|Pre-divider<sup>3</sup>
|Pre-divider<sup>3</sup>
|'''Hardware - 1''' to '''255 '''pulses – accept up to '''375 kpulses/sec'''
|Hardware - 1 to 255 pulses – accept up to 375 kpulses/sec


|-
|-
|Post-multiplier<sup>3</sup>
|Post-multiplier<sup>3</sup>
|'''1 to 50 pulses/trigger '''- Generate higher frequencies on low rate triggers. (ex. GPS)
|1 to 50 pulses/trigger - Generate higher frequencies on low rate triggers. (ex. GPS)


|-
|-
|Tachometer mode<sup>3</sup>
|Tachometer mode<sup>3</sup>
|'''Pulse per rev''': 0.5 to 4096 - '''average''' (n rev) - '''min''' (detect stopped shaft) and '''max RPM''' speed - '''hold off '''(in % of revolution) - '''rotation''' (clockwise/counterclockwise)
|Pulse per rev: 0.5 to 4096 - average (n rev) - min (detect stopped shaft) and max RPM speed - hold off (in % of revolution) - rotation (clockwise/counterclockwise)


|-
|-
|Torsional mode<sup>2</sup>: Instantaneous angular velocity converter  
|Torsional mode<sup>2</sup>: Instantaneous angular velocity converter  
|'''Pulse per rev''': 1 to 4096 - '''min''' (detect stopped shaft) and '''max RPM''' speed - '''sliding''' '''average''' (1 to 20 samples) - '''Missing pulse number''': 0 to 5, '''Missing pulses detection''' :1 to ±20%:  of previous pulses duration- '''Filters''' (dt, dt², 1/dt)
|Pulse per rev: 1 to 4096 - min (detect stopped shaft) and max RPM speed - sliding average (1 to 20 samples) - Missing pulse number: 0 to 5, Missing pulses detection :1 to ±20%:  of previous pulses duration- Filters (dt, dt², 1/dt)


|-
|-
|Angular sampling<sup>2</sup>: multi-pulse/rev clock used for signal sampling  
|Angular sampling<sup>2</sup>: multi-pulse/rev clock used for signal sampling  
|for '''SOA''' '''Pulse per rev''': 1 to 4096 continuously (no power of 2 limitation) - Real time, in line '''anti-aliasing''' - '''Missing pulse number''': 0 to 5, '''Missing pulses detection''': 1 to ±20%  of previous pulses duration – '''Phase reference''': Any tachometer including same input
|for SOA -  Pulse per rev: 1 to 4096 continuously (no power of 2 limitation) - Real time, in line anti-aliasing - Missing pulse number: 0 to 5, Missing pulses detection: 1 to ±20%  of previous pulses duration – Phase reference: Any tachometer including same input


|}
|}
Line 280: Line 268:




{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "10"|'''Outputs'''
! rowspan="10" | Outputs
|Sampling
|Sampling
|Same as dynamic inputs
|Same as dynamic inputs
Line 287: Line 275:
|-
|-
|Generated signals<sup>3</sup>
|Generated signals<sup>3</sup>
|Any '''outputs signals''', see §outputs - play-back of signal file tracks - DC levels - Inputs playback (Delay <nowiki>></nowiki> 256 samples)
|Any outputs signals, see §outputs - play-back of signal file tracks - DC levels - Inputs playback (Delay <nowiki>></nowiki> 256 samples)


|-
|-
|Filters<sup>3</sup>
|Filters<sup>3</sup>
|'''Any NVGate''' filter can independently be applied to each dynamic output (See Filters §)
|Any NVGate filter can independently be applied to each dynamic output (See Filters §)


|-
|-
|Impedance<sup>3</sup>
|Impedance<sup>3</sup>
|'''50 ohm''', '''600 ohm''' (only OR36 and OR38) or grounded
|50 ohm, 600 ohm (only OR36 and OR38) or grounded


|-
|-
|Gain<sup>3</sup>
|Gain<sup>3</sup>
|'''-100 dB''' to '''<nowiki>+</nowiki> 120 dB'''
|-100 dB to <nowiki>+</nowiki> 120 dB


|-
|-
|Synchronization<sup>3</sup>
|Synchronization<sup>3</sup>
|'''Free run '''or '''linked '''to acquisitions '''runs'''
|Free run or linked to acquisitions runs


|-
|-
|Level and phase transition<sup>3</sup>
|Level and phase transition<sup>3</sup>
|No ('''Steps''') - controlled ('''ramp''') - user selectable ramp time
|No (Steps) - controlled (ramp) - user selectable ramp time


|-
|-
|Clipping
|Clipping
|User selectable - '''protects shaker''' and '''amplifier'''.
|User selectable - protects shaker and amplifier.


|-
|-
|Events (can trig analysis or record)<sup>3</sup>
|Events (can trig analysis or record)<sup>3</sup>
|Outputs '''stabilized''' - new '''step reached''' - '''start & stop sweep'''
|Outputs stabilized - new step reached - start & stop sweep


|-
|-
|Controls
|Controls
|'''Emergency stop''' - '''mute'''/un-mute all
|Emergency stop - mute/un-mute all


|-
|-
|align = "center" rowspan = "7"|'''Parametric<sup>1</sup>(DC) inputs
! rowspan="7" | Parametric<sup>1</sup>(DC) inputs
|Sampling
|Sampling
|'''12.3 to 12.8 S/s - '''10 Hz to 100 Hz harmonics rejection
|12.3 to 12.8 S/s - 10 Hz to 100 Hz harmonics rejection


|-
|-
Line 340: Line 328:
|-
|-
|Range<sup>3</sup>
|Range<sup>3</sup>
|Pk to pk expressed in the associated unit (taking in account sensitivity)  - '''linear''' or '''dB'''
|Pk to pk expressed in the associated unit (taking in account sensitivity)  - linear or dB


|-
|-
|External conditioner compensation<sup>3</sup>
|External conditioner compensation<sup>3</sup>
|'''Gain '''(-120 dB  to <nowiki>+</nowiki>20 dB)''' ''' - '''polarity''' - '''offset '''(<nowiki><</nowiki> range).
|Gain (-120 dB  to <nowiki>+</nowiki>20 dB) - polarity - offset (<nowiki><</nowiki> range).


|-
|-
|Auto-range
|Auto-range
|'''Enable/disable''' auto-ranging '''independently on any input'''
|Enable/disable auto-ranging independently on any input


|-
|-
|align = "center" rowspan = "5"|'''Auto-ranging'''
! rowspan="5" | Auto-ranging
|Type
|Type
|'''Normal''' - on '''Peak''' detection  
|Normal - on Peak detection  


|-
|-
|Margin
|Margin
|Select the first highest range with '''0 dB - 3 dB''' or '''20 dB''' margin
|Select the first highest range with 0 dB - 3 dB or 20 dB margin


|-
|-
|Duration (apply for normal auto-range)
|Duration (apply for normal auto-range)
|'''0''' to '''10 sec'''.
|0 to 10 sec.


|-
|-
Line 369: Line 357:
|-
|-
|Peak parameters settings
|Peak parameters settings
|''' 1 to 10 '''peaks - '''sensitivity''' (low, normal, high) - '''rise''', '''fall''' or '''any edge'''.
| 1 to 10 peaks - sensitivity (low, normal, high) - rise, fall or any edge.


|-
|-
|align = "center" rowspan = "2"|'''Checking'''
! rowspan="2" | Checking
|ICP ''(check not available on OR34)''
|ICP ''(check not available on OR34)''
|Test and report ('''Open''', '''shortcut''', '''ok''') all enabled inputs.
|Test and report (Open, shortcut, ok) all enabled inputs.


|-
|-
|TEDS  
|TEDS  
|Transducers automatic recognition - Complies with '''IEEE 1451.4 2004  Rev. 1.0'''
|Transducers automatic recognition - Complies with IEEE 1451.4 2004  Rev. 1.0


|}
|}
Line 391: Line 379:
The following results are available for monitoring and connection to analysis mode (plug-in analyzers)
The following results are available for monitoring and connection to analysis mode (plug-in analyzers)


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Dynamic inputs & torsional<sup>1</sup> inputs'''
! rowspan="3" | Dynamic inputs & torsional<sup>1</sup> inputs
|Monitoring<sup>3</sup>
|Monitoring<sup>3</sup>
|Time domain instantaneous signal - Size '''256 samples'''
|Time domain instantaneous signal - Size 256 samples


|-
|-
|Status led<sup>3</sup> (docking tool bar)
|Status led<sup>3</sup> (docking tool bar)
|'''Overloaded''' = red - '''overload''' '''occurred since last start''' = red with yellow center - '''ok''' = green - '''under load''' (20 dB below range) = green with yellow center
|Overloaded = red - overload occurred since last start = red with yellow center - ok = green - under load (20 dB below range) = green with yellow center


|-
|-
|Connection<sup>3</sup>
|Connection<sup>3</sup>
|To any: '''plug-in analyzer''' '''channel, monitor''' '''channel, Virtual inputs''', '''recorder''' '''track''', '''edge event detector''' & '''tachometer '''
|To any: plug-in analyzer channel, monitor channel, Virtual inputs, recorder track, edge event detector & tachometer  


|-
|-
|align = "center" rowspan = "2"|'''Ext. Sync inputs '''
! rowspan="2" | Ext. Sync inputs
|Monitoring<sup>3</sup>
|Monitoring<sup>3</sup>
|Time domain '''instantaneous status''' (1 = threshold crossed, 0 = no event) - Size '''256 samples'''.
|Time domain instantaneous status (1 = threshold crossed, 0 = no event) - Size 256 samples.


|-
|-
|Connection<sup>3</sup>
|Connection<sup>3</sup>
|To any: '''start '''or''' stop averaging''' of plug-in analyzers, recorder '''start '''or''' stop recording''', '''waterfall start '''or''' stop acquisition''', '''new block trigger''' for FFT and SOA and '''new slice '''for''' waterfall''' acquisition.<br>To '''torsional''', '''tachometer''' & '''angular sampling'''
|To any: start or stop averaging of plug-in analyzers, recorder start or stop recording, waterfall start or stop acquisition, new block trigger for FFT and SOA and new slice for waterfall acquisition.<br>To torsional, tachometer & angular sampling


|-
|-
|align = "center" rowspan = "3"|'''Parametric<sup>1</sup> (DC) inputs & CAN<sup>1</sup>parameters'''
! rowspan="3" | Parametric<sup>1</sup> (DC) inputs & CAN<sup>1</sup>parameters
|Monitoring<sup>3</sup>
|Monitoring<sup>3</sup>
|'''Profiles''' versus time ('''160 ms to 163 sec''' user selectable), and '''digital'''/'''analog view meter'''.
|Profiles versus time (160 ms to 163 sec user selectable), and digital/analog view meter.


|-
|-
|Status led<sup>3</sup>
|Status led<sup>3</sup>
|'''Overloaded''' = red - '''ok''' = yellow - '''under load''' (20 dB below range) = Cyan
|Overloaded = red - ok = yellow - under load (20 dB below range) = Cyan


|-
|-
|Connection<sup>3</sup>
|Connection<sup>3</sup>
|To any: '''recorder track''', '''waterfall reference (Z/X axis)''' and '''profiles''', '''level''' '''delta level detectors''' and '''virtual DC'''
|To any: recorder track, waterfall reference (Z/X axis) and profiles, level delta level detectors and virtual DC


|}
|}
Line 441: Line 429:
* '''Time and tracks split''', the chosen recorded tracks (from signal file) are played back in the recorder using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be changed.
* '''Time and tracks split''', the chosen recorded tracks (from signal file) are played back in the recorder using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be changed.


'''Recorder settings'''
'''Recorder Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Bandwidths'''
! rowspan="3" | Bandwidths
|Dynamics inputs
|Dynamics inputs
|'''2 groups''' of user selectable sampling frequency - From '''102.4 kS/s '''to '''2.048 S/s''' ('''40 kHz''' to '''800 mHz''') - available simultaneously
|2 groups of user selectable sampling frequency - From 102.4 kS/s to 2.048 S/s (40 kHz to 800 mHz) - available simultaneously


|-
|-
|Ext. sync inputs
|Ext. sync inputs
|'''Automatic selection''' at Front-end sampling frequency - Resolution is '''64 time Front-end sampling frequency'''
|Automatic selection at Front-end sampling frequency - Resolution is 64 time Front-end sampling frequency


|-
|-
|Parametric (DC) inputs
|Parametric (DC) inputs
|'''12.5 S/s''' ('''50 Hz''' rejection) - '''15 S/s''' ('''60 Hz''' rejection)
|12.5 S/s (50 Hz rejection) - 15 S/s (60 Hz rejection)


|-
|-
|align = "center" rowspan = "2"|'''Tracks'''
! rowspan="2" | Tracks
|Number
|Number
|'''6 tracks For O4.''' or '''38 tracks For OR35, OR36, OR38 and Office licenses.''' (32 <nowiki>+</nowiki> 6 ext. Sync)<br>Sum of connected licenses using multiple hardware
|6 tracks For O4. or 38 tracks For OR35, OR36, OR38 and Office licenses. (32 <nowiki>+</nowiki> 6 ext. Sync)<br>Sum of connected licenses using multiple hardware


|-
|-
|Saved settings / track
|Saved settings / track
|'''Label''' - '''Coupling''' - '''External gain''' - Input '''range''' - '''Sampling frequency '''and''' signal bandwidth - Component''' - '''Node''' number - '''Direction''' - '''Type''' - Associated '''transducer''' - '''Unit -''' '''Sensitivity''' and '''Offset''' compensation.
|Label - Coupling - External gain - Input range - Sampling frequency and signal bandwidth - Component - Node number - Direction - Type - Associated transducer - Unit - Sensitivity and Offset compensation.


|-
|-
|align = "center" rowspan = "3"|'''Modes'''
! rowspan="3" | Modes
|Start to time
|Start to time
|Start recording on run or any activated event - '''Stop recording when duration ends '''- Duration '''10 ms''' to '''available space on target HDD.'''
|Start recording on run or any activated event - Stop recording when duration ends - Duration 10 ms to available space on target HDD.


|-
|-
|Start to stop
|Start to stop
|'''Start recording''' on '''run''' or '''any activated event''' - Stop recording on stop on any activated event - Start and stop event can be the same - Duration '''10 ms''' to '''available space on active drive '''(PC or OR3x).
|Start recording on run or any activated event - Stop recording on stop on any activated event - Start and stop event can be the same - Duration 10 ms to available space on active drive (PC or OR3x).


|-
|-
|Time to stop
|Time to stop
|Memorize the '''earliest x seconds''' - Stop = '''stop''' or any '''activated event''' - Duration '''10 ms''' to '''2 GSamples '''(limited by the drive available space)
|Memorize the earliest x seconds - Stop = stop or any activated event - Duration 10 ms to 2 GSamples (limited by the drive available space)


|-
|-
|align = "center" rowspan = "4"|'''Triggering'''
! rowspan="4" | Triggering
|Start recording (new record)
|Start recording (new record)
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Stop recording (end current record)
|Stop recording (end current record)
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Start delay
|Start delay
|Positive''' '''=''' unlimited''' - negative = '''128 kSamples'''
|Positive = unlimited - negative = 128 kSamples


|-
|-
|Stop delay
|Stop delay
|Positive''' '''=''' unlimited''' - negative = '''0'''
|Positive = unlimited - negative = 0


|-
|-
|align = "center"|'''Markers'''
! Markers
|Saved in the record files
|Saved in the record files
|Added by tool bar or shortcut - '''comment fields''' editable during record or at end of acquisition
|Added by tool bar or shortcut - comment fields editable during record or at end of acquisition


|-
|-
|align = "center" rowspan = "3"|'''Signal file'''
! rowspan="3" | Signal file
|Record on PC HDD
|Record on PC HDD
|Up to '''10''' dynamic inputs – Up to 4 parametric inputs - max total bandwidth '''512 kS/s''' - (2 MB/s)
|Up to 10 dynamic inputs – Up to 4 parametric inputs - max total bandwidth 512 kS/s - (2 MB/s)


|-
|-
|Record on Analyzer SSD
|Record on Analyzer SSD
|Up to '''40 channels''' (32 inputs <nowiki>+</nowiki> 6 ext. sync or DC)/instruments - max bandwidth '''3.2 MS/s''' (12.5 MB/s)
|Up to 40 channels (32 inputs <nowiki>+</nowiki> 6 ext. sync or DC)/instruments - max bandwidth 3.2 MS/s (12.5 MB/s)


|-
|-
|Format
|Format
|'''Normal''': '''32 bits'''/sample - '''Compressed''' '''16 bits'''/sample  
|Normal: 32 bits/sample - Compressed 16 bits/sample  


|}
|}


'''Recorder settings (continued)'''
'''Recorder Settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "4"|'''Management'''
! rowspan="4" | Management
|Download
|Download
|Discard Analyzer SSD on PC HDD''' - Batch download''' available - Rate '''4 MB/s'''
|Discard Analyzer SSD on PC HDD - Batch download available - Rate 4 MB/s


|-
|-
|Upload
|Upload
|Load PC HDD records on the analyzer SSD - Rate '''1.4 MB/s'''
|Load PC HDD records on the analyzer SSD - Rate 1.4 MB/s  


|-
|-
Line 532: Line 520:
|-
|-
|Hard drive management
|Hard drive management
|'''Format '''SSD - '''Selective delete''' - '''Sorting '''tools
|Format SSD - Selective delete - Sorting tools


|}
|}
Line 543: Line 531:
* USB 2.0 port for post processing and data management without the instrument.
* USB 2.0 port for post processing and data management without the instrument.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center"|'''Connection'''
! Connection
|1 to 8 Mobi-Disc<sup>TM</sup>
|1 to 8 Mobi-Disc<sup>TM</sup>
|Up to '''8 Mobi-disc''' connected - Active Mobi-Disc: '''1''' (user selectable) - '''hot swap'''
|Up to 8 Mobi-disc connected - Active Mobi-Disc: 1 (user selectable) - hot swap


|-
|-
|align = "center"|'''Transfer'''
! Transfer
|Based on the USB 3.0 link
|Based on the USB 3.0 link
|'''Download 15 MB/s'''
|Download 15 MB/s


|}
|}


'''Recorder monitoring'''
'''Recorder Monitoring'''


The following displays are available during real-time or post-analysis.
The following displays are available during real-time or post-analysis.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center"|'''RMS'''
! RMS
|Overall RMS / input
|Overall RMS / input
|'''Digital '''or '''analog view-meter'''
|Digital or analog view-meter


|-
|-
|align = "center"|'''Signal'''
! Signal
|Real-time
|Real-time
|'''Compressed view of entire''' recorded signals - Automatic update of time axis.
|Compressed view of entire recorded signals - Automatic update of time axis.


