Difference between revisions of "NVGate Specification"

From OROS Wiki
Jump to navigation Jump to search
(Created page with "==General description== The following specifications concern NVGate<sup>®</sup> software for OR34, OR35, OR36, OR38 and Mobi-Pack<sup>®</sup> multi-analyzer instruments. The...")
 
 
(61 intermediate revisions by 3 users not shown)
Line 1: Line 1:
==General description==
[[category: specification]]
The following specifications concern NVGate<sup>®</sup> software for OR34, OR35, OR36, OR38 and Mobi-Pack<sup>®</sup> multi-analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the instrument connected to the PC. This specification applies for NVGate version 9.00 and Teamwork instruments. The detailed features of analyzer hardware and metrological information are available in ''OROS 3-Series multi-analyzer/recorder instruments specifications ''documents.
The following official specifications concern NVGate<sup>®</sup> software for OR10, OR34, OR35, OR36, OR38 and Mobi-Pack<sup>®</sup> multi-analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the instrument connected to the PC. This specification applies for NVGate version 12 or upper and Teamwork instruments.


===Modules===
====Modules====
NVGate<sup>®</sup> software is based on a general-purpose platform where optional analysis modules called plug-in analyzers are added.
NVGate<sup>®</sup> software is based on a general-purpose platform where optional analysis modules called plug-in analyzers are added.


The FFT plug-in features specific add-on for specialized analysis.
The FFT plug-in features specific add-on for specialized analysis.


<font color="#4DA836">'''NVGate'''<sup>'''®</sup> ''''''Platform'''</font>
===NVGate'''<sup>'''®</sup>Platform===


NVGate<sup>®</sup> platform provides a comprehensive set of tools for noise and vibration acquisition, recording and analysis.
NVGate<sup>®</sup> platform provides a comprehensive set of tools for noise and vibration acquisition, recording and analysis.
Line 19: Line 19:


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Signal sources'''</font>
|align = "center" rowspan = "3"|'''Signal sources'''
|
|
=====Front-end =====
'''Front-end '''
|<font size = "1">Controls dynamic inputs, generators, external synchronization/tach. input and parametric (DC) inputs.</font>
|Controls dynamic inputs, generators, external synchronization/tach. input and parametric (DC) inputs.


|-
|-
|
|
=====Recorder=====
'''Recorder'''
|<font size = "1">Record and store time domain signal on analyzer or PC hard disc. Records dynamic, parametric (DC) and ext. synch/tach inputs with multiple frequencies.</font>
|Record and store time domain signal on analyzer or PC hard disc. Records dynamic, parametric (DC) and ext. synch/tach inputs with multiple frequencies.


|-
|-
|
|
=====Player=====
'''Player'''
|<font size = "1">Visualize, listen, playback on output or generate signals for post-analysis purpose. Based on recorded or imported time domain signal file.</font>
|Visualize, listen, playback on output or generate signals for post-analysis purpose. Based on recorded or imported time domain signal file.


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Shared resources'''</font>
|align = "center" rowspan = "4"|'''Shared resources'''
|
|
=====Filters=====
'''Filters'''
|<font size = "1">High/low pass, Band/Stop pass, single and double integrator, differentiator, A and C acoustic laws</font>
|High/low pass, Band/Stop pass, single and double integrator, differentiator, A and C acoustic laws


|-
|-
|
|
=====Events=====
'''Events'''
|<font size = "1">Edge detection, DC levels, Delta DC levels, RPM, Delta RPM, Manual, Time period, combinations.</font>
|Edge detection, DC levels, Delta DC levels, RPM, Delta RPM, Manual, Time period, combinations.


|-
|-
|
|
=====Weighting windows=====
'''Weighting windows'''
|<font size = "1">Programmable Force and response, Hanning, Hamming, Kaiser Bessel, flat-top, uniform. </font>
|Programmable Force and response, Hanning, Hamming, Kaiser Bessel, flat-top, uniform.  


|-
|-
|
|
=====Tachometers=====
'''Tachometers'''
|<font size = "1">Compute RPM from ext. sync and input pulses, DC, CAN. Fractional, simulated and combined tach.</font>
|Compute RPM from ext. sync and input pulses, DC, CAN. Fractional, simulated and combined tach.


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Operational tools'''</font>
|align = "center" rowspan = "4"|'''Operational tools'''
|
|
=====Transducers management =====
'''Transducers management '''
|<font size = "1">Automatic calibration, database management, calibration history, automatic setup of inputs, TEDS management.</font>
|Automatic calibration, database management, calibration history, automatic setup of inputs, TEDS management.


|-
|-
|
|
=====Masks=====
'''Masks'''
|<font size = "1">Mask editor for spectra, profiles, 1/n octave, order spectra</font>
|Mask editor for spectra, profiles, 1/n octave, order spectra


|-
|-
|
|
=====Report=====
'''Report'''
|<font size = "1">On-line report editor, automated report generation, word or Excel files.</font>
|On-line report editor, automated report generation, word or Excel files.


|-
|-
|<font size = "1">'''Automation''' '''tools'''</font>
|'''Automation''' '''tools'''
|<font size = "1">Macros, sequence from Excel<sup>®</sup> sheet, alarms based on mask comparison, customizable control panel, restricted user profiles, remote controller interface.</font>
|Macros, sequence from Excel<sup>®</sup> sheet, alarms based on mask comparison, customizable control panel, restricted user profiles, remote controller interface.


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''General purpose analysis'''</font>
|align = "center" rowspan = "2"|'''General purpose analysis'''
|
|
=====Monitor=====
'''Monitor'''
|<font size = "1">4 channels FFT analyzer with input hot swap. Statistical extraction (RMS, Max, Min, Kurtosis) from time domain with programmable band-pass filter. </font>
|4 channels FFT analyzer with input hot swap. Statistical extraction (RMS, Max, Min, Kurtosis) from time domain with programmable band-pass filter.  


|-
|-
|
|
=====Waterfall=====
'''Waterfall'''
|<font size = "1">Synchronized stack for 2D (spectra) and scalar (order, overall). 3D, color map and profiles displays. Extraction capabilities.</font>
|Synchronized stack for 2D (spectra) and scalar (order, overall). 3D, color map and profiles displays. Extraction capabilities.


|}
|}


<font color="#4DA836">'''NVGate'''<sup>'''®</sup> ''''''Plug-ins analyzers'''</font>
'''NVGate'''<sup>'''®</sup> Plug-ins analyzers'''


In addition to the standard functions featured in the software platform, NVGate<sup>®</sup> can receive additional plug-in analyzers that offer flexible configuration of independent analysis modes.  
In addition to the standard functions featured in the software platform, NVGate<sup>®</sup> can receive additional plug-in analyzers that offer flexible configuration of independent analysis modes.  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center"|<font size = "1">'''FFT'''</font>
|align = "center"|'''FFT'''
|<font size = "1">'''Fast Fourier Transform''' - Time to Frequency conversion with multiple average domains, combinable trigger, zooms and cross functions. </font>
|'''Fast Fourier Transform''' - Time to Frequency conversion with multiple average domains, combinable trigger, zooms and cross functions.  
|<font size = "1">General-purpose frequency analysis, machinery signature, modal analysis acquisition (hammer or shaker), machinery diagnostic.</font>
|General-purpose frequency analysis, machinery signature, modal analysis acquisition (hammer or shaker), machinery diagnostic.


|-
|-
|align = "center"|<font size = "1">'''FFT Add-ons'''</font>
|align = "center"|'''FFT Add-ons'''
|&nbsp;
|&nbsp;
|&nbsp;
|&nbsp;


|-
|-
|align = "center"|<font size = "1">'''CBT'''</font>
|align = "center"|'''CBT'''
|<font size = "1">'''Constant Band Tracking''' - Order extraction at constant bandwidth</font>
|'''Constant Band Tracking''' - Order extraction at constant bandwidth
|<font size = "1">Gears mesh analysis, gearboxes noise and vibration tracking.</font>
|Gears mesh analysis, gearboxes noise and vibration tracking.


|-
|-
|align = "center"|<font size = "1">'''FFTDiag'''</font>
|align = "center"|'''FFTDiag'''
|<font size = "1">'''Cepstrum, Auto '''and cross '''correlation''', '''DC''', '''Min/Max, Pk, pk/pk '''and '''crest factor'''.</font>
|'''Cepstrum, Auto '''and cross '''correlation''', '''DC''', '''Min/Max, Pk, pk/pk '''and '''crest factor'''.
|<font size = "1">Set of functions dedicated to the diagnostics of industrial machineries.</font>
|Set of functions dedicated to the diagnostics of industrial machineries.


|-
|-
|align = "center"|<font size = "1">'''TDA'''</font>
|align = "center"|'''TDA'''
|<font size = "1">'''Time Domain Analysis''' –Time view (oscilloscope). True statistical extraction, averaging.</font>
|'''Time Domain Analysis''' –Time view (oscilloscope). True statistical extraction, averaging.
|<font size = "1">Machinery diagnostic, long term acquisition monitoring, critical acquisition control, large structure damping control, shocks determination, machinery health control, trend analysis.</font>
|Machinery diagnostic, long term acquisition monitoring, critical acquisition control, large structure damping control, shocks determination, machinery health control, trend analysis.


|-
|-
|align = "center"|<font size = "1">'''SOA'''</font>
|align = "center"|'''SOA'''
|<font size = "1">'''Synchronous Order Analysis''' - Time to Order re-sampling with angle or order averaging, acoustic weightings, multiple tach management.</font>
|'''Synchronous Order Analysis''' - Time to Order re-sampling with angle or order averaging, acoustic weightings, multiple tach management.
|<font size = "1">Rotating machinery balancing. In vehicle order to noise correlation R&D. Pump, compressors, turbines and speed reducer/multiplier test. </font>
|Rotating machinery balancing. In vehicle order to noise correlation R&D. Pump, compressors, turbines and speed reducer/multiplier test.  


|-
|-
|align = "center"|<font size = "1">'''SOADiag <br>(SOA Add-on)'''</font>
|align = "center"|'''SOADiag <br>(SOA Add-on)'''
|<font size = "1">'''Copstrum™, '''Auto''' '''and cross '''correlation''', Revolution '''synchronous statistics''', Order transmission Function ('''ORF™'''), '''X''' functions</font>
|'''Copstrum™, '''Auto''' '''and cross '''correlation''', Revolution '''synchronous statistics''', Order transmission Function ('''ORF™'''), '''X''' functions
|<font size = "1">Set of diagnostic functions based on the synchronous order analysis for rotating part transmission and forced vibration extraction.</font>
|Set of diagnostic functions based on the synchronous order analysis for rotating part transmission and forced vibration extraction.


|-
|-
|align = "center"|<font size = "1">'''OCT'''</font>
|align = "center"|'''OCT'''
|<font size = "1">'''1/n Octave Constant Percentage Band filters''' - 1/n octave detectors set with multiple averaging modes.</font>
|'''1/n Octave Constant Percentage Band filters''' - 1/n octave detectors set with multiple averaging modes.
|<font size = "1">General-purpose industrial noise analysis. Vehicle acoustics R&D. Noise test.</font>
|General-purpose industrial noise analysis. Vehicle acoustics R&D. Noise test.


|-
|-
|align = "center"|<font size = "1">'''OVA'''</font>
|align = "center"|'''OVA'''
|<font size = "1">'''Overall Acoustics Detectors''' - 3 detectors <nowiki>+</nowiki> 1 peak detection per channel with parallel averaging and acoustic weightings.</font>
|'''Overall Acoustics Detectors''' - 3 detectors <nowiki>+</nowiki> 1 peak detection per channel with parallel averaging and acoustic weightings.
|<font size = "1">Industrial acoustic, sound power measurements. Machinery with cycles, R&D and test.</font>
|Industrial acoustic, sound power measurements. Machinery with cycles, R&D and test.


|}
|}


<font color="#4DA836">'''NVGate'''<sup>'''®</sup> ''''''Options'''</font>
===NVGate<sup>®</sup> Options===


Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software features options:  
Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software also features options:  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" bgcolor = "#D9D9D9"|<font size = "1">'''IVC'''</font>
|align = "center" bgcolor = "#D9D9D9"|'''IVC'''
|bgcolor = "#D9D9D9"|<font size = "1">Instantaneous angular velocity converter </font>
|bgcolor = "#D9D9D9"|Instantaneous angular velocity converter  
|bgcolor = "#D9D9D9"|<font size = "1">Torsional measurements from Ext. Synch inputs</font>
|bgcolor = "#D9D9D9"|Torsional measurements from Ext. Synch inputs


|-
|-
|align = "center"|<font size = "1">'''CTE'''</font>
|align = "center"|'''CTE'''
|<font size = "1">Combined Tachometer editor</font>
|Combined Tachometer editor
|<font size = "1">Tachometer math editor - 2 inputs, 1 output</font>
|Tachometer math editor - 2 inputs, 1 output


|-
|-
|align = "center" bgcolor = "#D9D9D9"|<font size = "1">'''VIN'''</font>
|align = "center" bgcolor = "#D9D9D9"|'''VIN'''
|bgcolor = "#D9D9D9"|<font size = "1">Virtual Inputs</font>
|bgcolor = "#D9D9D9"|Virtual Inputs
|bgcolor = "#D9D9D9"|<font size = "1">Real-time math combination of dynamic inputs</font>
|bgcolor = "#D9D9D9"|Real-time math combination of dynamic inputs


|-
|-
|align = "center"|<font size = "1">'''VDC'''</font>
|align = "center"|'''VDC'''
|<font size = "1">Virtual DC</font>
|Virtual DC
|<font size = "1">Real-time math combination of parametric inputs</font>
|Real-time math combination of parametric inputs


|-
|-
|align = "center"|<font size = "1">'''A-Samp'''</font>
|align = "center"|'''A-Samp'''
|<font size = "1">Angular sampling for SOA</font>
|Angular sampling for SOA
|<font size = "1">Real-time Angular sampling of inputs</font>
|Real-time Angular sampling of inputs


|}
|}
<font color="#4DA836">'''PC requirements'''</font>
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" bgcolor = "#D9D9D9"|<font size = "1">'''Minimum'''</font>
|bgcolor = "#D9D9D9"|<font size = "1">1 GB<ref>) Waterfall depth depends on available memory. </ref> of RAM / 250 MB free on HD <nowiki>+</nowiki> storage for measurements and signals / 1024 x 768 display </font>
|-
|align = "center"|<font size = "1">'''Recommended'''</font><br><font size = "1">'''(for laptop)'''</font>
|<font size = "1">'''Dual/quad core processor''' (e.g.: Intel Core i5) / <nowiki>></nowiki> 2.5 GHz / '''4 GB''' of RAM / GPU / 1368 x 768 display / 1 GB free on HD <nowiki>+</nowiki> storage for signals </font>
|-
|align = "center" bgcolor = "#D9D9D9"|<font size = "1">'''Recommended'''</font><br><font size = "1">'''(for desktop)'''</font>
|bgcolor = "#D9D9D9"|<font size = "1">'''Quad core processor''' (e.g.: Intel Core i7) / '''6 GB''' of RAM / GPU / '''1920 x 1080 display''' / 1 GB free on HD <nowiki>+</nowiki> storage for signals </font>
|-
|align = "center"|<font size = "1">'''Connections'''</font>
|<font size = "1">Type: '''Ethernet 100/1000 BASE-T''', 1 Gb/s : Connector: '''RJ45'''</font><br><font size = "1">For removable disk: '''USB 3.0''' / For dongle key: '''USB 2.0'''</font>
|-
|align = "center" bgcolor = "#D9D9D9"|<font size = "1">'''Operating systems'''</font>
|bgcolor = "#D9D9D9"|<font size = "1">Windows '''Vista '''Service Pack 2 / Windows''' 7 '''/ Windows''' 8''' & '''8.1 / MS Office 32 bits''' only</font>
|}
1) <font size = "1">Waterfall depth depends on available memory. Minimum configuration does not allow waterfall storage.</font>


==Signal sources==
==Signal sources==
Line 202: Line 176:
The front-end module gathers the different input and output settings available in the instrument unit. In addition, the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to the PC) which is useful for set-up.
The front-end module gathers the different input and output settings available in the instrument unit. In addition, the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to the PC) which is useful for set-up.


<font color="#4DA836">'''Front-end settings'''</font>
'''Front-end settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|rowspan = "10"|<font size = "1">'''Dynamic inputs'''</font>
|rowspan = "10"|'''Dynamic inputs'''
|<font size = "1">Sampling clock</font>
|Sampling clock
|<font size = "1">'''102.4 kS/s to 2,048 S/s''' or '''65.536 kS/s to 3,200 S/s'''</font>
|'''102.4 kS/s to 2,048 S/s''' or '''65.536 kS/s to 3,200 S/s'''


|-
|-
|<font size = "1">LEMO<ref>Optional features</ref> connectors management</font>
|LEMO<sup>2</sup> connectors management
|<font size = "1">'''200 V '''polarization''' on/off''' per block of 8 input</font>
|'''200 V '''polarization''' on/off''' per block of 8 input


|-
|-
|$$DeleteCell$$$$DeleteCell$$
|Coupling<sup>3,4</sup>
|<font size = "1">Coupling<ref>Independent for each input</ref><sup>,4</sup></font>
|'''AC''' (0.35 Hz) - '''DC''' - '''ICP''' - '''ICP <nowiki>+</nowiki> TEDS''' - '''AC float -''' '''DC float''' - '''Grounded'''
|<font size = "1">'''AC''' (0.35 Hz) - '''DC''' - '''ICP''' - '''ICP <nowiki>+</nowiki> TEDS''' - '''AC float -''' '''DC float''' - '''Grounded'''</font>


|-
|-
|<font size = "1">Node information<sup>3</sup></font>
|Node information<sup>3</sup>
|<font size = "1">'''Label''' - '''component''' - '''node''' - '''direction''' (± X ,Y,Z) - '''type''' (translation, rotation)</font>
|'''Label''' - '''component''' - '''node''' - '''direction''' (± X ,Y,Z) - '''type''' (translation, rotation)


|-
|-
|<font size = "1">Physical quantity<sup>3,4</sup></font>
|Physical quantity<sup>3,4</sup>
|<font size = "1">Any physical quantity can independently be associated to inputs.</font>
|Any physical quantity can independently be associated to inputs.


|-
|-
|$$DeleteCell$$$$DeleteCell$$
|Sensitivity<sup>3,4</sup>
|<font size = "1">Sensitivity<sup>3,<ref>Linked with the transducer database</ref></sup></font>
|User defined in '''V/unit.'''
|<font size = "1">User defined in '''V/unit.'''</font>


|-
|-
|<font size = "1">Range<sup>2</sup></font>
|Range<sup>2</sup>
|<font size = "1">'''Pk to pk''' expressed in the associated unit, (taking in account sensitivity) - '''linear '''or '''dB'''</font>
|'''Pk to pk''' expressed in the associated unit, (taking in account sensitivity) - '''linear '''or '''dB'''


|-
|-
|<font size = "1">External conditioner compensation<sup>3</sup></font>
|External conditioner compensation<sup>3</sup>
|<font size = "1">'''Gain''' - '''polarity''' - '''offset'''.</font>
|'''Gain''' - '''polarity''' - '''offset'''.


|-
|-


||<font size = "1">Filter<sup>3</sup></font>
||Filter<sup>3</sup>
|<font size = "1">'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)</font>
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)


|-
|-


||<font size = "1">Auto-range</font>
||Auto-range
|<font size = "1">Enable/disable auto-ranging independently on any input</font>
|Enable/disable auto-ranging independently on any input


|-
|-
|rowspan = "13"|<font size = "1">'''Ext. sync inputs'''</font>
|rowspan = "13"|'''Ext. sync inputs'''
|<font size = "1">Sampling</font>
|Sampling
|<font size = "1">over-sampled 64 time the dynamic inputs sampling</font>
|over-sampled 64 time the dynamic inputs sampling


|-
|-
|<font size = "1">Coupling<sup>3</sup></font>
|Coupling<sup>3</sup>
|<font size = "1">'''AC''' (0.35 Hz) - '''DC'''</font>
|'''AC''' (0.35 Hz) - '''DC'''


|-
|-
|<font size = "1">Information<sup>3</sup></font>
|Information<sup>3</sup>
|<font size = "1">Label</font>
|Label


|-
|-
|<font size = "1">Physical quantity<sup>3,4</sup></font>
|Physical quantity<sup>3,4</sup>
|<font size = "1">Any physical quantity can be associated to inputs.</font>
|Any physical quantity can be associated to inputs.


|-
|-
|<font size = "1">Sensitivity<sup>3,4</sup></font>
|Sensitivity<sup>3,4</sup>
|<font size = "1">User defined in '''V/unit'''.</font>
|User defined in '''V/unit'''.


|-
|-
|<font size = "1">Range<sup>3</sup></font>
|Range<sup>3</sup>
|<font size = "1">'''Pk to pk''' expressed in the associated unit, (taking in account sensitivity) -  '''linear or dB'''</font>
|'''Pk to pk''' expressed in the associated unit, (taking in account sensitivity) -  '''linear or dB'''


|-
|-
|<font size = "1">External conditioner compensation<sup>3</sup></font>
|External conditioner compensation<sup>3</sup>
|<font size = "1">'''Gain''', '''offset'''.</font>
|'''Gain''', '''offset'''.


|-
|-
|<font size = "1">Edge detection<sup>3</sup></font>
|Edge detection<sup>3</sup>
|<font size = "1">'''Threshold''' (in associated unit) - '''Edge''' (rise, fall) - '''Hystersis''' - '''hold off''' (sec)</font>
|'''Threshold''' (in associated unit) - '''Edge''' (rise, fall) - '''Hystersis''' - '''hold off''' (sec)


|-
|-
|<font size = "1">Pre-divider<sup>3</sup></font>
|Pre-divider<sup>3</sup>
|<font size = "1">'''Hardware - 1''' to '''255 '''pulses – accept up to '''375 kpulses/sec'''</font>
|'''Hardware - 1''' to '''255 '''pulses – accept up to '''375 kpulses/sec'''


|-
|-
|<font size = "1">Post-multiplier<sup>3</sup></font>
|Post-multiplier<sup>3</sup>
|<font size = "1">'''1 to 50 pulses/trigger '''- Generate higher frequencies on low rate triggers. (ex. GPS)</font>
|'''1 to 50 pulses/trigger '''- Generate higher frequencies on low rate triggers. (ex. GPS)


|-
|-
|<font size = "1">Tachometer mode<sup>3</sup></font>
|Tachometer mode<sup>3</sup>
|<font size = "1">'''Pulse per rev''': 0.5 to 4096 - '''average''' (n rev) - '''min''' (detect stopped shaft) and '''max RPM''' speed - '''hold off '''(in % of revolution) - '''rotation''' (clockwise/counterclockwise)</font>
|'''Pulse per rev''': 0.5 to 4096 - '''average''' (n rev) - '''min''' (detect stopped shaft) and '''max RPM''' speed - '''hold off '''(in % of revolution) - '''rotation''' (clockwise/counterclockwise)


|-
|-
|<font size = "1">Torsional mode<sup>2</sup>: Instantaneous angular velocity converter </font>
|Torsional mode<sup>2</sup>: Instantaneous angular velocity converter  
|<font size = "1">'''Pulse per rev''': 1 to 4096 - '''min''' (detect stopped shaft) and '''max RPM''' speed - '''sliding''' '''average''' (1 to 20 samples) - '''Missing pulse number''': 0 to 5, '''Missing pulses detection''' :1 to ±20%:  of previous pulses duration- '''Filters''' (dt, dt², 1/dt)</font>
|'''Pulse per rev''': 1 to 4096 - '''min''' (detect stopped shaft) and '''max RPM''' speed - '''sliding''' '''average''' (1 to 20 samples) - '''Missing pulse number''': 0 to 5, '''Missing pulses detection''' :1 to ±20%:  of previous pulses duration- '''Filters''' (dt, dt², 1/dt)


|-
|-
|<font size = "1">Angular sampling<sup>2</sup>: multi-pulse/rev clock used for signal sampling </font>
|Angular sampling<sup>2</sup>: multi-pulse/rev clock used for signal sampling  
|<font size = "1">for '''SOA''' -  '''Pulse per rev''': 1 to 4096 continuously (no power of 2 limitation) - Real time, in line '''anti-aliasing''' - '''Missing pulse number''': 0 to 5, '''Missing pulses detection''': 1 to ±20%  of previous pulses duration – '''Phase reference''': Any tachometer including same input</font>
|for '''SOA''' -  '''Pulse per rev''': 1 to 4096 continuously (no power of 2 limitation) - Real time, in line '''anti-aliasing''' - '''Missing pulse number''': 0 to 5, '''Missing pulses detection''': 1 to ±20%  of previous pulses duration – '''Phase reference''': Any tachometer including same input


|}
|}


----


<font color="#4DA836">'''Front-end (continued)'''</font>
 


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "10"|<font size = "1">'''Outputs'''</font>
|align = "center" rowspan = "10"|'''Outputs'''
|<font size = "1">Sampling</font>
|Sampling
|<font size = "1">Same as dynamic inputs</font>
|Same as dynamic inputs


|-
|-
|<font size = "1">Generated signals<sup>3</sup></font>
|Generated signals<sup>3</sup>
|<font size = "1">Any '''outputs signals''', see §outputs - play-back of signal file tracks - DC levels - Inputs playback (Delay <nowiki>></nowiki> 256 samples)</font>
|Any '''outputs signals''', see §outputs - play-back of signal file tracks - DC levels - Inputs playback (Delay <nowiki>></nowiki> 256 samples)


|-
|-
|<font size = "1">Filters<sup>3</sup></font>
|Filters<sup>3</sup>
|<font size = "1">'''Any NVGate''' filter can independently be applied to each dynamic output (See Filters §)</font>
|'''Any NVGate''' filter can independently be applied to each dynamic output (See Filters §)


|-
|-
|<font size = "1">Impedance<sup>3</sup></font>
|Impedance<sup>3</sup>
|<font size = "1">'''50 ''', '''600 ''' (only OR36 and 0R38) or grounded</font>
|'''50 ohm''', '''600 ohm''' (only OR36 and 0R38) or grounded


|-
|-
|<font size = "1">Gain<sup>3</sup></font>
|Gain<sup>3</sup>
|<font size = "1">'''-100 dB''' to '''<nowiki>+</nowiki> 120 dB'''</font>
|'''-100 dB''' to '''<nowiki>+</nowiki> 120 dB'''


|-
|-
|<font size = "1">Synchronization<sup>3</sup></font>
|Synchronization<sup>3</sup>
|<font size = "1">'''Free run '''or '''linked '''to acquisitions '''runs'''</font>
|'''Free run '''or '''linked '''to acquisitions '''runs'''


|-
|-
|<font size = "1">Level and phase transition<sup>3</sup></font>
|Level and phase transition<sup>3</sup>
|<font size = "1">No ('''Steps''') - controlled ('''ramp''') - user selectable ramp time</font>
|No ('''Steps''') - controlled ('''ramp''') - user selectable ramp time


|-
|-
|<font size = "1">Clipping</font>
|Clipping
|<font size = "1">User selectable - '''protects shaker''' and '''amplifier'''.</font>
|User selectable - '''protects shaker''' and '''amplifier'''.


