Difference between revisions of "NVGate 2021: Release note"

Jump to navigation Jump to search
Line 36: Line 36:
To guarantee the stability of the filter, there are certain constraints on the value of the cut-off frequency. Compared with previous version of NVGate, such constraints are relaxed significantly now.
To guarantee the stability of the filter, there are certain constraints on the value of the cut-off frequency. Compared with previous version of NVGate, such constraints are relaxed significantly now.


[[File:filter_cut_off.png|600px]]
[[File:image_2020-11-27_095451.png|600px]]


For the high pass and low pass filters, the maximum value of the cut-off frequency is the input frequency range '''FR'''. The minimum value is '''FR''' / 50000 in Office mode, and '''FR''' / 40000 in Connected mode. In previous version of NVGate, the minimum value is '''FR''' / 40 for the low pass filter, and '''FR''' / 400 for the high pass filter.  
For the high pass and low pass filters, the maximum value of the cut-off frequency is the input frequency range '''FR'''. The minimum value is '''FR''' / 50000 in Office mode, and '''FR''' / 40000 in Connected mode. In previous version of NVGate, the minimum value is '''FR''' / 40 for the low pass filter, and '''FR''' / 400 for the high pass filter.  