| This group manages D-rec (Direct recording) setups. D-rec allows recording signals with the analyzer without any PC or connection. | | This group manages D-rec (Direct recording) setups. D-rec allows recording signals with the analyzer without any PC or connection. |
|
| |
| *
| |
| [[Image:Reports_Tools_Ribbons_267.png|framed|none]]
| |
| New'':'' Create a new recording configuration and save it in the analyzer memory.
| |
| From this button, NVGate will propose a set of dialog boxes that guide the user through the necessary settings to build a D-rec setup. The displayed dialogs are the following:
| |
|
| |
| 1. Channels connection
| |
|
| |
| 2. Event connections
| |
|
| |
| 3. Recorder properties
| |
|
| |
| 4. Records setup management
| |
|
| |
| Item #1 and #3 are exactly the same as the standard ASB dialogs for the channels connection and the recorder setup.
| |
|
| |
| Item #2 is also the same but only a subset of the events is handled by D-rec: Recorder trigger on Edge on Ext. synch, Level and delta level on DC inputs and periodic event only.
| |
|
| |
| Item #4 allows managing the setups saved in the analyzer. (See ''Setup'' below for details)
| |
|
| |
| *
| |
| [[Image:Reports_Tools_Ribbons_268.png|framed|none]]
| |
| Setup: Manages the D-rec Setups through the following dialog box. The D-rec setups can be written, removed, imported and exported from this dialog box.
| |
| To write an NVGate configuration in the hardware, press the "Add current setup" button. The current front-end and recorder settings will be saved in the analyzer. Note that all active channels (IN, DC, EXT and CAN) will automatically be connected to the recorder tracks.
| |
|
| |
| The Delete button removes definitely the selected setup(s) from the analyzer memory".
| |
|
| |
| One of the D-rec setups saved in the analyzer can be set as the default setup when starting the D-rec. Check it in the "Default" column. The default setup is loaded when the operator
| |
|
| |
| Each setup can be exported in text file to be used as a reference of the D-rec settings when using it on-site to match the transducers and cables. To create a description file, use the "Export as txt" button.
| |
|
| |
| It is possible to recover, control or modify an existing D-rec setup which is already saved in the analyzer. The "Import in ASB" function will overwrite the current ASB with the configuration saved in the analyzer.
| |
|
| |
|
| |
|