|-
|-
|<font size = "1">Events (can trig analysis or record)<sup>3</sup></font>
|Events (can trig analysis or record)<sup>3</sup>
|<font size = "1">Outputs '''stabilized''' - new '''step reached''' - '''start & stop sweep'''</font>
|Outputs '''stabilized''' - new '''step reached''' - '''start & stop sweep'''


|-
|-
|<font size = "1">Controls</font>
|Controls
|<font size = "1">'''Emergency stop''' - '''mute'''/un-mute all</font>
|'''Emergency stop''' - '''mute'''/un-mute all


|-
|-
|align = "center" rowspan = "7"|<font size = "1">'''Parametric'''<sup>'''1</sup> ''''''(DC) inputs'''</font>
|align = "center" rowspan = "7"|'''Parametric<sup>1</sup>(DC) inputs
|<font size = "1">Sampling</font>
|Sampling
|<font size = "1">'''12.3 to 12,8 S/s - '''10 Hz to 100 Hz harmonics rejection</font>
|'''12.3 to 12,8 S/s - '''10 Hz to 100 Hz harmonics rejection


|-
|-
|<font size = "1">Information<sup>3</sup></font>
|Information<sup>3</sup>
|<font size = "1">Label</font>
|Label


|-
|-
|<font size = "1">Physical quantity<sup>3,4</sup></font>
|Physical quantity<sup>3,4</sup>
|<font size = "1">Any physical quantity can be associated to inputs.</font>
|Any physical quantity can be associated to inputs.


|-
|-
|<font size = "1">Sensitivity<sup>3,4</sup></font>
|Sensitivity<sup>3,4</sup>
|<font size = "1">User defined in V/unit.</font>
|User defined in V/unit.


|-
|-
|<font size = "1">Range<sup>3</sup></font>
|Range<sup>3</sup>
|<font size = "1">Pk to pk expressed in the associated unit (taking in account sensitivity)  - '''linear''' or '''dB'''</font>
|Pk to pk expressed in the associated unit (taking in account sensitivity)  - '''linear''' or '''dB'''


|-
|-
|<font size = "1">External conditioner compensation<sup>3</sup></font>
|External conditioner compensation<sup>3</sup>
|<font size = "1">'''Gain '''(-120 dB  to <nowiki>+</nowiki>20 dB)''' ''' - '''polarity''' - '''offset '''(<nowiki><</nowiki> range).</font>
|'''Gain '''(-120 dB  to <nowiki>+</nowiki>20 dB)''' ''' - '''polarity''' - '''offset '''(<nowiki><</nowiki> range).


|-
|-
|<font size = "1">Auto-range</font>
|Auto-range
|<font size = "1">'''Enable/disable''' auto-ranging '''independently on any input'''</font>
|'''Enable/disable''' auto-ranging '''independently on any input'''


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Auto-ranging'''</font>
|align = "center" rowspan = "5"|'''Auto-ranging'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Normal''' - on '''Peak''' detection </font>
|'''Normal''' - on '''Peak''' detection  


|-
|-
|<font size = "1">Margin</font>
|Margin
|<font size = "1">Select the first highest range with '''0 dB - 3 dB''' or '''20 dB''' margin</font>
|Select the first highest range with '''0 dB - 3 dB''' or '''20 dB''' margin


|-
|-
|<font size = "1">Duration (apply for normal auto-range)</font>
|Duration (apply for normal auto-range)
|<font size = "1">'''0''' to '''10 sec'''.</font>
|'''0''' to '''10 sec'''.


|-
|-
|<font size = "1">Peak detection</font>
|Peak detection
|<font size = "1">User select input were to detect peak, auto-range applies on all enabled input at each peak detection. </font>
|User select input were to detect peak, auto-range applies on all enabled input at each peak detection.  


|-
|-
|<font size = "1">Peak parameters settings</font>
|Peak parameters settings
|<font size = "1">''' 1 to 10 '''peaks - '''sensitivity''' (low, normal, high) - '''rise''', '''fall''' or '''any edge'''.</font>
|''' 1 to 10 '''peaks - '''sensitivity''' (low, normal, high) - '''rise''', '''fall''' or '''any edge'''.


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Checking'''</font>
|align = "center" rowspan = "2"|'''Checking'''
|<font size = "1">ICP<ref>For OR36, OR38 and Mobi-Pack only</ref></font>
|ICP ''(check not available on OR34)''
|<font size = "1">Test and report ('''Open''', '''shortcut''', '''ok''') all enabled inputs.</font>
|Test and report ('''Open''', '''shortcut''', '''ok''') all enabled inputs.


|-
|-
|<font size = "1">TEDS </font>
|TEDS  
|<font size = "1">Transducers automatic recognition - Complies with '''IEEE 1451.4 2004  Rev. 1.0'''</font>
|Transducers automatic recognition - Complies with '''IEEE 1451.4 2004  Rev. 1.0'''


|}
|}
2: Optional features<br>


<font color="#4DA836">'''Front-end results &''' connections''''''</font>
3: Independent for each input<br>
 
4: Linked with the transducer database
 
===Front-end results & connections===


The following results are available for monitoring and connection to analysis mode (plug-in analyzers)
The following results are available for monitoring and connection to analysis mode (plug-in analyzers)


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Dynamic inputs & torsional'''<sup>'''1</sup> ''''''inputs'''</font>
|align = "center" rowspan = "3"|'''Dynamic inputs & torsional<sup>1</sup> inputs'''
|<font size = "1">Monitoring<sup>3</sup></font>
|Monitoring<sup>3</sup>
|<font size = "1">Time domain instantaneous signal - Size '''256 samples'''</font>
|Time domain instantaneous signal - Size '''256 samples'''


|-
|-
|<font size = "1">Status led<sup>3</sup> (docking tool bar)</font>
|Status led<sup>3</sup> (docking tool bar)
|<font size = "1">'''Overloaded''' = red - '''overload''' '''occurred since last start''' = red with yellow center - '''ok''' = green - '''under load''' (20 dB below range) = green with yellow center</font>
|'''Overloaded''' = red - '''overload''' '''occurred since last start''' = red with yellow center - '''ok''' = green - '''under load''' (20 dB below range) = green with yellow center


|-
|-
|<font size = "1">Connection<sup>3</sup></font>
|Connection<sup>3</sup>
|<font size = "1">To any: '''plug-in analyzer''' '''channel, monitor''' '''channel, Virtual inputs''', '''recorder''' '''track''', '''edge event detector''' & '''tachometer '''</font>
|To any: '''plug-in analyzer''' '''channel, monitor''' '''channel, Virtual inputs''', '''recorder''' '''track''', '''edge event detector''' & '''tachometer '''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Ext. Sync inputs '''</font>
|align = "center" rowspan = "2"|'''Ext. Sync inputs '''
|<font size = "1">Monitoring<sup>3</sup></font>
|Monitoring<sup>3</sup>
|<font size = "1">Time domain '''instantaneous status''' (1 = threshold crossed, 0 = no event) - Size '''256 samples'''.</font>
|Time domain '''instantaneous status''' (1 = threshold crossed, 0 = no event) - Size '''256 samples'''.


|-
|-
|<font size = "1">Connection<sup>3</sup></font>
|Connection<sup>3</sup>
|<font size = "1">To any: '''start '''or''' stop averaging''' of plug-in analyzers, recorder '''start '''or''' stop recording''', '''waterfall start '''or''' stop acquisition''', '''new block trigger''' for FFT and SOA and '''new slice '''for''' waterfall''' acquisition.</font><br><font size = "1">To '''torsional''', '''tachometer''' & '''angular sampling'''</font>
|To any: '''start '''or''' stop averaging''' of plug-in analyzers, recorder '''start '''or''' stop recording''', '''waterfall start '''or''' stop acquisition''', '''new block trigger''' for FFT and SOA and '''new slice '''for''' waterfall''' acquisition.<br>To '''torsional''', '''tachometer''' & '''angular sampling'''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Parametric'''<sup>'''1</sup> ''''''(DC) inputs & CAN'''<sup>'''1</sup> ''''''parameters'''</font>
|align = "center" rowspan = "3"|'''Parametric<sup>1</sup> (DC) inputs & CAN<sup>1</sup>parameters'''
|<font size = "1">Monitoring<sup>3</sup></font>
|Monitoring<sup>3</sup>
|<font size = "1">'''Profiles''' versus time ('''160 ms to 163 sec''' user selectable), and '''digital'''/'''analog view meter'''.</font>
|'''Profiles''' versus time ('''160 ms to 163 sec''' user selectable), and '''digital'''/'''analog view meter'''.


|-
|-
|<font size = "1">Status led<sup>3</sup></font>
|Status led<sup>3</sup>
|<font size = "1">'''Overloaded''' = red - '''ok''' = yellow - '''under load''' (20 dB below range) = Cyan</font>
|'''Overloaded''' = red - '''ok''' = yellow - '''under load''' (20 dB below range) = Cyan


|-
|-
|<font size = "1">Connection<sup>3</sup></font>
|Connection<sup>3</sup>
|<font size = "1">To any: '''recorder track''', '''waterfall reference (Z/X axis)''' and '''profiles''', '''level''' '''delta level detectors''' and '''virtual DC'''</font>
|To any: '''recorder track''', '''waterfall reference (Z/X axis)''' and '''profiles''', '''level''' '''delta level detectors''' and '''virtual DC'''


|}
|}
Line 456: Line 432:
* Export or import time domain files.
* Export or import time domain files.
* Playback time domain files on analyzers outputs during analysis
* Playback time domain files on analyzers outputs during analysis
* Listen recorded tracks on the PC speakers.
* Listen to recorded tracks on the PC speakers.
* Post-analyze time domain files in the available plugs-in.
* Post-analyze time domain files using the available plugs-in.
 
===Recorder===
===Recorder===
The recorder module memorizes time domain signal into files located on 3-Series analyzer HD, Mobi-Disk<sup>TM</sup> or PC HD. This is available on 2 different modes:<br>
The recorder module saves time domain signal into files located on 3-Series analyzer HD, Mobi-Disk<sup>TM</sup> or PC HD. This is available on 2 different modes:<br>
* '''On-line record''', the chosen inputs (Dynamic, ext. sync, parametric) are recorded at selected sampling frequency during acquisition. Real-time analysis is available simultaneously.
* '''On-line record''', the chosen inputs (Dynamic, ext. sync, parametric) are recorded at selected sampling frequency during acquisition. Real-time analysis is available simultaneously.
* '''Time and tracks split''', the chosen recorded tracks (from signal file) are played back in the recorder using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be changed.
* '''Time and tracks split''', the chosen recorded tracks (from signal file) are played back in the recorder using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be changed.


<font color="#4DA836">'''Recorder settings'''</font>
'''Recorder settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Bandwidths'''</font>
|align = "center" rowspan = "3"|'''Bandwidths'''
|<font size = "1">Dynamics inputs</font>
|Dynamics inputs
|<font size = "1">'''2 groups''' of user selectable sampling frequency - From '''102.4 kS/s '''to '''2.048 S/s''' ('''40 kHz''' to '''800 mHz''') - available simultaneously</font>
|'''2 groups''' of user selectable sampling frequency - From '''102.4 kS/s '''to '''2.048 S/s''' ('''40 kHz''' to '''800 mHz''') - available simultaneously


|-
|-
|<font size = "1">Ext. sync inputs</font>
|Ext. sync inputs
|<font size = "1">'''Automatic selection''' at Front-end sampling frequency - Resolution is '''64 time Front-end sampling frequency'''</font>
|'''Automatic selection''' at Front-end sampling frequency - Resolution is '''64 time Front-end sampling frequency'''


|-
|-
|<font size = "1">Parametric (DC) inputs</font>
|Parametric (DC) inputs
|<font size = "1">'''12.5 S/s''' ('''50 Hz''' rejection) - '''15 S/s''' ('''60 Hz''' rejection)</font>
|'''12.5 S/s''' ('''50 Hz''' rejection) - '''15 S/s''' ('''60 Hz''' rejection)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Tracks'''</font>
|align = "center" rowspan = "2"|'''Tracks'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''14 tracks<ref>For OR34, OR35 and Office-8 licenses.</ref>''' (8 <nowiki>+</nowiki> 6 ext. sync) or '''38 tracks<ref>For OR36, OR38 and Office licenses.</ref>''' (32 <nowiki>+</nowiki> 6 ext. Sync)</font><br><font size = "1">Sum of connected licenses using multiple hardware</font>
|'''14 tracks<ref>For OR34, OR35 and Office-8 licenses.</ref>''' (8 <nowiki>+</nowiki> 6 ext. sync) or '''38 tracks<ref>For OR36, OR38 and Office licenses.</ref>''' (32 <nowiki>+</nowiki> 6 ext. Sync)<br>Sum of connected licenses using multiple hardware


|-
|-
|<font size = "1">Saved settings / track</font>
|Saved settings / track
|<font size = "1">'''Label''' - '''Coupling''' - '''External gain''' - Input '''range''' - '''Sampling frequency '''and''' signal bandwidth - Component''' - '''Node''' number - '''Direction''' - '''Type''' - Associated '''transducer''' - '''Unit -''' '''Sensitivity''' and '''Offset''' compensation.</font>
|'''Label''' - '''Coupling''' - '''External gain''' - Input '''range''' - '''Sampling frequency '''and''' signal bandwidth - Component''' - '''Node''' number - '''Direction''' - '''Type''' - Associated '''transducer''' - '''Unit -''' '''Sensitivity''' and '''Offset''' compensation.


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Modes'''</font>
|align = "center" rowspan = "3"|'''Modes'''
|<font size = "1">Start to time</font>
|Start to time
|<font size = "1">Start recording on run or any activated event - '''Stop recording when duration ends '''- Duration '''10 ms''' to '''available space on target HDD.'''</font>
|Start recording on run or any activated event - '''Stop recording when duration ends '''- Duration '''10 ms''' to '''available space on target HDD.'''


|-
|-
|<font size = "1">Start to stop</font>
|Start to stop
|<font size = "1">'''Start recording''' on '''run''' or '''any activated event''' - Stop recording on stop on any activated event - Start and stop event can be the same - Duration '''10 ms''' to '''available space on active drive '''(PC or OR3x).</font>
|'''Start recording''' on '''run''' or '''any activated event''' - Stop recording on stop on any activated event - Start and stop event can be the same - Duration '''10 ms''' to '''available space on active drive '''(PC or OR3x).


|-
|-
|<font size = "1">Time to stop</font>
|Time to stop
|<font size = "1">Memorize the '''earliest x seconds''' - Stop = '''stop''' or any '''activated event''' - Duration '''10 ms''' to '''2 GSamples '''(limited by the drive available space)</font>
|Memorize the '''earliest x seconds''' - Stop = '''stop''' or any '''activated event''' - Duration '''10 ms''' to '''2 GSamples '''(limited by the drive available space)


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Triggering'''</font>
|align = "center" rowspan = "4"|'''Triggering'''
|<font size = "1">Start recording (new record)</font>
|Start recording (new record)
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">Stop recording (end current record)</font>
|Stop recording (end current record)
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">Start delay</font>
|Start delay
|<font size = "1">Positive''' '''=''' unlimited''' - negative = '''128 kSamples'''</font>
|Positive''' '''=''' unlimited''' - negative = '''128 kSamples'''


|-
|-
|<font size = "1">Stop delay</font>
|Stop delay
|<font size = "1">Positive''' '''=''' unlimited''' - negative = '''0'''</font>
|Positive''' '''=''' unlimited''' - negative = '''0'''


|-
|-
|align = "center"|<font size = "1">'''Markers'''</font>
|align = "center"|'''Markers'''
|<font size = "1">Saved in the record files</font>
|Saved in the record files
|<font size = "1">Added by tool bar or shortcut - '''comment fields''' editable during record or at end of acquisition</font>
|Added by tool bar or shortcut - '''comment fields''' editable during record or at end of acquisition


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Signal file'''</font>
|align = "center" rowspan = "3"|'''Signal file'''
|<font size = "1">Record on PC HDD</font>
|Record on PC HDD
|<font size = "1">Up to '''10''' dynamic inputs – Up to 4 parametric inputs - max total bandwidth '''512 kS/s''' - (2 MB/s)</font>
|Up to '''10''' dynamic inputs – Up to 4 parametric inputs - max total bandwidth '''512 kS/s''' - (2 MB/s)


|-
|-
|<font size = "1">Record on Analyzer SSD</font>
|Record on Analyzer SSD
|<font size = "1">Up to '''40 channels''' (32 inputs <nowiki>+</nowiki> 6 ext. sync or DC)/instruments - max bandwidth </font>'''3.2 MS/s''' (12.5 MB/s)
|Up to '''40 channels''' (32 inputs <nowiki>+</nowiki> 6 ext. sync or DC)/instruments - max bandwidth '''3.2 MS/s''' (12.5 MB/s)


|-
|-
|<font size = "1">Format</font>
|Format
|<font size = "1">'''Normal''': '''32 bits'''/sample - '''Compressed''' '''16 bits'''/sample </font>
|'''Normal''': '''32 bits'''/sample - '''Compressed''' '''16 bits'''/sample  


|}
|}


<font color="#4DA836">'''Recorder settings (continued)'''</font>
'''Recorder settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "4"|<font size = "1">'''Management'''</font>
|align = "center" rowspan = "4"|'''Management'''
|<font size = "1">Download</font>
|Download
|<font size = "1">Discard Analyzer SSD on PC HDD''' - Batch download''' available - Rate '''4 MB/s'''</font>
|Discard Analyzer SSD on PC HDD''' - Batch download''' available - Rate '''4 MB/s'''


|-
|-
|<font size = "1">Upload</font>
|Upload
|<font size = "1">Load PC HDD records on the analyzer SSD - Rate '''1.4 MB/s''' </font>
|Load PC HDD records on the analyzer SSD - Rate '''1.4 MB/s'''  


|-
|-
|<font size = "1">Location</font>
|Location
|<font size = "1">All records accessible from the project manager tree </font>
|All records accessible from the project manager tree  


|-
|-
|<font size = "1">Hard drive management</font>
|Hard drive management
|<font size = "1">'''Format '''SSD - '''Selective delete''' - '''Sorting '''tools</font>
|'''Format '''SSD - '''Selective delete''' - '''Sorting '''tools


|}
|}


<font color="#4DA836">'''Mobi-Disk'''<sup>'''TM'''</sup></font>
'''Mobi-Disk'''<sup>'''TM'''</sup>


OR36 Mobi-Pack and OR38 multi analyzers/recorders feature a removable hard drive with 2 ports:
OR36 Mobi-Pack and OR38 multi analyzers/recorders feature a removable hard drive with 2 ports:
Line 566: Line 543:


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center"|<font size = "1">'''Connection'''</font>
|align = "center"|'''Connection'''
|<font size = "1">1 to 8 Mobi-Disc<sup>TM</sup></font>
|1 to 8 Mobi-Disc<sup>TM</sup>
|<font size = "1">Up to '''8 Mobi-disc''' connected - Active Mobi-Disc: '''1''' (user selectable) - '''hot swap'''</font>
|Up to '''8 Mobi-disc''' connected - Active Mobi-Disc: '''1''' (user selectable) - '''hot swap'''


|-
|-
|align = "center"|<font size = "1">'''Transfer'''</font>
|align = "center"|'''Transfer'''
|<font size = "1">Based on the USB 3.0 link</font>
|Based on the USB 3.0 link
|<font size = "1">'''Download 15 MB/s'''</font>
|'''Download 15 MB/s'''


|}
|}


<font color="#4DA836">'''Recorder monitoring'''</font>
'''Recorder monitoring'''


The following displays are available during real-time or post-analysis.
The following displays are available during real-time or post-analysis.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center"|<font size = "1">'''RMS'''</font>
|align = "center"|'''RMS'''
|<font size = "1">Overall RMS / input</font>
|Overall RMS / input
|<font size = "1">'''Digital '''or '''analog view-meter'''</font>
|'''Digital '''or '''analog view-meter'''


|-
|-
|align = "center"|<font size = "1">'''Signal'''</font>
|align = "center"|'''Signal'''
|<font size = "1">Real-time</font>
|Real-time
|<font size = "1">'''Compressed view of entire''' recorded signals - Automatic update of time axis.</font>
|'''Compressed view of entire''' recorded signals - Automatic update of time axis.


|}
|}


===Player===
===Player===
The player module plays the recorded signal files. This is achieved following 2 different ways:
The player module plays the recorded signal files. This is done one of the following two ways:


* '''Post-analysis''', the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
* '''Post-analysis''', the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
* '''Playbacks''', the recorded dynamic input are generated on the instrument outputs simultaneously with standard analysis of the inputs.
* '''Playbacks''', the recorded dynamic input are generated on the instrument outputs simultaneously with standard analysis of the inputs.
<font color="#4DA836">'''Player settings'''</font>
'''Player settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "5"|<font size = "1">'''Played section'''</font>
|align = "center" rowspan = "5"|'''Played section'''
|<font size = "1">File selection</font>
|File selection
|<font size = "1">NVGate<sup>®</sup> '''recorded files''' - '''Imported files''' - located on '''instrument SSD, Mobi-Dsic'''<sup>'''TM'''</sup> or on '''PC HDD'''.</font>
|NVGate<sup>®</sup> '''recorded files''' - '''Imported files''' - located on '''instrument SSD, Mobi-Dsic'''<sup>'''TM'''</sup> or on '''PC HDD'''.


|-
|-
|<font size = "1">Record selection</font>
|Record selection
|<font size = "1">'''Record number selection''' (for multi record files)</font>
|'''Record number selection''' (for multi record files)


|-
|-
|<font size = "1">Section definition</font>
|Section definition
|<font size = "1">User selectable '''Start''' and '''stop offsets''' (in sec.) - available in the '''file overview window''' (marker and slider)</font>
|User selectable '''Start''' and '''stop offsets''' (in sec.) - available in the '''file overview window''' (marker and slider)


|-
|-
|<font size = "1">Duration</font>
|Duration
|<font size = "1">'''Play backed duration''' (User information)</font>
|'''Play backed duration''' (User information)


|-
|-
|<font size = "1">Repeat</font>
|Repeat
|<font size = "1">'''On/off''' - repeat continuously the selection (available only for playback on analyzer output)</font>
|'''On/off''' - repeat continuously the selection (available only for playback on analyzer output)


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Mode'''</font>
|align = "center" rowspan = "3"|'''Mode'''
|<font size = "1">File swap</font>
|File swap
|<font size = "1">'''Swap loaded files with same tracks number''' - applies '''identical post-analysis''' setup to multiple files.</font>
|'''Swap loaded files with same tracks number''' - applies '''identical post-analysis''' setup to multiple files.


|-
|-
|<font size = "1">Playback</font>
|Playback
|<font size = "1">Continuous or step by step (5 ms to 360,000 s)</font>
|Continuous or step by step (5 ms to 360,000 s)


|-
|-
|<font size = "1">Bandwidth</font>
|Bandwidth
|<font size = "1">All tracks at lowest frequency (compatible with all plugs-in) or initial record frequencies (no track mix in plugs-in)</font>
|All tracks at lowest frequency (compatible with all plugs-in) or initial record frequencies (no track mix in plugs-in)


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Tracks'''</font>
|align = "center" rowspan = "5"|'''Tracks'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''Up to 352 '''according to user<nowiki>'</nowiki>s fleet</font>
|'''Up to 352 '''according to user<nowiki>'</nowiki>s fleet


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Dynamic input''' record (2 sampling frequencies), '''parametric '''(DC) input record and '''ext. sync''' record.</font>
|'''Dynamic input''' record (2 sampling frequencies), '''parametric '''(DC) input record and '''ext. sync''' record.


|-
|-
|<font size = "1">Fixed setting (information) / track</font>
|Fixed setting (information) / track
|<font size = "1">'''Label''', '''Coupling''', '''external gain''', input '''range''', '''sampling frequency '''and''' signal bandwidth'''</font>
|'''Label''', '''Coupling''', '''external gain''', input '''range''', '''sampling frequency '''and''' signal bandwidth'''


|-
|-
|<font size = "1">Modifiable measurement point information settings (apply for post analysis or re-recording) / track</font>
|Modifiable measurement point information settings (apply for post analysis or re-recording) / track
|<font size = "1">'''Component''' - '''node''' number - '''direction''' - '''type'''</font>
|'''Component''' - '''node''' number - '''direction''' - '''type'''


|-
|-
|<font size = "1">Modifiable settings (apply for post analysis or re-recording) / track</font>
|Modifiable settings (apply for post analysis or re-recording) / track
|<font size = "1">Associated '''transducer''' - '''unit''' - '''sensitivity''' - '''offset''' compensation</font>
|Associated '''transducer''' - '''unit''' - '''sensitivity''' - '''offset''' compensation


|}
|}


<font color="#4DA836">'''Player settings (continued)'''</font>
'''Player settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Listening'''</font>
|align = "center" rowspan = "2"|'''Listening'''
|<font size = "1">File location</font>
|File location
|<font size = "1">'''Analyzer HD''' - '''Mobi-Disk'''<sup>'''TM'''</sup> connected with USB or inside the analyzer - '''PC HD''' </font>
|'''Analyzer HD''' - '''Mobi-Disk'''<sup>'''TM'''</sup> connected with USB or inside the analyzer - '''PC HD'''  


|-
|-
|<font size = "1">Playback</font>
|Playback
|<font size = "1">One track - from cursor position - repeat displayed signal </font>
|One track - from cursor position - repeat displayed signal  


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Markers'''</font>
|align = "center" rowspan = "2"|'''Markers'''
|<font size = "1">Display</font>
|Display
|<font size = "1">Recorded markers are available in the '''file overview''' </font>
|Recorded markers are available in the '''file overview'''  


|-
|-
|<font size = "1">Use</font>
|Use
|<font size = "1">Set the start and stop playback offset (user selectable)</font>
|Set the start and stop playback offset (user selectable)


|}
|}


<font color="#4DA836">'''Player connections'''</font>
'''Player connections'''


The following table describes the available connections to  the analysis modes (plug-in analyzers and recorder) during post-analysis operation:
The following table describes the available connections to  the analysis modes (plug-in analyzers and recorder) during post-analysis operation:


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Dynamic inputs & torsional Tracks'''</font>
|align = "center" rowspan = "3"|'''Dynamic inputs & torsional Tracks'''
|<font size = "1">Preview (multiple file simultaneously)</font>
|Preview (multiple file simultaneously)
|<font size = "1">'''Entire file fast overview''' (pre-compressed at recording time) - '''Track contents''' preview, independently (multi-graph display)</font>
|'''Entire file fast overview''' (pre-compressed at recording time) - '''Track contents''' preview, independently (multi-graph display)


|-
|-
|<font size = "1">Monitoring (loaded file only)</font>
|Monitoring (loaded file only)
|<font size = "1">'''Zoom '''on selected '''play-back section'''</font>
|'''Zoom '''on selected '''play-back section'''


|-
|-
|<font size = "1">Connection3 (loaded file only)</font>
|Connection3 (loaded file only)
|<font size = "1">To any: '''plug-in analyzer''' '''channel, monitor''' '''channel,''' '''recorder''' '''track''', '''edge event detector''' and '''tachometer '''</font>
|To any: '''plug-in analyzer''' '''channel, monitor''' '''channel,''' '''recorder''' '''track''', '''edge event detector''' and '''tachometer '''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Ext. sync Tracks'''</font>
|align = "center" rowspan = "3"|'''Ext. sync Tracks'''
|<font size = "1">Preview (multiple file simultaneously)</font>
|Preview (multiple file simultaneously)
|<font size = "1">'''Entire file fast overview''' (pre-compressed at recording time) - '''Track contents preview''', independently (multi-graph display)</font>
|'''Entire file fast overview''' (pre-compressed at recording time) - '''Track contents preview''', independently (multi-graph display)


|-
|-
|<font size = "1">Monitoring (loaded file only)</font>
|Monitoring (loaded file only)
|<font size = "1">'''Zoom '''on selected '''play-back section'''</font>
|'''Zoom '''on selected '''play-back section'''


|-
|-
|<font size = "1">Connection<sup>3 </sup>(loaded file only)</font>
|Connection<sup>3 </sup>(loaded file only)
|<font size = "1">To any: '''start '''or''' stop averaging''' of plug-in analyzers, recorder '''start '''or''' stop recording''', '''waterfall start '''or''' stop acquisition''', '''new block trigger''' for FFT and SOA and '''new slice '''for''' waterfall''' acquisition.</font><br><font size = "1">To '''torsional''', '''tachometer''' & '''angular sampling'''</font>
|To any: '''start '''or''' stop averaging''' of plug-in analyzers, recorder '''start '''or''' stop recording''', '''waterfall start '''or''' stop acquisition''', '''new block trigger''' for FFT and SOA and '''new slice '''for''' waterfall''' acquisition.<br>To '''torsional''', '''tachometer''' & '''angular sampling'''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Recorded parametric'''<sup>'''1</sup> ''''''(DC) inputs & CAN parameters'''</font>
|align = "center" rowspan = "3"|'''Recorded parametric<sup>1</sup>(DC) inputs & CAN parameters'''
|<font size = "1">Preview (multiple file simultaneously)</font>
|Preview (multiple file simultaneously)
|<font size = "1">Entire file''' fast overview'''  - '''Track contents''' preview</font>
|Entire file''' fast overview'''  - '''Track contents''' preview


|-
|-
|<font size = "1">Monitoring (loaded file only)</font>
|Monitoring (loaded file only)
|<font size = "1">'''Profiles''' versus time and '''digital''' or '''analog view meter'''.</font>
|'''Profiles''' versus time and '''digital''' or '''analog view meter'''.


|-
|-
|<font size = "1">Connection<sup>3 </sup>(loaded file only)</font>
|Connection<sup>3 </sup>(loaded file only)
|<font size = "1">To any: '''recorder track''', '''waterfall reference (Z/X axis)''' and '''profiles''', '''level''' and '''delta level events'''.</font>
|To any: '''recorder track''', '''waterfall reference (Z/X axis)''' and '''profiles''', '''level''' and '''delta level events'''.


|}
|}
Line 730: Line 707:
The TDA plug-in analyzer provides time domain based analyses and visualizations. This plug-in analyzer computes statistical extractions and data compression of long duration oscilloscope views. All results are real time and operations are possible on-line or in post processing. The TDA plug-ins can operate free run or synchronized with the other plug-in analyzers.
The TDA plug-in analyzer provides time domain based analyses and visualizations. This plug-in analyzer computes statistical extractions and data compression of long duration oscilloscope views. All results are real time and operations are possible on-line or in post processing. The TDA plug-ins can operate free run or synchronized with the other plug-in analyzers.


<font color="#4DA836">'''TDA Settings'''</font>
'''TDA Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Bandwidths'''</font>
|align = "center" rowspan = "2"|'''Bandwidths'''
|<font size = "1">Oscillator 1</font>
|Oscillator 1
|<font size = "1">'''800 mHz''' to '''40 kHz''' <br>(sampling oscillator 102.4 kS/s to 2,048 S/s) </font>
|'''800 mHz''' to '''40 kHz''' <br>(sampling oscillator 102.4 kS/s to 2,048 S/s)  


|-
|-
|<font size = "1">Oscillator 2</font>
|Oscillator 2
|<font size = "1">'''512 mHz''' to '''25.6 kHz''' <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)</font>
|'''512 mHz''' to '''25.6 kHz''' <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Time views'''</font>
|align = "center" rowspan = "2"|'''Time views'''
|<font size = "1">Depth</font>
|Depth
|<font size = "1">'''7200/ABW<ref>ABW: the plug-in Analysis Bandwidth.</ref> '''sec to''' 100 000 '''sec (27 hours) - '''Independent''' on each channel</font>
|'''7200/ABW ABW: the plug-in Analysis Bandwidth. '''sec to''' 100 000 '''sec (27 hours) - '''Independent''' on each channel


|-
|-
|<font size = "1">Time base (resolution)</font>
|Time base (resolution)
|<font size = "1">'''Depth/2048'''- '''Independent''' on each channel</font>
|'''Depth/2048'''- '''Independent''' on each channel


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Time span for extraction'''</font><br><font size = "1">'''(averaging)'''</font>
|align = "center" rowspan = "2"|'''Time span for extraction'''<br>'''(averaging)'''
|<font size = "1">Type</font>
|Type
|<font size = "1">Continuous sliding with refresh each 256 sample'''s '''('''Exponential''') - One shot ('''Linear''') - Repeated shot ('''Repeated linear''')</font>
|Continuous sliding with refresh each 256 sample'''s '''('''Exponential''') - One shot ('''Linear''') - Repeated shot ('''Repeated linear''')


|-
|-
|<font size = "1">Duration</font>
|Duration
|<font size = "1">'''1/ABW '''to'''  2 10'''<sup>'''6'''</sup>'''/ABW''' (ex: 50 ms to 1000 sec @ 2 kHz)</font>
|'''1/ABW '''to'''  2 10'''<sup>'''6'''</sup>'''/ABW''' (ex: 50 ms to 1000 sec @ 2 kHz)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "2"|'''Channels'''
|<font size = "1">OR36 & OR38 analyzers</font>
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line''' 32 to 256 in '''post-analysis'''
|'''32 channels '''per plug-in for '''on line''' 32 to 256 in '''post-analysis'''


|-
|-
|<font size = "1">OR34 & OR35 analyzers</font>
|OR34 & OR35 analyzers
|<font size = "1">'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''</font>
|'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''


|-
|-
|align = "center"|<font size = "1">'''Filters'''</font>
|align = "center"|'''Filters'''
|<font size = "1">On each channel</font>
|On each channel
|<font size = "1">'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)</font>
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Triggering'''</font><br><br>
|align = "center" rowspan = "4"|'''Triggering'''<br><br>
|<font size = "1">Start analysis</font>
|Start analysis
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">End analysis</font>
|End analysis
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">Repeat analysis</font>
|Repeat analysis
|<font size = "1">On '''new start''' or '''end of averaging'''</font>
|On '''new start''' or '''end of averaging'''


|-
|-
|<font size = "1">Trigger delay</font>
|Trigger delay
|<font size = "1">Positive''' '''=''' unlimited''' - negative = '''32k Samples'''</font>
|Positive''' '''=''' unlimited''' - negative = '''32k Samples'''


|}
|}


<font color="#4DA836">'''TDA Results'''</font>
'''TDA Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center"|<font size = "1">'''Time Views'''</font>
|align = "center"|'''Time Views'''
|<font size = "1">Overview</font>
|Overview
|<font size = "1">'''Signal envelope: '''Min/Max line for each time step - </font>
|'''Signal envelope: '''Min/Max line for each time step -  


|-
|-
|align = "center"|&nbsp;
|align = "center"|&nbsp;
|<font size = "1">Statistical values</font>
|Statistical values
|<font size = "1">'''DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Skewness - Kurtosis - '''computed on a user defined area of the time view</font>
|'''DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Skewness - Kurtosis - '''computed on a user defined area of the time view


|-
|-
|align = "center"|<font size = "1">'''Extractions'''</font>
|align = "center"|'''Extractions'''
|<font size = "1">Statistical values</font>
|Statistical values
|<font size = "1">'''DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Kurtosis - '''computed on the Time span – All values available for '''View-meter''' and/or '''Waterfall profiles'''</font>
|'''DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Kurtosis - '''computed on the Time span – All values available for '''View-meter''' and/or '''Waterfall profiles'''


|}
|}
Line 818: Line 795:
The FFT plug-in analyzer features multiple vibrations results depending on applied setting. This type of plug-in analyzers computes real-time (on-line or post-processing) Fast Fourier Transform algorithms with multiple averaging domain and trigger capabilities. Up to 4 FFT plug-ins can operate at the same time.
The FFT plug-in analyzer features multiple vibrations results depending on applied setting. This type of plug-in analyzers computes real-time (on-line or post-processing) Fast Fourier Transform algorithms with multiple averaging domain and trigger capabilities. Up to 4 FFT plug-ins can operate at the same time.


<font color="#4DA836">'''FFT Settings'''</font>
'''FFT Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Bandwidths'''</font>
|align = "center" rowspan = "2"|'''Bandwidths'''
|<font size = "1">Oscillator 1</font>
|Oscillator 1
|<font size = "1">'''800 mHz''' to '''40 kHz''' <br>(sampling oscillator 102.4 kS/s to 2,048 S/s) </font>
|'''800 mHz''' to '''40 kHz''' <br>(sampling oscillator 102.4 kS/s to 2,048 S/s)  


|-
|-
|<font size = "1">Oscillator 2</font>
|Oscillator 2
|<font size = "1">'''512 mHz''' to '''25.6 kHz''' <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)</font>
|'''512 mHz''' to '''25.6 kHz''' <br>(sampling oscillator 65.536 kS/s to 3,277 S/s)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Resolution'''</font>
|align = "center" rowspan = "2"|'''Resolution'''
|<font size = "1">Lines</font>
|Lines
|<font size = "1">'''101''', '''201''', '''401''', '''801''', '''1601''', '''3201''', '''6401, 12801<ref>With force DSP</ref> and 25601 lines'''</font>
|'''101''', '''201''', '''401''', '''801''', '''1601''', '''3201''', '''6401, 12801 With force DSP and 25601 lines'''


|-
|-
|<font size = "1">Frequency resolution</font>
|Frequency resolution
|<font size = "1">'''80 µHz '''(512 mHz/6401 lines) <br>to '''400 Hz '''(40 kHz/101 lines)</font>
|'''80 µHz '''(512 mHz/6401 lines) <br>to '''400 Hz '''(40 kHz/101 lines)


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Averaging'''</font>
|align = "center" rowspan = "4"|'''Averaging'''
|<font size = "1">Domain </font>
|Domain  
|<font size = "1">'''Spectral''' (power) - '''time''' (with phase) - '''FDSA''' (Synchronized with one frequency)</font>
|'''Spectral''' (power) - '''time''' (with phase) - '''FDSA''' (Synchronized with one frequency)


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Exponential''' - '''linear''' - '''repeated linear''' - '''referenced peak hold''' (eq. to tracking filter locked on reference channel frequency) - '''peak hold'''</font>
|'''Exponential''' - '''linear''' - '''repeated linear''' - '''referenced peak hold''' (eq. to tracking filter locked on reference channel frequency) - '''peak hold'''


|-
|-
|<font size = "1">Size</font>
|Size
|<font size = "1">'''Duration '''or '''number of blocks''', unlimited size</font>
|'''Duration '''or '''number of blocks''', unlimited size


|-
|-
|<font size = "1">Overlap</font>
|Overlap
|<font size = "1">'''0%''' to '''99.9%''' depending on block size - '''retrigger'''</font>
|'''0%''' to '''99.9%''' depending on block size - '''retrigger'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "2"|'''Channels'''
|<font size = "1">OR36, MP & OR38 analyzers</font>
|OR36, MP & OR38 analyzers
|<font size = "1">'''32 to 256 ch. '''per plug-in for '''on line''' and '''post-analysis'''</font>
|'''32 to 256 ch. '''per plug-in for '''on line''' and '''post-analysis'''


|-
|-
|<font size = "1">OR34 & OR35 analyzers</font>
|OR34 & OR35 analyzers
|<font size = "1">'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''</font>
|'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Weighting'''</font>
|align = "center" rowspan = "3"|'''Weighting'''
|<font size = "1">General</font>
|General
|<font size = "1">'''Uniform - Hanning - Hamming - Kaiser Bessel'''</font>
|'''Uniform - Hanning - Hamming - Kaiser Bessel'''


|-
|-
|<font size = "1">Modal acquisition</font>
|Modal acquisition
|<font size = "1">User defined''' Force '''(rect.)''' '''and '''Response '''(Rect. <nowiki>+</nowiki> exp)</font>
|User defined''' Force '''(rect.)''' '''and '''Response '''(Rect. <nowiki>+</nowiki> exp)


|-
|-
|<font size = "1">Assignment</font>
|Assignment
|<font size = "1">Independently on any channels</font>
|Independently on any channels


|-
|-
|align = "center"|<font size = "1">'''Filters'''</font>
|align = "center"|'''Filters'''
|<font size = "1">On each channel</font>
|On each channel
|<font size = "1">'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)</font>
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)


|-
|-
|align = "center"|<font size = "1">'''Cross functions'''</font>
|align = "center"|'''Cross functions'''
|<font size = "1">Reference selection</font>
|Reference selection
|<font size = "1">'''Full''' or '''partial matrix''' available (all channels can be references)</font>
|'''Full''' or '''partial matrix''' available (all channels can be references)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Zoom'''</font>
|align = "center" rowspan = "2"|'''Zoom'''
|<font size = "1">Magnification factor</font>
|Magnification factor
|<font size = "1">'''2 to 128 '''by step power of 2</font>
|'''2 to 128 '''by step power of 2


|-
|-
|<font size = "1">Min., max and center frequencies</font>
|Min., max and center frequencies
|<font size = "1">Graphically selectable on wide band spectrum</font>
|Graphically selectable on wide band spectrum


|-
|-
|align = "center"|<font size = "1">'''Triggering'''</font>
|align = "center" rowspan = "6"|'''Triggering'''
|<font size = "1">Start averaging</font>
|Start averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|align = "center"|&nbsp;
|End averaging
|<font size = "1">End averaging</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>


|-
|-
|align = "center"|&nbsp;
|New block
|<font size = "1">New block</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>


|-
|-
|align = "center"|&nbsp;
|Repeat averaging
|<font size = "1">Repeat averaging</font>
|On '''new start''' or '''end of averaging'''
|<font size = "1">On '''new start''' or '''end of averaging'''</font>


|-
|-
|align = "center"|<br>
|Trigger delay
|<font size = "1">Trigger delay</font>
|Positive''' '''=''' unlimited''' - negative = '''32k Samples'''
|<font size = "1">Positive''' '''=''' unlimited''' - negative = '''32k Samples'''</font>


|-
|-
|align = "center"|&nbsp;
|Automatic
|<font size = "1">Automatic</font>
|Reject '''overloaded blocks'''
|<font size = "1">Reject '''overloaded blocks'''</font>


|-
|-
|align = "center"|<font size = "1">'''Blocks rejection'''</font><br><font size = "1">'''Overall'''</font>
|align = "center" rowspan = "2"|'''Blocks rejection'''<br>'''Overall'''
|<font size = "1">Manual</font>
|Manual
|<font size = "1">'''Accept''' or '''Reject '''after preview of averaged results (FRF, Coherence)</font>
|'''Accept''' or '''Reject '''after preview of averaged results (FRF, Coherence)


|-
|-
|align = "center"|&nbsp;
|Normal
|<font size = "1">Normal</font>
|'''Accept''' all blocks
|<font size = "1">'''Accept''' all blocks</font>


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Overall'''</font>
|align = "center" rowspan = "2"|'''Overall'''
|<font size = "1">Detector</font>
|Detector
|<font size = "1">'''Quadratic sum of spectra lines''' taking in '''account weighting window equivalent noise bandwidth'''</font>
|'''Quadratic sum of spectra lines''' taking in '''account weighting window equivalent noise bandwidth'''


|-
|-
|<font size = "1">Lower and upper frequencies</font>
|Lower and upper frequencies
|<font size = "1">Selectable into the current FFT bandwidth</font>
|Selectable into the current FFT bandwidth


|}
|}


<font color="#4DA836">'''FFT Results'''</font>
'''FFT Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "4"|<font size = "1">'''Time averaging'''</font>
|align = "center" rowspan = "4"|'''Time averaging'''
|<font size = "1">Time domain</font>
|Time domain
|<font size = "1">'''Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajoux of triggered blocks'''</font>
|'''Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajoux of triggered blocks'''


|-
|-
|<font size = "1">Spectra</font>
|Spectra
|<font size = "1">'''Complex spectra'''</font>
|'''Complex spectra'''


|-
|-
|<font size = "1">Cross-functions</font>
|Cross-functions
|<font size = "1">Instantaneous''' cross-spectra''' - averaged '''cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence'''</font>
|Instantaneous''' cross-spectra''' - averaged '''cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence'''


|-
|-
|<font size = "1">Overall</font>
|Overall
|<font size = "1">'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.</font>
|'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.


|-
|-
|align = "center" rowspan = "6"|<font size = "1">'''Spectral averaging'''</font>
|align = "center" rowspan = "6"|'''Spectral averaging'''
|<font size = "1">Time domain</font>
|Time domain
|<font size = "1">'''Filtered signal - Triggered block - Weighted Block - Lissajous of triggered blocks'''</font>
|'''Filtered signal - Triggered block - Weighted Block - Lissajous of triggered blocks'''


|-
|-
|<font size = "1">Spectra</font>
|Spectra
|<font size = "1">Instantaneous '''complex spectra''' - averaged '''power spectra'''</font>
|Instantaneous '''complex spectra''' - averaged '''power spectra'''


|-
|-
|<font size = "1">Zoomed spectra</font>
|Zoomed spectra
|<font size = "1">'''Zoomed''' instantaneous '''complex spectra''' - '''zoomed '''averaged '''power spectra'''</font>
|'''Zoomed''' instantaneous '''complex spectra''' - '''zoomed '''averaged '''power spectra'''


|-
|-
|<font size = "1">Cross-functions</font>
|Cross-functions
|<font size = "1">'''Instantaneous cross-spectra''' - averaged '''cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence'''</font>
|'''Instantaneous cross-spectra''' - averaged '''cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence'''


|-
|-
|<font size = "1">Zoomed cross-function</font>
|Zoomed cross-function
|<font size = "1">Instantaneous '''zoomed cross-spectra''' - '''averaged zoomed cross-spectra''' - '''zoomed FRF H1''' - '''zoomed FRF H2''' - '''zoomed Coherence'''</font>
|Instantaneous '''zoomed cross-spectra''' - '''averaged zoomed cross-spectra''' - '''zoomed FRF H1''' - '''zoomed FRF H2''' - '''zoomed Coherence'''


|-
|-
|<font size = "1">Overall</font>
|Overall
|<font size = "1">'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.</font>
|'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Synchronous averaging on one frequency (FDSA)'''</font>
|align = "center" rowspan = "3"|'''Synchronous averaging on one frequency (FDSA)'''
|<font size = "1">Time domain</font>
|Time domain
|<font size = "1">'''Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks'''</font>
|'''Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks'''


|-
|-
|<font size = "1">Spectra</font>
|Spectra
|<font size = "1">Instantaneous '''complex spectra''' - time domain averaged '''power spectra'''</font>
|Instantaneous '''complex spectra''' - time domain averaged '''power spectra'''


|-
|-
|<font size = "1">Overall</font>
|Overall
|<font size = "1">'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.</font>
|'''Overall RMS value''' in specified band - '''Overall RMS profile''' in specified band.


|}
|}


===Constant Band Order Tracking Add-on (CBT)===
===Constant Band Order Tracking Add-on (CBT)===
Constant band order tracking is an optional add-on to the FFT plug-in analyzers. This add-on computes spectrum lines power related to RPM speed. This option adds settings and results to the FFT Plug-ins. Up to 4 constant Band tracking can operate on 4 different tachometers at the same time.
Constant band order tracking is an optional add-on to the FFT plug-in analyzers. This add-on computes spectrum lines power related to RPM speed. This option adds settings and results to the FFT Plug-ins. Constant band tracking can operate on up to 4 different tachometers at the same time.


<font color="#4DA836">'''CBT Settings'''</font>
'''CBT Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Tracked Order'''</font>
|align = "center" rowspan = "3"|'''Tracked Order'''
|<font size = "1">Number </font>
|Number  
|<font size = "1">'''8 per channels'''</font>
|'''8 per channels'''


|-
|-
|<font size = "1">Max</font>
|Max
|<font size = "1">'''0.001''' to '''800'''</font>
|'''0.001''' to '''800'''


|-
|-
|<font size = "1">Constant bandwidth</font>
|Constant bandwidth
|<font size = "1">'''User selectable per channel '''- minimum depend on weighting windows</font>
|'''User selectable per channel '''- minimum depend on weighting windows


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Computation'''</font>
|align = "center" rowspan = "3"|'''Computation'''
|<font size = "1">Associated tachometer</font>
|Associated tachometer
|<font size = "1">'''Any front end''', '''recorded''' or '''virtual tachometer'''</font>
|'''Any front end''', '''recorded''' or '''virtual tachometer'''


|-
|-
|<font size = "1">Peak tracking</font>
|Peak tracking
|<font size = "1">On/Off '''center bandwidth on nearest peak'''</font>
|On/Off '''center bandwidth on nearest peak'''


|-
|-
|<font size = "1">Order amplitude</font>
|Order amplitude
|<font size = "1">Weighting windows eq. noise bandwidth correction</font>
|Weighting windows eq. noise bandwidth correction


|}
|}


<font color="#4DA836">'''CBT Results'''</font>
'''CBT Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Scalar'''</font>
|align = "center" rowspan = "3"|'''Scalar'''
|<font size = "1">Tracked order</font>
|Tracked order
|<font size = "1">'''Digital '''(magnitude and phase)''' '''or '''analog '''view-meter</font>
|'''Digital '''(magnitude and phase)''' '''or '''analog '''view-meter


|-
|-
|<font size = "1">Cross phase tracking</font>
|Cross phase tracking
|<font size = "1">'''Order phases '''are '''relative''' to the same order from a ref. channel</font>
|'''Order phases '''are '''relative''' to the same order from a ref. channel


|-
|-
|<font size = "1">Overall</font>
|Overall
|<font size = "1">'''Digital '''or '''analog view-meter'''</font>
|'''Digital '''or '''analog view-meter'''


|-
|-
|align = "center"|<font size = "1">'''Monitoring<ref>These results cannot be saved</ref>'''</font>
|align = "center"|'''Monitoring'''
|<font size = "1">Continuous profiles of tracked order</font>
|Continuous profiles of tracked order
|<font size = "1">'''Profiles '''vs.''' time''' - '''profiles '''vs.''' RPM''' - max depth '''2048 pts''' - user selectable '''delta time''' - user selectable '''delta RPM'''</font>
|'''Profiles '''vs.''' time''' - '''profiles '''vs.''' RPM''' - max depth '''2048 pts''' - user selectable '''delta time''' - user selectable '''delta RPM'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Profiles'''</font>
|align = "center" rowspan = "2"|'''Profiles'''
|<font size = "1">One shot acquisition</font>
|One shot acquisition
|<font size = "1">Tracked order (complex), cross phase orders and overall level can be collected by the ''waterfall ''profiles. </font>
|Tracked order (complex), cross phase orders and overall level can be collected by the ''waterfall ''profiles.  


|-
|-
|<font size = "1">References</font>
|References
|<font size = "1">Time, RPM and DC levels</font>
|Time, RPM and DC levels


|}
|}


===Diagnostic Add-on (FFTDiag)===
===Diagnostic Add-on (FFTDiag)===
FFT Diagnostic is an optional add-on to the FFT plug-in analyzers. This add-on computes a set of useful analysis for machinery diagnostic. Up to 4 FFTDiag can operate on 4 different FFT plug-in at the same time.
FFT Diagnostic is an optional add-on to the FFT plug-in analyzers. This add-on computes a set of useful analysis for machinery diagnostic. Up to 4 FFTDiag can operate on 4 different FFT plug-ins at the same time.


This option adds settings and results to the FFT Plug-ins.
This option adds settings and results to the FFT Plug-ins.


<font color="#4DA836">'''FFTDiag Settings'''</font>
'''FFTDiag Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Envelop demodulation'''</font>
|align = "center" rowspan = "3"|'''Envelop demodulation'''
|<font size = "1">Activation</font>
|Activation
|<font size = "1">Enable/disable on all channels''' - '''requires '''zoom activation '''</font>
|Enable/disable on all channels''' - '''requires '''zoom activation '''


|-
|-
|<font size = "1">Averaging</font>
|Averaging
|<font size = "1">Spectral domain - FDSA </font>
|Spectral domain - FDSA  


|-
|-
|<font size = "1">Bandwidth</font>
|Bandwidth
|<font size = "1">'''½ of zoom span - ¼ of FFT bandwidth'''</font>
|'''½ of zoom span - ¼ of FFT bandwidth'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Correlation'''</font>
|align = "center" rowspan = "2"|'''Correlation'''
|<font size = "1">Activation</font>
|Activation
|<font size = "1">Enable/disable on all channels''' - '''All other FFT results are affected</font>
|Enable/disable on all channels''' - '''All other FFT results are affected


|-
|-
|<font size = "1">Weighting window</font>
|Weighting window
|<font size = "1">'''Uniform - Left '''zero padding''' - Centred '''zero padding</font>
|'''Uniform - Left '''zero padding''' - Centred '''zero padding


|}
|}


<font color="#4DA836">'''FFTDiag Results'''</font>
'''FFTDiag Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Scalar'''</font>
|align = "center" rowspan = "2"|'''Scalar'''
|<font size = "1">Min/Max</font>
|Min/Max
|<font size = "1">Minimum and maximum amplitude per trigger block</font>
|Minimum and maximum amplitude per trigger block


|-
|-
|<font size = "1">Peak detectors</font>
|Peak detectors
|<font size = "1">'''Peak level - Peak to Peak level - Crest Factor  - '''per trigger block</font>
|'''Peak level - Peak to Peak level - Crest Factor  - '''per trigger block


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Others'''</font>
|align = "center" rowspan = "3"|'''Others'''
|<font size = "1">Cepstrum</font>
|Cepstrum
|<font size = "1">'''Cepstrum - Zoomed cepstrum'''</font>
|'''Cepstrum - Zoomed cepstrum'''


|-
|-
|<font size = "1">Correlation</font>
|Correlation
|<font size = "1">'''Auto-correlation '''block''' - Cross correlation '''block</font>
|'''Auto-correlation '''block''' - Cross correlation '''block


|-
|-
|<font size = "1">Shaft view</font>
|Shaft view
|<font size = "1">Time domain signal on first rotation - angular representation</font>
|Time domain signal on first rotation - angular representation


|}
|}
Line 1,136: Line 1,107:


===Synchronous Order Analysis (SOA)===
===Synchronous Order Analysis (SOA)===
The SOA plug-in analyzer features several results depending on applied settings. Up to 2 SOA plug-ins can operate in parallel.
The SOA plug-in analyzer features several results depending on applied settings. Up to 3 SOA plug-ins can operate in parallel.


