NVGate Analyzer Setting Browser (ASB)

From OROS Wiki
Jump to navigation Jump to search

ASB (Analyzer Setting Browser)

ASB.png

The analyzer settings (front-end and analysis) are gathered in a tree called the Analyzer Setting Browser (ASB). The ASB window is shown/hidden from the Home tab / View group with the Workspace check box.

The ASB is designed to offer access to all analyzer settings. The standard access to the analyzer setup offers a simplified and more intuitive general Vision interface (see Chapter 6). The Modules: The main entries of the ASB consist of modules corresponding to the main analyzer functions.

  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 10.png The sources that acquire, store and play-back signals.
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 11.png The resources that provide events, filters, weighting windows, and tachometers to the plug-in analyzers.
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 12.png The plug-in analyzers that process independent analysis modes.
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 13.png The Waterfall that synchronizes the results computed by the plug-ins.

The ASB has a maximum depth of 3 levels: module, sub-module, and settings. In the case of multiple similar sub-modules (inputs for example), the sub-modules are arranged in collections.

For each module, this document provides an array that summarizes the available results with detailed information. The sub-modules: The modules contain the function and their setting functions.

  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 14.png Sub-module = a function
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 15.png Inputs = incoming signals measured from the front-end
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 16.png Outputs = signals generated from the hardware
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 17.png Collection = A set of identical sub-modules (e.g; inputs, channels, events, ect..)

The Settings:

The settings are the function’s parameters. There are multiple type of settings identified by specific icons:

  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 18.png A scalar (integer or real) boundaries are displayed in the tooltip
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 19.png A list: the setting is one of the list items. Note: the list content may vary with the current configuration.
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 20.png A text e.g; the label of channel
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 21.png A signal. Usually a file containing signal(s)
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 22.png A check box. (e.g; mute/un-mute generator)
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 23.png An action. Temporary information. (e.g; a user event trigger)
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 24.png Special: multi-parameter setting (e.g; The tach ratio)
  • REF-MAN VOL1 Analyzer Settings Browser wiki partA 25.png Special: graphical multi-parameter setting (e.g; force/response weighting windows)

Note: A setting may be informative, i.e, it is not possible to modify it. In such case an I is displayed close to the icon: REF-MAN VOL1 Analyzer Settings Browser wiki partA 26.png

Note: Settings may be hidden or visible depending on the current configuration.

Considerations for post-processing : If the time signal is recorded on the PC or on the Mobi-Disk, it is highly recommended to be in office mode (hardware not connected) in order to avoid low performances and malfunctions.


Controls

This section explains how to access the different controls contained in the ASB (settings, custom display...) and what they mean.

Access

There are 4 way to access a setting:

From the module

The first access level is available by right clicking on a module (Front-end, Recorder, Tachometer, FFT...) and Properties where all the settings contained in this module can be adjusted.

From the collection

The second access level is available by right clicking on a collection (Inputs, Outputs, Channels, Tracks...) and Properties where all the settings contained in this collection can be adjusted.

REF-MAN VOL1 Analyzer Settings Browser wiki partA 27.png

It is also possible to copy and paste some items from a cell, a line, a column or a group of cell to another if they have compatible titles (Label, Transducer, Sensitivity…).

Copy/paste Front-end tables

Taking advantages from the Excel capabilities for tables’ creation and updating, NVGate V8.30 offers now to update the front-end/player tracks from Excel table.

This feature uses the Windows® clipboard.

REF-MAN VOL1 Analyzer Settings Browser wiki partA 28.png

Two entries are made available from the contextual menu of the front-end or player tracks properties:

Copy to Clipboard and Paste from Clipboard. These commands come in addition to the local copy & paste which allows duplicating cells locally. Note that it is also available from any NVGate table.

Using the new Copy to clipboard capability allows pasting your inputs/track configuration to Excel, modifying it and then return it back in to NVGate. This also works with any spreadsheet software accepting csv content.

Generally, it works as a WYSIWYG (What You See Is What You Get); Data are exported in the same format/unit as displayed in NVGate table. Import (Paste into NVGate) behaves like you are typing the values.

Nevertheless, some precautions apply transferring data from Excel to NVGate:

  • Headers (line and column) are added when pasting in Excel. This cell should not be copied prior to paste in NVGate.
  • For the purpose of correct reading, units’ columns are automatically added. These columns are ignored when pasting into NVGate. These columns may be absent from the Excel table.
  • The figure (sensitivity, offset, etc...) units are read according to the current NVGate user preference one - not the units in the Excel table.
  • Paste in NVGate starts from the upper-right cell of the area to update in the NVGate table. The cell should not be in editing mode.
  • If pasted value is not acceptable for NVGate, the corresponding cell will:
  • Adjust it to the closest upper value if it is possible (ex: max range)
  • Ignore and keep the current one if not acceptable (ex: unknown coupling)

From the sub-module

The third access level is available by right clicking on a sub-module (Input x, Output x, Track x, Channel x...) and Properties, where all the settings contained in this sub-module can be adjusted.

