Difference between revisions of "NVGate Ribbons: Acquisition Tab"

From OROS Wiki
Jump to navigation Jump to search
Line 153: Line 153:
Available: Display the maximum possible recording duration depending the current recorder setup. Read only. The max duration is computed from the free space of the selected hard disk, the number of tracks (and their corresponding sampling frequency) and the number of records
Available: Display the maximum possible recording duration depending the current recorder setup. Read only. The max duration is computed from the free space of the selected hard disk, the number of tracks (and their corresponding sampling frequency) and the number of records
Note in time to stop, the max duration may also be limited by the local available memory (PC/Analyzer) for the description blocks. ''See Chapter 1 ASB - § Recorder/Mode for details''
Note in time to stop, the max duration may also be limited by the local available memory (PC/Analyzer) for the description blocks. ''See Chapter 1 ASB - § Recorder/Mode for details''
*
[[Image:V10_release_33.png|framed|none]]
The available disk space (PC or analyzers disk(s)) is now showed in the ribbon on 2 ways from the ribbon Measurement/Record:
First one is the available recording duration according to the current configuration.
Second gives a visual overview of the disk occupation.


=====Outputs group=====
=====Outputs group=====

Revision as of 16:08, 20 April 2020

The "Acquisition" tab

This tab provides all the necessary entries to setup the front-end and the use of any channels in details. It separates the acquisition channels setup from the outputs, triggering, tachometer, and filters ones. A special group is available to manage the recorder setup.

While switching to post-analysis the inputs group is swapped to player track group which control the player operations, track connections and track setup.

Source dispatcher groups

The Inputs, Outputs, Tachometers, Events and Filters groups operates in the same way.

The External sync. Inputs benefit of special button due to their importance in rotating applications.

Inputs group

The inputs group allows dispatching the front-end channels to the plug-in analyzer and setting up the front-end. These settings are available in the on-line mode only.

Connect inputs: This button allows dispatching the front end inputs (all type) to the analysis modules. See Chapter 1 ASB - § Connection wizard for details

Inputs: Allows modifying the dynamic input properties. Available when at least one inputs is active. See Chapter 1 ASB - § Inputs for details

Parametric: Allows modifying the parametric input (DC) properties. Available when at least one DC input is active. See Chapter 1 ASB - § DC inputs for details

External sync: Allows modifying the digital trigger inputs properties. Available when at least one Ext synch input is active. See Chapter 1 ASB - § External synchs for details

CAN: Allows modifying the CAN parametric input properties. Available when at least one CAN input is active. See Chapter 1 ASB - § CAN for details

XPod: Allows modifying the XPod conditioner properties (Gauges, temperature). Available when at least one XPod is connected to the analyzer or in off-line to simulate it See Chapter 1 ASB - § Xpod for details

Tracks group

In the Post-analysis mode, the inputs group switches to the Player management with and is replaced by the Track settings.

The Track group allows dispatching the recorded tracks to the plug-in analyzer and setting up the signal post-process. These settings are available in the post-analysis mode only.

Note: Clicking on the bottom right icon (

) opens the player plug-in properties dialog allowing a full access to all the settings.

Reports Tools Ribbons 328.png

Connect Tracks: This button allows dispatching the signal tracks (all type) to the analysis modules. See Chapter 1 ASB - § Connection wizard for details

Reports Tools Ribbons 329.png

Record num: Define the record number (i.e. a section in the file defined during the acquisition) to be post-processed. See Chapter 1 ASB - § Player/Selected record for details

Start offset: Define the start position of the signal to post-process or playback on the generators: (0 <= Start Offset < Stop Offset). See Chapter 1 ASB - § Player for details

Stop offset: Define the end position of the signal to post-process or playback on the generators: (Start Offset < Stop Offset <= selected record duration). See Chapter 1 ASB - § Player for details Note: The Start, Stop and Record num can be handled directly from the Player file or Player Zoomed signal windows with the start and stop cursors and/or the bottom slide

r

Reports Tools Ribbons 333.png

Playback mode: Allows choosing between a continuous and a step/step post-process. .