<font color="#4DA836">'''SOA Settings'''</font>
'''SOA Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Bandwidths'''</font>
|align = "center" rowspan = "3"|'''Bandwidths'''
|<font size = "1">Order span</font>
|Order span
|<font size = "1">'''6.25''' to '''400''' – up to '''800 '''with '''''ForceDSP'''''</font>
|'''6.25''' to '''400''' – up to '''800 '''with '''''ForceDSP'''''


|-
|-
|<font size = "1">Frequency bandwidth</font>
|Frequency bandwidth
|<font size = "1">Up to '''40 kHz''' </font>
|Up to '''40 kHz'''  


|-
|-
|<font size = "1">RPM span</font>
|RPM span
|<font size = "1">Up to''' 192 000 RPM - Ratio of 64 '''between Min & Max RPM</font>
|Up to''' 384 000 RPM - Ratio of 64 '''between Min & Max RPM


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Resolution'''</font>
|align = "center" rowspan = "2"|'''Resolution'''
|<font size = "1">Lines</font>
|Lines
|<font size = "1">'''101 - 201 - 401 – 801 – 1601 '''with '''''ForceDSP'''''</font>
|'''101 - 201 - 401 – 801 – 1601 '''with '''''ForceDSP'''''


|-
|-
|<font size = "1">Order resolution</font>
|Order resolution
|<font size = "1">'''1 - '''½ - ¼''' - 1/8 - 1/16 - 1/32'''</font>
|'''1 - '''½ - ¼''' - 1/8 - 1/16 - 1/32'''


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Averaging'''</font>
|align = "center" rowspan = "4"|'''Averaging'''
|<font size = "1">Domain </font>
|Domain  
|<font size = "1">'''Spectral''' (power) - '''revolution''' (re-sampled blocs)</font>
|'''Spectral''' (power) - '''revolution''' (re-sampled blocs)


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Exponential''' - '''linear''' - '''repeated linear '''- '''peak hold''' </font>
|'''Exponential''' - '''linear''' - '''repeated linear '''- '''peak hold'''  


|-
|-
|<font size = "1">Size</font>
|Size
|<font size = "1">'''Number of blocks''', unlimited size</font>
|'''Number of blocks''', unlimited size


|-
|-
|<font size = "1">Overlap</font>
|Overlap
|<font size = "1">'''0 to 31 revolutions''' depending on resolution – '''0 to 359°''' into revolution – Retrigger on''' multi-pulse/rev'''</font>
|'''0 to 31 revolutions''' depending on resolution – '''0 to 359°''' into revolution – Retrigger on''' multi-pulse/rev'''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Tachometer'''</font>
|align = "center" rowspan = "3"|'''Tachometer'''
|<font size = "1">Computation base</font>
|Computation base
|<font size = "1">Any '''NVGate'''<sup>'''®</sup> ''''''tachometer '''(ext. sync, regular input, DC input, CAN, simulated, Fractional, and combined tach.)</font>
|Any '''NVGate'''<sup>'''®</sup> ''''''tachometer '''(ext. sync, regular input, DC input, CAN, simulated, Fractional, and combined tach.)


|-
|-
|<font size = "1">Phase reference</font>
|Phase reference
|<font size = "1">Tachometer pulse '''edge '''or pulse '''center '''requires to analyze the tachometer input</font>
|Tachometer pulse '''edge '''or pulse '''center '''requires to analyze the tachometer input


|-
|-
|<font size = "1">0° Phase reference</font>
|0° Phase reference
|<font size = "1">Cosine or sine</font>
|Cosine or sine


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''RPM'''</font>
|align = "center" rowspan = "2"|'''RPM'''
|<font size = "1">Max speed variation</font>
|Max speed variation
|<font size = "1">'''1''' to '''99%''' per analyzed block - blocks with higher variation are rejected. No control (set @ 100%)</font>
|'''1''' to '''99%''' per analyzed block - blocks with higher variation are rejected. No control (set @ 100%)


|-
|-
|<font size = "1">RPM range</font>
|RPM range
|<font size = "1">User selectable '''max''' & '''min RPM''', under and over speed blocks are rejected</font>
|User selectable '''max''' & '''min RPM''', under and over speed blocks are rejected


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Tracked order'''</font>
|align = "center" rowspan = "3"|'''Tracked order'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''Up to 8 '''per channels</font>
|'''Up to 8 '''per channels


|-
|-
|<font size = "1">Min-Max</font>
|Min-Max
|<font size = "1">'''1/32 '''to''' 400''' – up to '''800 '''with '''''ForceDSP'''''</font>
|'''1/32 '''to''' 400''' – up to '''800 '''with '''''ForceDSP'''''


|-
|-
|<font size = "1">Cross phase tracking</font>
|Cross phase tracking
|<font size = "1">'''Order phases '''are '''relative''' to the same order from a ref. channel</font>
|'''Order phases '''are '''relative''' to the same order from a ref. channel


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "2"|'''Channels'''
|<font size = "1">OR36 & OR38 analyzers</font>
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line''' - 32 to 256 in '''post-analysis'''
|'''32 channels '''per plug-in for '''on line''' - 32 to 256 in '''post-analysis'''


|-
|-
|<font size = "1">OR34 & OR35 analyzers</font>
|OR34 & OR35 analyzers
|<font size = "1">'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''</font>
|'''8 channels '''per plug-in''' '''for '''on line''' and '''post-analysis'''


|-
|-
|align = "center"|<font size = "1">'''Weighting'''</font>
|align = "center"|'''Weighting'''
|<font size = "1">Windows selection</font>
|Windows selection
|<font size = "1">'''Uniform - Hanning - Hamming - Kaiser Bessel''' Applied independently on any channels</font>
|'''Uniform - Hanning - Hamming - Kaiser Bessel''' Applied independently on any channels


|-
|-
|align = "center"|<font size = "1">'''Filters'''</font>
|align = "center"|'''Filters'''
|<font size = "1">On each channel</font>
|On each channel
|<font size = "1">'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)</font>
|'''Any NVGate''' filter can independently be applied to each dynamic input (See Filters §)


|-
|-
|align = "center" rowspan = "6"|<font size = "1">'''Triggering'''</font>
|align = "center" rowspan = "6"|'''Triggering'''
|<font size = "1">Start averaging</font>
|Start averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">End averaging</font>
|End averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">New block</font>
|New block
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run '''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run '''


|-
|-
|<font size = "1">Repeat averaging</font>
|Repeat averaging
|<font size = "1">On '''new start''' or '''end of averaging'''</font>
|On '''new start''' or '''end of averaging'''


|-
|-
|<font size = "1">Trigger delay</font>
|Trigger delay
|<font size = "1">Positive''' '''=''' unlimited''' - negative = '''1/order res '''revolution</font>
|Positive''' '''=''' unlimited''' - negative = '''1/order res '''revolution


|-
|-
|<font size = "1">Trigger block phase shift</font>
|Trigger block phase shift
|<font size = "1">'''±720°''' - independent for each channel</font>
|'''±720°''' - independent for each channel


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Overall'''</font>
|align = "center" rowspan = "2"|'''Overall'''
|<font size = "1">Detector</font>
|Detector
|<font size = "1">'''Quadratic sum of order spectra lines''' taking in account '''weighting window equivalent noise bandwidth'''</font>
|'''Quadratic sum of order spectra lines''' taking in account '''weighting window equivalent noise bandwidth'''


|-
|-
|<font size = "1">Lower and upper orders</font>
|Lower and upper orders
|<font size = "1">Selectable into order span - from '''0,03125 '''to''' 800''' with '''''ForceDSP'''''</font>
|Selectable into order span - from '''0,03125 '''to''' 800''' with '''''ForceDSP'''''


|}
|}
Line 1,271: Line 1,242:
----
----


<font color="#4DA836">'''SOA Results'''</font>
'''SOA Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "5"|<font size = "1">'''Revolution averaging'''</font>
|align = "center" rowspan = "5"|'''Revolution averaging'''
|<font size = "1">Time domain</font>
|Time domain
|<font size = "1">'''Filtered '''signal''' '''</font>
|'''Filtered '''signal''' '''


|-
|-
|<font size = "1">Angle domain (Revolution)</font>
|Angle domain (Revolution)
|<font size = "1">'''Triggered '''block '''- Averaged Triggered '''block''' - Weighted '''block</font>
|'''Triggered '''block '''- Averaged Triggered '''block''' - Weighted '''block


|-
|-
|<font size = "1">Order spectra</font>
|Order spectra
|<font size = "1">'''Complex spectra'''</font>
|'''Complex spectra'''


|-
|-
|<font size = "1">Tracked orders</font>
|Tracked orders
|<font size = "1">'''Digital '''(magnitude & phase) view-meter '''- Analog '''view -meter</font>
|'''Digital '''(magnitude & phase) view-meter '''- Analog '''view -meter


|-
|-
|<font size = "1">Overall</font>
|Overall
|<font size = "1">'''Overall RMS value''' in specified order band (digital or analog view-meter)</font>
|'''Overall RMS value''' in specified order band (digital or analog view-meter)


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Spectral (order) averaging'''</font>
|align = "center" rowspan = "5"|'''Spectral (order) averaging'''
|<font size = "1">Time domain</font>
|Time domain
|<font size = "1">'''Filtered '''signal</font>
|'''Filtered '''signal


|-
|-
|<font size = "1">Angle domain (Revolution)</font>
|Angle domain (Revolution)
|<font size = "1">'''Triggered block - Weighted Block '''</font>
|'''Triggered block - Weighted Block '''


|-
|-
|<font size = "1">Order spectra</font>
|Order spectra
|<font size = "1">Instantaneous''' complex order spectra '''- averaged '''power order spectra'''</font>
|Instantaneous''' complex order spectra '''- averaged '''power order spectra'''


|-
|-
|<font size = "1">Tracked orders</font>
|Tracked orders
|<font size = "1">'''Digital (module and phase) '''and/or''' Analog '''view-meter.</font>
|'''Digital (module and phase) '''and/or''' Analog '''view-meter.


|-
|-
|<font size = "1">Overall</font>
|Overall
|<font size = "1">'''Overall RMS value''' in specified order band (digital or analog view-meter)</font>
|'''Overall RMS value''' in specified order band (digital or analog view-meter)


|}
|}
Line 1,324: Line 1,295:
This option adds settings and results to the SOA Plug-ins.
This option adds settings and results to the SOA Plug-ins.


<font color="#4DA836">'''SOADiag Settings'''</font>
'''SOADiag Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center"|<font size = "1">'''Cross-functions'''</font>
|align = "center"|'''Cross-functions'''
|<font size = "1">Reference selection</font>
|Reference selection
|<font size = "1">'''Full''' or '''partial matrix''' available (all channels can be references) - order domain averaging only.</font>
|'''Full''' or '''partial matrix''' available (all channels can be references) - order domain averaging only.


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Angular Correlation'''</font>
|align = "center" rowspan = "2"|'''Angular Correlation'''
|<font size = "1">Activation</font>
|Activation
|<font size = "1">Enable/disable on all channels''' - '''All other SOA results are affected simultaneously</font>
|Enable/disable on all channels''' - '''All other SOA results are affected simultaneously


|-
|-
|<font size = "1">Weighting window</font>
|Weighting window
|<font size = "1">'''Uniform - Left '''zero padding''' - Centred '''zero padding</font>
|'''Uniform - Left '''zero padding''' - Centred '''zero padding


|}
|}


<font color="#4DA836">'''SOADiag Results'''</font>
'''SOADiag Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Rev. synchronous Scalars'''</font>
|align = "center" rowspan = "3"|'''Rev. synchronous Scalars'''
|<font size = "1">Min/Max</font>
|Min/Max
|<font size = "1">Minimum and maximum amplitude per trigger block (n rev.)</font>
|Minimum and maximum amplitude per trigger block (n rev.)


|-
|-
|<font size = "1">RMS, DC</font>
|RMS, DC
|<font size = "1">Per trigger block (n rev.)</font>
|Per trigger block (n rev.)


|-
|-
|<font size = "1">Peak detectors</font>
|Peak detectors
|<font size = "1">'''Peak level - Peak to Peak level - Crest Factor - '''per trigger block (n rev.)</font>
|'''Peak level - Peak to Peak level - Crest Factor - '''per trigger block (n rev.)


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Others'''</font>
|align = "center" rowspan = "4"|'''Others'''
|<font size = "1">Copstrum™</font>
|Copstrum™
|<font size = "1">'''Cepstrum''' of the order spectra</font>
|'''Cepstrum''' of the order spectra


|-
|-
|<font size = "1">Cross-functions</font>
|Cross-functions
|<font size = "1">'''Instantaneous order cross-spectra''' - averaged order '''cross-spectra''' - '''ORF™ H1''' / '''H2''' -'''Coherence – angular '''or '''order '''domain averaging'''.'''</font>
|'''Instantaneous order cross-spectra''' - averaged order '''cross-spectra''' - '''ORF™ H1''' / '''H2''' -'''Coherence – angular '''or '''order '''domain averaging'''.'''


|-
|-
|<font size = "1">Angular Correlation</font>
|Angular Correlation
|<font size = "1">'''Auto-correlation '''block''' - Cross correlation '''block - angular averaged</font>
|'''Auto-correlation '''block''' - Cross correlation '''block - angular averaged


|-
|-
|<font size = "1">Re-sampled Shaft view</font>
|Re-sampled Shaft view
|<font size = "1">Time domain signal on first rotation – '''angular representation''' - '''instantaneous''' or '''revolutions''' averaged</font>
|Time domain signal on first rotation – '''angular representation''' - '''instantaneous''' or '''revolutions''' averaged


|}
|}
Line 1,380: Line 1,351:
----
----


===/n octave constant percentage band filter analysis - OCT===
===1/n octave constant percentage band filter analysis - OCT===
The OCT plug-in analyzer features multiple acoustic results depending on applied setting. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters (CPB) and detectors.
The OCT plug-in analyzer features multiple acoustic results depending on applied setting. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters (CPB) and detectors.


<font color="#4DA836">'''OCT Settings'''</font>
'''OCT Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Bandwidths'''</font>
|align = "center" rowspan = "3"|'''Bandwidths'''
|<font size = "1">Sampling oscillator #1 <br>(102.4 kS/s to 2,048 S/s) </font>
|Sampling oscillator #1 <br>(102.4 kS/s to 2,048 S/s)  
|<font size = "1">Center of '''highest 1/3'''<sup>'''rd</sup> '''band = '''40 kHz'''<br>Center of '''lowest''' '''1/3'''<sup>'''rd</sup> '''band = '''100 mHz'''</font>
|Center of '''highest 1/3'''<sup>'''rd</sup> '''band = '''40 kHz'''<br>Center of '''lowest''' '''1/3'''<sup>'''rd</sup> '''band = '''100 mHz'''


|-
|-
|<font size = "1">Sampling oscillator #2<br>(65.536 kS/s to 3,277 S/s)</font>
|Sampling oscillator #2<br>(65.536 kS/s to 3,277 S/s)
|<font size = "1">Center of '''highest''' '''1/3'''<sup>'''rd</sup> '''band = '''25 kHz'''<br>Center of '''lowest''' '''1/3'''<sup>'''rd</sup> '''band = '''100 mHz'''</font>
|Center of '''highest''' '''1/3'''<sup>'''rd</sup> '''band = '''25 kHz'''<br>Center of '''lowest''' '''1/3'''<sup>'''rd</sup> '''band = '''100 mHz'''


|-
|-
|<font size = "1">Frequency span</font>
|Frequency span
|<font size = "1">Highest Band / lowest band''' <nowiki><</nowiki> 2000 '''(ex. 10Hz  - 20kHz)</font>
|Highest Band / lowest band''' <nowiki><</nowiki> 2000 '''(ex. 10Hz  - 20kHz)


|-
|-
|align = "center"|<font size = "1">'''Resolution'''</font>
|align = "center"|'''Resolution'''
|<font size = "1">1/n Octave</font>
|1/n Octave
|<font size = "1">'''1, 1/3<sup> ''''''rd'''</sup>''', 1/12'''<sup>'''th'''</sup>''', 1/24'''<sup>'''th '''</sup></font>
|'''1, 1/3<sup> ''''''rd'''</sup>''', 1/12'''<sup>'''th'''</sup>''', 1/24'''<sup>'''th '''</sup>


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Averaging'''</font>
|align = "center" rowspan = "3"|'''Averaging'''
|<font size = "1">Basic</font>
|Basic
|
|Linear & repeated linear (20 ms to 60,000 s) exponential (20 ms to 60s)
=====Linear & repeated linear (20 ms to 60,000 s) exponential (20 ms to 60s)=====


|-
|-
|<font size = "1">Acoustics </font>
|Acoustics  
|<font size = "1">'''Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse '''</font>
|'''Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse '''


|-
|-
|<font size = "1">Constant Bandwidth * Time</font>
|Constant Bandwidth * Time
|<font size = "1">B*T =''' 0.2dB, 0.5 dB, 1 dB, 2 dB'''</font>
|B*T =''' 0.2dB, 0.5 dB, 1 dB, 2 dB'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Standards'''</font>
|align = "center" rowspan = "2"|'''Standards'''
|<font size = "1">Detectors</font>
|Detectors
|<font size = "1">Complies with '''IEC 651, IEC 804, IEC 61-672 class 1'''</font>
|Complies with '''IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 class 1'''


|-
|-
|<font size = "1">CPB filters </font>
|CPB filters  
|<font size = "1">Complies with '''IEC 1260 class 1'''</font>
|Complies with '''IEC ANSI/ASA S1.11-2014 / Part 1 / IEC 61260:1-2014 and IEC 1260 class 1'''


|}
|}


<font color="#4DA836">'''OCT Settings (continued)'''</font>
 


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "2"|'''Channels'''
|<font size = "1">OR36 & OR38 analyzers</font>
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line -''' 32 to 256 in '''post-analysis'''
|'''32 channels '''per plug-in for '''on line -''' 32 to 256 in '''post-analysis'''


|-
|-
|<font size = "1">OR34 & OR35 analyzers</font>
|OR34 & OR35 analyzers
|<font size = "1">'''8 channels '''for '''on line''' and '''post-analysis'''</font>
|'''8 channels '''for '''on line''' and '''post-analysis'''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Triggering'''</font>
|align = "center" rowspan = "3"|'''Triggering'''
|<font size = "1">Start averaging</font>
|Start averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">End averaging</font>
|End averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">Repeat averaging</font>
|Repeat averaging
|<font size = "1">On '''new start''' or '''end of averaging'''</font>
|On '''new start''' or '''end of averaging'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Overall'''</font>
|align = "center" rowspan = "2"|'''Overall'''
|<font size = "1">Detectors</font>
|Detectors
|<font size = "1">'''2 parallel overall detectors''' linear and weighted (time domain)</font>
|'''2 parallel overall detectors''' linear and weighted (time domain)


|-
|-
|<font size = "1">Weighting</font>
|Weighting
|<font size = "1">'''A''', '''C''', or '''Z''' (none) applicable in '''10 kHz - 40 kHz''' bandwidth</font>
|'''A''', '''C''', or '''Z''' (none) applicable in '''10 kHz - 40 kHz''' bandwidth


|}
|}


<font color="#4DA836">'''OCT Results'''</font>
'''OCT Results'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Spectra'''</font>
|align = "center" rowspan = "2"|'''Spectra'''
|<font size = "1">1/n Octave spectra</font>
|1/n Octave spectra
|<font size = "1">'''Instantaneous '''spectra''', averaged '''spectra</font>
|'''Instantaneous '''spectra''', averaged '''spectra


|-
|-
|<font size = "1">Holden spectra</font>
|Holden spectra
|<font size = "1">'''Minimum '''&''' maximum '''spectra</font>
|'''Minimum '''&''' maximum '''spectra


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Overall'''</font>
|align = "center" rowspan = "2"|'''Overall'''
|<font size = "1">Linear</font>
|Linear
|<font size = "1">'''Digital '''&''' analog '''view meter of time domain overall detector, exact bandwidth is informed</font>
|'''Digital '''&''' analog '''view meter of time domain overall detector, exact bandwidth is informed


|-
|-
|<font size = "1">Weighted</font>
|Weighted
|<font size = "1">'''Digital '''&''' analog '''view meter''' '''of''' A '''or''' C '''weighted''' '''overall levels computed in time domain, exact bandwidth is informed</font>
|'''Digital '''&''' analog '''view meter''' '''of''' A '''or''' C '''weighted''' '''overall levels computed in time domain, exact bandwidth is informed


|}
|}
Line 1,490: Line 1,460:


===Overall acoustics sound level meter - OVA===
===Overall acoustics sound level meter - OVA===
The OVA plug-in analyzer features class sound pressure level measurement according to IEC 61-672 standard. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters and detectors.
The OVA plug-in analyzer features class 1 sound pressure level measurement according to IEC 61672:2013 standard. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters and detectors.


<font color="#4DA836">'''OVA Settings'''</font>
'''OVA Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Type'''</font>
|align = "center" rowspan = "3"|'''Type'''
|<font size = "1">Standard</font>
|Standard
|<font size = "1">'''IEC 61-672-1 - IEC 651 - IEC 804'''</font>
|'''IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 class 1'''


|-
|-
|<font size = "1">Class</font>
|Class
|<font size = "1">'''1'''</font>
|'''1'''


|-
|-
|<font size = "1">Bandwidths</font>
|Bandwidths
|<font size = "1">'''10 Hz '''to '''40 kHz''' - Adjustable</font>
|'''10 Hz '''to '''40 kHz''' - Adjustable


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "2"|'''Channels'''
|<font size = "1">OR36 & OR38 analyzers</font>
|OR36 & OR38 analyzers
|'''32 channels '''per plug-in for '''on line''' - 32 to 256 in '''post-analysis'''
|'''32 channels '''per plug-in for '''on line''' - 32 to 256 in '''post-analysis'''


|-
|-
|<font size = "1">OR34 & OR35 analyzers</font>
|OR34 & OR35 analyzers
|<font size = "1">'''8 channels '''for '''on line''' and '''post-analysis'''</font>
|'''8 channels '''for '''on line''' and '''post-analysis'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Averaging'''</font>
|align = "center" rowspan = "2"|'''Averaging'''
|<font size = "1">Linear</font>
|Linear
|<font size = "1">'''User selectable '''duration '''- repeat '''mode</font>
|'''User selectable '''duration '''- repeat '''mode


|-
|-
|<font size = "1">Short leq</font>
|Short leq
|<font size = "1">'''1/8 s - 1 s - '''User selectable duration</font>
|'''1/8 s - 1 s - '''User selectable duration


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Triggering'''</font>
|align = "center" rowspan = "3"|'''Triggering'''
|<font size = "1">Start averaging</font>
|Start averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">End averaging</font>
|End averaging
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">Repeat averaging</font>
|Repeat averaging
|<font size = "1">On '''new start''' or '''end of averaging'''</font>
|On '''new start''' or '''end of averaging'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Detectors'''</font>
|align = "center" rowspan = "2"|'''Detectors'''
|<font size = "1">Peak</font>
|Peak
|<font size = "1">1 '''peak detector/Ch'''. - '''A''', '''C''' or '''Z''' (none) weighting -'''Independent '''for each channel</font>
|1 '''peak detector/Ch'''. - '''A''', '''C''' or '''Z''' (none) weighting -'''Independent '''for each channel


|-
|-
|<font size = "1">Weighting</font>
|Weighting
|<font size = "1">'''3 overall detectors/Ch. - A''', '''C''' or '''Z''' (none) weighting - '''Fast''', '''Slow''', '''Impulse''' and '''linear''' time weighting - '''Independent''' for each channel</font>
|'''3 overall detectors/Ch. - A''', '''C''' or '''Z''' (none) weighting - '''Fast''', '''Slow''', '''Impulse''' and '''linear''' time weighting - '''Independent''' for each channel


|}
|}


<font color="#4DA836">'''OVA Results'''</font>
'''OVA Results'''


The following results are available for real time or post-analysis display, report and saving. The following levels can be tracked in profiles up to 100 000 pts. each.
The following results are available for real time or post-analysis display, report and saving. The following levels can be tracked in profiles up to 100 000 pts. each.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''SPL'''</font>
|align = "center" rowspan = "3"|'''SPL'''
|<font size = "1">Time weighted</font>
|Time weighted
|<font size = "1">Instant - max hold - min Hold</font>
|Instant - max hold - min Hold


|-
|-
|<font size = "1">Averaged</font>
|Averaged
|<font size = "1">Short leq - leq</font>
|Short leq - leq


|-
|-
|<font size = "1">Peak</font>
|Peak
|<font size = "1">Peak - Time weighted - max hold min hold</font>
|Peak - Time weighted - max hold min hold


|}
|}
Line 1,574: Line 1,544:
The IVC option converts frequency to voltage from the External synch inputs. It covers torsional and acyclism measurements. The converted signals are made available as standard inputs (or player tracks). This option operates real-time computation and operations are possible on-line or in post processing.
The IVC option converts frequency to voltage from the External synch inputs. It covers torsional and acyclism measurements. The converted signals are made available as standard inputs (or player tracks). This option operates real-time computation and operations are possible on-line or in post processing.


