NVGate Ribbons: Analysis Tab

From OROS Wiki
Jump to navigation Jump to search

The "Analyses" tab

In this tab, the software gathers the analyses controls. Each group corresponds to a plug-in analyzer. The content of this tab depends on the purchased optional plug-in.


Monitor group

The Monitor plug-in offers simple and powerful ways to monitor the measured signals. This 4 channels plug-in runs independently from the general analyzer status (running/Paused/stopped) with the capability of swapping connected channels at any time.

Reports Tools Ribbons 378.png

The second capability of the monitor is to extract statistical component of the signal in a specific bandwidth. Extracted levels can be used for triggering and waterfall references.

The monitor runs its analyses on a specific processor (DSP) allowing a totally independent computation for the regular analyses.

Note: Clicking on the bottom right icon (Reports Tools Ribbons 379.png) opens the monitor plug-in properties dialog allowing a full access to all the settings.

Read monitor page for more details.

Time Domain Analysis group (Optional)

The TDA plug-in is designed to get the most from the time domain signal completing the frequency/order analyses. This plug-in offers flexible signal overview and statistical extraction in parallel.

Note: Clicking on the bottom right icon (File:Reports Tools Ribbons 385.png) opens the TDA plug-in properties dialog allowing a full access to all the settings.

Read the Time domain page for more info.

FFTx group (Optional)

The FFT plug-in provide frequency distribution of the analyzed signal. It runs an FFT algorithm.

Reports Tools Ribbons 391.png

Note: Clicking on the bottom right icon ( Reports Tools Ribbons 392.png ) opens the FFTx plug-in properties dialog allowing a full access to all the settings.

Note: Up to 4 FFT plug-in can be present in the software configuration, each of them running independent analyses on the same or different channels.

Range: Setup the analysis range for the plug-in. This value is maximized by the front-end sampling frequency divided by 2.56.
Note: Each plug-in feature an independent bandwidth.

Lines: Select the analysis resolution in number of frequency lines. The resolution = Range / (Lines-1) in Hz

Average: Select the domain of averaging:

  • Spectral: the spectra module are averaged together
  • Time: the trigger block are averaged together -> provide absolute / relative phase
  • Frequency domain synchronous (FDSA): The complex spectra are averaged together (real/img) after phase synchronization on one spectral line -> allows synchronous time averaging (STA) without a trigger.

See Chapter 1 ASB - § Analysis plug-in/ TDA for details

Synchronous order group (Optional)

The SOA plug-in provide order distribution and extraction of the analyzed signal related to a tachometer. It runs re sampling + FFT algorithms

Reports Tools Ribbons 396.png

Note: Clicking on the bottom right icon ( File:Reports Tools Ribbons 397.png ) opens the SOAx plug-in properties dialog allowing a full access to all the settings.

Note: Up to 2 SOA plug-in can be present in the software configuration, each of them running independent analyses on the same or different channels.

Reports Tools Ribbons 398.png

Max order: Setup the order analysis range for the plug-in from 6.25 to 400.

Associated tach: Select the tachometer to be used for signal re-sampling. The order results provided by the SOAx plug-in are related to this tachometer speed

Reports Tools Ribbons 400.png

Order resolution: Select the analysis resolution in orders. Ranges from 1/32 (0.03125) to 1 order.

Average: Select the type of averaging. Linear, exponential peak hold and ref. peak hold See Chapter 1 ASB - § Analysis plug-in/ SOA for details

1/nth octave group (Optional)

The 1/n Octave plug-in provides constant percentage band spectra. It runs filter and detector. This plug-in complies the IEC 60 804 and 62 260 standards.

Reports Tools Ribbons 402.png

Note: Clicking on the bottom right icon ( File:Reports Tools Ribbons 403.png ) opens the 1/n OCT plug-in properties dialog allowing a full access to all the settings.

Reports Tools Ribbons 404.png

Upper band: Setup the center frequency of the highest frequency band

Reports Tools Ribbons 405.png

Lower band: Setup the center frequency of the lowest frequency band

Resolution: Select band resolution: from 1 band per octave to 24 band per octave.

Reports Tools Ribbons 407.png

Type: Select Averaging type. This setting features a mix of different averaging modes such as time weighting, classic, constant band* time and repeated Leq.

Average duration: Define the total average duration and the linear duration for exponential.

Short duration: Define the short average duration for the linear repeated and Leq. See Chapter 1 ASB - § Analysis plug-in/ 1/n Octave for details

Overall acoustic group (Optional)

The overall acoustic plug-in provides sound pressure level measurement. This plug-in complies with the IEC 61672 standard. It runs 3 RMS detectors + 1 peak per channel.

Reports Tools Ribbons 410.png

Note: Clicking on the bottom right icon (Reports Tools Ribbons 411.png) opens the 1/n OCT plug-in properties dialog allowing a full access to all the settings.


Read the OVA page for more info

Waterfall group

The so called waterfall is a stack for analysis results and reference. It allows storing results from different plug-ins in a synchronized database. These data are viewed with the 3D waterfall or profiles graphs.

Note: Clicking on the bottom right icon ( File:Reports Tools Ribbons 417.png ) opens the 1/n OCT plug-in properties dialog allowing a full access to all the settings.

Reports Tools Ribbons 418.png

Add/remove: Opens the waterfall add/remove result dialog. This dialog allows selecting the list of result to be collected by the waterfall plug-in.

  • Plug-in analyzer tab: A tab is displayed for each plug-in analyzer connected. This displays waterfall source results.
  • Result: List of results provided by the plug-in analyzer selected. This can be the source of the waterfall plug-in analyzer.
  • Channel: List of plug-in analyzer channels connected, available for the result selected.
  • Reference / X: List of plug-in analyzer channels connected. This can be the reference. Available for results that require a reference only (cross-spectrum, for example).
  • Tracked order: List of tracked orders on the channel selected. Available if result selected is an order result.
  • Waterfall: List of 3D Waterfall displays. These waterfalls can be displayed versus any one of the references selected in the "Reference and profiles" list.
  • References & Profiles: List of profiles and references provided by the waterfall plug-in. All scalar profiles selected can be used as references, except monitor scalars, sound level meter scalars, and orders.

Adds a new Waterfall. If the result selected is a scalar, a new profile is added to the "References and profiles" list. Otherwise, a new Waterfall is added to the "Waterfall" list.
If the Waterfall or profile already exists, the clicking on this button has no effect.
If a tracked order is selected, all orders tracked on the channel selected are added.

Removes the selected result from the corresponding list.

Reports Tools Ribbons 422.png

Mode: Select the data collection synchronization:

  • One shot: The waterfall is fulfilled in one shot from its start trigger to the stop one.
  • Continuous: the plug-in is fulfilled as a rolling buffer (FiFo) memorizing the last data only.
Reports Tools Ribbons 423.png

Depth: Defines the maximum number of slices to be collected. In the continuous mode it defines the depth of the rolling buffer. See Chapter 1 ASB - § Monitor and waterfall/ Waterfall for details