NVGate X-Pod: Temperature probe conditioner
This Xpod handles thermocouples and RTD transducers.
- PT100, PT1000 and J, K, T, N, E
- Integrated linearization
- Automatic cold junction compensation
- Standard fl at pin connectors
Expander modules
The XPod is a device that can be fixed on OR3X TW analyzer. Each XPod is associated to a block of 8 inputs.
In the connected mode, when and XPod is present each corresponding input can be routed to:
· The BNC input (Standard)
· The XPod connector
Tutorial
- First connect Xpod to the analyzer
- Start Analyzer
- Start NVGate
- Select a DC Input and put in in a "plug in" (recorder, waterfall (for profile)...)
- On settings, put "input type" : physical qty: temperature,the type of probe, range will be automatically set.
Settings
General Xpod settings
- Type: For acquisition setup preparation, the XPod presence can be simulated in the office mode:
One XPod type selector is available for each XPod position.
- Cold Compensation: Enable or disable the internal cold junction compensation for the corresponding active Temperature Xpod. This setting enables the compensation of the connectors metal junction thermocouple voltage at the Xpod side. It uses the Xpod internal temperature measurement (2 locations/Xpod) to correct the measured temperatures.
Input settings
� Probe: Define the type of temperature probe connected on the XPod: o Thermocouple : J, K, T, N, E o RTD: PT100, PT1000, PT100 3 wires For each type of probe the Temperature range is shown in the “Range” setting.
While using RTDs (PT100 PT1000), the applied current can be selected
(the standard is 500 uA)
Sensor Connector
This is a standard. If you buy temperature sensor it is already cable like this :
note : for 2 wire probe the 3thd pin is optional. it is workng with a 2 pins connector.