<font color="#4DA836">'''IVC Settings & specs'''</font>
'''IVC Settings & specs'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "4"|<font size = "1">'''Rate'''</font>
|align = "center" rowspan = "4"|'''Rate'''
|<font size = "1">Pulse/rev</font>
|Pulse/rev
|<font size = "1">'''0.5 '''to''' 4096 - '''Up to''' 1 E6 '''with pre-divider</font>
|'''0.5 '''to''' 4096 - '''Up to''' 1 E6 '''with pre-divider


|-
|-
|<font size = "1">Pulse/rev Frequency</font>
|Pulse/rev Frequency
|<font size = "1">'''<nowiki>></nowiki> 40 kpulse/sec''''' max''</font>
|'''<nowiki>></nowiki> 40 kpulse/sec''''' max''


|-
|-
|<font size = "1">Pre-divider</font>
|Pre-divider
|<font size = "1">'''1 '''to''' 255 - '''hardware decimation (pulse are not measured)</font>
|'''1 '''to''' 255 - '''hardware decimation (pulse are not measured)


|-
|-
|<font size = "1">Pre-divider frequency</font>
|Pre-divider frequency
|<font size = "1">'''375 kpulse/sec'' '''''max.</font>
|'''375 kpulse/sec'' '''''max.


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Missing pulses'''</font>
|align = "center" rowspan = "3"|'''Missing pulses'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''0 to 5''' consecutive pulses '''integer or fractional'''</font>
|'''0 to 5''' consecutive pulses '''integer or fractional'''


|-
|-
|<font size = "1">Hold off</font>
|Hold off
|<font size = "1">'''2 '''consecutive pulses</font>
|'''2 '''consecutive pulses


|-
|-
|<font size = "1">Generated signal</font>
|Generated signal
|<font size = "1">'''1'''<sup>'''st</sup> ''''''order interpolation'''</font>
|'''1'''<sup>'''st</sup> ''''''order interpolation'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Tach. phase ref'''</font>
|align = "center" rowspan = "2"|'''Tach. phase ref'''
|<font size = "1">No missing pulse</font>
|No missing pulse
|<font size = "1">First random - continuous offset further</font>
|First random - continuous offset further


|-
|-
|<font size = "1">missing pulse <nowiki>></nowiki> 0</font>
|missing pulse <nowiki>></nowiki> 0
|<font size = "1">Last known edge before missing pulse</font>
|Last known edge before missing pulse


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Resolution'''</font>
|align = "center" rowspan = "2"|'''Resolution'''
|<font size = "1">Time</font>
|Time
|<font size = "1">'''160 ns''' : 1/(SF x 64)</font>
|'''160 ns''' : 1/(SF x 64)


|-
|-
|<font size = "1">Angular (with SOA)</font>
|Angular (with SOA)
|<font size = "1">350 µRad (20 mdegres) @ 8000 RPM, up to order 10</font>
|350 µRad (20 mdegres) @ 8000 RPM, up to order 10


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Pre-processing'''</font>
|align = "center" rowspan = "2"|'''Pre-processing'''
|<font size = "1">Smoothing</font>
|Smoothing
|<font size = "1">Sliding average - '''1''' to '''32 '''samples</font>
|Sliding average - '''1''' to '''32 '''samples


|-
|-
|<font size = "1">Filtering</font>
|Filtering
|<font size = "1">'''Any NVGate''' filter (See Filters §) incl. dt and 1/dt </font>
|'''Any NVGate''' filter (See Filters §) incl. dt and 1/dt  


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Max Speed'''</font>
|align = "center" rowspan = "2"|'''Max Speed'''
|<font size = "1">Without pre-divider</font>
|Without pre-divider
|<font size = "1">Max RPM = 2.4 e6/''Pulse per rev'' - 12&nbsp;000 RPM @ 200 pulse/rev</font>
|Max RPM = 2.4 e6/''Pulse per rev'' - 12&nbsp;000 RPM @ 200 pulse/rev


|-
|-
|<font size = "1">With pre-divider</font>
|With pre-divider
|<font size = "1">Max RPM = 36 e6/''Pulse per rev'' - 12&nbsp;000 RPM @ 200 pulse/rev</font>
|Max RPM = 36 e6/''Pulse per rev'' - 12&nbsp;000 RPM @ 200 pulse/rev


|}
|}


<font color="#4DA836">'''IVC Results'''</font>
'''IVC Results'''


The following results are available for real time or post-analysis.
The following results are available for real time or post-analysis.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Signals'''</font>
|align = "center" rowspan = "3"|'''Signals'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''1 '''to '''6 '''(depends on the number of available ext. synch)</font>
|'''1 '''to '''6 '''(depends on the number of available ext. synch)


|-
|-
|<font size = "1">Use</font>
|Use
|<font size = "1">Same as '''dynamic inputs''' or '''recorded''' dynamic inputs</font>
|Same as '''dynamic inputs''' or '''recorded''' dynamic inputs


|-
|-
|<font size = "1">Magnitude</font>
|Magnitude
|<font size = "1">'''Torsional''' acceleration, velocity & angle – separated from angular magnitudes (RPM)</font>
|'''Torsional''' acceleration, velocity & angle – separated from angular magnitudes (RPM)


|-
|-
|align = "center"|<font size = "1">'''View'''</font>
|align = "center"|'''View'''
|<font size = "1">Type</font>
|Type
|<font size = "1">Signal - filtered signal - '''256 samples''' frame</font>
|Signal - filtered signal - '''256 samples''' frame


|}
|}


===Angular sampling (A-Samp)===
===Angular sampling (A-Samp)===
The angular sampling option allows localizing accurately the order phenomenon angle on cyclic rotating machineries. It uses the pulses delivered by and encoder (or zebra tape or gear teeth) to resample the inputs and torsional signal in the SOA plug-in. This option operates real-time computation and operations are possible on-line or in post processing.
The angular sampling option allows accurately localizing the order phenomenon angle on cyclic rotating machineries. It uses the pulses delivered by an encoder (or zebra tape or gear teeth) to resample the inputs and torsional signal in the SOA plug-in. This option operates real-time computation and operations are possible on-line or in post processing.


<font color="#4DA836">'''IVC Settings & specs'''</font>
'''IVC Settings & specs'''


See front-end/ext. Synch § for details
See front-end/ext. Synch § for details
Line 1,680: Line 1,650:
This option operates real-time computation on raw or recorded signal.
This option operates real-time computation on raw or recorded signal.


<font color="#4DA836">'''VIn Settings'''</font>
'''VIn Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "4"|<font size = "1">'''Output'''</font>
|align = "center" rowspan = "4"|'''Output'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''12''' Operators</font>
|'''12''' Operators


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''New item''' in the active '''inputs/tracks list'''</font>
|'''New item''' in the active '''inputs/tracks list'''


|-
|-
|<font size = "1">Synchronization</font>
|Synchronization
|<font size = "1">'''0° phase shift''' with sources and analyzed signal</font>
|'''0° phase shift''' with sources and analyzed signal


|-
|-
|<font size = "1">Magnitude</font>
|Magnitude
|<font size = "1">'''Automatic''' or '''voltage''' if unresolved</font>
|'''Automatic''' or '''voltage''' if unresolved


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Sources'''</font>
|align = "center" rowspan = "2"|'''Sources'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''1''' to '''32 channels'''</font>
|'''1''' to '''32 channels'''


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">Analyzer '''dynamic inputs''' or '''recorded''' dynamic inputs</font>
|Analyzer '''dynamic inputs''' or '''recorded''' dynamic inputs


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Output adjustment'''</font>
|align = "center" rowspan = "4"|'''Output adjustment'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''''(Input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki>power''''' - independent on each output</font>
|'''''(Input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki>power''''' - independent on each output


|-
|-
|<font size = "1">Offset</font>
|Offset
|<font size = "1">Any real value from '''-1 e9 to 1 e9''' - ex: -5.67</font>
|Any real value from '''-1 e9 to 1 e9''' - ex: -5.67


|-
|-
|<font size = "1">Coefficient</font>
|Coefficient
|<font size = "1">Any real value from '''-1 e9 to 1 e9''' - ex: 12</font>
|Any real value from '''-1 e9 to 1 e9''' - ex: 12


|-
|-
|<font size = "1">Power</font>
|Power
|<font size = "1">Any real value from '''-4 to 4''' - ex:  0.5</font>
|Any real value from '''-4 to 4''' - ex:  0.5


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Sources adjustment'''</font>
|align = "center" rowspan = "5"|'''Sources adjustment'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''''Filtering <nowiki>[</nowiki> (input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki> power<nowiki>]</nowiki>''''' - independent for each source</font>
|'''''Filtering <nowiki>[</nowiki> (input * Coeff <nowiki>+</nowiki> Offset ) <nowiki>^</nowiki> power<nowiki>]</nowiki>''''' - independent for each source


|-
|-
|<font size = "1">Offset</font>
|Offset
|<font size = "1">Any real value from '''-1 e9 to 1 e9''' - ex: -5.67 e-5</font>
|Any real value from '''-1 e9 to 1 e9''' - ex: -5.67 e-5


|-
|-
|<font size = "1">Coefficient</font>
|Coefficient
|<font size = "1">Any real value from '''-1 e9 to 1 e9''' - ex: 1200</font>
|Any real value from '''-1 e9 to 1 e9''' - ex: 1200


|-
|-
|<font size = "1">Power</font>
|Power
|<font size = "1">Any real value from '''-4 to 4''' - ex:  -2.4567</font>
|Any real value from '''-4 to 4''' - ex:  -2.4567


|-
|-
|<font size = "1">Filtering</font>
|Filtering
|<font size = "1">Any NVGate filter</font>
|Any NVGate filter


|-
|-
|align = "center"|<font size = "1">'''Operators'''</font>
|align = "center"|'''Operators'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Product''' - '''Sum''' - independent on each operator</font>
|'''Product''' - '''Sum''' - independent on each operator


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Twist'''</font>
|align = "center" rowspan = "3"|'''Twist'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Phase comparison''' – output in '''plane angle'''</font>
|'''Phase comparison''' – output in '''plane angle'''


|-
|-
|<font size = "1">Correction (zeroing) offset</font>
|Correction (zeroing) offset
|<font size = "1">'''-360° '''to''' <nowiki>+</nowiki>360°'''</font>
|'''-360° '''to''' <nowiki>+</nowiki>360°'''


|-
|-
|<font size = "1">Output</font>
|Output
|<font size = "1">'''Plane angle''' or '''torque''' with conversion factor</font>
|'''Plane angle''' or '''torque''' with conversion factor


|}
|}
Line 1,769: Line 1,739:
The virtual parameter (VDC) allows combining front-end parametric inputs (DC) with math operations to generate a calculated parameter (new DC). This option covers numerous applications such as averaged temperature or non linear response transducers. The virtual parameters are made available as standard DC input. This option operates real-time computation on raw or recorded signal.
The virtual parameter (VDC) allows combining front-end parametric inputs (DC) with math operations to generate a calculated parameter (new DC). This option covers numerous applications such as averaged temperature or non linear response transducers. The virtual parameters are made available as standard DC input. This option operates real-time computation on raw or recorded signal.


<font color="#4DA836">'''VDC Settings'''</font>
'''VDC Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "4"|<font size = "1">'''Output'''</font>
|align = "center" rowspan = "4"|'''Output'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''12''' Operators</font>
|'''12''' Operators


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''New item''' in the DC inputs/track list</font>
|'''New item''' in the DC inputs/track list


|-
|-
|<font size = "1">Synchronization</font>
|Synchronization
|<font size = "1">'''0° phase shift''' with sources and analyzed signal</font>
|'''0° phase shift''' with sources and analyzed signal


|-
|-
|<font size = "1">Magnitude</font>
|Magnitude
|<font size = "1">'''User defined'''</font>
|'''User defined'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Sources'''</font>
|align = "center" rowspan = "2"|'''Sources'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''1''' to '''32 channels'''</font>
|'''1''' to '''32 channels'''


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">Analyzer '''DC inputs''' or '''recorded''' DC inputs</font>
|Analyzer '''DC inputs''' or '''recorded''' DC inputs


|-
|-
|align = "center" rowspan = "7"|<font size = "1">'''Editor'''</font>
|align = "center" rowspan = "7"|'''Editor'''
|<font size = "1">Type</font>
|Type
|<font size = "1">Text</font>
|Text


|-
|-
|<font size = "1">Variables</font>
|Variables
|<font size = "1">'''Real''' values </font>
|'''Real''' values  


|-
|-
|<font size = "1">Operators</font>
|Operators
|<font size = "1">'''<nowiki>+</nowiki>''', '''-''', '''x''', '''/''', '''<nowiki>^</nowiki>''', '''=''' (affectation)</font>
|'''<nowiki>+</nowiki>''', '''-''', '''x''', '''/''', '''<nowiki>^</nowiki>''', '''=''' (affectation)


|-
|-
|<font size = "1">Predefined </font>
|Predefined  
|<font size = "1">'''Ch x''' (Channels signals), '''Pi''', '''e'''</font>
|'''Ch x''' (Channels signals), '''Pi''', '''e'''


|-
|-
|<font size = "1">Trigonometric functions</font>
|Trigonometric functions
|<font size = "1">'''Sine''', '''Cosine''', '''Tangent''' - with '''Arc''' and '''Hyperbolic''' combination </font>
|'''Sine''', '''Cosine''', '''Tangent''' - with '''Arc''' and '''Hyperbolic''' combination  


|-
|-
|<font size = "1">Logarithmic functions</font>
|Logarithmic functions
|<font size = "1">'''Log2''', '''Log10''', '''Ln''', '''Exp'''</font>
|'''Log2''', '''Log10''', '''Ln''', '''Exp'''


|-
|-
|<font size = "1">Miscellaneous</font>
|Miscellaneous
|<font size = "1">'''Sqrt''', '''Abs, Min, Max, Sum, Avg'''</font>
|'''Sqrt''', '''Abs, Min, Max, Sum, Avg'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Output characterization'''</font>
|align = "center" rowspan = "2"|'''Output characterization'''
|<font size = "1">Limits</font>
|Limits
|<font size = "1">Min and max: '''<nowiki>+</nowiki>/- 1 E9 '''- In current unit</font>
|Min and max: '''<nowiki>+</nowiki>/- 1 E9 '''- In current unit


|-
|-
|<font size = "1">Information</font>
|Information
|<font size = "1">'''Label''' (text)</font>
|'''Label''' (text)


|}
|}
Line 1,840: Line 1,810:
The CTE option combines 2 actual tachometers with math to generates a calculated angular velocity (new tachometer). It covers various unreachable rotating parts and more specially the CVT gear chain measurements. The combined tachometers are made available as standard tachometer. This option operates real-time computation and operations are possible on-line or post processing.
The CTE option combines 2 actual tachometers with math to generates a calculated angular velocity (new tachometer). It covers various unreachable rotating parts and more specially the CVT gear chain measurements. The combined tachometers are made available as standard tachometer. This option operates real-time computation and operations are possible on-line or post processing.


<font color="#4DA836">'''CTE Settings'''</font>
'''CTE Settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Sources'''</font>
|align = "center" rowspan = "2"|'''Sources'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2'''</font>
|'''2'''


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Any NVGate tachometer – '''excepted another CTE</font>
|'''Any NVGate tachometer – '''excepted another CTE


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Output'''</font>
|align = "center" rowspan = "2"|'''Output'''
|<font size = "1">Type</font>
|Type
|<font size = "1">New item in the active tachometers list</font>
|New item in the active tachometers list


|-
|-
|<font size = "1">Synchronization</font>
|Synchronization
|<font size = "1">'''0° phase shift''' with sources and analyzed signal</font>
|'''0° phase shift''' with sources and analyzed signal


|-
|-
|align = "center" rowspan = "7"|<font size = "1">'''Editor'''</font>
|align = "center" rowspan = "7"|'''Editor'''
|<font size = "1">Type</font>
|Type
|<font size = "1">Text</font>
|Text


|-
|-
|<font size = "1">Variables</font>
|Variables
|<font size = "1">'''Real''' values </font>
|'''Real''' values  


|-
|-
|<font size = "1">Operators</font>
|Operators
|<font size = "1">'''<nowiki>+</nowiki>''', '''-''', '''x''', '''/''', '''<nowiki>^</nowiki>''', '''=''' (affectation)</font>
|'''<nowiki>+</nowiki>''', '''-''', '''x''', '''/''', '''<nowiki>^</nowiki>''', '''=''' (affectation)


|-
|-
|<font size = "1">Predefined </font>
|Predefined  
|<font size = "1">Rmp1, Rpm2 (sources speed), '''Pi''', '''e'''</font>
|Rmp1, Rpm2 (sources speed), '''Pi''', '''e'''


|-
|-
|<font size = "1">Trigonometric functions</font>
|Trigonometric functions
|<font size = "1">'''Sine''', '''Cosine''', '''Tangent''' - with '''Arc''' and '''Hyperbolic''' combination </font>
|'''Sine''', '''Cosine''', '''Tangent''' - with '''Arc''' and '''Hyperbolic''' combination  


|-
|-
|<font size = "1">Logarithmic functions</font>
|Logarithmic functions
|<font size = "1">'''Log2''', '''Log10''', '''Ln''', '''Exp'''</font>
|'''Log2''', '''Log10''', '''Ln''', '''Exp'''


|-
|-
|<font size = "1">Miscellaneous</font>
|Miscellaneous
|<font size = "1">'''Sqrt''', '''Abs, Min, Max, Sum, Avg, Sign, If, Rint'''</font>
|'''Sqrt''', '''Abs, Min, Max, Sum, Avg, Sign, If, Rint'''


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Output characterization'''</font>
|align = "center" rowspan = "4"|'''Output characterization'''
|<font size = "1">Rotation</font>
|Rotation
|<font size = "1">Clock wise or counter clock wise</font>
|Clock wise or counter clock wise


|-
|-
|<font size = "1">Average</font>
|Average
|<font size = "1">'''Sliding''' - '''1''' to '''256 '''revolution</font>
|'''Sliding''' - '''1''' to '''256 '''revolution


|-
|-
|<font size = "1">Speed limits</font>
|Speed limits
|<font size = "1">Min: '''0 RPM''' - max: '''1&nbsp;200&nbsp;000 RPM'''</font>
|Min: '''0 RPM''' - max: '''1&nbsp;200&nbsp;000 RPM'''


|-
|-
|<font size = "1">Information</font>
|Information
|<font size = "1">Label (text)</font>
|Label (text)


|}
|}


<font color="#4DA836">'''CTE Results'''</font>
'''CTE Results'''


The following results are available for real time or post-analysis  
The following results are available for real time or post-analysis  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Signals'''</font>
|align = "center" rowspan = "3"|'''Signals'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''4'''</font>
|'''4'''


|-
|-
|<font size = "1">Use</font>
|Use
|<font size = "1">Same as on-line '''Tachometer '''or '''recorded''' tachometer </font>
|Same as on-line '''Tachometer '''or '''recorded''' tachometer  


|-
|-
|<font size = "1">Magnitude</font>
|Magnitude
|<font size = "1">Angular velocity (RPM)</font>
|Angular velocity (RPM)


|-
|-
|align = "center"|<font size = "1">'''View'''</font>
|align = "center"|'''View'''
|<font size = "1">Types</font>
|Types
|<font size = "1">'''View meter''' - '''1''' to '''100 s''' rolling '''profile'''</font>
|'''View meter''' - '''1''' to '''100 s''' rolling '''profile'''


|}
|}


==General purpose analysis==
==General purpose analysis==
The following modules are available as standard features of NVGate<sup>®</sup> software platform; they feature additional analysis capabilities to regular plug-in analysis module.
The following modules are available as standard features of NVGate<sup>®</sup> software platform; they feature additional analysis capabilities to regular plug-in analysis modules.


===Monitor===
===Monitor===
The monitor is an independent (dedicated processor) module that processes continuously FFT analysis on 4 channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.
The monitor is an independent (dedicated processor) module that continuously processes FFT analysis on 4 channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.


<font color="#4DA836">'''Monitor settings'''</font>
'''Monitor settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "7"|<font size = "1">'''Fixed setup'''</font>
|align = "center" rowspan = "7"|'''Fixed setup'''
|<font size = "1">Bandwidth</font>
|Bandwidth
|<font size = "1">'''ABW = Sampling freq / 2.56''' </font>
|'''ABW = Sampling freq / 2.56'''  


|-
|-
|<font size = "1">Resolution</font>
|Resolution
|<font size = "1">'''401 lines'''</font>
|'''401 lines'''


|-
|-
|<font size = "1">Average domain</font>
|Average domain
|<font size = "1">'''Spectral '''(power)</font>
|'''Spectral '''(power)


|-
|-
|<font size = "1">Overlap</font>
|Overlap
|<font size = "1">'''0%'''</font>
|'''0%'''


|-
|-
|<font size = "1">Average type</font>
|Average type
|<font size = "1">'''Exponential''' </font>
|'''Exponential'''  


|-
|-
|<font size = "1">Weighting windows</font>
|Weighting windows
|<font size = "1">'''Hanning'''</font>
|'''Hanning'''


|-
|-
|<font size = "1">Trigger</font>
|Trigger
|<font size = "1">'''Free run'''</font>
|'''Free run'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "2"|'''Channels'''
|<font size = "1">Number</font>
|Number
|<font size = "1">'''4 '''</font>
|'''4 '''


|-
|-
|<font size = "1">Swap</font>
|Swap
|<font size = "1">Between '''any active dynamic input - hot swap '''capable</font>
|Between '''any active dynamic input - hot swap '''capable


|-
|-
|align = "center"|<font size = "1">'''Average'''</font>
|align = "center"|'''Average'''
|<font size = "1">Duration</font>
|Duration
|<font size = "1">'''Instantaneous '''(20 ms) to''' unlimited'''</font>
|'''Instantaneous '''(20 ms) to''' unlimited'''


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Overall analysis'''</font>
|align = "center" rowspan = "4"|'''Overall analysis'''
|<font size = "1">Pass band filter</font>
|Pass band filter
|<font size = "1">'''Butterworth''' order '''2''' to '''10 -''' '''IIR''' type - '''Bypass''' function</font>
|'''Butterworth''' order '''2''' to '''10 -''' '''IIR''' type - '''Bypass''' function


|-
|-
|<font size = "1">Upper & lower frequency</font>
|Upper & lower frequency
|<font size = "1">'''User defined - '''can be graphically modified</font>
|'''User defined - '''can be graphically modified


|-
|-
|<font size = "1">Computed indicators</font>
|Computed indicators
|<font size = "1">'''DC - Max - Min - RMS - Skew - Kurtosis - '''Into defined pass band - '''overall''' if bypass is on</font>
|'''DC - Max - Min - RMS - Skew - Kurtosis - '''Into defined pass band - '''overall''' if bypass is on


|-
|-
|<font size = "1">Average</font>
|Average
|<font size = "1">'''User defined '''duration''' - independent '''from spectral averaging</font>
|'''User defined '''duration''' - independent '''from spectral averaging


|}
|}


<font color="#4DA836">'''Monitor results and connections'''</font>
'''Monitor results and connections'''


The following results are available for real time or post-analysis display, connection to other modules.
The following results are available for real time or post-analysis display, connection to other modules.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center"|<font size = "1">'''Time domain'''</font>
|align = "center"|'''Time domain'''
|<font size = "1">Play back</font>
|Play back
|<font size = "1">On any output generator - hot swappable</font>
|On any output generator - hot swappable


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''FFT analyzer'''</font>
|align = "center" rowspan = "2"|'''FFT analyzer'''
|<font size = "1">Trigger bloc</font>
|Trigger bloc
|<font size = "1">'''1024 samples''' - time domain '''analyzed bloc'''</font>
|'''1024 samples''' - time domain '''analyzed bloc'''


|-
|-
|<font size = "1">Spectra</font>
|Spectra
|<font size = "1">Instantaneous '''complex spectra''' - averaged '''power spectra'''</font>
|Instantaneous '''complex spectra''' - averaged '''power spectra'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Overall detectors'''</font>
|align = "center" rowspan = "2"|'''Overall detectors'''
|<font size = "1">Display (digital or analog view-meter)</font>
|Display (digital or analog view-meter)
|
|DC - Max - Min - RMS - Skew - Kurtosis
=====DC - Max - Min - RMS - Skew - Kurtosis=====


|-
|-
|<font size = "1">Connection</font>
|Connection
|<font size = "1">To '''level''' and '''delta level''' event detector - to '''waterfall profiles'''</font>
|To '''level''' and '''delta level''' event detector - to '''waterfall profiles'''


|}
|}
Line 2,035: Line 2,004:
The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in analyzers and sources in one result.
The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in analyzers and sources in one result.