  • Continuous: the signal analyses are processed as fast as possible between the start and stop offset.
  • Time step: the signal analyses are processed step by step. The step duration is defined by the Step setting (see hereafter). Each new step is manually (or automated by a macro) triggered through the Next step setting (see hereafter)

See Chapter 1 ASB - § Player for details

Reports Tools Ribbons 334.png

Step: defines the time duration if play back step in the "Time step" analysis mode. The player switches to "pause" every "time step" second.

See Chapter 1 ASB - § Player for details
Reports Tools Ribbons 335.png

Next step: defines the time duration if play back step in the "Time step" analysis mode. The player switches to "pause" every "time step".

See Chapter 1 ASB - § Player for details
Recorder group

This group controls the recorder plug-in setup.

Note: Clicking on the bottom right icon (

) opens the recorder plug-in properties dialog allowing a full access to all the settings.

Reports Tools Ribbons 338.png

Bandwidths: Opens the bandwidths definition dialog. The recorder support up to 4 different bandwidths:

Type Range Mode
FS 1 2.048 S/s to FSF User define
FS 2 2.048 S/s to FSF User define
Ext. Synch Front-end Sampling Freq (FSF) Automatic
Parametric inputs 15 Hz Fixed
See Chapter 1 ASB - § Recorder for details
Reports Tools Ribbons 339.png

Mode: defines the time frame recording type. It can be:

  • Start to stop: Records between a start and a stop event (can be run/stop)
  • Start to time: Records for a fixed duration after a start event (can be run)
  • Time to Stop: Records a fixed duration until a stop event (can be the manual stop)
See Chapter 1 ASB - § Recorder/Mode for details
Reports Tools Ribbons 340.png

Records: defines the maximum number of records in the signal file. See Chapter 1 ASB - § Recorder/Mode for details

Duration: Set the maximum duration of a record. This value cannot exceed the available duration (see here after) for Start to time and Time to stop mode only See Chapter 1 ASB - § Recorder/Mode for details

Format: allows selecting the recorded data length:

  • Normal: 32 bit floating point coding for each sample. Keep the full dynamic and precision.
  • Compressed: 16 bit floating point coding for each sample. Reduce the size of the file(s) but reduce also the dynamic to 96 dB. Not recommended for low amplitude signal without auto-ranging. Adapted for exportation of signal in 16 bit format (Wav, txt)

See Chapter 1 ASB - § Recorder/Mode for details

Reports Tools Ribbons 343.png

Available: Display the maximum possible recording duration depending the current recorder setup. Read only. The max duration is computed from the free space of the selected hard disk, the number of tracks (and their corresponding sampling frequency) and the number of records Note in time to stop, the max duration may also be limited by the local available memory (PC/Analyzer) for the description blocks. See Chapter 1 ASB - § Recorder/Mode for details


The available disk space (PC or analyzers disk(s)) is now showed in the ribbon on 2 ways from the ribbon Measurement/Record:

First one is the available recording duration according to the current configuration.

Second gives a visual overview of the disk occupation.

Outputs group

Manages the generated signal on the front-end outputs (generators)

Reports Tools Ribbons 344.png

The left button (signals) allows selecting the signal type from the list and connecting it to the available outputs (1 to 6) .The others buttons open the signal settings and manage the generators activity.

Reports Tools Ribbons 345.png

Signals: Shows the list of available signals and let the users connect it to the outputs. See Chapter 1 ASB - § Connection wizard for details

The available signals are the following:

Reports Tools Ribbons 346.png

Sine: Opens the pure sine properties dialog for adjustment. Pure sine = 1 fixed frequency, 1 amplitude. Adapted for single mode excitation. Possibility to generate a fixed voltage with F=0.

