| ==DC Dynamical sensor y = ax+b calibration ==
| |
|
| |
|
| |
| It is now possible to calibrate a dynamical sensor in "DC" or "DC floating" coupling using 2 values, then the software will automatically compute sensitivity and offset to obtain the "y = ax + b" formula.
| |
|
| |
|
| |
| This function is practical for 4-20 mA sensors or “quasi-static” sensors acquired on dynamic channels. Wire sensor displacement sensor calibrated with a rule, pressure sensor with a calibrated compressor, proximity probe calibrated with a micrometer.
| |
|
| |
| For using it, first create a DC sensor on the sensor database. Then apply this sensor to a channel.
| |
|
| |
| Now on the calibration part, you can calibrate it using 2 values. Then the software will automatically apply the sensitivity and offset.
| |
|
| |
| [[File:calibrator.png|600px]]
| |
|
| |
|
| |
|