Front end

Front-end Dynamic analog inputs 4 inputs (BNC)
Dynamic analog outputs 1 output (mini Lemo), delivered with mini-Lemo to BNC adaptor
External syncs. 2 trigger/tachometer inputs (BNC)
Weight 534 g (1.17 lb)
Dimensions Overall (L.W.H.)

Power supply

Power < 8 W
From PC Port format
USB-PD 3.0 technology compliant
From main Voltage
Power
USB Port USB-C x 2

PC requirements

Minimum 1 GB<ref>) Waterfall depth depends on available memory.

</ref> of RAM / 250 MB free on HD + storage for measurements and signals / 1024 x 768 display

Recommended Quad core processor (e.g.: Intel Core i5) / 6 GB of RAM / GPU / 1920 x 1080 display / 1 GB free on SSD + storage for signals
Connections USB 3.0 for data & Power > 7 watts
Operating systems Windows 10 / Windows 11 / MS Office: 32 bits only

Environmental / Compliance with standards

CE Indicates compliance with: EMC Directive 2014/30/EU
EMC IEC 61326-1:2013 Electrical equipment for measurement, control and laboratory use - EMC requirements
Materials ROHS 2011/65/EU
WEEE 2012/19/EU
Humidity Max 93 % RH at 40°C non-condensing
Temperature Operating 0 °C to 40 °C (14 °F to 113 °F)
Storage -20 °C to 65 °C (-4 °F to 149 °F)
Absolute maximum rating -35 °C to 70 °C (-31 °F to 158 °F)
Shocks & bump Operating 40 g (6 ms, ½ sine, 3 chocks, all axes, IEC 60068-2-27)
60 g (3 ms, ½ sine, 3 chocks, all axes, IEC 60068-2-27)
Storage 40 g (6 ms, ½ sine, 1k shocks, IEC 60068-2-29)
60 g (11 ms sawtooth, 3 shocks 3 axes MIL-STD-810F 516.5)
Vibrations Operating 2.5 g (sine, 15-500 Hz, all axes, IEC 60068-2-6)
5 mm (sine, 5-15 Hz, all axes, IEC 60068-2-6)
7.7 grms (random, 20-2k Hz, 30 min, MIL-STD-810F 514.5)
Altitude Operating, non-tested above ≤ 2000 m (6562 feet)
Enclosure IP 40

Front-end

Dynamic inputs

Sampling
Sampling frequencies

(Additional decimators allow analysis bandwidth down to 0.8 Hz)

256 kHz, 204.8 kHz, 131.072 kHz, 102.4 kHz, 65.536 kHz, 51.2 kHz, 37.768 kHz, 25.6 kHz, 16.384 kHz, 12.8 kHz, 8.192 kHz, 6.4 kHz, 5.12 kHz, 4.096 kHz, 3.2 kHz, 2.048 kHz
Converters One 24 bit sigma-delta ADC for each input
Frequency relative precision 0.5 10-4 (typical 1 10 –5)
Synchronization All inputs synchronized on the same sampling clock
Anti-aliasing filter
Type Over-sampled digital filters
Slope > 400 dB/octave
Pass band ripple < ± 0.005 dB
Rejection of parasites bands > 100 dB (@ frequency > 0.57 x FS)
Effective bandwidth 0.45 x FS (ex: 23.4 kHz @ 51.2 kS/s)
Range (peak)
With amplifier (included) ±100 mV, ±300 mV, ±1 V
Direct ±10 V
With attenuator (included) ±40 V
Absolute accuracy
Resolution 24 bits (144 dB)
All input ranges at 1 kHz ±0.05 dB (typical ±0.015 dB)
Temperature variability < 0.002 dB / 10 °C
DC offset
±100 mV, ±300 mV and ±1V ranges < ± 100 µV
±10 V range < ± 1 mV
±40 V range < ± 2 mV
Frequency flatness and phase response<ref>) Includes channel to channel match with different ranges

</ref>

±10 V range, DC to 20 kHz < ±0.02 dB / < ±0.02 °
±10 V range, 20 kHz to 40 kHz < ±0.05 dB / < ±0.05 °
±10 V range, 40 kHz to 100 kHz < ±0.05 dB / < ±0.08 °
±0.3 V , ±1 V ranges, DC - 20 kHz < ±0.02 dB / < ±0.1 °
±0.3 V, ±1 V ranges, 20 kHz - 100 kHz < ±0.05 dB / < ±0.2 °
±0.1 V range, DC to 20 kHz < ±0.02 dB / < ±0.4 °
±0.1 V range, 20 kHz to 40 kHz < ±0.1 dB / < ±0.6 °
±0.1 V range, 40 kHz to 100 kHz < ±0.15 dB / < ±0.8 °
±40 V range, DC - 20 kHz < ±0.1 dB / < ±0.4°
±40 V range, 20 kHz - 40 kHz < ±0.1 dB / < ±0.6 °
±40 V range, 40 kHz - 100 kHz < ±0.15 dB / < ±0.8 °
Cross-talk
Between N (N is odd) and N+1 inputs:

