* '''Type''': For acquisition setup preparation, the XPod presence can be simulated in the office mode:


One XPod type selector is available for each XPod position. This setting will include each new Xpod type according to their availability on the OROS catalog.
===General Xpod settings===


[[Image:front_end_07.png|framed|none]]
* '''Type''': For acquisition setup preparation, the XPod presence can be simulated in the office mode:


One XPod type selector is available for each XPod position.
[[File:xpod bridge.png|framed|none]]


* '''Bridge max current''': Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.
* '''Bridge max current''': Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.[[File:xpod bridge2.png|400px]]
* '''Bridge voltage''': Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.
* '''Bridge voltage''': Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.
* '''Cold Compensation''':''' '''Enable or disable the internal cold junction compensation for the corresponding active Temperature Xpod. This setting enables the compensation of the connectors metal junction thermocouple voltage at the Xpod side. It uses the Xpod internal temperature measurement (2 locations/Xpod) to correct the measured temperatures.
The XPod Bridge conditioner extension module is for strain gauges, dynamic pressure, force and acceleration measurements.


===input settings===
[[File:xpod bridge2.png|framed|right]]
*'''Type:''' bridge signal conditioning for strain gage. <br>
*'''Type:''' bridge signal conditioning for strain gage. <br>