|}
|}
Line 576: Line 564:
* '''Post-analysis''', the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
* '''Post-analysis''', the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
* '''Playbacks''', the recorded dynamic input are generated on the instrument outputs simultaneously with standard analysis of the inputs.
* '''Playbacks''', the recorded dynamic input are generated on the instrument outputs simultaneously with standard analysis of the inputs.
'''Player settings'''
'''Player Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "5"|'''Played section'''
! rowspan="5" | Played section
|File selection
|File selection
|NVGate<sup>®</sup> '''recorded files''' - '''Imported files''' - located on '''instrument SSD, Mobi-Disk'''<sup>'''TM'''</sup> or on '''PC HDD'''.
|NVGate<sup>®</sup> recorded files - Imported files - located on instrument SSD, Mobi-Disk<sup>TM</sup> or on PC HDD.


|-
|-
|Record selection
|Record selection
|'''Record number selection''' (for multi record files)
|Record number selection (for multi record files)


|-
|-
|Section definition
|Section definition
|User selectable '''Start''' and '''stop offsets''' (in sec.) - available in the '''file overview window''' (marker and slider)
|User selectable Start and stop offsets (in sec.) - available in the file overview window (marker and slider)


|-
|-
|Duration
|Duration
|'''Play backed duration''' (User information)
|Play backed duration (User information)


|-
|-
|Repeat
|Repeat
|'''On/off''' - repeat continuously the selection (available only for playback on analyzer output)
|On/off - repeat continuously the selection (available only for playback on analyzer output)


|-
|-
|align = "center" rowspan = "3"|'''Mode'''
! rowspan="3" | Mode
|File swap
|File swap
|'''Swap loaded files with same tracks number''' - applies '''identical post-analysis''' setup to multiple files.
|Swap loaded files with same tracks number - applies identical post-analysis setup to multiple files.


|-
|-
Line 613: Line 601:


|-
|-
|align = "center" rowspan = "5"|'''Tracks'''
! rowspan="5" | Tracks
|Number
|Number
|'''Up to 352 '''according to user<nowiki>'</nowiki>s fleet
|Up to 352 according to user<nowiki>'</nowiki>s fleet


|-
|-
|Type
|Type
|'''Dynamic input''' record (2 sampling frequencies), '''parametric '''(DC) input record and '''ext. sync''' record.
|Dynamic input record (2 sampling frequencies), parametric (DC) input record and ext. sync record.


|-
|-
|Fixed setting (information) / track
|Fixed setting (information) / track
|'''Label''', '''Coupling''', '''external gain''', input '''range''', '''sampling frequency '''and''' signal bandwidth'''
|Label, Coupling, external gain, input range, sampling frequency and signal bandwidth


|-
|-
|Modifiable measurement point information settings (apply for post analysis or re-recording) / track
|Modifiable measurement point information settings (apply for post analysis or re-recording) / track
|'''Component''' - '''node''' number - '''direction''' - '''type'''
|Component - node number - direction - type


|-
|-
|Modifiable settings (apply for post analysis or re-recording) / track
|Modifiable settings (apply for post analysis or re-recording) / track
|Associated '''transducer''' - '''unit''' - '''sensitivity''' - '''offset''' compensation
|Associated transducer - unit - sensitivity - offset compensation


|}
|}


'''Player settings (continued)'''
'''Player Settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Listening'''
! rowspan="2" | Listening
|File location
|File location
|'''Analyzer HD''' - '''Mobi-Disk'''<sup>'''TM'''</sup> connected with USB or inside the analyzer - '''PC HD'''
|Analyzer HD - Mobi-Disk<sup>TM</sup> connected with USB or inside the analyzer - PC HD  


|-
|-
Line 647: Line 635:


|-
|-
|align = "center" rowspan = "2"|'''Markers'''
! rowspan="2" | Markers
|Display
|Display
|Recorded markers are available in the '''file overview'''
|Recorded markers are available in the file overview  


|-
|-
Line 657: Line 645:
|}
|}


'''Player connections'''
'''Player Connections'''


The following table describes the available connections to  the analysis modes (plug-in analyzers and recorder) during post-analysis operation:
The following table describes the available connections to  the analysis modes (plug-in analyzers and recorder) during post-analysis operation:


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Dynamic inputs & torsional Tracks'''
! rowspan="3" | Dynamic inputs & torsional Tracks
|Preview (multiple file simultaneously)
|Preview (multiple file simultaneously)
|'''Entire file fast overview''' (pre-compressed at recording time) - '''Track contents''' preview, independently (multi-graph display)
|Entire file fast overview (pre-compressed at recording time) - Track contents preview, independently (multi-graph display)


|-
|-
|Monitoring (loaded file only)
|Monitoring (loaded file only)
|'''Zoom '''on selected '''play-back section'''
|Zoom on selected play-back section


|-
|-
|Connection3 (loaded file only)
|Connection3 (loaded file only)
|To any: '''plug-in analyzer''' '''channel, monitor''' '''channel,''' '''recorder''' '''track''', '''edge event detector''' and '''tachometer '''
|To any: plug-in analyzer channel, monitor channel, recorder track, edge event detector and tachometer  


|-
|-
|align = "center" rowspan = "3"|'''Ext. sync Tracks'''
! rowspan="3" | Ext. sync Tracks
|Preview (multiple file simultaneously)
|Preview (multiple file simultaneously)
|'''Entire file fast overview''' (pre-compressed at recording time) - '''Track contents preview''', independently (multi-graph display)
|Entire file fast overview (pre-compressed at recording time) - Track contents preview, independently (multi-graph display)


|-
|-
|Monitoring (loaded file only)
|Monitoring (loaded file only)
|'''Zoom '''on selected '''play-back section'''
|Zoom on selected play-back section


|-
|-
|Connection<sup>3 </sup>(loaded file only)
|Connection<sup>3 </sup>(loaded file only)
|To any: '''start '''or''' stop averaging''' of plug-in analyzers, recorder '''start '''or''' stop recording''', '''waterfall start '''or''' stop acquisition''', '''new block trigger''' for FFT and SOA and '''new slice '''for''' waterfall''' acquisition.<br>To '''torsional''', '''tachometer''' & '''angular sampling'''
|To any: start or stop averaging of plug-in analyzers, recorder start or stop recording, waterfall start or stop acquisition, new block trigger for FFT and SOA and new slice for waterfall acquisition.<br>To torsional, tachometer & angular sampling


|-
|-
|align = "center" rowspan = "3"|'''Recorded parametric<sup>1</sup>(DC) inputs & CAN parameters'''
! rowspan="3" | Recorded parametric<sup>1</sup>(DC) inputs & CAN parameters
|Preview (multiple file simultaneously)
|Preview (multiple file simultaneously)
|Entire file''' fast overview''' - '''Track contents''' preview
|Entire file fast overview  - Track contents preview


|-
|-
|Monitoring (loaded file only)
|Monitoring (loaded file only)
|'''Profiles''' versus time and '''digital''' or '''analog view meter'''.
|Profiles versus time and digital or analog view meter.


|-
|-
|Connection<sup>3 </sup>(loaded file only)
|Connection<sup>3 </sup>(loaded file only)
|To any: '''recorder track''', '''waterfall reference (Z/X axis)''' and '''profiles''', '''level''' and '''delta level events'''.
|To any: recorder track, waterfall reference (Z/X axis) and profiles, level and delta level events.


|}
|}
Line 705: Line 693:
The following plug-in analyzers are available as options of NVGate<sup>®</sup> software platform.
The following plug-in analyzers are available as options of NVGate<sup>®</sup> software platform.


===Time Domain Analysis - TDA===
===Time Domain Analysis (TDA)===
The TDA plug-in analyzer provides time domain based analyses and visualizations. This plug-in analyzer computes statistical extractions and data compression of long duration oscilloscope views. All results are real time and operations are possible on-line or in post processing. The TDA plug-ins can operate free run or synchronized with the other plug-in analyzers.
The TDA plug-in analyzer provides time domain based analyses and visualizations. This plug-in analyzer computes statistical extractions and data compression of long duration oscilloscope views. All results are real time and operations are possible on-line or in post processing. The TDA plug-ins can operate free run or synchronized with the other plug-in analyzers.


'''TDA Settings'''
'''TDA Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Bandwidths'''
! rowspan="2" | Bandwidths
|Oscillator 1
|Oscillator 1
|'''800 mHz''' to '''40 kHz''' <br>(sampling oscillator 102.4 kS/s to 2,048 S/s)  
|800 mHz to 40 kHz <br>(sampling oscillator 102.4 kS/s to 2.048 S/s)  


|-
|-
|Oscillator 2
|Oscillator 2
|'''512 mHz''' to '''25.6 kHz''' <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)
|512 mHz to 25.6 kHz <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)


|-
|-
|align = "center" rowspan = "2"|'''Time views'''
! rowspan="2" | Time views
|Depth
|Depth
|'''7200/ABW ABW: the plug-in Analysis Bandwidth. '''sec to''' 100 000 '''sec (27 hours) - '''Independent''' on each channel
|7200/ABW ABW: the plug-in Analysis Bandwidth. sec to 100 000 sec (27 hours) - Independent on each channel


|-
|-
|Time base (resolution)
|Time base (resolution)
|'''Depth/2048'''- '''Independent''' on each channel
|Depth/2048- Independent on each channel


|-
|-
|align = "center" rowspan = "2"|'''Time span for extraction'''<br>'''(averaging)'''
! rowspan="2" | Time span for extraction<br>(averaging)
|Type
|Type
|Continuous sliding with refresh each 256 sample'''s '''('''Exponential''') - One shot ('''Linear''') - Repeated shot ('''Repeated linear''')
|Continuous sliding with refresh each 256 samples (Exponential) - One shot (Linear) - Repeated shot (Repeated linear)


|-
|-
|Duration
|Duration
|'''1/ABW '''to''' 2 10'''<sup>'''6'''</sup>'''/ABW''' (ex: 50 ms to 1000 sec @ 2 kHz)
|1/ABW to  2 10<sup>6</sup>/ABW (ex: 50 ms to 1000 sec @ 2 kHz)


|-
|-
|align = "center" rowspan = "2"|'''Channels'''
! rowspan="2" | Channels
|OR36 & OR38 analyzers
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line''' 32 to 256 in '''post-analysis'''
|32 channels per plug-in for on line 32 to 256 in post-analysis


|-
|-
|OR34 & OR35 analyzers
|OR34 & OR35 analyzers
|'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''
|8 channels per plug-in for on line and post-analysis


|-
|-
|align = "center"|'''Filters'''
! Filters
|On each channel
|On each channel
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)
|Any NVGate filter can independently be applied to each dynamic input (See Filters §)


|-
|-
|align = "center" rowspan = "4"|'''Triggering'''<br><br>
! rowspan="4" | Triggering<br><br>
|Start analysis
|Start analysis
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|End analysis
|End analysis
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Repeat analysis
|Repeat analysis
|On '''new start''' or '''end of averaging'''
|On new start or end of averaging


|-
|-
|Trigger delay
|Trigger delay
|Positive''' '''=''' unlimited''' - negative = '''32k Samples'''
|Positive = unlimited - negative = 32k Samples


|}
|}
Line 774: Line 762:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center"|'''Time Views'''
! Time Views
|Overview
|Overview
|'''Signal envelope: '''Min/Max line for each time step -  
|Signal envelope: Min/Max line for each time step -  


|-
|-
|align = "center"|&nbsp;
! &nbsp;
|Statistical values
|Statistical values
|'''DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Skewness - Kurtosis - '''computed on a user defined area of the time view
|DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Skewness - Kurtosis - computed on a user defined area of the time view


|-
|-
|align = "center"|'''Extractions'''
! Extractions
|Statistical values
|Statistical values
|'''DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Kurtosis - '''computed on the Time span – All values available for '''View-meter''' and/or '''Waterfall profiles'''
|DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Kurtosis - computed on the Time span – All values available for View-meter and/or Waterfall profiles


|}
|}


----


===Fast Fourier Transform analysis - FFT===
===Fast Fourier Transform analysis (FFT)===
The FFT plug-in analyzer features multiple vibrations results depending on applied setting. This type of plug-in analyzers computes real-time (on-line or post-processing) Fast Fourier Transform algorithms with multiple averaging domain and trigger capabilities. Up to 4 FFT plug-ins can operate at the same time.
The FFT plug-in analyzer features multiple vibrations results depending on applied setting. This type of plug-in analyzers computes real-time (on-line or post-processing) Fast Fourier Transform algorithms with multiple averaging domain and trigger capabilities. Up to 4 FFT plug-ins can operate at the same time.


'''FFT Settings'''
'''FFT Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Bandwidths'''
! rowspan="2" | Bandwidths
|Oscillator 1
|Oscillator 1
|'''800 mHz''' to '''40 kHz''' <br>(sampling oscillator 102.4 kS/s to 2,048 S/s)  
|800 mHz to 40 kHz <br>(sampling oscillator 102.4 kS/s to 2.048 S/s)  


|-
|-
|Oscillator 2
|Oscillator 2
|'''512 mHz''' to '''25.6 kHz''' <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)
|512 mHz to 25.6 kHz <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)


|-
|-
|align = "center" rowspan = "2"|'''Resolution'''
! rowspan="2" | Resolution
|Lines
|Lines
|'''101''', '''201''', '''401''', '''801''', '''1601''', '''3201''', '''6401, 12801 With force DSP and 25601 lines'''
|101, 201, 401, 801, 1601, 3201, 6401, 12801 With force DSP and 25601 lines


|-
|-
|Frequency resolution
|Frequency resolution
|'''80 µHz '''(512 mHz/6401 lines) <br>to '''400 Hz '''(40 kHz/101 lines)
|80 µHz (512 mHz/6401 lines) <br>to 400 Hz (40 kHz/101 lines)


|-
|-
|align = "center" rowspan = "4"|'''Averaging'''
! rowspan="4" | Averaging
|Domain  
|Domain  
|'''Spectral''' (power) - '''time''' (with phase) - '''FDSA''' (Synchronized with one frequency)
|Spectral (power) - time (with phase) - FDSA (Synchronized with one frequency)


|-
|-
|Type
|Type
|'''Exponential''' - '''linear''' - '''repeated linear''' - '''referenced peak hold''' (eq. to tracking filter locked on reference channel frequency) - '''peak hold'''
|Exponential - linear - repeated linear - referenced peak hold (eq. to tracking filter locked on reference channel frequency) - peak hold


|-
|-
|Size
|Size
|'''Duration '''or '''number of blocks''', unlimited size
|Duration or number of blocks, unlimited size


|-
|-
|Overlap
|Overlap
|'''0%''' to '''99.9%''' depending on block size - '''retrigger'''
|0% to 99.9% depending on block size - retrigger


|-
|-
|align = "center" rowspan = "2"|'''Channels'''
! rowspan="2" | Channels
|OR36, MP & OR38 analyzers
|OR36, MP & OR38 analyzers
|'''32 to 256 ch. '''per plug-in for '''on line''' and '''post-analysis'''
|32 to 256 ch. per plug-in for on line and post-analysis


|-
|-
|OR34 & OR35 analyzers
|OR34 & OR35 analyzers
|'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''
|8 channels per plug-in for on line and post-analysis


|-
|-
|align = "center" rowspan = "3"|'''Weighting'''
! rowspan="3" | Weighting
|General
|General
|'''Uniform - Hanning - Hamming - Kaiser Bessel'''
|Uniform - Hanning - Hamming - Kaiser Bessel


|-
|-
|Modal acquisition
|Modal acquisition
|User defined''' Force '''(rect.)''' '''and '''Response '''(Rect. <nowiki>+</nowiki> exp)
|User defined Force (rect.) and Response (Rect. <nowiki>+</nowiki> exp)


|-
|-
Line 856: Line 843:


|-
|-
|align = "center"|'''Filters'''
! Filters
|On each channel
|On each channel
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)
|Any NVGate filter can independently be applied to each dynamic input (See Filters §)


|-
|-
|align = "center"|'''Cross functions'''
! Cross functions
|Reference selection
|Reference selection
|'''Full''' or '''partial matrix''' available (all channels can be references)
|Full or partial matrix available (all channels can be references)


|-
|-
|align = "center" rowspan = "2"|'''Zoom'''
! rowspan="2" | Zoom
|Magnification factor
|Magnification factor
|'''2 to 128 '''by step power of 2
|2 to 128 by step power of 2


|-
|-
Line 875: Line 862:


|-
|-
|align = "center" rowspan = "6"|'''Triggering'''
! rowspan="6" | Triggering
|Start averaging
|Start averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|End averaging
|End averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|New block
|New block
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Repeat averaging
|Repeat averaging
|On '''new start''' or '''end of averaging'''
|On new start or end of averaging


|-
|-
|Trigger delay
|Trigger delay
|Positive''' '''=''' unlimited''' - negative = '''32k Samples'''
|Positive = unlimited - negative = 32k Samples


|-
|-
|Automatic
|Automatic
|Reject '''overloaded blocks'''
|Reject overloaded blocks


|-
|-
|align = "center" rowspan = "2"|'''Blocks rejection'''<br>'''Overall'''
! rowspan="2" | Blocks rejection<br>Overall
|Manual
|Manual
|'''Accept''' or '''Reject '''after preview of averaged results (FRF, Coherence)
|Accept or Reject after preview of averaged results (FRF, Coherence)


|-
|-
|Normal
|Normal
|'''Accept''' all blocks
|Accept all blocks


|-
|-
|align = "center" rowspan = "2"|'''Overall'''
! rowspan="2" | Overall
|Detector
|Detector
|'''Quadratic sum of spectra lines''' taking in '''account weighting window equivalent noise bandwidth'''
|Quadratic sum of spectra lines taking in account weighting window equivalent noise bandwidth


|-
|-
Line 923: Line 910:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "4"|'''Time averaging'''
! rowspan="4" | Time averaging
|Time domain
|Time domain
|'''Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks'''
|Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks


|-
|-
|Spectra
|Spectra
|'''Complex spectra'''
|Complex spectra


|-
|-
|Cross-functions
|Cross-functions
|Instantaneous''' cross-spectra''' - averaged '''cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence'''
|Instantaneous cross-spectra - averaged cross-spectra - FRF H1 - FRF H2 - Coherence


|-
|-
|Overall
|Overall
|'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.
|Overall RMS value in specified band - Overall RMS profile in specified band.


|-
|-
|align = "center" rowspan = "6"|'''Spectral averaging'''
! rowspan="6" | Spectral averaging
|Time domain
|Time domain
|'''Filtered signal - Triggered block - Weighted Block - Lissajous of triggered blocks'''
|Filtered signal - Triggered block - Weighted Block - Lissajous of triggered blocks


|-
|-
|Spectra
|Spectra
|Instantaneous '''complex spectra''' - averaged '''power spectra'''
|Instantaneous complex spectra - averaged power spectra


|-
|-
|Zoomed spectra
|Zoomed spectra
|'''Zoomed''' instantaneous '''complex spectra''' - '''zoomed '''averaged '''power spectra'''
|Zoomed instantaneous complex spectra - zoomed averaged power spectra


|-
|-
|Cross-functions
|Cross-functions
|'''Instantaneous cross-spectra''' - averaged '''cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence'''
|Instantaneous cross-spectra - averaged cross-spectra - FRF H1 - FRF H2 - Coherence


|-
|-
|Zoomed cross-function
|Zoomed cross-function
|Instantaneous '''zoomed cross-spectra''' - '''averaged zoomed cross-spectra''' - '''zoomed FRF H1''' - '''zoomed FRF H2''' - '''zoomed Coherence'''
|Instantaneous zoomed cross-spectra - averaged zoomed cross-spectra - zoomed FRF H1 - zoomed FRF H2 - zoomed Coherence


|-
|-
|Overall
|Overall
|'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.
|Overall RMS value in specified band - Overall RMS profile in specified band.


