====DC power supply====
The DC power must fulfill the following conditions:


{|border="2" cellspacing="0" cellpadding="4" width="79%"
====O4 Powering on====
|Voltage
A beep is produced when the instrument is connected on PC'''.'''
|OR34 Power
|
[[Image:Usersmanual_53.gif|framed|none]]
Overload action
 
|-
|DC, 10 V to 28 V
|<nowiki>></nowiki>15 VA
|<nowiki>+</nowiki> and – poles are short circuited if Vpeak <nowiki>></nowiki> 31 V
 
|}
 
OR34 instrument is DC power supplied using the following plug:
 
[[Image:Usersmanual_54.png|framed|none]]
 
All the connections of OR34 instruments comply with SELV (Safety Extra Low Voltage) conditions.
 
====OR34 Powering on====
A beep is produced when the instrument is connected to the mains'''.'''
 
'''Power on OR34 Instrument''' by pressing the On/Off button of the back panel and wait until a beep is produced (2 sec) then release the button. The LED is fast flashing green (system is loading) and then slow flashing green (system OK).
 
'''Power off OR34 Instrument''' by pressing the On/Off button of the back panel (The LED is flashing red) Then wait until a beep is produced (2 sec) to release the button.
 
'''If OR34 Instrument does not respond''', unplug the main power supply and press the reset button [[Image:Usersmanual_55.png]] (use a paperclip) of the back panel.