REF-MAN VOL1 Analyzer Settings Browser wiki partA 04.png


From the setting

The fourth access level is available by right clicking on the setting itself (Label, Duration, Transducer, Coupling...) and Properties, where the new value for this setting can be entered.

REF-MAN VOL1 Analyzer Settings Browser wiki partA 29.png

Contextual menu

Contextual menus appear when right clicking on an area or an item.

On an item

Different controls are available according to the item you right click on.

  • Properties: displays the properties of the item.
  • Add/Remove: Available for collections (Channels, Event...), it enables or disables some items of the collection.
  • Set to default: Reset a setting, a sub-module or a module to the default values. All connections are removed.
  • Copy (Ctrl+C): Copy a setting, a sub-module, or a module. This copy is internal and does not use the Clipboard.
  • Paste (Ctrl+V): Available when a copy action has been previously performed. Only the settings common to both the copy and the paste item are affected.
  • 'IDN 'info: Available only when the Shift key is pressed. It displays an identification number used by NVDrive.
  • View state: Available only on plug-ins, it hides or shows the status of this plug-in in the Status window.
  • Add to control panel: Enable the corresponding setting to be modified from the Control Panel.
  • dB: Enable the corresponding setting to be displayed in dB. Only available for some scalar settings.
On ASB background
  • Auto width: This function adjusts the width of the ASB.

Plug-in analyzer status

An icon is displayed in the ASB when a plug-in is connected. For each plug-in the icon displays the analysis status (running, stopped, recording...).

  • Running REF-MAN VOL1 Analyzer Settings Browser wiki partA 30.png: the plug-in analyzer is processing the data acquisition.
  • Paused REF-MAN VOL1 Analyzer Settings Browser wiki partA 31.png: the plug-in analyzer is paused. The plug-in analyzer is not processing any data. The user must click on the "continue" button to continue the analysis. For the recorder plug-in the "Continue" action starts a new record because in the paused state no data is recorded.
  • StoppedREF-MAN VOL1 Analyzer Settings Browser wiki partA 32.png: the plug-in analyzer is stopped.
  • RecordingREF-MAN VOL1 Analyzer Settings Browser wiki partA 33.png: the plug-in recorder is processing the data acquisition.
  • Block TriggeringREF-MAN VOL1 Analyzer Settings Browser wiki partA 34.png: the plug-in analyzer is waiting for the Trigger event to occur in order to process the data.
  • WaitingREF-MAN VOL1 Analyzer Settings Browser wiki partA 35.png: in the manual rejection mode, the plug-in analyzer is waiting for the user to accept or to reject the current analyzed signal block.
  • Failure REF-MAN VOL1 Analyzer Settings Browser wiki partA 36.png: the plug-in analyzer has failed (the analyzer is no longer able to compute the data acquisition; a warning message will appear)

Copy/Paste settings between plug-ins

By right clicking on any settings of the ASB you can Copy values of settings, and then Paste it anywhere else. There is an automatic type matching, which allows pasting only value of the same settings (using the IDN number).

This Copy/Paste is available at each level of the ASB. This allows for multi-analysis on the same condition.

For example, you can see the same signal in the time domain and in the spectral domain, with exactly the same settings (for instance resolution). In the following example, the FFT1 have been set in the ’Spectral domain’, with a specific range and a resolution. For the FFT2, right click on FFT2 and paste the setting from FFT1, then you have just to change the domain from spectral to time in that case. You can compare the signal in, exactly, the same conditions.

ASB User preference

NVGate provides specialized modes that enable dedicated functionalities related to the application. You can enable them by selecting specialized mode functionalities such as:

  • Availability of results
  • Settings
  • User preferences

Unselecting specialized mode will mask these functionalities. There are 2 specialized modes:

  • Structural mode: this mode is dedicated to the acquisition of modal data through hammer acquisition or signals provided by up to 6 shakers. The following topics are available when structural mode is active:
  • Visibility of FRF (H1, H2) and Coherence results in FFT plug-ins.
  • Visibility of time windows (force and response) in the ASB, and control panel.
  • Availability of the inputs node settings in the inputs properties (ASB, Control Panel and Ribbon)
REF-MAN VOL1 Analyzer Settings Browser wiki partA 03.png

Availability of accept/reject in the bottom right:

  • Rotating mode: this mode is dedicated to rotating and reciprocating machines. The following topics are available when rotating mode is active:
  • FFT order extraction, constant band tracking or synchronous order.
  • Tachometer settings.
  • Plug-in analysis ’Sync Order’ will be unlocked.