<font color="#4DA836">'''Waterfall settings'''</font>
'''Waterfall settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "2"|<font size = "1">'''Mode'''</font>
|align = "center" rowspan = "2"|'''Mode'''
|<font size = "1">Continuous</font>
|Continuous
|<font size = "1">'''Circular buffer''' of results</font>
|'''Circular buffer''' of results


|-
|-
|<font size = "1">One shot</font>
|One shot
|<font size = "1">Fill stack and stop waterfall acquisition</font>
|Fill stack and stop waterfall acquisition


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Triggering'''</font>
|align = "center" rowspan = "3"|'''Triggering'''
|<font size = "1">Start acquisition</font>
|Start acquisition
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">Stop acquisition</font>
|Stop acquisition
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' plus '''manual '''and '''free run'''


|-
|-
|<font size = "1">New slice (point or spectra)</font>
|New slice (point or spectra)
|<font size = "1">Any '''NVGate'''<sup>'''® '''</sup>'''event''' - '''manual - free run - '''availability of '''connected results''' - '''periodic'''</font>
|Any '''NVGate'''<sup>'''® '''</sup>'''event''' - '''manual - free run - '''availability of '''connected results''' - '''periodic'''


|}
|}


<font color="#4DA836">'''Waterfall settings (continued)'''</font>
'''Waterfall settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Size'''</font>
|align = "center" rowspan = "3"|'''Size'''
|<font size = "1">Number of result per acquisition</font>
|Number of result per acquisition
|<font size = "1">Up to '''95 results <nowiki>+</nowiki> Time'''</font>
|Up to '''95 results <nowiki>+</nowiki> Time'''


|-
|-
|<font size = "1">Depth</font>
|Depth
|<font size = "1">'''2''' to '''100,000''' slices or points - depend on PC available memory and requested result - '''automatically adjusted '''before acquisition</font>
|'''2''' to '''100,000''' slices or points - depend on PC available memory and requested result - '''automatically adjusted '''before acquisition


|-
|-
|<font size = "1">Depth for stand-alone acquisition</font>
|Depth for stand-alone acquisition
|<font size = "1">''' '''Up to''' 3 MSamples '''per computation DSP''' '''</font>
|''' '''Up to''' 3 MSamples '''per computation DSP''' '''


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Channels'''</font>
|align = "center" rowspan = "4"|'''Channels'''
|<font size = "1">'''Type scalar''' (the following results are connectable to waterfall channels)</font>
|'''Type scalar''' (the following results are connectable to waterfall channels)
|<font size = "1">'''Overall levels''' (lin and weighted) from OvA and 1/n OCT - '''Monitor indicators''' (DC, Max, Min, RMS, Kurtosis) - '''Orders''' (from CBT and SOA) - '''Complex orders''' (magnitude & phase) - '''Overall levels '''in selected BW from FFT & SOA (order or frequency) - '''TDA''' scalar: '''DC, Min/max, RMS, Kurtosis, peak, peak-peak, crest factor'''</font>
|'''Overall levels''' (lin and weighted) from OvA and 1/n OCT - '''Monitor indicators''' (DC, Max, Min, RMS, Kurtosis) - '''Orders''' (from CBT and SOA) - '''Complex orders''' (magnitude & phase) - '''Overall levels '''in selected BW from FFT & SOA (order or frequency) - '''TDA''' scalar: '''DC, Min/max, RMS, Kurtosis, peak, peak-peak, crest factor'''


|-
|-
|<font size = "1">'''Type 2D - FFT''' (the following results are connectable to waterfall channels)</font>
|'''Type 2D - FFT''' (the following results are connectable to waterfall channels)
|<font size = "1">'''Triggered '''block''' - Averaged Triggered '''block''' - Weighted '''Block''' - Complex '''spectra''' - Power '''spectra''' -Cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence - Zoomed''' complex spectra - '''zoomed power '''spectra</font>
|'''Triggered '''block''' - Averaged Triggered '''block''' - Weighted '''Block''' - Complex '''spectra''' - Power '''spectra''' -Cross-spectra''' - '''FRF H1''' - '''FRF H2''' - '''Coherence - Zoomed''' complex spectra - '''zoomed power '''spectra


|-
|-
|<font size = "1">'''Type 2D - OCT''' (the following results are connectable to waterfall channels)</font>
|'''Type 2D - OCT''' (the following results are connectable to waterfall channels)
|<font size = "1">'''Instantaneous''' spectra - '''averaged '''spectra - '''max  & min''' '''hold '''spectra</font>
|'''Instantaneous''' spectra - '''averaged '''spectra - '''max  & min''' '''hold '''spectra


|-
|-
|<font size = "1">'''Type 2D - SOA''' (the following results are connectable to waterfall channels)</font>
|'''Type 2D - SOA''' (the following results are connectable to waterfall channels)
|<font size = "1">'''Triggered '''block''' - Averaged Triggered '''block''' - Weighted '''block''' - Complex order '''spectra''' - Power order '''spectra</font>
|'''Triggered '''block''' - Averaged Triggered '''block''' - Weighted '''block''' - Complex order '''spectra''' - Power order '''spectra


|}
|}


<font color="#4DA836">'''Waterfall results'''</font>
'''Waterfall results'''


The following results are available for real time or post-analysis display, report and saving. All stacked result can be saved.
The following results are available for real time or post-analysis display, report and saving. All stacked results can be saved.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''3D'''</font>
|align = "center" rowspan = "3"|'''3D'''
|<font size = "1">Display</font>
|Display
|<font size = "1">'''1 pane''' (3D) - '''2 panes''' (3D <nowiki>+</nowiki> YZ view or 3D <nowiki>+</nowiki> XY view) , '''3 panes''' (3D <nowiki>+</nowiki> YZ <nowiki>+</nowiki> XY <nowiki>+</nowiki> Extraction view) windows - automatic or user selectable '''pane arrangement - Real or imaginary '''part and '''module or phase''' for complex results</font>
|'''1 pane''' (3D) - '''2 panes''' (3D <nowiki>+</nowiki> YZ view or 3D <nowiki>+</nowiki> XY view) , '''3 panes''' (3D <nowiki>+</nowiki> YZ <nowiki>+</nowiki> XY <nowiki>+</nowiki> Extraction view) windows - automatic or user selectable '''pane arrangement - Real or imaginary '''part and '''module or phase''' for complex results


|-
|-
|<font size = "1">Z axis (X axis for extraction and YZ view)</font>
|Z axis (X axis for extraction and YZ view)
|<font size = "1">'''Any reference''' - '''time''' - independent for any window - swap reference at any time </font>
|'''Any reference''' - '''time''' - independent for any window - swap reference at any time  


|-
|-
|<font size = "1">Saving selection</font>
|Saving selection
|<font size = "1">'''Entire 3D data '''and / or any active''' section '''(YZ, XY or Extraction)</font>
|'''Entire 3D data '''and / or any active''' section '''(YZ, XY or Extraction)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Profiles'''</font>
|align = "center" rowspan = "2"|'''Profiles'''
|<font size = "1">Display</font>
|Display
|<font size = "1">Profile of any scalar - '''Real or imaginary '''part and '''module or phase''' for complex results</font>
|Profile of any scalar - '''Real or imaginary '''part and '''module or phase''' for complex results


|-
|-
|<font size = "1">X axis</font>
|X axis
|<font size = "1">'''Any reference''' - '''Time''' – '''Slice''' - independent for any window - swap reference at any time</font>
|'''Any reference''' - '''Time''' – '''Slice''' - independent for any window - swap reference at any time


|}
|}


<font color="#4DA836">'''Waterfall Tools'''</font>
'''Waterfall Tools'''


The following results are available for real time or post-analysis display, report and saving.
The following results are available for real time or post-analysis display, report and saving.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|rowspan = "4"|<font size = "1">Sections</font>
|rowspan = "4"|Sections
|<font size = "1">YZ Sections (Profiles vs. Ref)</font>
|YZ Sections (Profiles vs. Ref)
|<font size = "1">Any order/frequency - '''Power''' - '''Peak''' - on selected bandwidth</font>
|Any order/frequency - '''Power''' - '''Peak''' - on selected bandwidth


|-
|-
|<font size = "1">XY Sections (Spectra)</font>
|XY Sections (Spectra)
|<font size = "1">Any position in the current ref. - '''Min''' - '''Max''' - '''Average''' - on selected range or all slices </font>
|Any position in the current ref. - '''Min''' - '''Max''' - '''Average''' - on selected range or all slices  


|-
|-
|<font size = "1">Order/Frequency extraction (profiles vs. Ref.)</font>
|Order/Frequency extraction (profiles vs. Ref.)
|<font size = "1">Any order/frequency extraction - '''user selectable tach'''. - '''Power''' - '''Peak''' - on selectable bandwidth - '''Max order '''</font>
|Any order/frequency extraction - '''user selectable tach'''. - '''Power''' - '''Peak''' - on selectable bandwidth - '''Max order '''


|-
|-
|<font size = "1">Number of sections</font>
|Number of sections
|<font size = "1">'''Unlimited''' - available on result or real-time waterfalls</font>
|'''Unlimited''' - available on result or real-time waterfalls


|-
|-
|rowspan = "3"|<font size = "1">Linked cursors</font>
|rowspan = "3"|Linked cursors
|<font size = "1">General</font>
|General
|
|Single or dual cursor in each view - linked with other graphs
=====Single or dual cursor in each view - linked with other graphs=====


|-
|-
|<font size = "1">Linked by value</font>
|Linked by value
|
|Linked cursors track the same X value in different windows.
=====Linked cursors track the same X value in different windows.=====


|-
|-
|<font size = "1">Linked by reference</font>
|Linked by reference
|
|Linked cursors track the same acquisition slice in different waterfall windows with different X or Z-axis.
=====Linked cursors track the same acquisition slice in different waterfall windows with different X or Z-axis.=====


|}
|}
Line 2,170: Line 2,136:
Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the Event module.
Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the Event module.


<font color="#4DA836">'''Event settings'''</font>
'''Event settings'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "8"|<font size = "1">'''Edge detector'''</font>
|align = "center" rowspan = "8"|'''Edge detector'''
|<font size = "1">Source</font>
|Source
|<font size = "1">'''Any dynamic input''' - Any '''recorded '''dynamic input</font>
|'''Any dynamic input''' - Any '''recorded '''dynamic input


|-
|-
|<font size = "1">Label</font>
|Label
|<font size = "1">String for '''event identification''' (Ex. "Impact" for a hammer impact detection)</font>
|String for '''event identification''' (Ex. "Impact" for a hammer impact detection)


|-
|-
|<font size = "1">Threshold</font>
|Threshold
|<font size = "1">Between '''min and max range''' - use '''source unit''' (ex. g for an accelerometer)</font>
|Between '''min and max range''' - use '''source unit''' (ex. g for an accelerometer)


|-
|-
|<font size = "1">Pre-filtering</font>
|Pre-filtering
|<font size = "1">'''A '''law - '''C''' law - any '''NVGate filter'''</font>
|'''A '''law - '''C''' law - any '''NVGate filter'''


|-
|-
|<font size = "1">Slope</font>
|Slope
|<font size = "1">'''Rise '''-''' fall '''</font>
|'''Rise '''-''' fall '''


|-
|-
|<font size = "1">Hold off</font>
|Hold off
|<font size = "1">'''0 '''to '''36000''' s</font>
|'''0 '''to '''36000''' s


|-
|-
|<font size = "1">Hystersis</font>
|Hystersis
|<font size = "1">Into source''' input range''' - use '''source unit''' (ex. g for an accelerometer)</font>
|Into source''' input range''' - use '''source unit''' (ex. g for an accelerometer)


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2 '''edge detectors</font>
|'''2 '''edge detectors


|-
|-
|align = "center" rowspan = "7"|<font size = "1">'''RPM speed detector'''</font>
|align = "center" rowspan = "7"|'''RPM speed detector'''
|<font size = "1">Source</font>
|Source
|<font size = "1">'''Any tachometer - '''computed from '''dynamic input''' or '''ext.sync '''-''' '''any r'''ecorded pulses '''(through tachometer)</font>
|'''Any tachometer - '''computed from '''dynamic input''' or '''ext.sync '''-''' '''any r'''ecorded pulses '''(through tachometer)


|-
|-
|<font size = "1">Label</font>
|Label
|<font size = "1">String for '''event identification''' (Ex. "start speed" for a run up initial triggering)</font>
|String for '''event identification''' (Ex. "start speed" for a run up initial triggering)


|-
|-
|<font size = "1">Threshold</font>
|Threshold
|<font size = "1">Detection into''' '''source''' '''tachometer''' '''RPM range </font>
|Detection into''' '''source''' '''tachometer''' '''RPM range  


|-
|-
|<font size = "1">Slope</font>
|Slope
|<font size = "1">'''Run up '''-''' Run down '''</font>
|'''Run up '''-''' Run down '''


|-
|-
|<font size = "1">Hysteresis </font>
|Hysteresis  
|<font size = "1">'''0''' to '''source '''max RPM</font>
|'''0''' to '''source '''max RPM


|-
|-
|<font size = "1">Interpolation</font>
|Interpolation
|<font size = "1">'''On''' interpolate event occurrence '''into revolution''' - '''Off''' event '''occur at revolution ends''' (new pulse)</font>
|'''On''' interpolate event occurrence '''into revolution''' - '''Off''' event '''occur at revolution ends''' (new pulse)


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2 to 6 '''RPM speed detector</font>
|'''2 to 6 '''RPM speed detector


|-
|-
|align = "center" rowspan = "8"|<font size = "1">'''Delta RPM speed detector'''</font>
|align = "center" rowspan = "8"|'''Delta RPM speed detector'''
|<font size = "1">Source</font>
|Source
|<font size = "1">'''Any tachometer - '''computed from '''dynamic input''' or '''ext.sync '''-''' '''any r'''ecorded pulses '''(through tachometer)</font>
|'''Any tachometer - '''computed from '''dynamic input''' or '''ext.sync '''-''' '''any r'''ecorded pulses '''(through tachometer)


|-
|-
|<font size = "1">Label</font>
|Label
|<font size = "1">String for '''event identification''' (Ex. "Each 100 RPM" for a run up triggering)</font>
|String for '''event identification''' (Ex. "Each 100 RPM" for a run up triggering)


|-
|-
|<font size = "1">Lower velocity</font>
|Lower velocity
|<font size = "1">'''Minimum''' RPM speed - Events occurs only for higher source speed''' '''</font>
|'''Minimum''' RPM speed - Events occurs only for higher source speed''' '''


|-
|-
|<font size = "1">Higher velocity</font>
|Higher velocity
|<font size = "1">'''Maximum''' RPM speed - Events occurs only for lower source speed</font>
|'''Maximum''' RPM speed - Events occurs only for lower source speed


|-
|-
|<font size = "1">Delta velocity</font>
|Delta velocity
|<font size = "1">Define '''velocity step''' - '''event occur each time''' source speed increase or decrease by delta velocity</font>
|Define '''velocity step''' - '''event occur each time''' source speed increase or decrease by delta velocity


|-
|-
|<font size = "1">Slope</font>
|Slope
|<font size = "1">'''Run up''' - '''run down''' - '''first,''' first slope is automatically selected - '''any,''' event occur on any slope</font>
|'''Run up''' - '''run down''' - '''first,''' first slope is automatically selected - '''any,''' event occur on any slope


|-
|-
|<font size = "1">Interpolation</font>
|Interpolation
|<font size = "1">'''On,''' interpolate event occurrence '''into revolution''' - '''Off,''' event '''occur at revolution end''' (new pulse)</font>
|'''On,''' interpolate event occurrence '''into revolution''' - '''Off,''' event '''occur at revolution end''' (new pulse)


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2 to 6 delta RPM''' speed detector</font>
|'''2 to 6 delta RPM''' speed detector


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Level detector'''</font>
|align = "center" rowspan = "5"|'''Level detector'''
|<font size = "1">Source</font>
|Source
|<font size = "1">Any '''parametric (DC)''' input - '''Monitor scalar''' - '''Filtered''' monitor scalar (Band Pass)</font>
|Any '''parametric (DC)''' input - '''Monitor scalar''' - '''Filtered''' monitor scalar (Band Pass)


|-
|-
|<font size = "1">Label</font>
|Label
|<font size = "1">String for '''event identification''' (Ex. "temp A" for a recording trigger)</font>
|String for '''event identification''' (Ex. "temp A" for a recording trigger)


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">Monitor scalar - DC - RMS - Max - Min - Kurtosis</font>
|Monitor scalar - DC - RMS - Max - Min - Kurtosis


|-
|-
|<font size = "1">Status</font>
|Status
|<font size = "1">'''Above''' /''' below''' - Detection level - available in '''lin''' or '''dB ''' </font>
|'''Above''' /''' below''' - Detection level - available in '''lin''' or '''dB '''  


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''1''' to '''4 '''level detectors</font>
|'''1''' to '''4 '''level detectors


|}
|}


<font color="#4DA836">'''Event settings (continued)'''</font>
'''Event settings (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''Periodic event'''</font>
|align = "center" rowspan = "3"|'''Periodic event'''
|<font size = "1">Label</font>
|Label
|<font size = "1">String for '''event identification''' (Ex. "Hourly" for 1 hour time interval)</font>
|String for '''event identification''' (Ex. "Hourly" for 1 hour time interval)


|-
|-
|<font size = "1">Time interval</font>
|Time interval
|<font size = "1">'''2 ms''' to '''11 days''' - '''Synchronized''' with analyzer.</font>
|'''2 ms''' to '''11 days''' - '''Synchronized''' with analyzer.


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2 '''periodic events</font>
|'''2 '''periodic events


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Combination'''</font>
|align = "center" rowspan = "4"|'''Combination'''
|<font size = "1">Sources</font>
|Sources
|<font size = "1">'''Any event - 2 '''different sources''' '''defined as''' A '''and''' B'''</font>
|'''Any event - 2 '''different sources''' '''defined as''' A '''and''' B'''


|-
|-
|<font size = "1">Label</font>
|Label
|<font size = "1">String for '''event identification''' (Ex. "Time/RPM variation" for a mix between periodic and delta RPM)</font>
|String for '''event identification''' (Ex. "Time/RPM variation" for a mix between periodic and delta RPM)


|-
|-
|<font size = "1">Combination </font>
|Combination  
|<font size = "1">A '''OR''' B - A '''AFTER''' B - A '''AND''' B (occur in the same bloc)</font>
|A '''OR''' B - A '''AFTER''' B - A '''AND''' B (occur in the same bloc)


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2 '''event combination''' '''</font>
|'''2 '''event combination''' '''


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Plug-in synchronization'''</font>
|align = "center" rowspan = "5"|'''Plug-in synchronization'''
|<font size = "1">FFTx result available </font>
|FFTx result available  
|<font size = "1">Occurs at each '''new spectra availability''' - Occurs at '''end of linear averaging''' (repeat and linear) - One event per active FFT plug in (FFT1 to FFT4)</font>
|Occurs at each '''new spectra availability''' - Occurs at '''end of linear averaging''' (repeat and linear) - One event per active FFT plug in (FFT1 to FFT4)


|-
|-
|<font size = "1">OCT result available </font>
|OCT result available  
|<font size = "1">Occurs at each new 1/n spectra availability - Occur at '''end of linear averaging''' (leq, short leq, linear repeat) </font>
|Occurs at each new 1/n spectra availability - Occur at '''end of linear averaging''' (leq, short leq, linear repeat)  


|-
|-
|<font size = "1">SOAx result available </font>
|SOAx result available  
|<font size = "1">Occurs at each '''new spectra availability''' - Occurs at '''end of linear averaging''' (repeat and linear) - One event per active SOA plug in (up to 2)</font>
|Occurs at each '''new spectra availability''' - Occurs at '''end of linear averaging''' (repeat and linear) - One event per active SOA plug in (up to 2)


|-
|-
|<font size = "1">TDA result available </font>
|TDA result available  
|<font size = "1">Occurs at each new level set ( RMS, DC, etc) availability - Occur at '''end of linear averaging''' (repeat and linear) </font>
|Occurs at each new level set ( RMS, DC, etc) availability - Occur at '''end of linear averaging''' (repeat and linear)  


|-
|-
|<font size = "1">OVA result available </font>
|OVA result available  
|<font size = "1">Occurs at each new set of overall level (short LEQ) availability - Occur at '''end of short  averaging''' (repeat and linear)</font>
|Occurs at each new set of overall level (short LEQ) availability - Occur at '''end of short  averaging''' (repeat and linear)


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Generators Synchronization'''</font>
|align = "center" rowspan = "2"|'''Generators Synchronization'''
|<font size = "1">Event synchronized with '''blocks of''':</font>
|Event synchronized with '''blocks of''':
|<font size = "1">'''Multi-sine''' - '''random noise''' - '''chirps'''</font>
|'''Multi-sine''' - '''random noise''' - '''chirps'''


|-
|-
|<font size = "1">'''Swept sine''' generator event:</font>
|'''Swept sine''' generator event:
|<font size = "1">'''Swept sine stabilized''' (output amplitude established) - '''step sine stabilized''' (occur x sec after step frequency is reached) - '''pure tone stabilized''' (output amplitude established)</font>
|'''Swept sine stabilized''' (output amplitude established) - '''step sine stabilized''' (occur x sec after step frequency is reached) - '''pure tone stabilized''' (output amplitude established)


|}
|}
Line 2,357: Line 2,323:
NVGate<sup>®</sup> proposes a set of generator signals.
NVGate<sup>®</sup> proposes a set of generator signals.


<font color="#4DA836">'''Output signals'''</font>
'''Output signals'''


The following output signals are available as standard and can be independently applied to the front-end outputs.
The following output signals are available as standard and can be independently applied to the front-end outputs.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "10"|<font size = "1">'''Sine'''</font>
|align = "center" rowspan = "10"|'''Sine'''
|<font size = "1">Type</font>
|Type
|<font size = "1">Pure tone - Swept sine - Step sine</font>
|Pure tone - Swept sine - Step sine


|-
|-
|<font size = "1">Frequency</font>
|Frequency
|<font size = "1">'''100 µHz''' to '''40 kHz''' - smooth variation between step - resolution '''10 µHz''' </font>
|'''100 µHz''' to '''40 kHz''' - smooth variation between step - resolution '''10 µHz'''  


|-
|-
|<font size = "1">Frequency control</font>
|Frequency control
|<font size = "1">Start & stop frequencies - pause/release during sweep - adjustment during pause - step  </font>
|Start & stop frequencies - pause/release during sweep - adjustment during pause - step   


|-
|-
|<font size = "1">Cycle control</font>
|Cycle control
|<font size = "1">One shot - One cycle - continuous sweep between boundaries</font>
|One shot - One cycle - continuous sweep between boundaries


|-
|-
|<font size = "1">Amplitude control</font>
|Amplitude control
|<font size = "1">'''Settling time''' 100 µs to 10 s - '''Stabilization time''' 0 s to 1000 s</font>
|'''Settling time''' 100 µs to 10 s - '''Stabilization time''' 0 s to 1000 s


|-
|-
|<font size = "1">Sweep speed</font>
|Sweep speed
|<font size = "1">Linear: '''300 mHz/s''' to '''20 kH/s''' - Log: '''5 mOct/s''' to '''330 Oct/s'''</font>
|Linear: '''300 mHz/s''' to '''20 kH/s''' - Log: '''5 mOct/s''' to '''330 Oct/s'''


|-
|-
|<font size = "1">Step control</font>
|Step control
|<font size = "1">'''Synchronized with analysis end''' - free run </font>
|'''Synchronized with analysis end''' - free run  


|-
|-
|<font size = "1">Gain control</font>
|Gain control
|<font size = "1">'''-15 dB''' to '''<nowiki>+</nowiki>60 dB''' - '''Independent''' for each output <br>- Amplitude '''variation controlled''' by settling time (1 ms to 1000 s)</font>
|'''-15 dB''' to '''<nowiki>+</nowiki>60 dB''' - '''Independent''' for each output <br>- Amplitude '''variation controlled''' by settling time (1 ms to 1000 s)


|-
|-
|<font size = "1">Phase control</font>
|Phase control
|<font size = "1">Offset ±360°  - '''Independent''' for each output - phase '''variation controlled''' by phase speed (1.5°/s to 360°/s)</font>
|Offset ±360°  - '''Independent''' for each output - phase '''variation controlled''' by phase speed (1.5°/s to 360°/s)


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2 to 6''' sine generated simultaneously - with '''independent phases and amplitudes''' - '''synchronized''' frequencies</font>
|'''2 to 6''' sine generated simultaneously - with '''independent phases and amplitudes''' - '''synchronized''' frequencies


|-
|-
|align = "center" rowspan = "6"|<font size = "1">'''Multi-sine'''</font>
|align = "center" rowspan = "6"|'''Multi-sine'''
|<font size = "1">Frequency span</font>
|Frequency span
|<font size = "1">From '''125 mHz''' to '''40 kHz''' </font>
|From '''125 mHz''' to '''40 kHz'''  


|-
|-
|<font size = "1">Amplitude</font>
|Amplitude
|<font size = "1">'''0 '''to''' 2.5 Vrms''' - Settling time''' 100 µs '''to''' 10 s'''</font>
|'''0 '''to''' 2.5 Vrms''' - Settling time''' 100 µs '''to''' 10 s'''


|-
|-
|<font size = "1">Resolution</font>
|Resolution
|<font size = "1">'''125 mHz''' to '''400 Hz''' - '''101''' lines to '''6401''' lines</font>
|'''125 mHz''' to '''400 Hz''' - '''101''' lines to '''6401''' lines


|-
|-
|<font size = "1">Phase</font>
|Phase
|<font size = "1">'''Fixed''' (all sine have same phase) - '''Random'''</font>
|'''Fixed''' (all sine have same phase) - '''Random'''


|-
|-
|<font size = "1">Burst</font>
|Burst
|<font size = "1">'''0 '''to '''100% '''- Step''' 1%'''</font>
|'''0 '''to '''100% '''- Step''' 1%'''


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2''' to '''4''' '''independents''' multi-sine</font>
|'''2''' to '''4''' '''independents''' multi-sine


|}
|}


<font color="#4DA836">'''Output signals (continued)'''</font>
'''Output signals (continued)'''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "5"|<font size = "1">'''Random noise'''</font>
|align = "center" rowspan = "5"|'''Random noise'''
|<font size = "1">Frequency span</font>
|Frequency span
|<font size = "1">From '''125 mHz''' to '''40 kHz''' - independent lower and upper frequencies</font>
|From '''125 mHz''' to '''40 kHz''' - independent lower and upper frequencies


|-
|-
|<font size = "1">Amplitude</font>
|Amplitude
|<font size = "1">'''0 '''to''' 2.5 Vrms''' - Settling time''' 100 µs '''to''' 10 s '''</font>
|'''0 '''to''' 2.5 Vrms''' - Settling time''' 100 µs '''to''' 10 s '''


|-
|-
|<font size = "1">Type</font>
|Type
|<font size = "1">'''White '''-''' pink '''</font>
|'''White '''-''' pink '''


|-
|-
|<font size = "1">Burst</font>
|Burst
|<font size = "1">'''0 '''to '''100% '''- Step''' 1% - '''Bloc duration''': 2.5 ms to 100 s'''</font>
|'''0 '''to '''100% '''- Step''' 1% - '''Bloc duration''': 2.5 ms to 100 s'''


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2''' to '''6 independent''' and '''uncorrelated''' random noises</font>
|'''2''' to '''6 independent''' and '''uncorrelated''' random noises


|-
|-
|align = "center" rowspan = "5"|<font size = "1">'''Chirp'''</font>
|align = "center" rowspan = "5"|'''Chirp'''
|<font size = "1">Frequency span</font>
|Frequency span
|<font size = "1">From '''3.125 Hz''' to '''40 kHz''' - independent lower and upper frequencies</font>
|From '''3.125 Hz''' to '''40 kHz''' - independent lower and upper frequencies


|-
|-
|<font size = "1">Amplitude</font>
|Amplitude
|<font size = "1">'''0 '''to''' 7,07 Vrms''' - Settling time''' 100 µs '''to''' 10 s'''</font>
|'''0 '''to''' 7,07 Vrms''' - Settling time''' 100 µs '''to''' 10 s'''


|-
|-
|<font size = "1">Size</font>
|Size
|<font size = "1">'''256 '''- '''512 '''-''' 1024 '''-''' 2048 '''-''' 4096 '''-''' 8192 '''-''' 16384  '''</font>
|'''256 '''- '''512 '''-''' 1024 '''-''' 2048 '''-''' 4096 '''-''' 8192 '''-''' 16384  '''


|-
|-
|<font size = "1">Burst</font>
|Burst
|<font size = "1">'''0 '''to '''100% '''- Step''' 1%'''</font>
|'''0 '''to '''100% '''- Step''' 1%'''


|-
|-
|<font size = "1">Number</font>
|Number
|<font size = "1">'''2''' to '''6''' '''independents''' chirps</font>
|'''2''' to '''6''' '''independents''' chirps


|}
|}
Line 2,478: Line 2,444:


