NVGate 2022: Release note

From OROS Wiki
Jump to navigation Jump to search

NVGate 2022


Ergonomics

Tab: new icone disposition

Go to results: new one click icones

RPM buttons create automatic events

New settings by default

The settings by default in NVGate have been update to the one below. With this default seetings, you can do rotating analysis measurement directly.. Be aware than if you load an exiting project , this is NOT affected you previous settings are keep. You can also easily change this settings.

FFT Overlap: 66%

FFT lines: 3200

FFT Overall low freq : 2 Hz


SOA resolution: 1/8

SOA overlap: 5

SOA OVL lower order: 0.2


Event Delta RPM: Delta velocity 10 rpm

Waterfall Depth: 1000

Tachometer/RPM Profiles/ duration: 120 sec

NVGate reader

Use NVGate in reader mode : no need usb dongle or hardware.

Fractionnal tach : ration over 100

Reports Tools Ribbons 356.png: Fractional tachometer: computes RPM speed for a non accessible shaft by using gear ratio setting. Adapted for gear boxes and transmissions. The ration ca be now set up to 100 ! (previously it was 2)

Used to define up to 4 fractional tachs using data from the tach or the Ext Tach.


Tacho ratio.png
  • Tach ratio: this is the ratio between the output angular speed and the input angular speed. This setting is defined by the product of 2 fractions: N1/D1 * N2/D2 where N1, D1, N2 and D2 are integer values. Tach.1 / Tach.2 ratio maximum value can be set up to 100.
Le100.png

Up to 3 real time: plug in SOA

Oros , leader on rotating vibration analysis, is the only one in the world to let you do at the smae time: 3 real time synchronous order analysis (this is also available in post analysis.)

3SOA.PNG

Add on : PV (pressure volume) diagram

How to use :

  • Define one (or more) channel as a pressure.
  • Add this channel into Sync. order (SOA) plug in.
  • On SOA plug in , average, domain, select: "Angular"
  • Open PV_Diagram.exe
  • Click on refresh (right click) in available results if you do not see anything.
  • Select the input.
  • Compute once will display the PV diagram base once based the last triggerd block.
  • "background computation, will continuously compute the diagram.

NVDrive : Read Write NVGate properties (Project / Measurement)

Orbigate

Tach ratio over 100

Teamwork: compatibility with several OR3X