The instant averaging is always running and is based on the exponential averaging with a time constant equal to 1/fc where fc is the center frequency of corresponding filter. So each detector has its own time constant and the output fluctuation in the worst case is limited to <nowiki>+</nowiki>/- 0.4 dB with a sine input signal.
The instant averaging is always running and is based on the exponential averaging with a time constant equal to 1/fc where fc is the center frequency of corresponding filter. So each detector has its own time constant and the output fluctuation in the worst case is limited to <nowiki>+</nowiki>/- 0.4 dB with a sine input signal.


The averaging process provides a lot of modes (linear, exponential and dedicated to acoustical measurements) which are described in the <font color="#FF0000">'''8.2.14.5''' </font>'''Settings <nowiki>|</nowiki> Analysis - Average Mode''' chapter.
The averaging process provides a lot of modes (linear, exponential and dedicated to acoustical measurements) which are described in averaging part.


A hold box allows to get Maximum and Minimum spectra during one measure.
A hold box allows to get Maximum and Minimum spectra during one measure.