|-
|-
|align = "center" rowspan = "3"|'''Synchronous averaging on one frequency (FDSA)'''
! rowspan="3" | Synchronous averaging on one frequency (FDSA)
|Time domain
|Time domain
|'''Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks'''
|Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks


|-
|-
|Spectra
|Spectra
|Instantaneous '''complex spectra''' - time domain averaged '''power spectra'''
|Instantaneous complex spectra - time domain averaged power spectra


|-
|-
|Overall
|Overall
|'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.
|Overall RMS value in specified band - Overall RMS profile in specified band.


|}
|}
Line 985: Line 972:
'''CBT Settings'''
'''CBT Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Tracked Order'''
! rowspan="3" | Tracked Order
|Number  
|Number  
|'''8 per channels'''
|8 per channels


|-
|-
|Max
|Max
|'''0.001''' to '''800'''
|0.001 to 800


|-
|-
|Constant bandwidth
|Constant bandwidth
|'''User selectable per channel '''- minimum depend on weighting windows
|User selectable per channel - minimum depend on weighting windows


|-
|-
|align = "center" rowspan = "3"|'''Computation'''
! rowspan="3" | Computation
|Associated tachometer
|Associated tachometer
|'''Any front end''', '''recorded''' or '''virtual tachometer'''
|Any front end, recorded or virtual tachometer


|-
|-
|Peak tracking
|Peak tracking
|On/Off '''center bandwidth on nearest peak'''
|On/Off center bandwidth on nearest peak


|-
|-
Line 1,017: Line 1,004:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Scalar'''
! rowspan="3" | Scalar
|Tracked order
|Tracked order
|'''Digital '''(magnitude and phase)''' '''or '''analog '''view-meter
|Digital (magnitude and phase) or analog view-meter


|-
|-
|Cross phase tracking
|Cross phase tracking
|'''Order phases '''are '''relative''' to the same order from a ref. channel
|Order phases are relative to the same order from a ref. channel


|-
|-
|Overall
|Overall
|'''Digital '''or '''analog view-meter'''
|Digital or analog view-meter


|-
|-
|align = "center"|'''Monitoring'''
! Monitoring
|Continuous profiles of tracked order
|Continuous profiles of tracked order
|'''Profiles '''vs.''' time''' - '''profiles '''vs.''' RPM''' - max depth '''2048 pts''' - user selectable '''delta time''' - user selectable '''delta RPM'''
|Profiles vs. time - profiles vs. RPM - max depth 2048 pts - user selectable delta time - user selectable delta RPM


|-
|-
|align = "center" rowspan = "2"|'''Profiles'''
! rowspan="2" | Profiles
|One shot acquisition
|One shot acquisition
|Tracked order (complex), cross phase orders and overall level can be collected by the ''waterfall ''profiles.  
|Tracked order (complex), cross phase orders and overall level can be collected by the ''waterfall ''profiles.  
Line 1,053: Line 1,040:
'''FFTDiag Settings'''
'''FFTDiag Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Envelop demodulation'''
! rowspan="3" | Envelop demodulation
|Activation
|Activation
|Enable/disable on all channels''' - '''requires '''zoom activation '''
|Enable/disable on all channels - requires zoom activation  


|-
|-
Line 1,064: Line 1,051:
|-
|-
|Bandwidth
|Bandwidth
|'''½ of zoom span - ¼ of FFT bandwidth'''
|½ of zoom span - ¼ of FFT bandwidth


|-
|-
|align = "center" rowspan = "2"|'''Correlation'''
! rowspan="2" | Correlation
|Activation
|Activation
|Enable/disable on all channels''' - '''All other FFT results are affected
|Enable/disable on all channels - All other FFT results are affected


|-
|-
|Weighting window
|Weighting window
|'''Uniform - Left '''zero padding''' - Centred '''zero padding
|Uniform - Left zero padding - Centred zero padding


|}
|}
Line 1,081: Line 1,068:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Scalar'''
! rowspan="2" | Scalar
|Min/Max
|Min/Max
|Minimum and maximum amplitude per trigger block
|Minimum and maximum amplitude per trigger block
Line 1,088: Line 1,075:
|-
|-
|Peak detectors
|Peak detectors
|'''Peak level - Peak to Peak level - Crest Factor  - '''per trigger block
|Peak level - Peak to Peak level - Crest Factor  - per trigger block


|-
|-
|align = "center" rowspan = "3"|'''Others'''
! rowspan="3" | Others
|Cepstrum
|Cepstrum
|'''Cepstrum - Zoomed cepstrum'''
|Cepstrum - Zoomed cepstrum


|-
|-
|Correlation
|Correlation
|'''Auto-correlation '''block''' - Cross correlation '''block
|Auto-correlation block - Cross correlation block


|-
|-
Line 1,105: Line 1,092:
|}
|}


----


===Synchronous Order Analysis (SOA)===
===Synchronous Order Analysis (SOA)===
Line 1,112: Line 1,098:
'''SOA Settings'''
'''SOA Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Bandwidths'''
! rowspan="3" | Bandwidths
|Order span
|Order span
|'''6.25''' to '''400''' – up to '''800 '''with '''''ForceDSP'''''
|6.25 to 400 – up to 800 with ''ForceDSP''


|-
|-
|Frequency bandwidth
|Frequency bandwidth
|Up to '''40 kHz'''
|Up to 40 kHz  


|-
|-
|RPM span
|RPM span
|Up to''' 384 000 RPM - Ratio of 64 '''between Min & Max RPM
|Up to 384 000 RPM - Ratio of 64 between Min & Max RPM


|-
|-
|align = "center" rowspan = "2"|'''Resolution'''
! rowspan="2" | Resolution
|Lines
|Lines
|'''101 - 201 - 401 – 801 – 1601 '''with '''''ForceDSP'''''
|101 - 201 - 401 – 801 – 1601 with ''ForceDSP''


|-
|-
|Order resolution
|Order resolution
|'''1 - '''½ - ¼''' - 1/8 - 1/16 - 1/32'''
|1 - ½ - ¼ - 1/8 - 1/16 - 1/32


|-
|-
|align = "center" rowspan = "4"|'''Averaging'''
! rowspan="4" | Averaging
|Domain  
|Domain  
|'''Spectral''' (power) - '''revolution''' (re-sampled blocs)
|Spectral (power) - revolution (re-sampled blocs)


|-
|-
|Type
|Type
|'''Exponential''' - '''linear''' - '''repeated linear '''- '''peak hold'''
|Exponential - linear - repeated linear - peak hold  


|-
|-
|Size
|Size
|'''Number of blocks''', unlimited size
|Number of blocks, unlimited size


|-
|-
|Overlap
|Overlap
|'''0 to 31 revolutions''' depending on resolution – '''0 to 359°''' into revolution – Retrigger on''' multi-pulse/rev'''
|0 to 31 revolutions depending on resolution – 0 to 359° into revolution – Retrigger on multi-pulse/rev


|-
|-
|align = "center" rowspan = "3"|'''Tachometer'''
! rowspan="3" | Tachometer
|Computation base
|Computation base
|Any '''NVGate'''<sup>'''®</sup> ''''''tachometer '''(ext. sync, regular input, DC input, CAN, simulated, Fractional, and combined tach.)
|Any NVGate<sup>®</sup> tachometer (ext. sync, regular input, DC input, CAN, simulated, Fractional, and combined tach.)


|-
|-
|Phase reference
|Phase reference
|Tachometer pulse '''edge '''or pulse '''center '''requires to analyze the tachometer input
|Tachometer pulse edge or pulse center requires to analyze the tachometer input


|-
|-
Line 1,165: Line 1,151:


|-
|-
|align = "center" rowspan = "2"|'''RPM'''
! rowspan="2" | RPM
|Max speed variation
|Max speed variation
|'''1''' to '''99%''' per analyzed block - blocks with higher variation are rejected. No control (set @ 100%)
|1 to 99% per analyzed block - blocks with higher variation are rejected. No control (set @ 100%)


|-
|-
|RPM range
|RPM range
|User selectable '''max''' & '''min RPM''', under and over speed blocks are rejected
|User selectable max & min RPM, under and over speed blocks are rejected


|-
|-
|align = "center" rowspan = "3"|'''Tracked order'''
! rowspan="3" | Tracked order
|Number
|Number
|'''Up to 8 '''per channels
|Up to 8 per channels


|-
|-
|Min-Max
|Min-Max
|'''1/32 '''to''' 400''' – up to '''800 '''with '''''ForceDSP'''''
|1/32 to 400 – up to 800 with ''ForceDSP''


|-
|-
|Cross phase tracking
|Cross phase tracking
|'''Order phases '''are '''relative''' to the same order from a ref. channel
|Order phases are relative to the same order from a ref. channel


|-
|-
|align = "center" rowspan = "2"|'''Channels'''
! rowspan="2" | Channels
|OR36 & OR38 analyzers
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line''' - 32 to 256 in '''post-analysis'''
|32 channels per plug-in for on line - 32 to 256 in post-analysis


|-
|-
|OR34 & OR35 analyzers
|OR34 & OR35 analyzers
|'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''
|8 channels per plug-in for on line and post-analysis


|-
|-
|align = "center"|'''Weighting'''
! Weighting
|Windows selection
|Windows selection
|'''Uniform - Hanning - Hamming - Kaiser Bessel''' Applied independently on any channels
|Uniform - Hanning - Hamming - Kaiser Bessel Applied independently on any channels


|-
|-
|align = "center"|'''Filters'''
! Filters
|On each channel
|On each channel
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)
|Any NVGate filter can independently be applied to each dynamic input (See Filters §)


|-
|-
|align = "center" rowspan = "6"|'''Triggering'''
! rowspan="6" | Triggering
|Start averaging
|Start averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|End averaging
|End averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|New block
|New block
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run '''
|Any NVGate<sup>® </sup>event plus manual and free run  


|-
|-
|Repeat averaging
|Repeat averaging
|On '''new start''' or '''end of averaging'''
|On new start or end of averaging


|-
|-
|Trigger delay
|Trigger delay
|Positive''' '''=''' unlimited''' - negative = '''1/order res '''revolution
|Positive = unlimited - negative = 1/order res revolution


|-
|-
|Trigger block phase shift
|Trigger block phase shift
|'''±720°''' - independent for each channel
|±720° - independent for each channel


|-
|-
|align = "center" rowspan = "2"|'''Overall'''
! rowspan="2" | Overall
|Detector
|Detector
|'''Quadratic sum of order spectra lines''' taking in account '''weighting window equivalent noise bandwidth'''
|Quadratic sum of order spectra lines taking in account weighting window equivalent noise bandwidth


|-
|-
|Lower and upper orders
|Lower and upper orders
|Selectable into order span - from '''0,03125 '''to''' 800''' with '''''ForceDSP'''''
|Selectable into order span - from 0.03125 to 800 with ''ForceDSP''


|}
|}


----


'''SOA Results'''
'''SOA Results'''
Line 1,247: Line 1,232:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "5"|'''Revolution averaging'''
! rowspan="5" | Revolution averaging
|Time domain
|Time domain
|'''Filtered '''signal''' '''
|Filtered signal  


|-
|-
|Angle domain (Revolution)
|Angle domain (Revolution)
|'''Triggered '''block '''- Averaged Triggered '''block''' - Weighted '''block
|Triggered block - Averaged Triggered block - Weighted block


|-
|-
|Order spectra
|Order spectra
|'''Complex spectra'''
|Complex spectra


|-
|-
|Tracked orders
|Tracked orders
|'''Digital '''(magnitude & phase) view-meter '''- Analog '''view -meter
|Digital (magnitude & phase) view-meter - Analog view -meter


|-
|-
|Overall
|Overall
|'''Overall RMS value''' in specified order band (digital or analog view-meter)
|Overall RMS value in specified order band (digital or analog view-meter)


|-
|-
|align = "center" rowspan = "5"|'''Spectral (order) averaging'''
! rowspan="5" | Spectral (order) averaging
|Time domain
|Time domain
|'''Filtered '''signal
|Filtered signal


|-
|-
|Angle domain (Revolution)
|Angle domain (Revolution)
|'''Triggered block - Weighted Block '''
|Triggered block - Weighted Block  


|-
|-
|Order spectra
|Order spectra
|Instantaneous''' complex order spectra '''- averaged '''power order spectra'''
|Instantaneous complex order spectra - averaged power order spectra


|-
|-
|Tracked orders
|Tracked orders
|'''Digital (module and phase) '''and/or''' Analog '''view-meter.
|Digital (module and phase) and/or Analog view-meter.


|-
|-
|Overall
|Overall
|'''Overall RMS value''' in specified order band (digital or analog view-meter)
|Overall RMS value in specified order band (digital or analog view-meter)


|}
|}
Line 1,298: Line 1,283:
'''SOADiag Settings'''
'''SOADiag Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center"|'''Cross-functions'''
! Cross-functions
|Reference selection
|Reference selection
|'''Full''' or '''partial matrix''' available (all channels can be references) - order domain averaging only.
|Full or partial matrix available (all channels can be references) - order domain averaging only.


|-
|-
|align = "center" rowspan = "2"|'''Angular Correlation'''
! rowspan="2" | Angular Correlation
|Activation
|Activation
|Enable/disable on all channels''' - '''All other SOA results are affected simultaneously
|Enable/disable on all channels - All other SOA results are affected simultaneously


|-
|-
|Weighting window
|Weighting window
|'''Uniform - Left '''zero padding''' - Centred '''zero padding
|Uniform - Left zero padding - Centred zero padding


|}
|}
Line 1,318: Line 1,303:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Rev. synchronous Scalars'''
! rowspan="3" | Rev. synchronous Scalars
|Min/Max
|Min/Max
|Minimum and maximum amplitude per trigger block (n rev.)
|Minimum and maximum amplitude per trigger block (n rev.)
Line 1,329: Line 1,314:
|-
|-
|Peak detectors
|Peak detectors
|'''Peak level - Peak to Peak level - Crest Factor - '''per trigger block (n rev.)
|Peak level - Peak to Peak level - Crest Factor - per trigger block (n rev.)


|-
|-
|align = "center" rowspan = "4"|'''Others'''
! rowspan="4" | Others
|Cepstrum
|Cepstrum
|'''Cepstrum''' of the order spectra
|Cepstrum of the order spectra


|-
|-
|Cross-functions
|Cross-functions
|'''Instantaneous order cross-spectra''' - averaged order '''cross-spectra''' - '''ORF™ H1''' / '''H2''' -'''Coherence – angular '''or '''order '''domain averaging'''.'''
|Instantaneous order cross-spectra - averaged order cross-spectra - ORF™ H1 / H2 -Coherence – angular or order domain averaging.


|-
|-
|Angular Correlation
|Angular Correlation
|'''Auto-correlation '''block''' - Cross correlation '''block - angular averaged
|Auto-correlation block - Cross correlation block - angular averaged


|-
|-
|Re-sampled Shaft view
|Re-sampled Shaft view
|Time domain signal on first rotation – '''angular representation''' - '''instantaneous''' or '''revolutions''' averaged
|Time domain signal on first rotation – angular representation - instantaneous or revolutions averaged


|}
|}


----


===1/n octave constant percentage band filter analysis - OCT===
===1/n octave constant percentage band filter analysis (OCT)===
The OCT plug-in analyzer features multiple acoustic results depending on applied setting. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters (CPB) and detectors.
The OCT plug-in analyzer features multiple acoustic results depending on applied setting. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters (CPB) and detectors.


'''OCT Settings'''
'''OCT Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Bandwidths'''
! rowspan="3" | Bandwidths
|Sampling oscillator #1 <br>(102.4 kS/s to 2,048 S/s)  
|Sampling oscillator #1 <br>(102.4 kS/s to 2.048 S/s)  
|Center of '''highest 1/3'''<sup>'''rd</sup> '''band = '''40 kHz'''<br>Center of '''lowest''' '''1/3'''<sup>'''rd</sup> '''band = '''100 mHz'''
|Center of highest 1/3<sup>rd</sup> band = 40 kHz<br>Center of lowest 1/3<sup>rd</sup> band = 100 mHz


|-
|-
|Sampling oscillator #2<br>(65.536 kS/s to 3,277 S/s)
|Sampling oscillator #2<br>(65.536 kS/s to 3,277 S/s)
|Center of '''highest''' '''1/3'''<sup>'''rd</sup> '''band = '''25 kHz'''<br>Center of '''lowest''' '''1/3'''<sup>'''rd</sup> '''band = '''100 mHz'''
|Center of highest 1/3<sup>rd</sup> band = 25 kHz<br>Center of lowest 1/3<sup>rd</sup> band = 100 mHz


|-
|-
|Frequency span
|Frequency span
|Highest Band / lowest band''' <nowiki><</nowiki> 2000 '''(ex. 10Hz  - 20kHz)
|Highest Band / lowest band <nowiki><</nowiki> 2000 (ex. 10 Hz - 20 kHz)


|-
|-
|align = "center"|'''Resolution'''
! Resolution
|1/n Octave
|1/n Octave
|'''1, 1/3<sup> ''''''rd'''</sup>''', 1/12'''<sup>'''th'''</sup>''', 1/24'''<sup>'''th '''</sup>
|1, 1/3<sup> rd</sup>, 1/12<sup>th</sup>, 1/24<sup>th </sup>


|-
|-
|align = "center" rowspan = "3"|'''Averaging'''
! rowspan="3" | Averaging
|Basic
|Basic
|Linear & repeated linear (20 ms to 60,000 s) exponential (20 ms to 60s)
|Linear & repeated linear (20 ms to 60,000 s) exponential (20 ms to 60s)
Line 1,382: Line 1,366:
|-
|-
|Acoustics  
|Acoustics  
|'''Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse '''
|Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse  


|-
|-
|Constant Bandwidth * Time
|Constant Bandwidth * Time
|B*T =''' 0.2dB, 0.5 dB, 1 dB, 2 dB'''
|B*T = 0.2 dB, 0.5 dB, 1 dB, 2 dB


|-
|-
|align = "center" rowspan = "2"|'''Standards'''
! rowspan="2" | Standards
|Detectors
|Detectors
|Complies with '''IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 compatible: class 1'''
|Complies with IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 compatible: class 1


|-
|-
|CPB filters  
|CPB filters  
|Complies with '''IEC ANSI/ASA S1.11-2014 / Part 1 / IEC 61260:1-2014 and IEC 1260 compatible: class 1'''
|Complies with IEC ANSI/ASA S1.11-2014 / Part 1 / IEC 61260:1-2014 and IEC 1260 compatible: class 1


|}
|}
Line 1,401: Line 1,385:




{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Channels'''
! rowspan="2" | Channels
|OR36 & OR38 analyzers
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line -''' 32 to 256 in '''post-analysis'''
|32 channels per plug-in for on line - 32 to 256 in post-analysis


|-
|-
|OR34 & OR35 analyzers
|OR34 & OR35 analyzers
|'''8 channels '''for '''on line''' and '''post-analysis'''
|8 channels for on line and post-analysis


|-
|-
|align = "center" rowspan = "3"|'''Triggering'''
! rowspan="3" | Triggering
|Start averaging
|Start averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|End averaging
|End averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Repeat averaging
|Repeat averaging
|On '''new start''' or '''end of averaging'''
|On new start or end of averaging


|-
|-
|align = "center" rowspan = "2"|'''Overall'''
! rowspan="2" | Overall
|Detectors
|Detectors
|'''2 parallel overall detectors''' linear and weighted (time domain)
|2 parallel overall detectors linear and weighted (time domain)


|-
|-
|Weighting
|Weighting
|'''A''', '''C''', or '''Z''' (none) applicable in '''10 kHz - 40 kHz''' bandwidth
|A, C, or Z (none) applicable in 10 kHz - 40 kHz bandwidth


|}
|}
Line 1,438: Line 1,422:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Spectra'''
! rowspan="2" | Spectra
|1/n Octave spectra
|1/n Octave spectra
|'''Instantaneous '''spectra''', averaged '''spectra
|Instantaneous spectra, averaged spectra


|-
|-
|Holden spectra
|Holden spectra
|'''Minimum '''&''' maximum '''spectra
|Minimum & maximum spectra


|-
|-
|align = "center" rowspan = "2"|'''Overall'''
! rowspan="2" | Overall
|Linear
|Linear
|'''Digital '''&''' analog '''view meter of time domain overall detector, exact bandwidth is informed
|Digital & analog view meter of time domain overall detector, exact bandwidth is informed


|-
|-
|Weighted
|Weighted
|'''Digital '''&''' analog '''view meter''' '''of''' A '''or''' C '''weighted''' '''overall levels computed in time domain, exact bandwidth is informed
|Digital & analog view meter of A or C weighted overall levels computed in time domain, exact bandwidth is informed


|}
|}


----


===Overall acoustics sound level meter - OVA===
===Overall acoustics sound level meter (OVA)===
The OVA plug-in analyzer features compatible class 1 sound pressure level measurement according to IEC 61672:2013 standard. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters and detectors.
The OVA plug-in analyzer features compatible class 1 sound pressure level measurement according to IEC 61672:2013 standard. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters and detectors.