===Filters===
===Filters===
The following filters apply on front-end inputs, plug-ins analyzers and output generators.
The following filters can be applied on front-end inputs, plug-ins analyzers and output generators.<br>
 
'''ABW:''' the analysis bandwidth, of the plug-in or front end where the filter is in use.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|align = "center" rowspan = "3"|<font size = "1">'''High pass'''</font>
|align = "center" rowspan = "3"|'''High pass'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Butterworth''' – '''IIR '''type</font>
|'''Butterworth''' – '''IIR '''type


|-
|-
|<font size = "1">Order</font>
|Order
|<font size = "1">'''1''' to '''6'''</font>
|'''1''' to '''6'''


|-
|-
|<font size = "1">Cutoff frequency</font>
|Cutoff frequency
|<font size = "1">'''ABW<ref>ABW&nbsp;: the analysis bandwidth, of the plug-in or front end where the filter is in use.</ref> '''to '''ABW/1024''' </font>
|'''ABW'''to '''ABW/1024'''  


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Low  pass'''</font>
|align = "center" rowspan = "3"|'''Low  pass'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Butterworth''' – '''IIR '''type</font>
|'''Butterworth''' – '''IIR '''type


|-
|-
|<font size = "1">Order</font>
|Order
|<font size = "1">'''1''' to '''6'''</font>
|'''1''' to '''6'''


|-
|-
|<font size = "1">Cutoff frequency</font>
|Cutoff frequency
|<font size = "1">'''ABW '''to '''ABW /102.4''' </font>
|'''ABW '''to '''ABW /102.4'''  


|-
|-
|align = "center" rowspan = "4"|<font size = "1">'''Stop/pass band'''</font>
|align = "center" rowspan = "4"|'''Stop/pass band'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Butterworth''' – '''IIR '''type</font>
|'''Butterworth''' – '''IIR '''type


|-
|-
|<font size = "1">Order</font>
|Order
|<font size = "1">'''2''' to '''10'''</font>
|'''2''' to '''10'''


|-
|-
|<font size = "1">Cutoff frequencies</font>
|Cutoff frequencies
|<font size = "1">'''ABW '''to''' 0.055 * ABW '''</font>
|'''ABW '''to''' 0.055 * ABW '''


|-
|-
|<font size = "1">Bandwidth</font>
|Bandwidth
|<font size = "1">'''ABW/2 '''to''' 0.0075 * ABW '''</font>
|'''ABW/2 '''to''' 0.0075 * ABW '''


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''Integrators'''</font>
|align = "center" rowspan = "3"|'''Integrators'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''HP, Single '''or''' double '''</font>
|'''HP, Single '''or''' double '''


|-
|-
|<font size = "1">High pass frequency</font>
|High pass frequency
|<font size = "1">'''ABW''' '''/10 000 '''or ''' ABW''' /'''2 000 '''</font>
|'''ABW''' '''/10 000 '''or ''' ABW''' /'''2 000 '''


|-
|-
|<font size = "1">Integration time</font>
|Integration time
|<font size = "1">'''2 ms '''to''' 500 s'''</font>
|'''2 ms '''to''' 500 s'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Differentiators'''</font>
|align = "center" rowspan = "2"|'''Differentiators'''
|<font size = "1">Type</font>
|Type
|<font size = "1">'''Single'''</font>
|'''Single'''


|-
|-
|<font size = "1">Average</font>
|Average
|<font size = "1">'''Sliding '''-''' 0 '''to''' 2 s'''</font>
|'''Sliding '''-''' 0 '''to''' 2 s'''


|-
|-
|align = "center" rowspan = "2"|<font size = "1">'''Weightings'''</font>
|align = "center" rowspan = "2"|'''Weightings'''
|<font size = "1">Laws</font>
|Laws
|<font size = "1">'''A''' & '''C''' laws </font>
|'''A''' & '''C''' laws  


|-
|-
|<font size = "1">Bandwidths </font>
|Bandwidths  
|<font size = "1">'''10''' kHz to '''40''' kHz </font>
|'''10''' kHz to '''40''' kHz  


|-
|-
|align = "center" rowspan = "3"|<font size = "1">'''All'''</font>
|align = "center" rowspan = "3"|'''All'''
|<font size = "1">By-Pass</font>
|By-Pass
|<font size = "1">'''Apply/bypass''' without stabilisation time - independently on each filter</font>
|'''Apply/bypass''' without stabilisation time - independently on each filter


|-
|-
|<font size = "1">Application</font>
|Application
|<font size = "1">Same filters set for each plug-in</font>
|Same filters set for each plug-in


|-
|-
|<font size = "1">Label</font>
|Label
|<font size = "1">Each filter features a user define name</font>
|Each filter features a user define name


|}
|}
Line 2,572: Line 2,540:


==Notes==
==Notes==
The above specifications describe all the guaranteed capacities and performances of the NVGate V8.00 software installed in a PC running Windows 7 pro (recommended configuration). Functionalities may change depending on operation mode (connected to a 3-Series unit or office). Plug-in analyzers, options and channel number availability depend on purchased options.
The above specifications describe all the guaranteed capacities and performances of the NVGate V12 or upper. Functionalities may change depending on operation mode (connected to a 3-Series unit or office). Plug-in analyzers, options and channel number availability depend on purchased options.
 
The instruments hardware are described separately in the "Instruments specifications" data sheet for OR34 & OR35 and OR36, MP & OR38.


The instruments hardware are described separately in the "Instruments specifications page"
OROS reserves the right to modify the specifications without notification.
OROS reserves the right to modify the specifications without notification.
[[Image:NVGate_TechSpe_01.png|framed|none]]

Latest revision as of 13:16, 19 July 2022

The following official specifications concern NVGate® software for OR10, OR34, OR35, OR36, OR38 and Mobi-Pack® multi-analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the instrument connected to the PC. This specification applies for NVGate version 12 or upper and Teamwork instruments.

Modules

NVGate® software is based on a general-purpose platform where optional analysis modules called plug-in analyzers are added.

The FFT plug-in features specific add-on for specialized analysis.

NVGate®Platform

NVGate® platform provides a comprehensive set of tools for noise and vibration acquisition, recording and analysis.

These functions are arranged in 4 parts:

  • Signal sources that condition, provide and store time domain signals.
  • Shared resources that apply simultaneously or independently on distributed signal to plug-in analyzers.
  • Operational tools that increase measurement efficiency and reliability.
  • General-purpose analysis for monitoring and result tracking.
Signal sources

Front-end

Controls dynamic inputs, generators, external synchronization/tach. input and parametric (DC) inputs.

Recorder

Record and store time domain signal on analyzer or PC hard disc. Records dynamic, parametric (DC) and ext. synch/tach inputs with multiple frequencies.

Player

Visualize, listen, playback on output or generate signals for post-analysis purpose. Based on recorded or imported time domain signal file.
Shared resources

Filters

High/low pass, Band/Stop pass, single and double integrator, differentiator, A and C acoustic laws

Events

Edge detection, DC levels, Delta DC levels, RPM, Delta RPM, Manual, Time period, combinations.

Weighting windows

Programmable Force and response, Hanning, Hamming, Kaiser Bessel, flat-top, uniform.

Tachometers

Compute RPM from ext. sync and input pulses, DC, CAN. Fractional, simulated and combined tach.
Operational tools

Transducers management

Automatic calibration, database management, calibration history, automatic setup of inputs, TEDS management.

Masks

Mask editor for spectra, profiles, 1/n octave, order spectra

Report

On-line report editor, automated report generation, word or Excel files.
Automation tools Macros, sequence from Excel® sheet, alarms based on mask comparison, customizable control panel, restricted user profiles, remote controller interface.
General purpose analysis

Monitor

4 channels FFT analyzer with input hot swap. Statistical extraction (RMS, Max, Min, Kurtosis) from time domain with programmable band-pass filter.

Waterfall

Synchronized stack for 2D (spectra) and scalar (order, overall). 3D, color map and profiles displays. Extraction capabilities.

NVGate® Plug-ins analyzers

In addition to the standard functions featured in the software platform, NVGate® can receive additional plug-in analyzers that offer flexible configuration of independent analysis modes.

FFT Fast Fourier Transform - Time to Frequency conversion with multiple average domains, combinable trigger, zooms and cross functions. General-purpose frequency analysis, machinery signature, modal analysis acquisition (hammer or shaker), machinery diagnostic.
FFT Add-ons    
CBT Constant Band Tracking - Order extraction at constant bandwidth Gears mesh analysis, gearboxes noise and vibration tracking.
FFTDiag Cepstrum, Auto and cross correlation, DC, Min/Max, Pk, pk/pk and crest factor. Set of functions dedicated to the diagnostics of industrial machineries.
TDA Time Domain Analysis –Time view (oscilloscope). True statistical extraction, averaging. Machinery diagnostic, long term acquisition monitoring, critical acquisition control, large structure damping control, shocks determination, machinery health control, trend analysis.
SOA Synchronous Order Analysis - Time to Order re-sampling with angle or order averaging, acoustic weightings, multiple tach management. Rotating machinery balancing. In vehicle order to noise correlation R&D. Pump, compressors, turbines and speed reducer/multiplier test.
SOADiag
(SOA Add-on)
Copstrum™, Auto and cross correlation, Revolution synchronous statistics, Order transmission Function (ORF™), X functions Set of diagnostic functions based on the synchronous order analysis for rotating part transmission and forced vibration extraction.
OCT 1/n Octave Constant Percentage Band filters - 1/n octave detectors set with multiple averaging modes. General-purpose industrial noise analysis. Vehicle acoustics R&D. Noise test.
OVA Overall Acoustics Detectors - 3 detectors + 1 peak detection per channel with parallel averaging and acoustic weightings. Industrial acoustic, sound power measurements. Machinery with cycles, R&D and test.

NVGate® Options

Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software also features options:

IVC Instantaneous angular velocity converter Torsional measurements from Ext. Synch inputs
CTE Combined Tachometer editor Tachometer math editor - 2 inputs, 1 output
VIN Virtual Inputs Real-time math combination of dynamic inputs
VDC Virtual DC Real-time math combination of parametric inputs
A-Samp Angular sampling for SOA Real-time Angular sampling of inputs

Signal sources

NVGate®/OROS 3-Series multi-analyzers platform can process signal coming from different sources. These sources are detailed in this section as modules.

Front-end

The front-end module gathers the different input and output settings available in the instrument unit. In addition, the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to the PC) which is useful for set-up.

Front-end settings

Dynamic inputs Sampling clock 102.4 kS/s to 2,048 S/s or 65.536 kS/s to 3,200 S/s
LEMO2 connectors management 200 V polarization on/off per block of 8 input
Coupling3,4 AC (0.35 Hz) - DC - ICP - ICP + TEDS - AC float - DC float - Grounded
Node information3 Label - component - node - direction (± X ,Y,Z) - type (translation, rotation)
Physical quantity3,4 Any physical quantity can independently be associated to inputs.
Sensitivity3,4 User defined in V/unit.
Range2 Pk to pk expressed in the associated unit, (taking in account sensitivity) - linear or dB
External conditioner compensation3 Gain - polarity - offset.
Filter3 Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Auto-range Enable/disable auto-ranging independently on any input
Ext. sync inputs Sampling over-sampled 64 time the dynamic inputs sampling
Coupling3 AC (0.35 Hz) - DC
Information3 Label
Physical quantity3,4 Any physical quantity can be associated to inputs.
Sensitivity3,4 User defined in V/unit.
Range3 Pk to pk expressed in the associated unit, (taking in account sensitivity) - linear or dB
External conditioner compensation3 Gain, offset.
Edge detection3 Threshold (in associated unit) - Edge (rise, fall) - Hystersis - hold off (sec)
Pre-divider3 Hardware - 1 to 255 pulses – accept up to 375 kpulses/sec
Post-multiplier3 1 to 50 pulses/trigger - Generate higher frequencies on low rate triggers. (ex. GPS)
Tachometer mode3 Pulse per rev: 0.5 to 4096 - average (n rev) - min (detect stopped shaft) and max RPM speed - hold off (in % of revolution) - rotation (clockwise/counterclockwise)
Torsional mode2: Instantaneous angular velocity converter Pulse per rev: 1 to 4096 - min (detect stopped shaft) and max RPM speed - sliding average (1 to 20 samples) - Missing pulse number: 0 to 5, Missing pulses detection :1 to ±20%: of previous pulses duration- Filters (dt, dt², 1/dt)
Angular sampling2: multi-pulse/rev clock used for signal sampling for SOA - Pulse per rev: 1 to 4096 continuously (no power of 2 limitation) - Real time, in line anti-aliasing - Missing pulse number: 0 to 5, Missing pulses detection: 1 to ±20% of previous pulses duration – Phase reference: Any tachometer including same input



Outputs Sampling Same as dynamic inputs
Generated signals3 Any outputs signals, see §outputs - play-back of signal file tracks - DC levels - Inputs playback (Delay > 256 samples)
Filters3 Any NVGate filter can independently be applied to each dynamic output (See Filters §)
Impedance3 50 ohm, 600 ohm (only OR36 and 0R38) or grounded
Gain3 -100 dB to + 120 dB
Synchronization3 Free run or linked to acquisitions runs
Level and phase transition3 No (Steps) - controlled (ramp) - user selectable ramp time
Clipping User selectable - protects shaker and amplifier.
Events (can trig analysis or record)3 Outputs stabilized - new step reached - start & stop sweep
Controls Emergency stop - mute/un-mute all
Parametric1(DC) inputs Sampling 12.3 to 12,8 S/s - 10 Hz to 100 Hz harmonics rejection
Information3 Label
Physical quantity3,4 Any physical quantity can be associated to inputs.
Sensitivity3,4 User defined in V/unit.
Range3 Pk to pk expressed in the associated unit (taking in account sensitivity) - linear or dB
External conditioner compensation3 Gain (-120 dB to +20 dB) - polarity - offset (< range).
Auto-range Enable/disable auto-ranging independently on any input
Auto-ranging Type Normal - on Peak detection
Margin Select the first highest range with 0 dB - 3 dB or 20 dB margin
Duration (apply for normal auto-range) 0 to 10 sec.
Peak detection User select input were to detect peak, auto-range applies on all enabled input at each peak detection.
Peak parameters settings 1 to 10 peaks - sensitivity (low, normal, high) - rise, fall or any edge.
Checking ICP (check not available on OR34) Test and report (Open, shortcut, ok) all enabled inputs.
TEDS Transducers automatic recognition - Complies with IEEE 1451.4 2004 Rev. 1.0

2: Optional features

3: Independent for each input

4: Linked with the transducer database

Front-end results & connections

The following results are available for monitoring and connection to analysis mode (plug-in analyzers)

Dynamic inputs & torsional1 inputs Monitoring3 Time domain instantaneous signal - Size 256 samples
Status led3 (docking tool bar) Overloaded = red - overload occurred since last start = red with yellow center - ok = green - under load (20 dB below range) = green with yellow center
Connection3 To any: plug-in analyzer channel, monitor channel, Virtual inputs, recorder track, edge event detector & tachometer
Ext. Sync inputs Monitoring3 Time domain instantaneous status (1 = threshold crossed, 0 = no event) - Size 256 samples.
Connection3 To any: start or stop averaging of plug-in analyzers, recorder start or stop recording, waterfall start or stop acquisition, new block trigger for FFT and SOA and new slice for waterfall acquisition.
To torsional, tachometer & angular sampling
Parametric1 (DC) inputs & CAN1parameters Monitoring3 Profiles versus time (160 ms to 163 sec user selectable), and digital/analog view meter.
Status led3 Overloaded = red - ok = yellow - under load (20 dB below range) = Cyan
Connection3 To any: recorder track, waterfall reference (Z/X axis) and profiles, level delta level detectors and virtual DC

Recorder / Player

OROS 3-Series/NVGate instruments feature a recorder and a player module that allows users to:

  • Record time domain signal, torsional inputs, trigger and parametric input.
  • Export or import time domain files.
  • Playback time domain files on analyzers outputs during analysis
  • Listen to recorded tracks on the PC speakers.
  • Post-analyze time domain files using the available plugs-in.

Recorder

The recorder module saves time domain signal into files located on 3-Series analyzer HD, Mobi-DiskTM or PC HD. This is available on 2 different modes:

  • On-line record, the chosen inputs (Dynamic, ext. sync, parametric) are recorded at selected sampling frequency during acquisition. Real-time analysis is available simultaneously.
  • Time and tracks split, the chosen recorded tracks (from signal file) are played back in the recorder using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be changed.

Recorder settings

Bandwidths Dynamics inputs 2 groups of user selectable sampling frequency - From 102.4 kS/s to 2.048 S/s (40 kHz to 800 mHz) - available simultaneously
Ext. sync inputs Automatic selection at Front-end sampling frequency - Resolution is 64 time Front-end sampling frequency
Parametric (DC) inputs 12.5 S/s (50 Hz rejection) - 15 S/s (60 Hz rejection)
Tracks Number 14 tracks<ref>For OR34, OR35 and Office-8 licenses.</ref> (8 + 6 ext. sync) or 38 tracks<ref>For OR36, OR38 and Office licenses.</ref> (32 + 6 ext. Sync)
Sum of connected licenses using multiple hardware
Saved settings / track Label - Coupling - External gain - Input range - Sampling frequency and signal bandwidth - Component - Node number - Direction - Type - Associated transducer - Unit - Sensitivity and Offset compensation.
Modes Start to time Start recording on run or any activated event - Stop recording when duration ends - Duration 10 ms to available space on target HDD.
Start to stop Start recording on run or any activated event - Stop recording on stop on any activated event - Start and stop event can be the same - Duration 10 ms to available space on active drive (PC or OR3x).
Time to stop Memorize the earliest x seconds - Stop = stop or any activated event - Duration 10 ms to 2 GSamples (limited by the drive available space)
Triggering Start recording (new record) Any NVGate® event plus manual and free run
Stop recording (end current record) Any NVGate® event plus manual and free run
Start delay Positive = unlimited - negative = 128 kSamples
Stop delay Positive = unlimited - negative = 0
Markers Saved in the record files Added by tool bar or shortcut - comment fields editable during record or at end of acquisition
Signal file Record on PC HDD Up to 10 dynamic inputs – Up to 4 parametric inputs - max total bandwidth 512 kS/s - (2 MB/s)
Record on Analyzer SSD Up to 40 channels (32 inputs + 6 ext. sync or DC)/instruments - max bandwidth 3.2 MS/s (12.5 MB/s)
Format Normal: 32 bits/sample - Compressed 16 bits/sample

Recorder settings (continued)

Management Download Discard Analyzer SSD on PC HDD - Batch download available - Rate 4 MB/s
Upload Load PC HDD records on the analyzer SSD - Rate 1.4 MB/s
Location All records accessible from the project manager tree
Hard drive management Format SSD - Selective delete - Sorting tools

Mobi-DiskTM

OR36 Mobi-Pack and OR38 multi analyzers/recorders feature a removable hard drive with 2 ports:

  • Parallel High speed port for direct data throughput to the Mobi-DiskTM.
  • USB 2.0 port for post processing and data management without the instrument.
Connection 1 to 8 Mobi-DiscTM Up to 8 Mobi-disc connected - Active Mobi-Disc: 1 (user selectable) - hot swap
Transfer Based on the USB 3.0 link Download 15 MB/s

Recorder monitoring

The following displays are available during real-time or post-analysis.

RMS Overall RMS / input Digital or analog view-meter
Signal Real-time Compressed view of entire recorded signals - Automatic update of time axis.

Player

The player module plays the recorded signal files. This is done one of the following two ways:

  • Post-analysis, the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
  • Playbacks, the recorded dynamic input are generated on the instrument outputs simultaneously with standard analysis of the inputs.

Player settings

Played section File selection NVGate® recorded files - Imported files - located on instrument SSD, Mobi-DsicTM or on PC HDD.
Record selection Record number selection (for multi record files)
Section definition User selectable Start and stop offsets (in sec.) - available in the file overview window (marker and slider)
Duration Play backed duration (User information)
Repeat On/off - repeat continuously the selection (available only for playback on analyzer output)
Mode File swap Swap loaded files with same tracks number - applies identical post-analysis setup to multiple files.
Playback Continuous or step by step (5 ms to 360,000 s)
Bandwidth All tracks at lowest frequency (compatible with all plugs-in) or initial record frequencies (no track mix in plugs-in)
Tracks Number Up to 352 according to user's fleet
Type Dynamic input record (2 sampling frequencies), parametric (DC) input record and ext. sync record.
Fixed setting (information) / track Label, Coupling, external gain, input range, sampling frequency and signal bandwidth
Modifiable measurement point information settings (apply for post analysis or re-recording) / track Component - node number - direction - type
Modifiable settings (apply for post analysis or re-recording) / track Associated transducer - unit - sensitivity - offset compensation

Player settings (continued)

Listening File location Analyzer HD - Mobi-DiskTM connected with USB or inside the analyzer - PC HD
Playback One track - from cursor position - repeat displayed signal
Markers Display Recorded markers are available in the file overview
Use Set the start and stop playback offset (user selectable)

Player connections

The following table describes the available connections to the analysis modes (plug-in analyzers and recorder) during post-analysis operation:

Dynamic inputs & torsional Tracks Preview (multiple file simultaneously) Entire file fast overview (pre-compressed at recording time) - Track contents preview, independently (multi-graph display)
Monitoring (loaded file only) Zoom on selected play-back section
Connection3 (loaded file only) To any: plug-in analyzer channel, monitor channel, recorder track, edge event detector and tachometer
Ext. sync Tracks Preview (multiple file simultaneously) Entire file fast overview (pre-compressed at recording time) - Track contents preview, independently (multi-graph display)
Monitoring (loaded file only) Zoom on selected play-back section
Connection3 (loaded file only) To any: start or stop averaging of plug-in analyzers, recorder start or stop recording, waterfall start or stop acquisition, new block trigger for FFT and SOA and new slice for waterfall acquisition.
To torsional, tachometer & angular sampling
Recorded parametric1(DC) inputs & CAN parameters Preview (multiple file simultaneously) Entire file fast overview - Track contents preview
Monitoring (loaded file only) Profiles versus time and digital or analog view meter.
Connection3 (loaded file only) To any: recorder track, waterfall reference (Z/X axis) and profiles, level and delta level events.

Plug-ins analyzers

The following plug-in analyzers are available as options of NVGate® software platform.

Time Domain Analysis - TDA

The TDA plug-in analyzer provides time domain based analyses and visualizations. This plug-in analyzer computes statistical extractions and data compression of long duration oscilloscope views. All results are real time and operations are possible on-line or in post processing. The TDA plug-ins can operate free run or synchronized with the other plug-in analyzers.

TDA Settings

Bandwidths Oscillator 1 800 mHz to 40 kHz
(sampling oscillator 102.4 kS/s to 2,048 S/s)
Oscillator 2 512 mHz to 25.6 kHz
(sampling oscillator 65.536 kS/s to 3,277 S/s)
Time views Depth 7200/ABW ABW: the plug-in Analysis Bandwidth. sec to 100 000 sec (27 hours) - Independent on each channel
Time base (resolution) Depth/2048- Independent on each channel
Time span for extraction
(averaging)
Type Continuous sliding with refresh each 256 samples (Exponential) - One shot (Linear) - Repeated shot (Repeated linear)
Duration 1/ABW to 2 106/ABW (ex: 50 ms to 1000 sec @ 2 kHz)
Channels OR36 & OR38 analyzers 32 channels per plug-in for on line 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Filters On each channel Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Triggering

Start analysis Any NVGate® event plus manual and free run
End analysis Any NVGate® event plus manual and free run
Repeat analysis On new start or end of averaging
Trigger delay Positive = unlimited - negative = 32k Samples

TDA Results

The following results are available for real time or post-analysis display, report and saving.

Time Views Overview Signal envelope: Min/Max line for each time step -
  Statistical values DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Skewness - Kurtosis - computed on a user defined area of the time view
Extractions Statistical values DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor - Kurtosis - computed on the Time span – All values available for View-meter and/or Waterfall profiles

Fast Fourier Transform analysis - FFT

The FFT plug-in analyzer features multiple vibrations results depending on applied setting. This type of plug-in analyzers computes real-time (on-line or post-processing) Fast Fourier Transform algorithms with multiple averaging domain and trigger capabilities. Up to 4 FFT plug-ins can operate at the same time.

FFT Settings

Bandwidths Oscillator 1 800 mHz to 40 kHz
(sampling oscillator 102.4 kS/s to 2,048 S/s)
Oscillator 2 512 mHz to 25.6 kHz
(sampling oscillator 65.536 kS/s to 3,277 S/s)
Resolution Lines 101, 201, 401, 801, 1601, 3201, 6401, 12801 With force DSP and 25601 lines
Frequency resolution 80 µHz (512 mHz/6401 lines)
to 400 Hz (40 kHz/101 lines)
Averaging Domain Spectral (power) - time (with phase) - FDSA (Synchronized with one frequency)
Type Exponential - linear - repeated linear - referenced peak hold (eq. to tracking filter locked on reference channel frequency) - peak hold
Size Duration or number of blocks, unlimited size
Overlap 0% to 99.9% depending on block size - retrigger
Channels OR36, MP & OR38 analyzers 32 to 256 ch. per plug-in for on line and post-analysis
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Weighting General Uniform - Hanning - Hamming - Kaiser Bessel
Modal acquisition User defined Force (rect.) and Response (Rect. + exp)
Assignment Independently on any channels
Filters On each channel Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Cross functions Reference selection Full or partial matrix available (all channels can be references)
Zoom Magnification factor 2 to 128 by step power of 2
Min., max and center frequencies Graphically selectable on wide band spectrum
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
New block Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Trigger delay Positive = unlimited - negative = 32k Samples
Automatic Reject overloaded blocks
Blocks rejection
Overall
Manual Accept or Reject after preview of averaged results (FRF, Coherence)
Normal Accept all blocks
Overall Detector Quadratic sum of spectra lines taking in account weighting window equivalent noise bandwidth
Lower and upper frequencies Selectable into the current FFT bandwidth

FFT Results

The following results are available for real time or post-analysis display, report and saving.

