OR34 Compact analyzer
OR34 Compact analyzer
OR34 is a compact analyzer. It offers 2 or 4 dynamic inputs with 2 ext. sync and 1 generator.
Back Panel
The back panel includes:
- Ethernet connector: RJ-45 connector, Network environment 100 Mbps (100 BASE-TX), cable required Category 5 unshielded twisted-pair
- Interface connector: RJ-11 for remote control and maintenance
- DC power connector: Connector DIN female
DC power supply
The DC power must fulfill the following conditions:
Voltage | OR34 Power |
Overload action |
DC, 10 V to 28 V | >15 VA | + and – poles are short circuited if Vpeak > 31 V |
OR34 instrument is DC power supplied using the following plug:
All the connections of OR34 instruments comply with SELV (Safety Extra Low Voltage) conditions.
OR34 Powering on
A beep is produced when the instrument is connected to the mains.
Power on OR34 Instrument by pressing the On/Off button of the back panel and wait until a beep is produced (2 sec) then release the button. The LED is fast flashing green (system is loading) and then slow flashing green (system OK).
Power off OR34 Instrument by pressing the On/Off button of the back panel (The LED is flashing red) Then wait until a beep is produced (2 sec) to release the button.
If OR34 Instrument does not respond, unplug the main power supply and press the reset button (use a paperclip) of the back panel.
OR35/OR36/ORMP/OR38 Teamwork instruments
OR35, OR36, ORMP & OR38 cover high end applications with multi-channels chassis. They offer from 4 to 32 universal inputs, 2 ext. synch, 2 generators and optional auxiliary channels supporting generators, trigger/tach and parametric inputs.