'''OVA Settings'''
'''OVA Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Type'''
! rowspan="3" | Type
|Standard
|Standard
|'''IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 compatible: class 1'''
|IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 compatible: class 1


|-
|-
|Class
|Class
|'''1'''
|1


|-
|-
|Bandwidths
|Bandwidths
|'''10 Hz '''to '''40 kHz''' - Adjustable
|10 Hz to 40 kHz - Adjustable


|-
|-
|align = "center" rowspan = "2"|'''Channels'''
! rowspan="2" | Channels
|OR36 & OR38 analyzers
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line''' - 32 to 256 in '''post-analysis'''
|32 channels per plug-in for on line - 32 to 256 in post-analysis


|-
|-
|OR34 & OR35 analyzers
|OR34 & OR35 analyzers
|'''8 channels '''for '''on line''' and '''post-analysis'''
|8 channels for on line and post-analysis


|-
|-
|align = "center" rowspan = "2"|'''Averaging'''
! rowspan="2" | Averaging
|Linear
|Linear
|'''User selectable '''duration '''- repeat '''mode
|User selectable duration - repeat mode


|-
|-
|Short leq
|Short leq
|'''1/8 s - 1 s - '''User selectable duration
|1/8 s - 1 s - User selectable duration


|-
|-
|align = "center" rowspan = "3"|'''Triggering'''
! rowspan="3" | Triggering
|Start averaging
|Start averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|End averaging
|End averaging
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Repeat averaging
|Repeat averaging
|On '''new start''' or '''end of averaging'''
|On new start or end of averaging


|-
|-
|align = "center" rowspan = "2"|'''Detectors'''
! rowspan="2" | Detectors
|Peak
|Peak
|1 '''peak detector/Ch'''. - '''A''', '''C''' or '''Z''' (none) weighting -'''Independent '''for each channel
|1 peak detector/Ch. - A, C or Z (none) weighting -Independent for each channel


|-
|-
|Weighting
|Weighting
|'''3 overall detectors/Ch. - A''', '''C''' or '''Z''' (none) weighting - '''Fast''', '''Slow''', '''Impulse''' and '''linear''' time weighting - '''Independent''' for each channel
|3 overall detectors/Ch. - A, C or Z (none) weighting - Fast, Slow, Impulse and linear time weighting - Independent for each channel


|}
|}
Line 1,524: Line 1,507:
The following results are available for real time or post-analysis display, report and saving. The following levels can be tracked in profiles up to 100 000 pts. each.
The following results are available for real time or post-analysis display, report and saving. The following levels can be tracked in profiles up to 100 000 pts. each.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''SPL'''
! rowspan="3" | SPL
|Time weighted
|Time weighted
|Instant - max hold - min Hold
|Instant - max hold - min Hold
Line 1,539: Line 1,522:
|}
|}


----


==Options==
==Options==
===Instantaneous angular Velocity Converter (IVC)===
===Instantaneous Angular Velocity Converter (IVC)===
The IVC option converts frequency to voltage from the External synch inputs. It covers torsional and acyclism measurements. The converted signals are made available as standard inputs (or player tracks). This option operates real-time computation and operations are possible on-line or in post processing.
The IVC option converts frequency to voltage from the External synch inputs. It covers torsional and acyclism measurements. The converted signals are made available as standard inputs (or player tracks). This option operates real-time computation and operations are possible on-line or in post processing.


'''IVC Settings & specs'''
'''IVC Settings & specs'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "4"|'''Rate'''
! rowspan="4" | Rate
|Pulse/rev
|Pulse/rev
|'''0.5 '''to''' 4096 - '''Up to''' 1 E6 '''with pre-divider
|0.5 to 4096 - Up to 1 E6 with pre-divider


|-
|-
|Pulse/rev Frequency
|Pulse/rev Frequency
|'''<nowiki>></nowiki> 40 kpulse/sec''''' max''
|<nowiki>></nowiki> 40 kpulse/sec max


|-
|-
|Pre-divider
|Pre-divider
|'''1 '''to''' 255 - '''hardware decimation (pulse are not measured)
|1 to 255 - hardware decimation (pulse are not measured)


|-
|-
|Pre-divider frequency
|Pre-divider frequency
|'''375 kpulse/sec'' '''''max.
|375 kpulse/sec max.


|-
|-
|align = "center" rowspan = "3"|'''Missing pulses'''
! rowspan="3" | Missing pulses
|Number
|Number
|'''0 to 5''' consecutive pulses '''integer or fractional'''
|0 to 5 consecutive pulses integer or fractional


|-
|-
|Hold off
|Hold off
|'''2 '''consecutive pulses
|2 consecutive pulses


|-
|-
|Generated signal
|Generated signal
|'''1'''<sup>'''st</sup> ''''''order interpolation'''
|1<sup>st</sup> order interpolation


|-
|-
|align = "center" rowspan = "2"|'''Tach. phase ref'''
! rowspan="2" | Tach. phase ref
|No missing pulse
|No missing pulse
|First random - continuous offset further
|First random - continuous offset further
Line 1,587: Line 1,569:


|-
|-
|align = "center" rowspan = "2"|'''Resolution'''
! rowspan="2" | Resolution
|Time
|Time
|'''160 ns''' : 1/(SF x 64)
|160 ns : 1/(SF x 64)


|-
|-
Line 1,596: Line 1,578:


|-
|-
|align = "center" rowspan = "2"|'''Pre-processing'''
! rowspan="2" | Pre-processing
|Smoothing
|Smoothing
|Sliding average - '''1''' to '''32 '''samples
|Sliding average - 1 to 32 samples


|-
|-
|Filtering
|Filtering
|'''Any NVGate''' filter (See Filters §) incl. dt and 1/dt  
|Any NVGate filter (See Filters §) incl. dt and 1/dt  


|-
|-
|align = "center" rowspan = "2"|'''Max Speed'''
! rowspan="2" | Max Speed
|Without pre-divider
|Without pre-divider
|Max RPM = 2.4 e6/''Pulse per rev'' - 12&nbsp;000 RPM @ 200 pulse/rev
|Max RPM = 2.4 e6/''Pulse per rev'' - 12&nbsp;000 RPM @ 200 pulse/rev
Line 1,619: Line 1,601:
The following results are available for real time or post-analysis.
The following results are available for real time or post-analysis.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Signals'''
! rowspan="3" | Signals
|Number
|Number
|'''1 '''to '''6 '''(depends on the number of available ext. synch)
|1 to 6 (depends on the number of available ext. synch)


|-
|-
|Use
|Use
|Same as '''dynamic inputs''' or '''recorded''' dynamic inputs
|Same as dynamic inputs or recorded dynamic inputs


|-
|-
|Magnitude
|Magnitude
|'''Torsional''' acceleration, velocity & angle – separated from angular magnitudes (RPM)
|Torsional acceleration, velocity & angle – separated from angular magnitudes (RPM)


|-
|-
|align = "center"|'''View'''
! View
|Type
|Type
|Signal - filtered signal - '''256 samples''' frame
|Signal - filtered signal - 256 samples frame


|}
|}
Line 1,653: Line 1,635:
'''VIn Settings'''
'''VIn Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "4"|'''Output'''
! rowspan="4" | Output
|Number
|Number
|'''12''' Operators
|12 Operators


|-
|-
|Type
|Type
|'''New item''' in the active '''inputs/tracks list'''
|New item in the active inputs/tracks list


|-
|-
|Synchronization
|Synchronization
|'''0° phase shift''' with sources and analyzed signal
|0° phase shift with sources and analyzed signal


|-
|-
|Magnitude
|Magnitude
|'''Automatic''' or '''voltage''' if unresolved
|Automatic or voltage if unresolved


|-
|-
|align = "center" rowspan = "2"|'''Sources'''
! rowspan="2" | Sources
|Number
|Number
|'''1''' to '''32 channels'''
|1 to 32 channels


|-
|-
|Type
|Type
|Analyzer '''dynamic inputs''' or '''recorded''' dynamic inputs
|Analyzer dynamic inputs or recorded dynamic inputs


|-
|-
|align = "center" rowspan = "4"|'''Output adjustment'''
! rowspan="4" | Output adjustment
|Type
|Type
|'''''(Input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki>power''''' - independent on each output
|''(Input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki>power'' - independent on each output


|-
|-
|Offset
|Offset
|Any real value from '''-1 e9 to 1 e9''' - ex: -5.67
|Any real value from -1 e9 to 1 e9 - ex: -5.67


|-
|-
|Coefficient
|Coefficient
|Any real value from '''-1 e9 to 1 e9''' - ex: 12
|Any real value from -1 e9 to 1 e9 - ex: 12


|-
|-
|Power
|Power
|Any real value from '''-4 to 4''' - ex:  0.5
|Any real value from -4 to 4 - ex:  0.5


|-
|-
|align = "center" rowspan = "5"|'''Sources adjustment'''
! rowspan="5" | Sources adjustment
|Type
|Type
|'''''Filtering <nowiki>[</nowiki> (input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki> power<nowiki>]</nowiki>''''' - independent for each source
|''Filtering <nowiki>[</nowiki> (input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki> power<nowiki>]</nowiki>'' - independent for each source


|-
|-
|Offset
|Offset
|Any real value from '''-1 e9 to 1 e9''' - ex: -5.67 e-5
|Any real value from -1 e9 to 1 e9 - ex: -5.67 e-5


|-
|-
|Coefficient
|Coefficient
|Any real value from '''-1 e9 to 1 e9''' - ex: 1200
|Any real value from -1 e9 to 1 e9 - ex: 1200


|-
|-
|Power
|Power
|Any real value from '''-4 to 4''' - ex:  -2.4567
|Any real value from -4 to 4 - ex:  -2.4567


|-
|-
Line 1,718: Line 1,700:


|-
|-
|align = "center"|'''Operators'''
! Operators
|Type
|Type
|'''Product''' - '''Sum''' - independent on each operator
|Product - Sum - independent on each operator


|-
|-
|align = "center" rowspan = "3"|'''Twist'''
! rowspan="3" | Twist
|Type
|Type
|'''Phase comparison''' – output in '''plane angle'''
|Phase comparison – output in plane angle


|-
|-
|Correction (zeroing) offset
|Correction (zeroing) offset
|'''-360° '''to''' <nowiki>+</nowiki>360°'''
|-360° to <nowiki>+</nowiki>360°


|-
|-
|Output
|Output
|'''Plane angle''' or '''torque''' with conversion factor
|Plane angle or torque with conversion factor


|}
|}
Line 1,742: Line 1,724:
'''VDC Settings'''
'''VDC Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "4"|'''Output'''
! rowspan="4" | Output
|Number
|Number
|'''12''' Operators
|12 Operators


|-
|-
|Type
|Type
|'''New item''' in the DC inputs/track list
|New item in the DC inputs/track list


|-
|-
|Synchronization
|Synchronization
|'''0° phase shift''' with sources and analyzed signal
|0° phase shift with sources and analyzed signal


|-
|-
|Magnitude
|Magnitude
|'''User defined'''
|User defined


|-
|-
|align = "center" rowspan = "2"|'''Sources'''
! rowspan="2" | Sources
|Number
|Number
|'''1''' to '''32 channels'''
|1 to 32 channels


|-
|-
|Type
|Type
|Analyzer '''DC inputs''' or '''recorded''' DC inputs
|Analyzer DC inputs or recorded DC inputs


|-
|-
|align = "center" rowspan = "7"|'''Editor'''
! rowspan="7" | Editor
|Type
|Type
|Text
|Text
Line 1,775: Line 1,757:
|-
|-
|Variables
|Variables
|'''Real''' values  
|Real values  


|-
|-
|Operators
|Operators
|'''<nowiki>+</nowiki>''', '''-''', '''x''', '''/''', '''<nowiki>^</nowiki>''', '''=''' (affectation)
|<nowiki>+</nowiki>, -, x, /, <nowiki>^</nowiki>, = (affectation)


|-
|-
|Predefined  
|Predefined  
|'''Ch x''' (Channels signals), '''Pi''', '''e'''
|Ch x (Channels signals), Pi, e


|-
|-
|Trigonometric functions
|Trigonometric functions
|'''Sine''', '''Cosine''', '''Tangent''' - with '''Arc''' and '''Hyperbolic''' combination  
|Sine, Cosine, Tangent - with Arc and Hyperbolic combination  


|-
|-
|Logarithmic functions
|Logarithmic functions
|'''Log2''', '''Log10''', '''Ln''', '''Exp'''
|Log2, Log10, Ln, Exp


|-
|-
|Miscellaneous
|Miscellaneous
|'''Sqrt''', '''Abs, Min, Max, Sum, Avg'''
|Sqrt, Abs, Min, Max, Sum, Avg


|-
|-
|align = "center" rowspan = "2"|'''Output characterization'''
! rowspan="2" | Output characterization
|Limits
|Limits
|Min and max: '''<nowiki>+</nowiki>/- 1 E9 '''- In current unit
|Min and max: <nowiki>+</nowiki>/- 1 E9 - In current unit


|-
|-
|Information
|Information
|'''Label''' (text)
|Label (text)


|}
|}
Line 1,813: Line 1,795:
'''CTE Settings'''
'''CTE Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Sources'''
! rowspan="2" | Sources
|Number
|Number
|'''2'''
|2


|-
|-
|Type
|Type
|'''Any NVGate tachometer – '''excepted another CTE
|Any NVGate tachometer – excepted another CTE


|-
|-
|align = "center" rowspan = "2"|'''Output'''
! rowspan="2" | Output
|Type
|Type
|New item in the active tachometers list
|New item in the active tachometers list
Line 1,829: Line 1,811:
|-
|-
|Synchronization
|Synchronization
|'''0° phase shift''' with sources and analyzed signal
|0° phase shift with sources and analyzed signal


|-
|-
|align = "center" rowspan = "7"|'''Editor'''
! rowspan="7" | Editor
|Type
|Type
|Text
|Text
Line 1,838: Line 1,820:
|-
|-
|Variables
|Variables
|'''Real''' values  
|Real values  


|-
|-
|Operators
|Operators
|'''<nowiki>+</nowiki>''', '''-''', '''x''', '''/''', '''<nowiki>^</nowiki>''', '''=''' (affectation)
|<nowiki>+</nowiki>, -, x, /, <nowiki>^</nowiki>, = (affectation)


|-
|-
|Predefined  
|Predefined  
|Rpm1, Rpm2 (sources speed), '''Pi''', '''e'''
|Rpm1, Rpm2 (sources speed), Pi, e


|-
|-
|Trigonometric functions
|Trigonometric functions
|'''Sine''', '''Cosine''', '''Tangent''' - with '''Arc''' and '''Hyperbolic''' combination  
|Sine, Cosine, Tangent - with Arc and Hyperbolic combination  


|-
|-
|Logarithmic functions
|Logarithmic functions
|'''Log2''', '''Log10''', '''Ln''', '''Exp'''
|Log2, Log10, Ln, Exp


|-
|-
|Miscellaneous
|Miscellaneous
|'''Sqrt''', '''Abs, Min, Max, Sum, Avg, Sign, If, Rint'''
|Sqrt, Abs, Min, Max, Sum, Avg, Sign, If, Rint


|-
|-
|align = "center" rowspan = "4"|'''Output characterization'''
! rowspan="4" | Output characterization
|Rotation
|Rotation
|Clock wise or counter clock wise
|Clock wise or counter clock wise
Line 1,867: Line 1,849:
|-
|-
|Average
|Average
|'''Sliding''' - '''1''' to '''256 '''revolution
|Sliding - 1 to 256 revolution


|-
|-
|Speed limits
|Speed limits
|Min: '''0 RPM''' - max: '''1&nbsp;200&nbsp;000 RPM'''
|Min: 0 RPM - max: 1&nbsp;200&nbsp;000 RPM


|-
|-
Line 1,883: Line 1,865:
The following results are available for real time or post-analysis  
The following results are available for real time or post-analysis  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Signals'''
! rowspan="3" | Signals
|Number
|Number
|'''4'''
|4


|-
|-
|Use
|Use
|Same as on-line '''Tachometer '''or '''recorded''' tachometer  
|Same as on-line Tachometer or recorded tachometer  


|-
|-
Line 1,897: Line 1,879:


|-
|-
|align = "center"|'''View'''
! View
|Types
|Types
|'''View meter''' - '''1''' to '''100 s''' rolling '''profile'''
|View meter - 1 to 100 s rolling profile


|}
|}
Line 1,909: Line 1,891:
The monitor is an independent (dedicated processor) module that continuously processes FFT analysis on 4 channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.
The monitor is an independent (dedicated processor) module that continuously processes FFT analysis on 4 channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.