Time averaging Time domain Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajoux of triggered blocks
Spectra Complex spectra
Cross-functions Instantaneous cross-spectra - averaged cross-spectra - FRF H1 - FRF H2 - Coherence
Overall Overall RMS value in specified band - Overall RMS profile in specified band.
Spectral averaging Time domain Filtered signal - Triggered block - Weighted Block - Lissajous of triggered blocks
Spectra Instantaneous complex spectra - averaged power spectra
Zoomed spectra Zoomed instantaneous complex spectra - zoomed averaged power spectra
Cross-functions Instantaneous cross-spectra - averaged cross-spectra - FRF H1 - FRF H2 - Coherence
Zoomed cross-function Instantaneous zoomed cross-spectra - averaged zoomed cross-spectra - zoomed FRF H1 - zoomed FRF H2 - zoomed Coherence
Overall Overall RMS value in specified band - Overall RMS profile in specified band.
Synchronous averaging on one frequency (FDSA) Time domain Filtered signal - Triggered block - Averaged Triggered block - Weighted Block - Lissajous of triggered blocks
Spectra Instantaneous complex spectra - time domain averaged power spectra
Overall Overall RMS value in specified band - Overall RMS profile in specified band.

Constant Band Order Tracking Add-on (CBT)

Constant band order tracking is an optional add-on to the FFT plug-in analyzers. This add-on computes spectrum lines power related to RPM speed. This option adds settings and results to the FFT Plug-ins. Constant band tracking can operate on up to 4 different tachometers at the same time.

CBT Settings

Tracked Order Number 8 per channels
Max 0.001 to 800
Constant bandwidth User selectable per channel - minimum depend on weighting windows
Computation Associated tachometer Any front end, recorded or virtual tachometer
Peak tracking On/Off center bandwidth on nearest peak
Order amplitude Weighting windows eq. noise bandwidth correction

CBT Results

The following results are available for real time or post-analysis display, report and saving.

Scalar Tracked order Digital (magnitude and phase) or analog view-meter
Cross phase tracking Order phases are relative to the same order from a ref. channel
Overall Digital or analog view-meter
Monitoring Continuous profiles of tracked order Profiles vs. time - profiles vs. RPM - max depth 2048 pts - user selectable delta time - user selectable delta RPM
Profiles One shot acquisition Tracked order (complex), cross phase orders and overall level can be collected by the waterfall profiles.
References Time, RPM and DC levels

Diagnostic Add-on (FFTDiag)

FFT Diagnostic is an optional add-on to the FFT plug-in analyzers. This add-on computes a set of useful analysis for machinery diagnostic. Up to 4 FFTDiag can operate on 4 different FFT plug-ins at the same time.

This option adds settings and results to the FFT Plug-ins.

FFTDiag Settings

Envelop demodulation Activation Enable/disable on all channels - requires zoom activation
Averaging Spectral domain - FDSA
Bandwidth ½ of zoom span - ¼ of FFT bandwidth
Correlation Activation Enable/disable on all channels - All other FFT results are affected
Weighting window Uniform - Left zero padding - Centred zero padding

FFTDiag Results

The following results are available for real time or post-analysis display, report and saving.

Scalar Min/Max Minimum and maximum amplitude per trigger block
Peak detectors Peak level - Peak to Peak level - Crest Factor - per trigger block
Others Cepstrum Cepstrum - Zoomed cepstrum
Correlation Auto-correlation block - Cross correlation block
Shaft view Time domain signal on first rotation - angular representation

Synchronous Order Analysis (SOA)

The SOA plug-in analyzer features several results depending on applied settings. Up to 3 SOA plug-ins can operate in parallel.

SOA Settings

Bandwidths Order span 6.25 to 400 – up to 800 with ForceDSP
Frequency bandwidth Up to 40 kHz
RPM span Up to 384 000 RPM - Ratio of 64 between Min & Max RPM
Resolution Lines 101 - 201 - 401 – 801 – 1601 with ForceDSP
Order resolution 1 - ½ - ¼ - 1/8 - 1/16 - 1/32
Averaging Domain Spectral (power) - revolution (re-sampled blocs)
Type Exponential - linear - repeated linear - peak hold
Size Number of blocks, unlimited size
Overlap 0 to 31 revolutions depending on resolution – 0 to 359° into revolution – Retrigger on multi-pulse/rev
Tachometer Computation base Any NVGate'® 'tachometer (ext. sync, regular input, DC input, CAN, simulated, Fractional, and combined tach.)
Phase reference Tachometer pulse edge or pulse center requires to analyze the tachometer input
0° Phase reference Cosine or sine
RPM Max speed variation 1 to 99% per analyzed block - blocks with higher variation are rejected. No control (set @ 100%)
RPM range User selectable max & min RPM, under and over speed blocks are rejected
Tracked order Number Up to 8 per channels
Min-Max 1/32 to 400 – up to 800 with ForceDSP
Cross phase tracking Order phases are relative to the same order from a ref. channel
Channels OR36 & OR38 analyzers 32 channels per plug-in for on line - 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Weighting Windows selection Uniform - Hanning - Hamming - Kaiser Bessel Applied independently on any channels
Filters On each channel Any NVGate filter can independently be applied to each dynamic input (See Filters §)
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
New block Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Trigger delay Positive = unlimited - negative = 1/order res revolution
Trigger block phase shift ±720° - independent for each channel
Overall Detector Quadratic sum of order spectra lines taking in account weighting window equivalent noise bandwidth
Lower and upper orders Selectable into order span - from 0,03125 to 800 with ForceDSP

SOA Results

The following results are available for real time or post-analysis display, report and saving.

Revolution averaging Time domain Filtered signal
Angle domain (Revolution) Triggered block - Averaged Triggered block - Weighted block
Order spectra Complex spectra
Tracked orders Digital (magnitude & phase) view-meter - Analog view -meter
Overall Overall RMS value in specified order band (digital or analog view-meter)
Spectral (order) averaging Time domain Filtered signal
Angle domain (Revolution) Triggered block - Weighted Block
Order spectra Instantaneous complex order spectra - averaged power order spectra
Tracked orders Digital (module and phase) and/or Analog view-meter.
Overall Overall RMS value in specified order band (digital or analog view-meter)

Diagnostic Add-on (SOADiag)

Order based Diagnostic is an optional add-on to the SOA plug-in analyzers. This add-on computes a set of useful synchronous order transmission extraction tools and analyses. Up to 2 SOADiag can operate on 2 different SOA plug-in at the same time.

This option adds settings and results to the SOA Plug-ins.

SOADiag Settings

Cross-functions Reference selection Full or partial matrix available (all channels can be references) - order domain averaging only.
Angular Correlation Activation Enable/disable on all channels - All other SOA results are affected simultaneously
Weighting window Uniform - Left zero padding - Centred zero padding

SOADiag Results

The following results are available for real time or post-analysis display, report and saving.

Rev. synchronous Scalars Min/Max Minimum and maximum amplitude per trigger block (n rev.)
RMS, DC Per trigger block (n rev.)
Peak detectors Peak level - Peak to Peak level - Crest Factor - per trigger block (n rev.)
Others Copstrum™ Cepstrum of the order spectra
Cross-functions Instantaneous order cross-spectra - averaged order cross-spectra - ORF™ H1 / H2 -Coherence – angular or order domain averaging.
Angular Correlation Auto-correlation block - Cross correlation block - angular averaged
Re-sampled Shaft view Time domain signal on first rotation – angular representation - instantaneous or revolutions averaged

1/n octave constant percentage band filter analysis - OCT

The OCT plug-in analyzer features multiple acoustic results depending on applied setting. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters (CPB) and detectors.

OCT Settings

Bandwidths Sampling oscillator #1
(102.4 kS/s to 2,048 S/s)
Center of highest 1/3rd band = 40 kHz
Center of lowest 1/3rd band = 100 mHz
Sampling oscillator #2
(65.536 kS/s to 3,277 S/s)
Center of highest 1/3rd band = 25 kHz
Center of lowest 1/3rd band = 100 mHz
Frequency span Highest Band / lowest band < 2000 (ex. 10Hz - 20kHz)
Resolution 1/n Octave '1, 1/3 'rd, 1/12th, 1/24th
Averaging Basic Linear & repeated linear (20 ms to 60,000 s) exponential (20 ms to 60s)
Acoustics Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse
Constant Bandwidth * Time B*T = 0.2dB, 0.5 dB, 1 dB, 2 dB
Standards Detectors Complies with IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 class 1
CPB filters Complies with IEC ANSI/ASA S1.11-2014 / Part 1 / IEC 61260:1-2014 and IEC 1260 class 1


Channels OR36 & OR38 analyzers 32 channels per plug-in for on line - 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels for on line and post-analysis
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Overall Detectors 2 parallel overall detectors linear and weighted (time domain)
Weighting A, C, or Z (none) applicable in 10 kHz - 40 kHz bandwidth

OCT Results

The following results are available for real time or post-analysis display, report and saving.

Spectra 1/n Octave spectra Instantaneous spectra, averaged spectra
Holden spectra Minimum & maximum spectra
Overall Linear Digital & analog view meter of time domain overall detector, exact bandwidth is informed
Weighted Digital & analog view meter of A or C weighted overall levels computed in time domain, exact bandwidth is informed

Overall acoustics sound level meter - OVA

The OVA plug-in analyzer features class 1 sound pressure level measurement according to IEC 61672:2013 standard. This plug-in analyzer computes real-time signal (on-line or post-processing analysis) based on digital filters and detectors.

OVA Settings

Type Standard IEC 651, IEC 804, ANSI/ASA S1.4-2014 / Part 1 / IEC 61672-1:2013 class 1
Class 1
Bandwidths 10 Hz to 40 kHz - Adjustable
Channels OR36 & OR38 analyzers 32 channels per plug-in for on line - 32 to 256 in post-analysis
OR34 & OR35 analyzers 8 channels for on line and post-analysis
Averaging Linear User selectable duration - repeat mode
Short leq 1/8 s - 1 s - User selectable duration
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Detectors Peak 1 peak detector/Ch. - A, C or Z (none) weighting -Independent for each channel
Weighting 3 overall detectors/Ch. - A, C or Z (none) weighting - Fast, Slow, Impulse and linear time weighting - Independent for each channel

OVA Results

The following results are available for real time or post-analysis display, report and saving. The following levels can be tracked in profiles up to 100 000 pts. each.

SPL Time weighted Instant - max hold - min Hold
Averaged Short leq - leq
Peak Peak - Time weighted - max hold min hold

Options

Instantaneous angular Velocity Converter (IVC)

The IVC option converts frequency to voltage from the External synch inputs. It covers torsional and acyclism measurements. The converted signals are made available as standard inputs (or player tracks). This option operates real-time computation and operations are possible on-line or in post processing.

IVC Settings & specs

Rate Pulse/rev 0.5 to 4096 - Up to 1 E6 with pre-divider
Pulse/rev Frequency > 40 kpulse/sec max
Pre-divider 1 to 255 - hardware decimation (pulse are not measured)
Pre-divider frequency 375 kpulse/sec max.
Missing pulses Number 0 to 5 consecutive pulses integer or fractional
Hold off 2 consecutive pulses
Generated signal '1st 'order interpolation
Tach. phase ref No missing pulse First random - continuous offset further
missing pulse > 0 Last known edge before missing pulse
Resolution Time 160 ns : 1/(SF x 64)
Angular (with SOA) 350 µRad (20 mdegres) @ 8000 RPM, up to order 10
Pre-processing Smoothing Sliding average - 1 to 32 samples
Filtering Any NVGate filter (See Filters §) incl. dt and 1/dt
Max Speed Without pre-divider Max RPM = 2.4 e6/Pulse per rev - 12 000 RPM @ 200 pulse/rev
With pre-divider Max RPM = 36 e6/Pulse per rev - 12 000 RPM @ 200 pulse/rev

IVC Results

The following results are available for real time or post-analysis.

Signals Number 1 to 6 (depends on the number of available ext. synch)
Use Same as dynamic inputs or recorded dynamic inputs
Magnitude Torsional acceleration, velocity & angle – separated from angular magnitudes (RPM)
View Type Signal - filtered signal - 256 samples frame

Angular sampling (A-Samp)

The angular sampling option allows accurately localizing the order phenomenon angle on cyclic rotating machineries. It uses the pulses delivered by an encoder (or zebra tape or gear teeth) to resample the inputs and torsional signal in the SOA plug-in. This option operates real-time computation and operations are possible on-line or in post processing.

IVC Settings & specs

See front-end/ext. Synch § for details

Virtual inputs (VIn)

The virtual input allows combining front end inputs with polynomial operations to generate a signal (new input). This option covers numerous applications such as vector strain calculation with rosette or dynamic twist measurement on machinery transmission. The virtual inputs are made available as standard dynamic input. A static twist computation based on phase comparison is also available with the Vin + IVC options.

This option operates real-time computation on raw or recorded signal.

VIn Settings

Output Number 12 Operators
Type New item in the active inputs/tracks list
Synchronization 0° phase shift with sources and analyzed signal
Magnitude Automatic or voltage if unresolved
Sources Number 1 to 32 channels
Type Analyzer dynamic inputs or recorded dynamic inputs
Output adjustment Type (Input * Coeff + Offset ) ^power - independent on each output
Offset Any real value from -1 e9 to 1 e9 - ex: -5.67
Coefficient Any real value from -1 e9 to 1 e9 - ex: 12
Power Any real value from -4 to 4 - ex: 0.5
Sources adjustment Type Filtering [ (input * Coeff + Offset ) ^ power] - independent for each source
Offset Any real value from -1 e9 to 1 e9 - ex: -5.67 e-5
Coefficient Any real value from -1 e9 to 1 e9 - ex: 1200
Power Any real value from -4 to 4 - ex: -2.4567
Filtering Any NVGate filter
Operators Type Product - Sum - independent on each operator
Twist Type Phase comparison – output in plane angle
Correction (zeroing) offset -360° to +360°
Output Plane angle or torque with conversion factor

Virtual parameter (VDC)

The virtual parameter (VDC) allows combining front-end parametric inputs (DC) with math operations to generate a calculated parameter (new DC). This option covers numerous applications such as averaged temperature or non linear response transducers. The virtual parameters are made available as standard DC input. This option operates real-time computation on raw or recorded signal.

VDC Settings

Output Number 12 Operators
Type New item in the DC inputs/track list
Synchronization 0° phase shift with sources and analyzed signal
Magnitude User defined
Sources Number 1 to 32 channels
Type Analyzer DC inputs or recorded DC inputs
Editor Type Text
Variables Real values
Operators +, -, x, /, ^, = (affectation)
Predefined Ch x (Channels signals), Pi, e
Trigonometric functions Sine, Cosine, Tangent - with Arc and Hyperbolic combination
Logarithmic functions Log2, Log10, Ln, Exp
Miscellaneous Sqrt, Abs, Min, Max, Sum, Avg
Output characterization Limits Min and max: +/- 1 E9 - In current unit
Information Label (text)

Combined tachometer editor (CTE)

The CTE option combines 2 actual tachometers with math to generates a calculated angular velocity (new tachometer). It covers various unreachable rotating parts and more specially the CVT gear chain measurements. The combined tachometers are made available as standard tachometer. This option operates real-time computation and operations are possible on-line or post processing.

CTE Settings

Sources Number 2
Type Any NVGate tachometer – excepted another CTE
Output Type New item in the active tachometers list
Synchronization 0° phase shift with sources and analyzed signal
Editor Type Text
Variables Real values
Operators +, -, x, /, ^, = (affectation)
Predefined Rmp1, Rpm2 (sources speed), Pi, e
Trigonometric functions Sine, Cosine, Tangent - with Arc and Hyperbolic combination
Logarithmic functions Log2, Log10, Ln, Exp
Miscellaneous Sqrt, Abs, Min, Max, Sum, Avg, Sign, If, Rint
Output characterization Rotation Clock wise or counter clock wise
Average Sliding - 1 to 256 revolution
Speed limits Min: 0 RPM - max: 1 200 000 RPM
Information Label (text)

CTE Results

The following results are available for real time or post-analysis

Signals Number 4
Use Same as on-line Tachometer or recorded tachometer
Magnitude Angular velocity (RPM)
View Types View meter - 1 to 100 s rolling profile

General purpose analysis

The following modules are available as standard features of NVGate® software platform; they feature additional analysis capabilities to regular plug-in analysis modules.

Monitor

The monitor is an independent (dedicated processor) module that continuously processes FFT analysis on 4 channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.

Monitor settings

Fixed setup Bandwidth ABW = Sampling freq / 2.56
Resolution 401 lines
Average domain Spectral (power)
Overlap 0%
Average type Exponential
Weighting windows Hanning
Trigger Free run
Channels Number 4
Swap Between any active dynamic input - hot swap capable
Average Duration Instantaneous (20 ms) to unlimited
Overall analysis Pass band filter Butterworth order 2 to 10 - IIR type - Bypass function
Upper & lower frequency User defined - can be graphically modified
Computed indicators DC - Max - Min - RMS - Skew - Kurtosis - Into defined pass band - overall if bypass is on
Average User defined duration - independent from spectral averaging

Monitor results and connections

The following results are available for real time or post-analysis display, connection to other modules.

Time domain Play back On any output generator - hot swappable
FFT analyzer Trigger bloc 1024 samples - time domain analyzed bloc
Spectra Instantaneous complex spectra - averaged power spectra
Overall detectors Display (digital or analog view-meter) DC - Max - Min - RMS - Skew - Kurtosis
Connection To level and delta level event detector - to waterfall profiles

Waterfall

The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in analyzers and sources in one result.

Waterfall settings

Mode Continuous Circular buffer of results
One shot Fill stack and stop waterfall acquisition
Triggering Start acquisition Any NVGate® event plus manual and free run
Stop acquisition Any NVGate® event plus manual and free run
New slice (point or spectra) Any NVGate® event - manual - free run - availability of connected results - periodic

Waterfall settings (continued)

Size Number of result per acquisition Up to 95 results + Time
Depth 2 to 100,000 slices or points - depend on PC available memory and requested result - automatically adjusted before acquisition
Depth for stand-alone acquisition Up to 3 MSamples per computation DSP
Channels Type scalar (the following results are connectable to waterfall channels) Overall levels (lin and weighted) from OvA and 1/n OCT - Monitor indicators (DC, Max, Min, RMS, Kurtosis) - Orders (from CBT and SOA) - Complex orders (magnitude & phase) - Overall levels in selected BW from FFT & SOA (order or frequency) - TDA scalar: DC, Min/max, RMS, Kurtosis, peak, peak-peak, crest factor
Type 2D - FFT (the following results are connectable to waterfall channels) Triggered block - Averaged Triggered block - Weighted Block - Complex spectra - Power spectra -Cross-spectra - FRF H1 - FRF H2 - Coherence - Zoomed complex spectra - zoomed power spectra
Type 2D - OCT (the following results are connectable to waterfall channels) Instantaneous spectra - averaged spectra - max & min hold spectra
Type 2D - SOA (the following results are connectable to waterfall channels) Triggered block - Averaged Triggered block - Weighted block - Complex order spectra - Power order spectra

Waterfall results

The following results are available for real time or post-analysis display, report and saving. All stacked results can be saved.

3D Display 1 pane (3D) - 2 panes (3D + YZ view or 3D + XY view) , 3 panes (3D + YZ + XY + Extraction view) windows - automatic or user selectable pane arrangement - Real or imaginary part and module or phase for complex results
Z axis (X axis for extraction and YZ view) Any reference - time - independent for any window - swap reference at any time
Saving selection Entire 3D data and / or any active section (YZ, XY or Extraction)
Profiles Display Profile of any scalar - Real or imaginary part and module or phase for complex results
X axis Any reference - TimeSlice - independent for any window - swap reference at any time

Waterfall Tools

The following results are available for real time or post-analysis display, report and saving.

Sections YZ Sections (Profiles vs. Ref) Any order/frequency - Power - Peak - on selected bandwidth
XY Sections (Spectra) Any position in the current ref. - Min - Max - Average - on selected range or all slices
Order/Frequency extraction (profiles vs. Ref.) Any order/frequency extraction - user selectable tach. - Power - Peak - on selectable bandwidth - Max order
Number of sections Unlimited - available on result or real-time waterfalls
Linked cursors General Single or dual cursor in each view - linked with other graphs
Linked by value Linked cursors track the same X value in different windows.
Linked by reference Linked cursors track the same acquisition slice in different waterfall windows with different X or Z-axis.

References are: Time - Slice number - DC channels (expressed in physical quantity) - TDA scalars - Monitor levels- any Tachometer

Shared resources

The following resources are available for each plug-in analyzer (when it is compatible). For example once a trigger is activated it can be applied to FFT and SOA: Then modifying the trigger setting in the resources will apply on all corresponding plug-in.

Triggering

Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the Event module.

Event settings

Edge detector Source Any dynamic input - Any recorded dynamic input
Label String for event identification (Ex. "Impact" for a hammer impact detection)
Threshold Between min and max range - use source unit (ex. g for an accelerometer)
Pre-filtering A law - C law - any NVGate filter
Slope Rise - fall
Hold off 0 to 36000 s
Hystersis Into source input range - use source unit (ex. g for an accelerometer)
Number 2 edge detectors
RPM speed detector Source Any tachometer - computed from dynamic input or ext.sync - any recorded pulses (through tachometer)
Label String for event identification (Ex. "start speed" for a run up initial triggering)
Threshold Detection into source tachometer RPM range
Slope Run up - Run down
Hysteresis 0 to source max RPM
Interpolation On interpolate event occurrence into revolution - Off event occur at revolution ends (new pulse)
Number 2 to 6 RPM speed detector
Delta RPM speed detector Source Any tachometer - computed from dynamic input or ext.sync - any recorded pulses (through tachometer)
Label String for event identification (Ex. "Each 100 RPM" for a run up triggering)
Lower velocity Minimum RPM speed - Events occurs only for higher source speed
Higher velocity Maximum RPM speed - Events occurs only for lower source speed
Delta velocity Define velocity step - event occur each time source speed increase or decrease by delta velocity
Slope Run up - run down - first, first slope is automatically selected - any, event occur on any slope
Interpolation On, interpolate event occurrence into revolution - Off, event occur at revolution end (new pulse)
Number 2 to 6 delta RPM speed detector
Level detector Source Any parametric (DC) input - Monitor scalar - Filtered monitor scalar (Band Pass)
Label String for event identification (Ex. "temp A" for a recording trigger)
Type Monitor scalar - DC - RMS - Max - Min - Kurtosis
Status Above / below - Detection level - available in lin or dB
Number 1 to 4 level detectors

Event settings (continued)

Periodic event Label String for event identification (Ex. "Hourly" for 1 hour time interval)
Time interval 2 ms to 11 days - Synchronized with analyzer.
Number 2 periodic events
Combination Sources Any event - 2 different sources defined as A and B
Label String for event identification (Ex. "Time/RPM variation" for a mix between periodic and delta RPM)
Combination A OR B - A AFTER B - A AND B (occur in the same bloc)
Number 2 event combination
Plug-in synchronization FFTx result available Occurs at each new spectra availability - Occurs at end of linear averaging (repeat and linear) - One event per active FFT plug in (FFT1 to FFT4)
OCT result available Occurs at each new 1/n spectra availability - Occur at end of linear averaging (leq, short leq, linear repeat)
SOAx result available Occurs at each new spectra availability - Occurs at end of linear averaging (repeat and linear) - One event per active SOA plug in (up to 2)
TDA result available Occurs at each new level set ( RMS, DC, etc) availability - Occur at end of linear averaging (repeat and linear)
OVA result available Occurs at each new set of overall level (short LEQ) availability - Occur at end of short averaging (repeat and linear)
Generators Synchronization Event synchronized with blocks of: Multi-sine - random noise - chirps
Swept sine generator event: Swept sine stabilized (output amplitude established) - step sine stabilized (occur x sec after step frequency is reached) - pure tone stabilized (output amplitude established)

Output signals

NVGate® proposes a set of generator signals.

Output signals

The following output signals are available as standard and can be independently applied to the front-end outputs.

Sine Type Pure tone - Swept sine - Step sine
Frequency 100 µHz to 40 kHz - smooth variation between step - resolution 10 µHz
Frequency control Start & stop frequencies - pause/release during sweep - adjustment during pause - step
Cycle control One shot - One cycle - continuous sweep between boundaries
Amplitude control Settling time 100 µs to 10 s - Stabilization time 0 s to 1000 s
Sweep speed Linear: 300 mHz/s to 20 kH/s - Log: 5 mOct/s to 330 Oct/s
Step control Synchronized with analysis end - free run
Gain control -15 dB to +60 dB - Independent for each output
- Amplitude variation controlled by settling time (1 ms to 1000 s)
Phase control Offset ±360° - Independent for each output - phase variation controlled by phase speed (1.5°/s to 360°/s)
Number 2 to 6 sine generated simultaneously - with independent phases and amplitudes - synchronized frequencies
Multi-sine Frequency span From 125 mHz to 40 kHz
Amplitude 0 to 2.5 Vrms - Settling time 100 µs to 10 s
Resolution 125 mHz to 400 Hz - 101 lines to 6401 lines
Phase Fixed (all sine have same phase) - Random
Burst 0 to 100% - Step 1%
Number 2 to 4 independents multi-sine

Output signals (continued)

Random noise Frequency span From 125 mHz to 40 kHz - independent lower and upper frequencies
Amplitude 0 to 2.5 Vrms - Settling time 100 µs to 10 s
Type White - pink
Burst 0 to 100% - Step 1% - Bloc duration: 2.5 ms to 100 s
Number 2 to 6 independent and uncorrelated random noises
Chirp Frequency span From 3.125 Hz to 40 kHz - independent lower and upper frequencies
Amplitude 0 to 7,07 Vrms - Settling time 100 µs to 10 s
Size 256 - 512 - 1024 - 2048 - 4096 - 8192 - 16384
Burst 0 to 100% - Step 1%
Number 2 to 6 independents chirps

In addition to these predefined signal, any recorder or imported signal file can be generated synchronously on front-end the outputs.

Filters

The following filters can be applied on front-end inputs, plug-ins analyzers and output generators.

ABW: the analysis bandwidth, of the plug-in or front end where the filter is in use.

High pass Type ButterworthIIR type
Order 1 to 6
Cutoff frequency ABWto ABW/1024
Low pass Type ButterworthIIR type
Order 1 to 6
Cutoff frequency ABW to ABW /102.4
Stop/pass band Type ButterworthIIR type
Order 2 to 10
Cutoff frequencies ABW to 0.055 * ABW
Bandwidth ABW/2 to 0.0075 * ABW
Integrators Type HP, Single or double
High pass frequency ABW /10 000 or ABW /2 000
Integration time 2 ms to 500 s
Differentiators Type Single
Average Sliding - 0 to 2 s
Weightings Laws A & C laws
Bandwidths 10 kHz to 40 kHz
All By-Pass Apply/bypass without stabilisation time - independently on each filter
Application Same filters set for each plug-in
Label Each filter features a user define name

Notes

The above specifications describe all the guaranteed capacities and performances of the NVGate V12 or upper. Functionalities may change depending on operation mode (connected to a 3-Series unit or office). Plug-in analyzers, options and channel number availability depend on purchased options.

The instruments hardware are described separately in the "Instruments specifications page" OROS reserves the right to modify the specifications without notification.