Difference between revisions of "Sound Intensity Manual"

Jump to navigation Jump to search
m
Line 335: Line 335:
The '''Probe''' tab allows the user to select the probe type, the model, to enter the serial number, to set up the probe serial remote control and the keyboard shortcuts.
The '''Probe''' tab allows the user to select the probe type, the model, to enter the serial number, to set up the probe serial remote control and the keyboard shortcuts.


The '''Transducers''' tab enables the user to select the microphone pair by mean of a drop down menu linked to a transducer database. The value for the sensitivity will be loaded in the relevant fields of the table at the bottom of the window. All these values can be edited by clicking the table fields.
The '''Transducers''' tab enables the user to select the microphone pair by means of a drop down menu linked to a transducer database. The value for the sensitivity will be loaded in the relevant fields of the table at the bottom of the window. All these values can be edited by clicking the table fields.


The table shown in the '''Spacer''' tab must be used to set-up the spacer length and the measurement frequency span for each one. The spacers defined in this tab are the only one visible and selectable during the editing and arrangement of the spacers in the Analysis Set-up.
The table shown in the '''Spacer''' tab must be used to set-up the spacer length and the measurement frequency span for each one. The spacers defined in this tab are the only ones visible and selectable during the editing and arrangement of the spacers in the Analysis Set-up.


The <font color="#0000FF">'''PRI index''' </font>'''Tab''' shows the '''''Pressure Residual Intensity Index''''' for the measurement chain in use (both numerically and graphically) and the comparison with the IEC 61043 Standard Limits. Using the Load and Save buttons, the user can retrieve from or write to the disk several PRI profiles. Pushing the Set IEC 61043 PRI limits button, the default limits specified in the standard are loaded. The PRI here used is referred to the actual measurement chain and it follows the advices of the standards: the PRI is the one for the 25mm spacer. During the measurement the software automatically computes, depending on the spacer in use and its frequency range, the corrected values for each active frequency band. The values of the PRI spectrum can be edited typing in the Lp and Li cells (useful if the user has the probe/system calibration chart from the producer or the Calibration Laboratory).
The <font color="#0000FF">'''PRI index''' </font>'''Tab''' shows the '''''Pressure Residual Intensity Index''''' for the measurement chain in use (both numerically and graphically) and the comparison with the IEC 61043 Standard Limits. Using the Load and Save buttons, the user can retrieve from or write to the disk several PRI profiles. Pushing the Set IEC 61043 PRI limits button, the default limits specified in the standard are loaded. The PRI here used is referred to the actual measurement chain and it follows the advices of the standards: the PRI is the one for the 25mm spacer. During the measurement the software automatically computes, depending on the spacer in use and its frequency range, the corrected values for each active frequency band. The values of the PRI spectrum can be edited typing in the Lp and Li cells (useful if the user has the probe/system calibration chart from the producer or the Calibration Laboratory).
writer
240

edits

Navigation menu