'''Monitor settings'''
'''Monitor Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "7"|'''Fixed setup'''
! rowspan="7" | Fixed setup
|Bandwidth
|Bandwidth
|'''ABW = Sampling freq / 2.56'''
|ABW = Sampling freq / 2.56  


|-
|-
|Resolution
|Resolution
|'''401 lines'''
|401 lines


|-
|-
|Average domain
|Average domain
|'''Spectral '''(power)
|Spectral (power)


|-
|-
|Overlap
|Overlap
|'''0%'''
|0%


|-
|-
|Average type
|Average type
|'''Exponential'''
|Exponential  


|-
|-
|Weighting windows
|Weighting windows
|'''Hanning'''
|Hanning


|-
|-
|Trigger
|Trigger
|'''Free run'''
|Free run


|-
|-
|align = "center" rowspan = "2"|'''Channels'''
! rowspan="2" | Channels
|Number
|Number
|'''4 '''
|4  


|-
|-
|Swap
|Swap
|Between '''any active dynamic input - hot swap '''capable
|Between any active dynamic input - hot swap capable


|-
|-
|align = "center"|'''Average'''
! Average
|Duration
|Duration
|'''Instantaneous '''(20 ms) to''' unlimited'''
|Instantaneous (20 ms) to unlimited


|-
|-
|align = "center" rowspan = "4"|'''Overall analysis'''
! rowspan="4" | Overall analysis
|Pass band filter
|Pass band filter
|'''Butterworth''' order '''2''' to '''10 -''' '''IIR''' type - '''Bypass''' function
|Butterworth order 2 to 10 - IIR type - Bypass function


|-
|-
|Upper & lower frequency
|Upper & lower frequency
|'''User defined - '''can be graphically modified
|User defined - can be graphically modified


|-
|-
|Computed indicators
|Computed indicators
|'''DC - Max - Min - RMS - Skew - Kurtosis - '''Into defined pass band - '''overall''' if bypass is on
|DC - Max - Min - RMS - Skew - Kurtosis - Into defined pass band - overall if bypass is on


|-
|-
|Average
|Average
|'''User defined '''duration''' - independent '''from spectral averaging
|User defined duration - independent from spectral averaging


|}
|}


'''Monitor results and connections'''
'''Monitor Results and Connections'''


The following results are available for real time or post-analysis display, connection to other modules.
The following results are available for real time or post-analysis display, connection to other modules.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center"|'''Time domain'''
! Time domain
|Play back
|Play back
|On any output generator - hot swappable
|On any output generator - hot swappable


|-
|-
|align = "center" rowspan = "2"|'''FFT analyzer'''
! rowspan="2" | FFT analyzer
|Trigger bloc
|Trigger bloc
|'''1024 samples''' - time domain '''analyzed bloc'''
|1024 samples - time domain analyzed bloc


|-
|-
|Spectra
|Spectra
|Instantaneous '''complex spectra''' - averaged '''power spectra'''
|Instantaneous complex spectra - averaged power spectra


|-
|-
|align = "center" rowspan = "2"|'''Overall detectors'''
! rowspan="2" | Overall detectors
|Display (digital or analog view-meter)
|Display (digital or analog view-meter)
|DC - Max - Min - RMS - Skew - Kurtosis
|DC - Max - Min - RMS - Skew - Kurtosis
Line 1,998: Line 1,980:
|-
|-
|Connection
|Connection
|To '''level''' and '''delta level''' event detector - to '''waterfall profiles'''
|To level and delta level event detector - to waterfall profiles


|}
|}
Line 2,005: Line 1,987:
The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in analyzers and sources in one result.
The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in analyzers and sources in one result.


'''Waterfall settings'''
'''Waterfall Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "2"|'''Mode'''
! rowspan="2" | Mode
|Continuous
|Continuous
|'''Circular buffer''' of results
|Circular buffer of results


|-
|-
Line 2,017: Line 1,999:


|-
|-
|align = "center" rowspan = "3"|'''Triggering'''
! rowspan="3" | Triggering
|Start acquisition
|Start acquisition
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|Stop acquisition
|Stop acquisition
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|Any NVGate<sup>® </sup>event plus manual and free run


|-
|-
|New slice (point or spectra)
|New slice (point or spectra)
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' - '''manual - free run - '''availability of '''connected results''' - '''periodic'''
|Any NVGate<sup>® </sup>event - manual - free run - availability of connected results - periodic


|}
|}


'''Waterfall settings (continued)'''
'''Waterfall Settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Size'''
! rowspan="3" | Size
|Number of result per acquisition
|Number of result per acquisition
|Up to '''95 results <nowiki>+</nowiki> Time'''
|Up to 95 results <nowiki>+</nowiki> Time


|-
|-
|Depth
|Depth
|'''2''' to '''100,000''' slices or points - depend on PC available memory and requested result - '''automatically adjusted '''before acquisition
|2 to 100,000 slices or points - depend on PC available memory and requested result - automatically adjusted before acquisition


|-
|-
|Depth for stand-alone acquisition
|Depth for stand-alone acquisition
|''' '''Up to''' 3 MSamples '''per computation DSP''' '''
| Up to 3 MSamples per computation DSP  


|-
|-
|align = "center" rowspan = "4"|'''Channels'''
! rowspan="4" | Channels
|'''Type scalar''' (the following results are connectable to waterfall channels)
|Type scalar (the following results are connectable to waterfall channels)
|'''Overall levels''' (lin and weighted) from OvA and 1/n OCT - '''Monitor indicators''' (DC, Max, Min, RMS, Kurtosis) - '''Orders''' (from CBT and SOA) - '''Complex orders''' (magnitude & phase) - '''Overall levels '''in selected BW from FFT & SOA (order or frequency) - '''TDA''' scalar: '''DC, Min/max, RMS, Kurtosis, peak, peak-peak, crest factor'''
|Overall levels (lin and weighted) from OvA and 1/n OCT - Monitor indicators (DC, Max, Min, RMS, Kurtosis) - Orders (from CBT and SOA) - Complex orders (magnitude & phase) - Overall levels in selected BW from FFT & SOA (order or frequency) - TDA scalar: DC, Min/max, RMS, Kurtosis, peak, peak-peak, crest factor


|-
|-
|'''Type 2D - FFT''' (the following results are connectable to waterfall channels)
|Type 2D - FFT (the following results are connectable to waterfall channels)
|'''Triggered '''block''' - Averaged Triggered '''block''' - Weighted '''Block''' - Complex '''spectra''' - Power '''spectra''' -Cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence - Zoomed''' complex spectra - '''zoomed power '''spectra
|Triggered block - Averaged Triggered block - Weighted Block - Complex spectra - Power spectra -Cross-spectra - FRF H1 - FRF H2 - Coherence - Zoomed complex spectra - zoomed power spectra


|-
|-
|'''Type 2D - OCT''' (the following results are connectable to waterfall channels)
|Type 2D - OCT (the following results are connectable to waterfall channels)
|'''Instantaneous''' spectra - '''averaged '''spectra - '''max  & min''' '''hold '''spectra
|Instantaneous spectra - averaged spectra - max  & min hold spectra


|-
|-
|'''Type 2D - SOA''' (the following results are connectable to waterfall channels)
|Type 2D - SOA (the following results are connectable to waterfall channels)
|'''Triggered '''block''' - Averaged Triggered '''block''' - Weighted '''block''' - Complex order '''spectra''' - Power order '''spectra
|Triggered block - Averaged Triggered block - Weighted block - Complex order spectra - Power order spectra


|}
|}


'''Waterfall results'''
'''Waterfall Results'''


The following results are available for real time or post-analysis display, report and saving. All stacked results can be saved.
The following results are available for real time or post-analysis display, report and saving. All stacked results can be saved.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''3D'''
! rowspan="3" | 3D
|Display
|Display
|'''1 pane''' (3D) - '''2 panes''' (3D <nowiki>+</nowiki> YZ view or 3D <nowiki>+</nowiki> XY view) , '''3 panes''' (3D <nowiki>+</nowiki> YZ <nowiki>+</nowiki> XY <nowiki>+</nowiki> Extraction view) windows - automatic or user selectable '''pane arrangement - Real or imaginary '''part and '''module or phase''' for complex results
|1 pane (3D) - 2 panes (3D <nowiki>+</nowiki> YZ view or 3D <nowiki>+</nowiki> XY view) , 3 panes (3D <nowiki>+</nowiki> YZ <nowiki>+</nowiki> XY <nowiki>+</nowiki> Extraction view) windows - automatic or user selectable pane arrangement - Real or imaginary part and module or phase for complex results


|-
|-
|Z axis (X axis for extraction and YZ view)
|Z axis (X axis for extraction and YZ view)
|'''Any reference''' - '''time''' - independent for any window - swap reference at any time  
|Any reference - time - independent for any window - swap reference at any time  


|-
|-
|Saving selection
|Saving selection
|'''Entire 3D data '''and / or any active''' section '''(YZ, XY or Extraction)
|Entire 3D data and / or any active section (YZ, XY or Extraction)


|-
|-
|align = "center" rowspan = "2"|'''Profiles'''
! rowspan="2" | Profiles
|Display
|Display
|Profile of any scalar - '''Real or imaginary '''part and '''module or phase''' for complex results
|Profile of any scalar - Real or imaginary part and module or phase for complex results


|-
|-
|X axis
|X axis
|'''Any reference''' - '''Time''' '''Slice''' - independent for any window - swap reference at any time
|Any reference - Time – Slice - independent for any window - swap reference at any time


|}
|}
Line 2,097: Line 2,079:
The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|rowspan = "4"|Sections
! rowspan="4" | Sections
|YZ Sections (Profiles vs. Ref)
|YZ Sections (Profiles vs. Ref)
|Any order/frequency - '''Power''' - '''Peak''' - on selected bandwidth
|Any order/frequency - Power - Peak - on selected bandwidth


|-
|-
|XY Sections (Spectra)
|XY Sections (Spectra)
|Any position in the current ref. - '''Min''' - '''Max''' - '''Average''' - on selected range or all slices  
|Any position in the current ref. - Min - Max - Average - on selected range or all slices  


|-
|-
|Order/Frequency extraction (profiles vs. Ref.)
|Order/Frequency extraction (profiles vs. Ref.)
|Any order/frequency extraction - '''user selectable tach'''. - '''Power''' - '''Peak''' - on selectable bandwidth - '''Max order '''
|Any order/frequency extraction - user selectable tach. - Power - Peak - on selectable bandwidth - Max order  


|-
|-
|Number of sections
|Number of sections
|'''Unlimited''' - available on result or real-time waterfalls
|Unlimited - available on result or real-time waterfalls


|-
|-
|rowspan = "3"|Linked cursors
! rowspan="3" | Linked cursors
|General
|General
|Single or dual cursor in each view - linked with other graphs
|Single or dual cursor in each view - linked with other graphs
Line 2,137: Line 2,119:
Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the Event module.
Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the Event module.


'''Event settings'''
'''Event Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "8"|'''Edge detector'''
! rowspan="8" | Edge detector
|Source
|Source
|'''Any dynamic input''' - Any '''recorded '''dynamic input
|Any dynamic input - Any recorded dynamic input


|-
|-
|Label
|Label
|String for '''event identification''' (Ex. "Impact" for a hammer impact detection)
|String for event identification (Ex. "Impact" for a hammer impact detection)


|-
|-
|Threshold
|Threshold
|Between '''min and max range''' - use '''source unit''' (ex. g for an accelerometer)
|Between min and max range - use source unit (ex. g for an accelerometer)


|-
|-
|Pre-filtering
|Pre-filtering
|'''A '''law - '''C''' law - any '''NVGate filter'''
|A law - C law - any NVGate filter


|-
|-
|Slope
|Slope
|'''Rise '''-''' fall '''
|Rise - fall  


|-
|-
|Hold off
|Hold off
|'''0 '''to '''36000''' s
|0 to 36000 s


|-
|-
|Hysteresis
|Hysteresis
|Into source''' input range''' - use '''source unit''' (ex. g for an accelerometer)
|Into source input range - use source unit (ex. g for an accelerometer)


|-
|-
|Number
|Number
|'''2 '''edge detectors
|2 edge detectors


|-
|-
|align = "center" rowspan = "7"|'''RPM speed detector'''
! rowspan="7" | RPM speed detector
|Source
|Source
|'''Any tachometer - '''computed from '''dynamic input''' or '''ext.sync '''-''' '''any r'''ecorded pulses '''(through tachometer)
|Any tachometer - computed from dynamic input or ext.sync - any recorded pulses (through tachometer)


|-
|-
|Label
|Label
|String for '''event identification''' (Ex. "start speed" for a run up initial triggering)
|String for event identification (Ex. "start speed" for a run up initial triggering)


|-
|-
|Threshold
|Threshold
|Detection into''' '''source''' '''tachometer''' '''RPM range  
|Detection into source tachometer RPM range  


|-
|-
|Slope
|Slope
|'''Run up '''-''' Run down '''
|Run up - Run down  


|-
|-
|Hysteresis  
|Hysteresis  
|'''0''' to '''source '''max RPM
|0 to source max RPM


|-
|-
|Interpolation
|Interpolation
|'''On''' interpolate event occurrence '''into revolution''' - '''Off''' event '''occur at revolution ends''' (new pulse)
|On interpolate event occurrence into revolution - Off event occur at revolution ends (new pulse)


|-
|-
|Number
|Number
|'''2 to 6 '''RPM speed detector
|2 to 6 RPM speed detector


|-
|-
|align = "center" rowspan = "8"|'''Delta RPM speed detector'''
! rowspan="8" | Delta RPM speed detector
|Source
|Source
|'''Any tachometer - '''computed from '''dynamic input''' or '''ext.sync '''-''' '''any r'''ecorded pulses '''(through tachometer)
|Any tachometer - computed from dynamic input or ext.sync - any recorded pulses (through tachometer)


|-
|-
|Label
|Label
|String for '''event identification''' (Ex. "Each 100 RPM" for a run up triggering)
|String for event identification (Ex. "Each 100 RPM" for a run up triggering)


|-
|-
|Lower velocity
|Lower velocity
|'''Minimum''' RPM speed - Events occurs only for higher source speed''' '''
|Minimum RPM speed - Events occurs only for higher source speed  


|-
|-
|Higher velocity
|Higher velocity
|'''Maximum''' RPM speed - Events occurs only for lower source speed
|Maximum RPM speed - Events occurs only for lower source speed


|-
|-
|Delta velocity
|Delta velocity
|Define '''velocity step''' - '''event occur each time''' source speed increase or decrease by delta velocity
|Define velocity step - event occur each time source speed increase or decrease by delta velocity


|-
|-
|Slope
|Slope
|'''Run up''' - '''run down''' - '''first,''' first slope is automatically selected - '''any,''' event occur on any slope
|Run up - run down - first, first slope is automatically selected - any, event occur on any slope


|-
|-
|Interpolation
|Interpolation
|'''On,''' interpolate event occurrence '''into revolution''' - '''Off,''' event '''occur at revolution end''' (new pulse)
|On, interpolate event occurrence into revolution - Off, event occur at revolution end (new pulse)


|-
|-
|Number
|Number
|'''2 to 6 delta RPM''' speed detector
|2 to 6 delta RPM speed detector


|-
|-
|align = "center" rowspan = "5"|'''Level detector'''
! rowspan="5" | Level detector
|Source
|Source
|Any '''parametric (DC)''' input - '''Monitor scalar''' - '''Filtered''' monitor scalar (Band Pass)
|Any parametric (DC) input - Monitor scalar - Filtered monitor scalar (Band Pass)


|-
|-
|Label
|Label
|String for '''event identification''' (Ex. "temp A" for a recording trigger)
|String for event identification (Ex. "temp A" for a recording trigger)


|-
|-
Line 2,249: Line 2,231:
|-
|-
|Status
|Status
|'''Above''' /''' below''' - Detection level - available in '''lin''' or '''dB '''
|Above / below - Detection level - available in lin or dB


|-
|-
|Number
|Number
|'''1''' to '''4 '''level detectors
|1 to 4 level detectors


|}
|}


'''Event settings (continued)'''
'''Event Settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''Periodic event'''
! rowspan="3" | Periodic event
|Label
|Label
|String for '''event identification''' (Ex. "Hourly" for 1 hour time interval)
|String for event identification (Ex. "Hourly" for 1 hour time interval)


|-
|-
|Time interval
|Time interval
|'''2 ms''' to '''11 days''' - '''Synchronized''' with analyzer.
|2 ms to 11 days - Synchronized with analyzer.


