Line 282: | Line 282: | ||
The XPod Bridge conditioner extension module is for strain gauges, dynamic pressure, force and acceleration measurements. | The XPod Bridge conditioner extension module is for strain gauges, dynamic pressure, force and acceleration measurements. | ||
Type: bridge signal conditioning for strain | *'''Type:''' bridge signal conditioning for strain gage. <br> | ||
Gauge type: Full, Half or Quarter bridge mount. The completion resistors are included in the Xpod | |||
Resistance: Bridge completion resistor: 120 or 350 Ohms for quarter bridge mount. | *'''Gauge type''': Full, Half or Quarter bridge mount. The completion resistors are included in the Xpod.<br> | ||
Gain: Bridge gain: 10 or 100 depending on the required precision and range. | |||
Max current: the provided current can be limited to 30 mA up to 4 V and 12 mA up to 10 V. | *'''Resistance''': Bridge completion resistor: 120 or 350 Ohms for quarter bridge mount.<br> | ||
Excitation voltage: Each Xpod provides an excitation voltage from 0 to 10 V. | |||
Offset Comp.: Bridge offset comp: Can be used for manual balance of the bridge | *'''Gain''': Bridge gain: 10 or 100 depending on the required precision and range.<br> | ||
*'''Max current''': the provided current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.<br> | |||
*'''Excitation voltage''': Each Xpod provides an excitation voltage from 0 to 10 V.<br> | |||
*'''Offset Comp.''': Bridge offset comp: Can be used for manual balance of the bridge.<br> | |||
===X-Pod: Temperature probe conditioner=== | ===X-Pod: Temperature probe conditioner=== |