Difference between revisions of "Wifi operation"

From OROS Wiki
Jump to navigation Jump to search
(Created page with "Wi-Fi operations '''Hardware:''' OR3X TW '''Software NVGate :''' V10 - V11 - V2021 (incompatibility with NVgate V12) All Teamwork instruments support Wi-Fi operation with N...")
 
Line 2: Line 2:


'''Hardware:''' OR3X TW  
'''Hardware:''' OR3X TW  
'''Software NVGate :''' V10 - V11 - V2021 (incompatibility with NVgate V12)
'''Software NVGate :''' V10 - V11 - V2021 (incompatibility with NVgate V12)


Line 7: Line 8:


1. Connect your PC to the analyzer with the Ethernet cable
1. Connect your PC to the analyzer with the Ethernet cable
2. From the Connection dialog
2. From the Connection dialog
3. Select DHCP server
3. Select DHCP server
4. Enable the Wi-Fi.
4. Enable the Wi-Fi.
5. Enter the password and click on Ok  
5. Enter the password and click on Ok  
6. Plug the Wi-Fi dongle to the USB port on the back plane of the analyzer and restart the analyzer
6. Plug the Wi-Fi dongle to the USB port on the back plane of the analyzer and restart the analyzer
7. Check your PC's Wi-Fi configuration application: The configured analyzer SSID should appear.
7. Check your PC's Wi-Fi configuration application: The configured analyzer SSID should appear.
8. Connect to the Analyzer by Wi-Fi: use the configured pass-phrase.
8. Connect to the Analyzer by Wi-Fi: use the configured pass-phrase.
9. Unplug the network cable BEFORE you launch NVGate  
9. Unplug the network cable BEFORE you launch NVGate  
10. Launch NVGate  
10. Launch NVGate  
11. In the startup screen NVGate connection the analyzer is available for running NVGate through Wi-Fi
11. In the startup screen NVGate connection the analyzer is available for running NVGate through Wi-Fi


default password is : 12345678 (if not working try orosoros)
default password is : 12345678 (if not working try orosoros)

Revision as of 15:28, 4 February 2021

Wi-Fi operations

Hardware: OR3X TW

Software NVGate : V10 - V11 - V2021 (incompatibility with NVgate V12)

All Teamwork instruments support Wi-Fi operation with NVGate V10.10. It is necessary to purchase the Wi-Fi dongle (ORNV-WIF-D) form OROS. Wi-Fi operation is not compatible with other Wi-Fi dongles type. The analyzer must be configured once prior to access it through Wi-Fi.

1. Connect your PC to the analyzer with the Ethernet cable

2. From the Connection dialog

3. Select DHCP server

4. Enable the Wi-Fi.

5. Enter the password and click on Ok

6. Plug the Wi-Fi dongle to the USB port on the back plane of the analyzer and restart the analyzer

7. Check your PC's Wi-Fi configuration application: The configured analyzer SSID should appear.

8. Connect to the Analyzer by Wi-Fi: use the configured pass-phrase.

9. Unplug the network cable BEFORE you launch NVGate

10. Launch NVGate

11. In the startup screen NVGate connection the analyzer is available for running NVGate through Wi-Fi


default password is : 12345678 (if not working try orosoros)