|-
|-
|Number
|Number
|'''2 '''periodic events
|2 periodic events


|-
|-
|align = "center" rowspan = "4"|'''Combination'''
! rowspan="4" | Combination
|Sources
|Sources
|'''Any event - 2 '''different sources''' '''defined as''' A '''and''' B'''
|Any event - 2 different sources defined as A and B


|-
|-
|Label
|Label
|String for '''event identification''' (Ex. "Time/RPM variation" for a mix between periodic and delta RPM)
|String for event identification (Ex. "Time/RPM variation" for a mix between periodic and delta RPM)


|-
|-
|Combination  
|Combination  
|A '''OR''' B - A '''AFTER''' B - A '''AND''' B (occur in the same bloc)
|A OR B - A AFTER B - A AND B (occur in the same bloc)


|-
|-
|Number
|Number
|'''2 '''event combination''' '''
|2 event combination  


|-
|-
|align = "center" rowspan = "5"|'''Plug-in synchronization'''
! rowspan="5" | Plug-in synchronization
|FFTx result available  
|FFTx result available  
|Occurs at each '''new spectra availability''' - Occurs at '''end of linear averaging''' (repeat and linear) - One event per active FFT plug in (FFT1 to FFT4)
|Occurs at each new spectra availability - Occurs at end of linear averaging (repeat and linear) - One event per active FFT plug in (FFT1 to FFT4)


|-
|-
|OCT result available  
|OCT result available  
|Occurs at each new 1/n spectra availability - Occur at '''end of linear averaging''' (leq, short leq, linear repeat)  
|Occurs at each new 1/n spectra availability - Occur at end of linear averaging (leq, short leq, linear repeat)  


|-
|-
|SOAx result available  
|SOAx result available  
|Occurs at each '''new spectra availability''' - Occurs at '''end of linear averaging''' (repeat and linear) - One event per active SOA plug in (up to 2)
|Occurs at each new spectra availability - Occurs at end of linear averaging (repeat and linear) - One event per active SOA plug in (up to 2)


|-
|-
|TDA result available  
|TDA result available  
|Occurs at each new level set ( RMS, DC, etc) availability - Occur at '''end of linear averaging''' (repeat and linear)  
|Occurs at each new level set ( RMS, DC, etc) availability - Occur at end of linear averaging (repeat and linear)  


|-
|-
|OVA result available  
|OVA result available  
|Occurs at each new set of overall level (short LEQ) availability - Occur at '''end of short  averaging''' (repeat and linear)
|Occurs at each new set of overall level (short LEQ) availability - Occur at end of short  averaging (repeat and linear)


|-
|-
|align = "center" rowspan = "2"|'''Generators Synchronization'''
! rowspan="2" | Generators Synchronization
|Event synchronized with '''blocks of''':
|Event synchronized with blocks of:
|'''Multi-sine''' - '''random noise''' - '''chirps'''
|Multi-sine - random noise - chirps


|-
|-
|'''Swept sine''' generator event:
|Swept sine generator event:
|'''Swept sine stabilized''' (output amplitude established) - '''step sine stabilized''' (occur x sec after step frequency is reached) - '''pure tone stabilized''' (output amplitude established)
|Swept sine stabilized (output amplitude established) - step sine stabilized (occur x sec after step frequency is reached) - pure tone stabilized (output amplitude established)


|}
|}
Line 2,328: Line 2,310:
The following output signals are available as standard and can be independently applied to the front-end outputs.
The following output signals are available as standard and can be independently applied to the front-end outputs.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "10"|'''Sine'''
! rowspan="10" | Sine
|Type
|Type
|Pure tone - Swept sine - Step sine
|Pure tone - Swept sine - Step sine
Line 2,335: Line 2,317:
|-
|-
|Frequency
|Frequency
|'''100 µHz''' to '''40 kHz''' - smooth variation between step - resolution '''10 µHz'''
|100 µHz to 40 kHz - smooth variation between step - resolution 10 µHz  


|-
|-
Line 2,347: Line 2,329:
|-
|-
|Amplitude control
|Amplitude control
|'''Settling time''' 100 µs to 10 s - '''Stabilization time''' 0 s to 1000 s
|Settling time 100 µs to 10 s - Stabilization time 0 s to 1000 s


|-
|-
|Sweep speed
|Sweep speed
|Linear: '''300 mHz/s''' to '''20 kH/s''' - Log: '''5 mOct/s''' to '''330 Oct/s'''
|Linear: 300 mHz/s to 20 kHz/s - Log: 5 mOct/s to 330 Oct/s


|-
|-
|Step control
|Step control
|'''Synchronized with analysis end''' - free run  
|Synchronized with analysis end - free run  


|-
|-
|Gain control
|Gain control
|'''-15 dB''' to '''<nowiki>+</nowiki>60 dB''' - '''Independent''' for each output <br>- Amplitude '''variation controlled''' by settling time (1 ms to 1000 s)
|-15 dB to <nowiki>+</nowiki>60 dB - Independent for each output <br>- Amplitude variation controlled by settling time (1 ms to 1000 s)


|-
|-
|Phase control
|Phase control
|Offset ±360°  - '''Independent''' for each output - phase '''variation controlled''' by phase speed (1.5°/s to 360°/s)
|Offset ±360°  - Independent for each output - phase variation controlled by phase speed (1.5°/s to 360°/s)


|-
|-
|Number
|Number
|'''2 to 6''' sine generated simultaneously - with '''independent phases and amplitudes''' - '''synchronized''' frequencies
|2 to 6 sine generated simultaneously - with independent phases and amplitudes - synchronized frequencies


|-
|-
|align = "center" rowspan = "6"|'''Multi-sine'''
! rowspan="6" | Multi-sine
|Frequency span
|Frequency span
|From '''125 mHz''' to '''40 kHz'''
|From 125 mHz to 40 kHz  


|-
|-
|Amplitude
|Amplitude
|'''0 '''to''' 2.5 Vrms''' - Settling time''' 100 µs '''to''' 10 s'''
|0 to 2.5 Vrms - Settling time 100 µs to 10 s


|-
|-
|Resolution
|Resolution
|'''125 mHz''' to '''400 Hz''' - '''101''' lines to '''6401''' lines
|125 mHz to 400 Hz - 101 lines to 6401 lines


|-
|-
|Phase
|Phase
|'''Fixed''' (all sine have same phase) - '''Random'''
|Fixed (all sine have same phase) - Random


|-
|-
|Burst
|Burst
|'''0 '''to '''100% '''- Step''' 1%'''
|0 to 100% - Step 1%


|-
|-
|Number
|Number
|'''2''' to '''4''' '''independents''' multi-sine
|2 to 4 independents multi-sine


|}
|}
Line 2,398: Line 2,380:
'''Output signals (continued)'''
'''Output signals (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "5"|'''Random noise'''
! rowspan="5" | Random noise
|Frequency span
|Frequency span
|From '''125 mHz''' to '''40 kHz''' - independent lower and upper frequencies
|From 125 mHz to 40 kHz - independent lower and upper frequencies


|-
|-
|Amplitude
|Amplitude
|'''0 '''to''' 2.5 Vrms''' - Settling time''' 100 µs '''to''' 10 s '''
|0 to 2.5 Vrms - Settling time 100 µs to 10 s  


|-
|-
|Type
|Type
|'''White '''-''' pink '''
|White - pink  


|-
|-
|Burst
|Burst
|'''0 '''to '''100% '''- Step''' 1% - '''Bloc duration''': 2.5 ms to 100 s'''
|0 to 100% - Step 1% - Bloc duration: 2.5 ms to 100 s


|-
|-
|Number
|Number
|'''2''' to '''6 independent''' and '''uncorrelated''' random noises
|2 to 6 independent and uncorrelated random noises


|-
|-
|align = "center" rowspan = "5"|'''Chirp'''
! rowspan="5" | Chirp
|Frequency span
|Frequency span
|From '''3.125 Hz''' to '''40 kHz''' - independent lower and upper frequencies
|From 3.125 Hz to 40 kHz - independent lower and upper frequencies


|-
|-
|Amplitude
|Amplitude
|'''0 '''to''' 7,07 Vrms''' - Settling time''' 100 µs '''to''' 10 s'''
|0 to 7.07 Vrms - Settling time 100 µs to 10 s


|-
|-
|Size
|Size
|'''256 '''- '''512 '''-''' 1024 '''-''' 2048 '''-''' 4096 '''-''' 8192 '''-''' 16384  '''
|256 - 512 - 1024 - 2048 - 4096 - 8192 - 16384   


|-
|-
|Burst
|Burst
|'''0 '''to '''100% '''- Step''' 1%'''
|0 to 100% - Step 1%


|-
|-
|Number
|Number
|'''2''' to '''6''' '''independents''' chirps
|2 to 6 independents chirps


|}
|}
Line 2,449: Line 2,431:
'''ABW:''' the analysis bandwidth, of the plug-in or front end where the filter is in use.
'''ABW:''' the analysis bandwidth, of the plug-in or front end where the filter is in use.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{| class="wikitable" width="100%"
|align = "center" rowspan = "3"|'''High pass'''
! rowspan="3" | High pass
|Type
|Type
|'''Butterworth''' '''IIR '''type
|Butterworth – IIR type


|-
|-
|Order
|Order
|'''1''' to '''6'''
|1 to 6


|-
|-
|Cutoff frequency
|Cutoff frequency
|'''ABW'''to '''ABW/1024'''
|ABWto ABW/1024  


|-
|-
|align = "center" rowspan = "3"|'''Low  pass'''
! rowspan="3" | Low  pass
|Type
|Type
|'''Butterworth''' '''IIR '''type
|Butterworth – IIR type


|-
|-
|Order
|Order
|'''1''' to '''6'''
|1 to 6


|-
|-
|Cutoff frequency
|Cutoff frequency
|'''ABW '''to '''ABW /102.4'''
|ABW to ABW /102.4  


|-
|-
|align = "center" rowspan = "4"|'''Stop/pass band'''
! rowspan="4" | Stop/pass band
|Type
|Type
|'''Butterworth''' '''IIR '''type
|Butterworth – IIR type


|-
|-
|Order
|Order
|'''2''' to '''10'''
|2 to 10


|-
|-
|Cutoff frequencies
|Cutoff frequencies
|'''ABW '''to''' 0.055 * ABW '''
|ABW to 0.055 * ABW  


|-
|-
|Bandwidth
|Bandwidth
|'''ABW/2 '''to''' 0.0075 * ABW '''
|ABW/2 to 0.0075 * ABW  


|-
|-
|align = "center" rowspan = "3"|'''Integrators'''
! rowspan="3" | Integrators
|Type
|Type
|'''HP, Single '''or''' double '''
|HP, Single or double  


|-
|-
|High pass frequency
|High pass frequency
|'''ABW''' '''/10 000 '''or ''' ABW''' /'''2 000 '''
|ABW /10 000 or ABW /2 000  


|-
|-
|Integration time
|Integration time
|'''2 ms '''to''' 500 s'''
|2 ms to 500 s


|-
|-
|align = "center" rowspan = "2"|'''Differentiators'''
! rowspan="2" | Differentiators
|Type
|Type
|'''Single'''
|Single


|-
|-
|Average
|Average
|'''Sliding '''-''' 0 '''to''' 2 s'''
|Sliding - 0 to 2 s


|-
|-
|align = "center" rowspan = "2"|'''Weightings'''
! rowspan="2" | Weightings
|Laws
|Laws
|'''A''' & '''C''' laws  
|A & C laws  


|-
|-
|Bandwidths  
|Bandwidths  
|'''10''' kHz to '''40''' kHz  
|10 kHz to 40 kHz  


|-
|-
|align = "center" rowspan = "3"|'''All'''
! rowspan="3" | All
|By-Pass
|By-Pass
|'''Apply/bypass''' without stabilisation time - independently on each filter
|Apply/bypass without stabilisation time - independently on each filter


|-
|-
Line 2,538: Line 2,520:
|}
|}


----


==Notes==
==Notes==

Latest revision as of 12:24, 22 July 2026

The following official specifications concern NVGate® software for OR10, OR34, OR35, OR36, OR38 and Mobi-Pack® multi-analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the instrument connected to the PC. This specification applies for NVGate version 12 or upper and Teamwork instruments.

Modules

NVGate® software is based on a general-purpose platform where optional analysis modules called plug-in analyzers are added.

The FFT plug-in features specific add-on for specialized analysis.

NVGate® Platform

NVGate® platform provides a comprehensive set of tools for noise and vibration acquisition, recording and analysis.

These functions are arranged in 4 parts:

  • Signal sources that condition, provide and store time domain signals.
  • Shared resources that apply simultaneously or independently on distributed signal to plug-in analyzers.
  • Operational tools that increase measurement efficiency and reliability.
  • General-purpose analysis for monitoring and result tracking.
Signal sources Front-end Controls dynamic inputs, generators, external synchronization/tach. input and parametric (DC) inputs.
Recorder Record and store time domain signal on analyzer or PC hard disc. Records dynamic, parametric (DC) and ext. synch/tach inputs with multiple frequencies.
Player Visualize, listen, playback on output or generate signals for post-analysis purpose. Based on recorded or imported time domain signal file.
Shared resources Filters High/low pass, Band/Stop pass, single and double integrator, differentiator, A and C acoustic laws
Events Edge detection, DC levels, Delta DC levels, RPM, Delta RPM, Manual, Time period, combinations.
Weighting windows Programmable Force and response, Hanning, Hamming, Kaiser Bessel, flat-top, uniform.
Tachometers Compute RPM from ext. sync and input pulses, DC, CAN. Fractional, simulated and combined tach.
Operational tools Transducers management Automatic calibration, database management, calibration history, automatic setup of inputs, TEDS management.
Masks Mask editor for spectra, profiles, 1/n octave, order spectra
Report On-line report editor, automated report generation, word or Excel files.
Automation tools Macros, sequence from Excel® sheet, alarms based on mask comparison, customizable control panel, restricted user profiles, remote controller interface.
General purpose analysis Monitor 4 channels FFT analyzer with input hot swap. Statistical extraction (RMS, Max, Min, Kurtosis) from time domain with programmable band-pass filter.
Waterfall Synchronized stack for 2D (spectra) and scalar (order, overall). 3D, color map and profiles displays. Extraction capabilities.

NVGate® Plug-ins analyzers

In addition to the standard functions featured in the software platform, NVGate® can receive additional plug-in analyzers that offer flexible configuration of independent analysis modes.

FFT Fast Fourier Transform - Time to Frequency conversion with multiple average domains, combinable trigger, zooms and cross functions. General-purpose frequency analysis, machinery signature, modal analysis acquisition (hammer or shaker), machinery diagnostic.
FFT Add-ons    
CBT Constant Band Tracking - Order extraction at constant bandwidth Gears mesh analysis, gearboxes noise and vibration tracking.
FFTDiag Cepstrum, Auto and cross correlation, DC, Min/Max, Pk, pk/pk and crest factor. Set of functions dedicated to the diagnostics of industrial machineries.
TDA Time Domain Analysis –Time view (oscilloscope). True statistical extraction, averaging. Machinery diagnostic, long term acquisition monitoring, critical acquisition control, large structure damping control, shocks determination, machinery health control, trend analysis.
SOA Synchronous Order Analysis - Time to Order re-sampling with angle or order averaging, acoustic weightings, multiple tach management. Rotating machinery balancing. In vehicle order to noise correlation R&D. Pump, compressors, turbines and speed reducer/multiplier test.
SOADiag
(SOA Add-on)
Cepstrum, Auto and cross correlation, Revolution synchronous statistics, Order transmission Function (ORF™), X functions Set of diagnostic functions based on the synchronous order analysis for rotating part transmission and forced vibration extraction.
OCT 1/n Octave Constant Percentage Band filters - 1/n octave detectors set with multiple averaging modes. General-purpose industrial noise analysis. Vehicle acoustics R&D. Noise test.
OVA Overall Acoustics Detectors - 3 detectors + 1 peak detection per channel with parallel averaging and acoustic weightings. Industrial acoustic, sound power measurements. Machinery with cycles, R&D and test.

NVGate® Options

Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software also features options:

IVC Instantaneous angular velocity converter Torsional measurements from Ext. Synch inputs
CTE Combined Tachometer editor Tachometer math editor - 2 inputs, 1 output
VIN Virtual Inputs Real-time math combination of dynamic inputs
VDC Virtual DC Real-time math combination of parametric inputs
A-Samp Angular sampling for SOA Real-time Angular sampling of inputs

Signal sources

NVGate®/OROS 3-Series multi-analyzers platform can process signal coming from different sources. These sources are detailed in this section as modules.

Front-end

The front-end module gathers the different input and output settings available in the instrument unit. In addition, the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to the PC) which is useful for set-up.

Front-end Settings

Dynamic inputs Sampling clock 102.4 kS/s to 2.048 S/s or 65.536 kS/s to 3,200 S/s
LEMO2 connectors management 200 V polarization on/off per block of 8 input
Coupling3,4 AC (0.35 Hz) - DC - ICP - ICP + TEDS - AC float - DC float - Grounded
Node information3 Label - component - node - direction (± X ,Y,Z) - type (translation, rotation)
Physical quantity3,4 Any physical quantity can independently be associated to inputs.
Sensitivity3,4 User defined in V/unit.
Range2 Pk to pk expressed in the associated unit, (taking in account sensitivity) - linear or dB
External conditioner compensation3 Gain - polarity - offset.
Filter3 Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Auto-range Enable/disable auto-ranging independently on any input
Ext. sync inputs Sampling over-sampled 64 times the dynamic inputs sampling
Coupling3 AC (0.35 Hz) - DC
Information3 Label
Physical quantity3,4 Any physical quantity can be associated to inputs.
Sensitivity3,4 User defined in V/unit.
Range3 Pk to pk expressed in the associated unit, (taking in account sensitivity) - linear or dB
External conditioner compensation3 Gain, offset.
Edge detection3 Threshold (in associated unit) - Edge (rise, fall) - Hysteresis - hold off (sec)
Pre-divider3 Hardware - 1 to 255 pulses – accept up to 375 kpulses/sec
Post-multiplier3 1 to 50 pulses/trigger - Generate higher frequencies on low rate triggers. (ex. GPS)
Tachometer mode3 Pulse per rev: 0.5 to 4096 - average (n rev) - min (detect stopped shaft) and max RPM speed - hold off (in % of revolution) - rotation (clockwise/counterclockwise)
Torsional mode2: Instantaneous angular velocity converter Pulse per rev: 1 to 4096 - min (detect stopped shaft) and max RPM speed - sliding average (1 to 20 samples) - Missing pulse number: 0 to 5, Missing pulses detection :1 to ±20%: of previous pulses duration- Filters (dt, dt², 1/dt)
Angular sampling2: multi-pulse/rev clock used for signal sampling for SOA - Pulse per rev: 1 to 4096 continuously (no power of 2 limitation) - Real time, in line anti-aliasing - Missing pulse number: 0 to 5, Missing pulses detection: 1 to ±20% of previous pulses duration – Phase reference: Any tachometer including same input



Outputs Sampling Same as dynamic inputs
Generated signals3 Any outputs signals, see §outputs - play-back of signal file tracks - DC levels - Inputs playback (Delay > 256 samples)
Filters3 Any NVGate filter can independently be applied to each dynamic output (See Filters §)
Impedance3 50 ohm, 600 ohm (only OR36 and OR38) or grounded
Gain3
Synchronization3 Free run or linked to acquisitions runs
Level and phase transition3 No (Steps) - controlled (ramp) - user selectable ramp time
Clipping User selectable - protects shaker and amplifier.
Events (can trig analysis or record)3 Outputs stabilized - new step reached - start & stop sweep
Controls Emergency stop - mute/un-mute all
Parametric1(DC) inputs Sampling 12.3 to 12.8 S/s - 10 Hz to 100 Hz harmonics rejection
Information3 Label
Physical quantity3,4 Any physical quantity can be associated to inputs.
Sensitivity3,4 User defined in V/unit.
Range3 Pk to pk expressed in the associated unit (taking in account sensitivity) - linear or dB
External conditioner compensation3 Gain (-120 dB to +20 dB) - polarity - offset (< range).
Auto-range Enable/disable auto-ranging independently on any input
Auto-ranging Type Normal - on Peak detection
Margin Select the first highest range with 0 dB - 3 dB or 20 dB margin
Duration (apply for normal auto-range) 0 to 10 sec.
Peak detection User select input where to detect peak, auto-range applies on all enabled input at each peak detection.
Peak parameters settings 1 to 10 peaks - sensitivity (low, normal, high) - rise, fall or any edge.
Checking ICP (check not available on OR34) Test and report (Open, shortcut, ok) all enabled inputs.
TEDS Transducers automatic recognition - Complies with IEEE 1451.4 2004 Rev. 1.0

2: Optional features

3: Independent for each input

4: Linked with the transducer database

Front-end results & connections

The following results are available for monitoring and connection to analysis mode (plug-in analyzers)

Dynamic inputs & torsional1 inputs Monitoring3 Time domain instantaneous signal - Size 256 samples
Status led3 (docking tool bar) Overloaded = red - overload occurred since last start = red with yellow center - ok = green - under load (20 dB below range) = green with yellow center
Connection3 To any: plug-in analyzer channel, monitor channel, Virtual inputs, recorder track, edge event detector & tachometer
Ext. Sync inputs Monitoring3 Time domain instantaneous status (1 = threshold crossed, 0 = no event) - Size 256 samples.
Connection3 To any: start or stop averaging of plug-in analyzers, recorder start or stop recording, waterfall start or stop acquisition, new block trigger for FFT and SOA and new slice for waterfall acquisition.
To torsional, tachometer & angular sampling
Parametric1 (DC) inputs & CAN1parameters Monitoring3 Profiles versus time (160 ms to 163 sec user selectable), and digital/analog view meter.
Status led3 Overloaded = red - ok = yellow - under load (20 dB below range) = Cyan
Connection3 To any: recorder track, waterfall reference (Z/X axis) and profiles, level delta level detectors and virtual DC

Recorder / Player

OROS 3-Series/NVGate instruments feature a recorder and a player module that allows users to:

  • Record time domain signal, torsional inputs, trigger and parametric input.
  • Export or import time domain files.
  • Playback time domain files on analyzers outputs during analysis
  • Listen to recorded tracks on the PC speakers.
  • Post-analyze time domain files using the available plug-ins.