@ 1 kHz: < -120 dB, @ 20 kHz: < -96 dB, @ 40 kHz: < -90 dB

Between any inputs excluding: N (N is odd) and N+1 inputs:

@ 1 kHz: < -140 dB, @ 20 kHz: < -114 dB, @ 40 kHz: < -108 dB

Signal to noise ratio
With 50 Ω terminators:

±10 V range, 100 kHz bandwidth: > 100 dB, spurious lines < -115 dB of full scale

±10 V range, 20 kHz bandwidth: > 104 dB, spurious lines < -125 dB of full scale

Input noise
With 50 Ω terminators:
±100 mV range 20 kHz BW < 3.5 µV rms, 80 kHz BW < 5 µV rms, 100 kHz BW < 6 µV rms
Impedance 1 MΩ ±1 %, < 100 pF
Protection Overvoltage ±60 V peak without damage - On any inputi
Dynamic Spectral domain > 140 dB<ref>) 25601 lines / 30 sec. averaging

</ref>

Coupling AC Cut-off frequency 1.13 Hz ±10% (analog filter)
DC
ICP 2 mA or 4 mA power supply with AC coupling (±10%)
ICP + TEDS ICP + reverse current on TEDS reading operations
GND Shortcut to ground - Automatic current limitation to 50 mA
Floating Coupling AC or DC / All ranges / Overall voltage < ±40 V
TEDS Standards IEEE 1451.4 2001 revision 1
Supported templates Accelerometer/Force meter (25)

Microphones (27, 28 and 29)

Dynamic outputs

Sampling Converters One 24 bit DAC for each output
Synchronization Same sampling clock as the dynamic inputs

Range

Direct ±10 V peak
With attenuator (included) ±1 V peak
Clipping User selectable in the output range
Digital gain From 10-5 to 103
Absolute accuracy Resolution 24 bits (144 dB)
All output ranges at 1 kHz ±0.05 dB
Temperature variability < 0.1 dB / 10 °C
Frequency response Variation relative to 0 dB @ 1kHz
All ranges, at 10 kHz < ±0.05 dB
All ranges, at 20 kHz < ±0.15 dB
All ranges, at 40 kHz < ±0.8 dB
All ranges, at 80 kHz < ±2 dB
All ranges, at 100 kHz < ±3 dB
Noise floor level 10 V range, 20 kHz bandwidth -110 dB of full scale, spurious lines < -125 dB of full scale
10 V range, 100 kHz bandwidth -105 dB of full scale, spurious lines < -125 dB of full scale
1 V range, 20 kHz bandwidth -99 dB of full scale, spurious lines < -110 dB of full scale
1 V range, 40 kHz bandwidth -94 dB of full scale, spurious lines < -110 dB of full scale
1 V range, 100 kHz bandwidth -90 dB of full scale, spurious lines < -107 dB of full scale
Impedance User selectable 50 Ω or Grounded
Current Max  ±10 mA
Protection Sum of injected + generated voltages ±15 V peak, On any outputi

Permanent short circuit supported

Total harmonic distortion THD @ 1 kHz < 0.002% or -94dB at 20 kHz BW
THD @ 5 kHz < 0.005% or -86dB at 20 kHz BW
Cross-talk Output 0 dBV to 50 Ω terminated input Lower than measurable noise

External sync

Sampling

Frequencies 128 times over-sampling of the current input sampling

(up to 32.8 MHz)

Converters High speed voltage comparator and time counter
Range (peak) Direct ±300 mV, ±1 V, ±3 V, ±10 V, ±40 V
threshold Amplitude precision ±1 % of range
Setting Hysteresis 1% (of input range) to input range
Hold off 0 s to 500 s
Slope Rise or fall
Hardwired pre-divider From 1 to 255
== Time resolution ==
> 160 ns (0.06° at 1kHz and 1.2 ° at 20kHz)
Pulse rate Max 375k pulse/s
Coupling AC Cut-off frequency 0.35 Hz ±10% (analog filter)

DC

Impedance 1 MΩ, < 100 pF
Protection on any external synci ±60 V peak without damage

Notes

The above specifications describe all the guaranteed capacities and performances of the instrument and are applicable to an O4 hardware, powered through USB port, at a stabilized room temperature of 23°C ±5°C and calibrated since less than one year.

The adapted control software NVGate is described separately.


i Exceeding absolute maximum ratings damages the system and voids guarantee.

File:Media/image4.png

<references />