Recorder

The recorder module saves time domain signal into files located on 3-Series analyzer HD, Mobi-DiskTM or PC HD. This is available on 2 different modes:

  • On-line record, the chosen inputs (Dynamic, ext. sync, parametric) are recorded at selected sampling frequency during acquisition. Real-time analysis is available simultaneously.
  • Time and tracks split, the chosen recorded tracks (from signal file) are played back in the recorder using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be changed.

Recorder Settings

Bandwidths Dynamics inputs 2 groups of user selectable sampling frequency - From 102.4 kS/s to 2.048 S/s (40 kHz to 800 mHz) - available simultaneously
Ext. sync inputs Automatic selection at Front-end sampling frequency - Resolution is 64 time Front-end sampling frequency
Parametric (DC) inputs 12.5 S/s (50 Hz rejection) - 15 S/s (60 Hz rejection)
Tracks Number 6 tracks For O4. or 38 tracks For OR35, OR36, OR38 and Office licenses. (32 + 6 ext. Sync)
Sum of connected licenses using multiple hardware
Saved settings / track Label - Coupling - External gain - Input range - Sampling frequency and signal bandwidth - Component - Node number - Direction - Type - Associated transducer - Unit - Sensitivity and Offset compensation.
Modes Start to time Start recording on run or any activated event - Stop recording when duration ends - Duration 10 ms to available space on target HDD.
Start to stop Start recording on run or any activated event - Stop recording on stop on any activated event - Start and stop event can be the same - Duration 10 ms to available space on active drive (PC or OR3x).
Time to stop Memorize the earliest x seconds - Stop = stop or any activated event - Duration 10 ms to 2 GSamples (limited by the drive available space)
Triggering Start recording (new record) Any NVGate® event plus manual and free run
Stop recording (end current record) Any NVGate® event plus manual and free run
Start delay Positive = unlimited - negative = 128 kSamples
Stop delay Positive = unlimited - negative = 0
Markers Saved in the record files Added by tool bar or shortcut - comment fields editable during record or at end of acquisition
Signal file Record on PC HDD Up to 10 dynamic inputs – Up to 4 parametric inputs - max total bandwidth 512 kS/s - (2 MB/s)
Record on Analyzer SSD Up to 40 channels (32 inputs + 6 ext. sync or DC)/instruments - max bandwidth 3.2 MS/s (12.5 MB/s)
Format Normal: 32 bits/sample - Compressed 16 bits/sample

Recorder Settings (continued)

Management Download Discard Analyzer SSD on PC HDD - Batch download available - Rate 4 MB/s
Upload Load PC HDD records on the analyzer SSD - Rate 1.4 MB/s
Location All records accessible from the project manager tree
Hard drive management Format SSD - Selective delete - Sorting tools

Mobi-DiskTM

OR36 Mobi-Pack and OR38 multi analyzers/recorders feature a removable hard drive with 2 ports:

  • Parallel High speed port for direct data throughput to the Mobi-DiskTM.
  • USB 2.0 port for post processing and data management without the instrument.
Connection 1 to 8 Mobi-DiscTM Up to 8 Mobi-disc connected - Active Mobi-Disc: 1 (user selectable) - hot swap
Transfer Based on the USB 3.0 link Download 15 MB/s

Recorder Monitoring

The following displays are available during real-time or post-analysis.

RMS Overall RMS / input Digital or analog view-meter
Signal Real-time Compressed view of entire recorded signals - Automatic update of time axis.

Player

The player module plays the recorded signal files. This is done one of the following two ways:

  • Post-analysis, the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
  • Playbacks, the recorded dynamic input are generated on the instrument outputs simultaneously with standard analysis of the inputs.

Player Settings

Played section File selection NVGate® recorded files - Imported files - located on instrument SSD, Mobi-DiskTM or on PC HDD.
Record selection Record number selection (for multi record files)
Section definition User selectable Start and stop offsets (in sec.) - available in the file overview window (marker and slider)
Duration Play backed duration (User information)
Repeat On/off - repeat continuously the selection (available only for playback on analyzer output)
Mode File swap Swap loaded files with same tracks number - applies identical post-analysis setup to multiple files.
Playback Continuous or step by step (5 ms to 360,000 s)
Bandwidth All tracks at lowest frequency (compatible with all plug-ins) or initial record frequencies (no track mix in plug-ins)
Tracks Number Up to 352 according to user's fleet
Type Dynamic input record (2 sampling frequencies), parametric (DC) input record and ext. sync record.
Fixed setting (information) / track Label, Coupling, external gain, input range, sampling frequency and signal bandwidth
Modifiable measurement point information settings (apply for post analysis or re-recording) / track Component - node number - direction - type
Modifiable settings (apply for post analysis or re-recording) / track Associated transducer - unit - sensitivity - offset compensation

Player Settings (continued)

Listening File location Analyzer HD - Mobi-DiskTM connected with USB or inside the analyzer - PC HD
Playback One track - from cursor position - repeat displayed signal
Markers Display Recorded markers are available in the file overview
Use Set the start and stop playback offset (user selectable)

Player Connections

The following table describes the available connections to the analysis modes (plug-in analyzers and recorder) during post-analysis operation:

Dynamic inputs & torsional Tracks Preview (multiple file simultaneously) Entire file fast overview (pre-compressed at recording time) - Track contents preview, independently (multi-graph display)
Monitoring (loaded file only) Zoom on selected play-back section
Connection3 (loaded file only) To any: plug-in analyzer channel, monitor channel, recorder track, edge event detector and tachometer
Ext. sync Tracks Preview (multiple file simultaneously) Entire file fast overview (pre-compressed at recording time) - Track contents preview, independently (multi-graph display)
Monitoring (loaded file only) Zoom on selected play-back section
Connection3 (loaded file only) To any: start or stop averaging of plug-in analyzers, recorder start or stop recording, waterfall start or stop acquisition, new block trigger for FFT and SOA and new slice for waterfall acquisition.
To torsional, tachometer & angular sampling
Recorded parametric1(DC) inputs & CAN parameters Preview (multiple file simultaneously) Entire file fast overview - Track contents preview
Monitoring (loaded file only) Profiles versus time and digital or analog view meter.
Connection3 (loaded file only) To any: recorder track, waterfall reference (Z/X axis) and profiles, level and delta level events.

Plug-ins analyzers

The following plug-in analyzers are available as options of NVGate® software platform.

Time Domain Analysis (TDA)

The TDA plug-in analyzer provides time domain based analyses and visualizations. This plug-in analyzer computes statistical extractions and data compression of long duration oscilloscope views. All results are real time and operations are possible on-line or in post processing. The TDA plug-ins can operate free run or synchronized with the other plug-in analyzers.

TDA Settings

Bandwidths Oscillator 1 800 mHz to 40 kHz
(sampling oscillator 102.4 kS/s to 2.048 S/s)
Oscillator 2 512 mHz to 25.6 kHz
(sampling oscillator 65.536 kS/s to 3,277 S/s)
Time views Depth 7200/ABW ABW: the plug-in Analysis Bandwidth. sec to 100 000 sec (27 hours) - Independent on each channel
Time base (resolution) Depth/2048- Independent on each channel
Time span for extraction
(averaging)
Type Continuous sliding with refresh each 256 samples (Exponential) - One shot (Linear) - Repeated shot (Repeated linear)
Duration 1/ABW to 2 106/ABW (ex: 50 ms to 1000 sec @ 2 kHz)
Channels OR36 & OR38 analyzers 32 channels per plug-in for on line 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Filters On each channel Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Triggering

Start analysis Any NVGate® event plus manual and free run
End analysis Any NVGate® event plus manual and free run
Repeat analysis On new start or end of averaging
Trigger delay Positive = unlimited - negative = 32k Samples

TDA Results

The following results are available for real time or post-analysis display, report and saving.

Time Views Overview Signal envelope: Min/Max line for each time step -
  Statistical values DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Skewness - Kurtosis - computed on a user defined area of the time view
Extractions Statistical values DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Kurtosis - computed on the Time span – All values available for View-meter and/or Waterfall profiles


Fast Fourier Transform analysis (FFT)

The FFT plug-in analyzer features multiple vibrations results depending on applied setting. This type of plug-in analyzers computes real-time (on-line or post-processing) Fast Fourier Transform algorithms with multiple averaging domain and trigger capabilities. Up to 4 FFT plug-ins can operate at the same time.

FFT Settings

Bandwidths Oscillator 1 800 mHz to 40 kHz
(sampling oscillator 102.4 kS/s to 2.048 S/s)
Oscillator 2 512 mHz to 25.6 kHz
(sampling oscillator 65.536 kS/s to 3,277 S/s)
Resolution Lines 101, 201, 401, 801, 1601, 3201, 6401, 12801 With force DSP and 25601 lines
Frequency resolution 80 µHz (512 mHz/6401 lines)
to 400 Hz (40 kHz/101 lines)
Averaging Domain Spectral (power) - time (with phase) - FDSA (Synchronized with one frequency)
Type Exponential - linear - repeated linear - referenced peak hold (eq. to tracking filter locked on reference channel frequency) - peak hold
Size Duration or number of blocks, unlimited size
Overlap 0% to 99.9% depending on block size - retrigger
Channels OR36, MP & OR38 analyzers 32 to 256 ch. per plug-in for on line and post-analysis
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Weighting General Uniform - Hanning - Hamming - Kaiser Bessel
Modal acquisition User defined Force (rect.) and Response (Rect. + exp)
Assignment Independently on any channels
Filters On each channel Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Cross functions Reference selection Full or partial matrix available (all channels can be references)
Zoom Magnification factor 2 to 128 by step power of 2
Min., max and center frequencies Graphically selectable on wide band spectrum
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
New block Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Trigger delay Positive = unlimited - negative = 32k Samples
Automatic Reject overloaded blocks
Blocks rejection
Overall
Manual Accept or Reject after preview of averaged results (FRF, Coherence)
Normal Accept all blocks
Overall Detector Quadratic sum of spectra lines taking in account weighting window equivalent noise bandwidth
Lower and upper frequencies Selectable into the current FFT bandwidth

FFT Results

The following results are available for real time or post-analysis display, report and saving.

Time averaging Time domain Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks
Spectra Complex spectra
Cross-functions Instantaneous cross-spectra - averaged cross-spectra - FRF H1 - FRF H2 - Coherence
Overall Overall RMS value in specified band - Overall RMS profile in specified band.
Spectral averaging Time domain Filtered signal - Triggered block - Weighted Block - Lissajous of triggered blocks
Spectra Instantaneous complex spectra - averaged power spectra
Zoomed spectra Zoomed instantaneous complex spectra - zoomed averaged power spectra
Cross-functions Instantaneous cross-spectra - averaged cross-spectra - FRF H1 - FRF H2 - Coherence
Zoomed cross-function Instantaneous zoomed cross-spectra - averaged zoomed cross-spectra - zoomed FRF H1 - zoomed FRF H2 - zoomed Coherence
Overall Overall RMS value in specified band - Overall RMS profile in specified band.
Synchronous averaging on one frequency (FDSA) Time domain Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks
Spectra Instantaneous complex spectra - time domain averaged power spectra
Overall Overall RMS value in specified band - Overall RMS profile in specified band.

Constant Band Order Tracking Add-on (CBT)

Constant band order tracking is an optional add-on to the FFT plug-in analyzers. This add-on computes spectrum lines power related to RPM speed. This option adds settings and results to the FFT Plug-ins. Constant band tracking can operate on up to 4 different tachometers at the same time.

CBT Settings

Tracked Order Number 8 per channels
Max 0.001 to 800
Constant bandwidth User selectable per channel - minimum depend on weighting windows
Computation Associated tachometer Any front end, recorded or virtual tachometer
Peak tracking On/Off center bandwidth on nearest peak
Order amplitude Weighting windows eq. noise bandwidth correction

CBT Results

The following results are available for real time or post-analysis display, report and saving.

Scalar Tracked order Digital (magnitude and phase) or analog view-meter
Cross phase tracking Order phases are relative to the same order from a ref. channel
Overall Digital or analog view-meter
Monitoring Continuous profiles of tracked order Profiles vs. time - profiles vs. RPM - max depth 2048 pts - user selectable delta time - user selectable delta RPM
Profiles One shot acquisition Tracked order (complex), cross phase orders and overall level can be collected by the waterfall profiles.
References Time, RPM and DC levels

Diagnostic Add-on (FFTDiag)

FFT Diagnostic is an optional add-on to the FFT plug-in analyzers. This add-on computes a set of useful analysis for machinery diagnostic. Up to 4 FFTDiag can operate on 4 different FFT plug-ins at the same time.

This option adds settings and results to the FFT Plug-ins.

FFTDiag Settings

Envelop demodulation Activation Enable/disable on all channels - requires zoom activation
Averaging Spectral domain - FDSA
Bandwidth ½ of zoom span - ¼ of FFT bandwidth
Correlation Activation Enable/disable on all channels - All other FFT results are affected
Weighting window Uniform - Left zero padding - Centred zero padding

FFTDiag Results

The following results are available for real time or post-analysis display, report and saving.

Scalar Min/Max Minimum and maximum amplitude per trigger block
Peak detectors Peak level - Peak to Peak level - Crest Factor - per trigger block
Others Cepstrum Cepstrum - Zoomed cepstrum
Correlation Auto-correlation block - Cross correlation block
Shaft view Time domain signal on first rotation - angular representation


Synchronous Order Analysis (SOA)

The SOA plug-in analyzer features several results depending on applied settings. Up to 3 SOA plug-ins can operate in parallel.

SOA Settings

Bandwidths Order span 6.25 to 400 – up to 800 with ForceDSP
Frequency bandwidth Up to 40 kHz
RPM span Up to 384 000 RPM - Ratio of 64 between Min & Max RPM
Resolution Lines 101 - 201 - 401 – 801 – 1601 with ForceDSP
Order resolution 1 - ½ - ¼ - 1/8 - 1/16 - 1/32
Averaging Domain Spectral (power) - revolution (re-sampled blocs)
Type Exponential - linear - repeated linear - peak hold
Size Number of blocks, unlimited size
Overlap 0 to 31 revolutions depending on resolution – 0 to 359° into revolution – Retrigger on multi-pulse/rev
Tachometer Computation base Any NVGate® tachometer (ext. sync, regular input, DC input, CAN, simulated, Fractional, and combined tach.)
Phase reference Tachometer pulse edge or pulse center requires to analyze the tachometer input
0° Phase reference Cosine or sine
RPM Max speed variation 1 to 99% per analyzed block - blocks with higher variation are rejected. No control (set @ 100%)
RPM range User selectable max & min RPM, under and over speed blocks are rejected
Tracked order Number Up to 8 per channels
Min-Max 1/32 to 400 – up to 800 with ForceDSP
Cross phase tracking Order phases are relative to the same order from a ref. channel
Channels OR36 & OR38 analyzers 32 channels per plug-in for on line - 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Weighting Windows selection Uniform - Hanning - Hamming - Kaiser Bessel Applied independently on any channels
Filters On each channel Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
New block Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Trigger delay Positive = unlimited - negative = 1/order res revolution
Trigger block phase shift ±720° - independent for each channel
Overall Detector Quadratic sum of order spectra lines taking in account weighting window equivalent noise bandwidth
Lower and upper orders Selectable into order span - from 0.03125 to 800 with ForceDSP


SOA Results

The following results are available for real time or post-analysis display, report and saving.

Revolution averaging Time domain Filtered signal
Angle domain (Revolution) Triggered block - Averaged Triggered block - Weighted block
Order spectra Complex spectra
Tracked orders Digital (magnitude & phase) view-meter - Analog view -meter
Overall Overall RMS value in specified order band (digital or analog view-meter)
Spectral (order) averaging Time domain Filtered signal
Angle domain (Revolution) Triggered block - Weighted Block
Order spectra Instantaneous complex order spectra - averaged power order spectra
Tracked orders Digital (module and phase) and/or Analog view-meter.
Overall Overall RMS value in specified order band (digital or analog view-meter)

Diagnostic Add-on (SOADiag)

Order based Diagnostic is an optional add-on to the SOA plug-in analyzers. This add-on computes a set of useful synchronous order transmission extraction tools and analyses. Up to 2 SOADiag can operate on 2 different SOA plug-in at the same time.

This option adds settings and results to the SOA Plug-ins.

SOADiag Settings

Cross-functions Reference selection Full or partial matrix available (all channels can be references) - order domain averaging only.
Angular Correlation Activation Enable/disable on all channels - All other SOA results are affected simultaneously
Weighting window Uniform - Left zero padding - Centred zero padding

SOADiag Results

The following results are available for real time or post-analysis display, report and saving.

Rev. synchronous Scalars Min/Max Minimum and maximum amplitude per trigger block (n rev.)
RMS, DC Per trigger block (n rev.)
Peak detectors Peak level - Peak to Peak level - Crest Factor - per trigger block (n rev.)
Others Cepstrum Cepstrum of the order spectra
Cross-functions Instantaneous order cross-spectra - averaged order cross-spectra - ORF™ H1 / H2 -Coherence – angular or order domain averaging.
Angular Correlation Auto-correlation block - Cross correlation block - angular averaged
Re-sampled Shaft view Time domain signal on first rotation – angular representation - instantaneous or revolutions averaged


1/n octave constant percentage band filter analysis (OCT)

The OCT plug-in analyzer features multiple acoustic results depending on applied setting. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters (CPB) and detectors.

OCT Settings

Bandwidths Sampling oscillator #1
(102.4 kS/s to 2.048 S/s)
Center of highest 1/3rd band = 40 kHz
Center of lowest 1/3rd band = 100 mHz
Sampling oscillator #2
(65.536 kS/s to 3,277 S/s)
Center of highest 1/3rd band = 25 kHz
Center of lowest 1/3rd band = 100 mHz
Frequency span Highest Band / lowest band < 2000 (ex. 10 Hz - 20 kHz)
Resolution 1/n Octave 1, 1/3 rd, 1/12th, 1/24th
Averaging Basic Linear & repeated linear (20 ms to 60,000 s) exponential (20 ms to 60s)
Acoustics Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse
Constant Bandwidth * Time B*T = 0.2 dB, 0.5 dB, 1 dB, 2 dB
Standards Detectors Complies with IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 compatible: class 1
CPB filters Complies with IEC ANSI/ASA S1.11-2014 / Part 1 / IEC 61260:1-2014 and IEC 1260 compatible: class 1


Channels OR36 & OR38 analyzers 32 channels per plug-in for on line - 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels for on line and post-analysis
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Overall Detectors 2 parallel overall detectors linear and weighted (time domain)
Weighting A, C, or Z (none) applicable in 10 kHz - 40 kHz bandwidth

OCT Results

The following results are available for real time or post-analysis display, report and saving.

Spectra 1/n Octave spectra Instantaneous spectra, averaged spectra
Holden spectra Minimum & maximum spectra
Overall Linear Digital & analog view meter of time domain overall detector, exact bandwidth is informed
Weighted Digital & analog view meter of A or C weighted overall levels computed in time domain, exact bandwidth is informed


Overall acoustics sound level meter (OVA)

The OVA plug-in analyzer features compatible class 1 sound pressure level measurement according to IEC 61672:2013 standard. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters and detectors.

OVA Settings

Type Standard IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 compatible: class 1
Class 1
Bandwidths 10 Hz to 40 kHz - Adjustable
Channels OR36 & OR38 analyzers 32 channels per plug-in for on line - 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels for on line and post-analysis
Averaging Linear User selectable duration - repeat mode
Short leq 1/8 s - 1 s - User selectable duration
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Detectors Peak 1 peak detector/Ch. - A, C or Z (none) weighting -Independent for each channel
Weighting 3 overall detectors/Ch. - A, C or Z (none) weighting - Fast, Slow, Impulse and linear time weighting - Independent for each channel

OVA Results

The following results are available for real time or post-analysis display, report and saving. The following levels can be tracked in profiles up to 100 000 pts. each.

SPL Time weighted Instant - max hold - min Hold
Averaged Short leq - leq
Peak Peak - Time weighted - max hold min hold


Options

Instantaneous Angular Velocity Converter (IVC)

The IVC option converts frequency to voltage from the External synch inputs. It covers torsional and acyclism measurements. The converted signals are made available as standard inputs (or player tracks). This option operates real-time computation and operations are possible on-line or in post processing.

IVC Settings & specs

Rate Pulse/rev 0.5 to 4096 - Up to 1 E6 with pre-divider
Pulse/rev Frequency > 40 kpulse/sec max
Pre-divider 1 to 255 - hardware decimation (pulse are not measured)
Pre-divider frequency 375 kpulse/sec max.
Missing pulses Number 0 to 5 consecutive pulses integer or fractional
Hold off 2 consecutive pulses
Generated signal 1st order interpolation
Tach. phase ref No missing pulse First random - continuous offset further
missing pulse > 0 Last known edge before missing pulse
Resolution Time 160 ns : 1/(SF x 64)
Angular (with SOA) 350 µRad (20 mdegrees) @ 8000 RPM, up to order 10
Pre-processing Smoothing Sliding average - 1 to 32 samples
Filtering Any NVGate filter (See Filters §) incl. dt and 1/dt
Max Speed Without pre-divider Max RPM = 2.4 e6/Pulse per rev - 12 000 RPM @ 200 pulse/rev
With pre-divider Max RPM = 36 e6/Pulse per rev - 12 000 RPM @ 200 pulse/rev

IVC Results

The following results are available for real time or post-analysis.

Signals Number 1 to 6 (depends on the number of available ext. synch)
Use Same as dynamic inputs or recorded dynamic inputs
Magnitude Torsional acceleration, velocity & angle – separated from angular magnitudes (RPM)
View Type Signal - filtered signal - 256 samples frame

Angular sampling (A-Samp)

The angular sampling option allows accurately localizing the order phenomenon angle on cyclic rotating machineries. It uses the pulses delivered by an encoder (or zebra tape or gear teeth) to resample the inputs and torsional signal in the SOA plug-in. This option operates real-time computation and operations are possible on-line or in post processing.

IVC Settings & specs

See front-end/ext. Synch § for details

Virtual inputs (VIn)

The virtual input allows combining front end inputs with polynomial operations to generate a signal (new input). This option covers numerous applications such as vector strain calculation with rosette or dynamic twist measurement on machinery transmission. The virtual inputs are made available as standard dynamic input. A static twist computation based on phase comparison is also available with the Vin + IVC options.

This option operates real-time computation on raw or recorded signal.

VIn Settings

Output Number 12 Operators
Type New item in the active inputs/tracks list
Synchronization 0° phase shift with sources and analyzed signal
Magnitude Automatic or voltage if unresolved
Sources Number 1 to 32 channels
Type Analyzer dynamic inputs or recorded dynamic inputs
Output adjustment Type (Input * Coeff + Offset ) ^power - independent on each output
Offset Any real value from -1 e9 to 1 e9 - ex: -5.67
Coefficient Any real value from -1 e9 to 1 e9 - ex: 12
Power Any real value from -4 to 4 - ex: 0.5
Sources adjustment Type Filtering [ (input * Coeff + Offset ) ^ power] - independent for each source
Offset Any real value from -1 e9 to 1 e9 - ex: -5.67 e-5
Coefficient Any real value from -1 e9 to 1 e9 - ex: 1200
Power Any real value from -4 to 4 - ex: -2.4567
Filtering Any NVGate filter
Operators Type Product - Sum - independent on each operator
Twist Type Phase comparison – output in plane angle
Correction (zeroing) offset
Output Plane angle or torque with conversion factor

Virtual parameter (VDC)

The virtual parameter (VDC) allows combining front-end parametric inputs (DC) with math operations to generate a calculated parameter (new DC). This option covers numerous applications such as averaged temperature or non linear response transducers. The virtual parameters are made available as standard DC input. This option operates real-time computation on raw or recorded signal.

VDC Settings

Output Number 12 Operators
Type New item in the DC inputs/track list
Synchronization 0° phase shift with sources and analyzed signal
Magnitude User defined
Sources Number 1 to 32 channels
Type Analyzer DC inputs or recorded DC inputs
Editor Type Text
Variables Real values
Operators +, -, x, /, ^, = (affectation)
Predefined Ch x (Channels signals), Pi, e
Trigonometric functions Sine, Cosine, Tangent - with Arc and Hyperbolic combination
Logarithmic functions Log2, Log10, Ln, Exp
Miscellaneous Sqrt, Abs, Min, Max, Sum, Avg
Output characterization Limits Min and max: +/- 1 E9 - In current unit
Information Label (text)

Combined tachometer editor (CTE)

The CTE option combines 2 actual tachometers with math to generate a calculated angular velocity (new tachometer). It covers various unreachable rotating parts and more specially the CVT gear chain measurements. The combined tachometers are made available as standard tachometer. This option operates real-time computation and operations are possible on-line or post processing.

CTE Settings

Sources Number 2
Type Any NVGate tachometer – excepted another CTE
Output Type New item in the active tachometers list
Synchronization 0° phase shift with sources and analyzed signal
Editor Type Text
Variables Real values
Operators +, -, x, /, ^, = (affectation)
Predefined Rpm1, Rpm2 (sources speed), Pi, e
Trigonometric functions Sine, Cosine, Tangent - with Arc and Hyperbolic combination
Logarithmic functions Log2, Log10, Ln, Exp
Miscellaneous Sqrt, Abs, Min, Max, Sum, Avg, Sign, If, Rint
Output characterization Rotation Clock wise or counter clock wise
Average Sliding - 1 to 256 revolution
Speed limits Min: 0 RPM - max: 1 200 000 RPM
Information Label (text)

CTE Results

The following results are available for real time or post-analysis

Signals Number 4
Use Same as on-line Tachometer or recorded tachometer
Magnitude Angular velocity (RPM)
View Types View meter - 1 to 100 s rolling profile

General purpose analysis

The following modules are available as standard features of NVGate® software platform; they feature additional analysis capabilities to regular plug-in analysis modules.

Monitor

The monitor is an independent (dedicated processor) module that continuously processes FFT analysis on 4 channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.

Monitor Settings

Fixed setup Bandwidth ABW = Sampling freq / 2.56
Resolution 401 lines
Average domain Spectral (power)
Overlap 0%
Average type Exponential
Weighting windows Hanning
Trigger Free run
Channels Number 4
Swap Between any active dynamic input - hot swap capable
Average Duration Instantaneous (20 ms) to unlimited
Overall analysis Pass band filter Butterworth order 2 to 10 - IIR type - Bypass function
Upper & lower frequency User defined - can be graphically modified
Computed indicators DC - Max - Min - RMS - Skew - Kurtosis - Into defined pass band - overall if bypass is on
Average User defined duration - independent from spectral averaging

Monitor Results and Connections

The following results are available for real time or post-analysis display, connection to other modules.

Time domain Play back On any output generator - hot swappable
FFT analyzer Trigger bloc 1024 samples - time domain analyzed bloc
Spectra Instantaneous complex spectra - averaged power spectra
Overall detectors Display (digital or analog view-meter) DC - Max - Min - RMS - Skew - Kurtosis
Connection To level and delta level event detector - to waterfall profiles

Waterfall

The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in analyzers and sources in one result.

Waterfall Settings

Mode Continuous Circular buffer of results
One shot Fill stack and stop waterfall acquisition
Triggering Start acquisition Any NVGate® event plus manual and free run
Stop acquisition Any NVGate® event plus manual and free run
New slice (point or spectra) Any NVGate® event - manual - free run - availability of connected results - periodic

Waterfall Settings (continued)

Size Number of result per acquisition Up to 95 results + Time
Depth 2 to 100,000 slices or points - depend on PC available memory and requested result - automatically adjusted before acquisition
Depth for stand-alone acquisition Up to 3 MSamples per computation DSP
Channels Type scalar (the following results are connectable to waterfall channels) Overall levels (lin and weighted) from OvA and 1/n OCT - Monitor indicators (DC, Max, Min, RMS, Kurtosis) - Orders (from CBT and SOA) - Complex orders (magnitude & phase) - Overall levels in selected BW from FFT & SOA (order or frequency) - TDA scalar: DC, Min/max, RMS, Kurtosis, peak, peak-peak, crest factor
Type 2D - FFT (the following results are connectable to waterfall channels) Triggered block - Averaged Triggered block - Weighted Block - Complex spectra - Power spectra -Cross-spectra - FRF H1 - FRF H2 - Coherence - Zoomed complex spectra - zoomed power spectra
Type 2D - OCT (the following results are connectable to waterfall channels) Instantaneous spectra - averaged spectra - max & min hold spectra
Type 2D - SOA (the following results are connectable to waterfall channels) Triggered block - Averaged Triggered block - Weighted block - Complex order spectra - Power order spectra

Waterfall Results

The following results are available for real time or post-analysis display, report and saving. All stacked results can be saved.

3D Display 1 pane (3D) - 2 panes (3D + YZ view or 3D + XY view) , 3 panes (3D + YZ + XY + Extraction view) windows - automatic or user selectable pane arrangement - Real or imaginary part and module or phase for complex results
Z axis (X axis for extraction and YZ view) Any reference - time - independent for any window - swap reference at any time
Saving selection Entire 3D data and / or any active section (YZ, XY or Extraction)
Profiles Display Profile of any scalar - Real or imaginary part and module or phase for complex results
X axis Any reference - Time – Slice - independent for any window - swap reference at any time

Waterfall Tools

The following results are available for real time or post-analysis display, report and saving.

Sections YZ Sections (Profiles vs. Ref) Any order/frequency - Power - Peak - on selected bandwidth
XY Sections (Spectra) Any position in the current ref. - Min - Max - Average - on selected range or all slices
Order/Frequency extraction (profiles vs. Ref.) Any order/frequency extraction - user selectable tach. - Power - Peak - on selectable bandwidth - Max order
Number of sections Unlimited - available on result or real-time waterfalls
Linked cursors General Single or dual cursor in each view - linked with other graphs
Linked by value Linked cursors track the same X value in different windows.
Linked by reference Linked cursors track the same acquisition slice in different waterfall windows with different X or Z-axis.

References are: Time - Slice number - DC channels (expressed in physical quantity) - TDA scalars - Monitor levels- any Tachometer

Shared resources

The following resources are available for each plug-in analyzer (when it is compatible). For example once a trigger is activated it can be applied to FFT and SOA: Then modifying the trigger setting in the resources will apply on all corresponding plug-in.

Triggering

Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the Event module.

Event Settings

Edge detector Source Any dynamic input - Any recorded dynamic input
Label String for event identification (Ex. "Impact" for a hammer impact detection)
Threshold Between min and max range - use source unit (ex. g for an accelerometer)
Pre-filtering A law - C law - any NVGate filter
Slope Rise - fall
Hold off 0 to 36000 s
Hysteresis Into source input range - use source unit (ex. g for an accelerometer)
Number 2 edge detectors
RPM speed detector Source Any tachometer - computed from dynamic input or ext.sync - any recorded pulses (through tachometer)
Label String for event identification (Ex. "start speed" for a run up initial triggering)
Threshold Detection into source tachometer RPM range
Slope Run up - Run down
Hysteresis 0 to source max RPM
Interpolation On interpolate event occurrence into revolution - Off event occur at revolution ends (new pulse)
Number 2 to 6 RPM speed detector
Delta RPM speed detector Source Any tachometer - computed from dynamic input or ext.sync - any recorded pulses (through tachometer)
Label String for event identification (Ex. "Each 100 RPM" for a run up triggering)
Lower velocity Minimum RPM speed - Events occurs only for higher source speed
Higher velocity Maximum RPM speed - Events occurs only for lower source speed
Delta velocity Define velocity step - event occur each time source speed increase or decrease by delta velocity
Slope Run up - run down - first, first slope is automatically selected - any, event occur on any slope
Interpolation On, interpolate event occurrence into revolution - Off, event occur at revolution end (new pulse)
Number 2 to 6 delta RPM speed detector
Level detector Source Any parametric (DC) input - Monitor scalar - Filtered monitor scalar (Band Pass)
Label String for event identification (Ex. "temp A" for a recording trigger)
Type Monitor scalar - DC - RMS - Max - Min - Kurtosis
Status Above / below - Detection level - available in lin or dB
Number 1 to 4 level detectors

Event Settings (continued)

Periodic event Label String for event identification (Ex. "Hourly" for 1 hour time interval)
Time interval 2 ms to 11 days - Synchronized with analyzer.
Number 2 periodic events
Combination Sources Any event - 2 different sources defined as A and B
Label String for event identification (Ex. "Time/RPM variation" for a mix between periodic and delta RPM)
Combination A OR B - A AFTER B - A AND B (occur in the same bloc)
Number 2 event combination
Plug-in synchronization FFTx result available Occurs at each new spectra availability - Occurs at end of linear averaging (repeat and linear) - One event per active FFT plug in (FFT1 to FFT4)
OCT result available Occurs at each new 1/n spectra availability - Occur at end of linear averaging (leq, short leq, linear repeat)
SOAx result available Occurs at each new spectra availability - Occurs at end of linear averaging (repeat and linear) - One event per active SOA plug in (up to 2)
TDA result available Occurs at each new level set ( RMS, DC, etc) availability - Occur at end of linear averaging (repeat and linear)
OVA result available Occurs at each new set of overall level (short LEQ) availability - Occur at end of short averaging (repeat and linear)
Generators Synchronization Event synchronized with blocks of: Multi-sine - random noise - chirps
Swept sine generator event: Swept sine stabilized (output amplitude established) - step sine stabilized (occur x sec after step frequency is reached) - pure tone stabilized (output amplitude established)

Output signals

NVGate® proposes a set of generator signals.

Output signals

The following output signals are available as standard and can be independently applied to the front-end outputs.

Sine Type Pure tone - Swept sine - Step sine
Frequency 100 µHz to 40 kHz - smooth variation between step - resolution 10 µHz
Frequency control Start & stop frequencies - pause/release during sweep - adjustment during pause - step
Cycle control One shot - One cycle - continuous sweep between boundaries
Amplitude control Settling time 100 µs to 10 s - Stabilization time 0 s to 1000 s
Sweep speed Linear: 300 mHz/s to 20 kHz/s - Log: 5 mOct/s to 330 Oct/s
Step control Synchronized with analysis end - free run
Gain control
Phase control Offset ±360° - Independent for each output - phase variation controlled by phase speed (1.5°/s to 360°/s)
Number 2 to 6 sine generated simultaneously - with independent phases and amplitudes - synchronized frequencies
Multi-sine Frequency span From 125 mHz to 40 kHz
Amplitude 0 to 2.5 Vrms - Settling time 100 µs to 10 s
Resolution 125 mHz to 400 Hz - 101 lines to 6401 lines
Phase Fixed (all sine have same phase) - Random
Burst 0 to 100% - Step 1%
Number 2 to 4 independents multi-sine

Output signals (continued)

Random noise Frequency span From 125 mHz to 40 kHz - independent lower and upper frequencies
Amplitude 0 to 2.5 Vrms - Settling time 100 µs to 10 s
Type White - pink
Burst 0 to 100% - Step 1% - Bloc duration: 2.5 ms to 100 s
Number 2 to 6 independent and uncorrelated random noises
Chirp Frequency span From 3.125 Hz to 40 kHz - independent lower and upper frequencies
Amplitude 0 to 7.07 Vrms - Settling time 100 µs to 10 s
Size 256 - 512 - 1024 - 2048 - 4096 - 8192 - 16384
Burst 0 to 100% - Step 1%
Number 2 to 6 independents chirps

In addition to these predefined signal, any recorder or imported signal file can be generated synchronously on front-end the outputs.

Filters

The following filters can be applied on front-end inputs, plug-ins analyzers and output generators.

ABW: the analysis bandwidth, of the plug-in or front end where the filter is in use.

High pass Type Butterworth – IIR type
Order 1 to 6
Cutoff frequency ABWto ABW/1024
Low pass Type Butterworth – IIR type
Order 1 to 6
Cutoff frequency ABW to ABW /102.4
Stop/pass band Type Butterworth – IIR type
Order 2 to 10
Cutoff frequencies ABW to 0.055 * ABW
Bandwidth ABW/2 to 0.0075 * ABW
Integrators Type HP, Single or double
High pass frequency ABW /10 000 or ABW /2 000
Integration time 2 ms to 500 s
Differentiators Type Single
Average Sliding - 0 to 2 s
Weightings Laws A & C laws
Bandwidths 10 kHz to 40 kHz
All By-Pass Apply/bypass without stabilisation time - independently on each filter
Application Same filters set for each plug-in
Label Each filter features a user define name


Notes

The above specifications describe all the guaranteed capacities and performances of the NVGate V12 or upper. Functionalities may change depending on operation mode (connected to a 3-Series unit or office). Plug-in analyzers, options and channel number availability depend on purchased options.

The instruments hardware are described separately in the "Instruments specifications page" OROS reserves the right to modify the specifications without notification.