Reports Tools Ribbons 347.png

Multi-sine: Opens the Multi-sine properties dialog for adjustment. Multi sine = n fixed frequencies, 1 amplitude Adapted for multi-mode excitation and FFT analysis

Reports Tools Ribbons 348.png

Swept-sine: Opens the Swept-sine properties dialog for adjustment. Swept-sine = 1 sweeping frequency, 1 to 6 proportional amplitudes Adapted for swept sine and MiMo excitation

Reports Tools Ribbons 349.png

Chirp: Opens the Chirp properties dialog for adjustment. Chirp = Continuous short term variable frequency (1 analysis block), 1 amplitude. Adapted for damping measurement and FFT analysis

Reports Tools Ribbons 350.png

Random Noise: Opens the Random noise properties dialog for adjustment. True Random noise = Infinity of frequencies, 1 amplitude, generated randomly. Adapted for non linear responses measurement. In addition to these predefined signals, the content of a file and the replication of an input is also available to be connected on the generators. See Chapter 1 ASB - § Resources/Output signals for details

Reports Tools Ribbons 351.png

Outputs settings: Manages the generated signal settling, (Mute, transition time) See Chapter 1 ASB - § Front-end/Output settings for details

Tachometers group

Manage the tachometer resources. This group allows selecting the tach type (source, setup) and associates it with the plug-in analyzers and waterfall

The left button (Select) allows dispatching the different tachometer type to the plug-in analyzers. The others items open the corresponding event detection setup.

Select: Shows the list of available tachometer sources and allows plugging it to the plug-in and waterfall. See Chapter 1 ASB - § Connection wizard for details

Reports Tools Ribbons 354.png

Inputs: Opens the properties dialog for the tachometers based on a dynamic input

Reports Tools Ribbons 355.png

Ext. Synch: Opens the properties dialog for the tachometers based on a high speed oversampled Ext Synch input.

Reports Tools Ribbons 356.png

Fractional: Opens the properties dialog for the tachometers that derives from another one. Fractional tach. computes RPM speed for a non accessible shaft by using gear ratio setting. Adapted for gear boxes and transmissions. Note: the fractional tach. cannot be settled from the Vision interface, use the ASB for it.

Reports Tools Ribbons 357.png

Combined: Opens the properties dialog for the tachometers computed from other ones. Adapted for CVT. Note: the combined tach cannot be settled from the Vision interface, use the ASB for it.

Reports Tools Ribbons 358.png

Profile: Set up the tachometer profiles duration. These graphs are available for the Tachometer module in the Add/Remove graph dialog. The profile displays continuously the tachometer speeds with a memory depth defined by the profile setting.

Reports Tools Ribbons 359.png

Resolution: Set up the tachometer profiles update rate. See Chapter 1 ASB - § Resources/Tachometer for details

Events group
Reports Tools Ribbons 17.png

Manage the event and corresponding triggers. This group allows selecting the event type (source, setup) and associates it with the plug-in analyzers triggers

The left button (Associate to) allows dispatching the different possible event (internal or external) to the plug-in analyzers.

The 2 others items open the corresponding event detection setup.

Associate to: This button opens the corresponding Event properties. The opened dialog shows both the source signal setup and the detection settings (may be spitted in different tabs) See Chapter 1 ASB - § Connection wizard for details

  • Events properties: This button opens the corresponding Events properties dialog for setup.

The available event types are:

Reports Tools Ribbons 361.png

Edge: Detect when the signal from an input or a recorded track cross a threshold.

Reports Tools Ribbons 362.png

Level: Detect when a signal (from parametric input or statistical extraction form the TDA or monitor) is below or above a level

Reports Tools Ribbons 363.png

Delta-Level: Same as the Level but generates an event at each step

Reports Tools Ribbons 364.png

RPM: Detect when a tachometer is below or above a specified angular speed

Reports Tools Ribbons 365.png

Delta-RPM: Same as RPM but generates an event at each step

Reports Tools Ribbons 366.png

Periodic: Generates events periodically at a fixed rate

Reports Tools Ribbons 367.png

Combined: Generates an event which is the result of a combination of 2 events. Possible combinations are: OR, AND and AFTER.

Reports Tools Ribbons 368.png

User: Generates an event when the operator presses the corresponding event button in the software/remote controller interface.

Reports Tools Ribbons 369.png

External synch: Properties of the Ext synch trigger inputs. See Chapter 1 ASB - § Resources/Events for details

Filters group

Manage the filters definition and position. These filters operate on the time domain series. They can be applied in various locations onto the signal process (Inputs, Plug-in channels, Outputs, Player tracks)

The left button (Apply to) allows dispatching the filters to the different possible location in the signal process..

The second button allows selecting a filter type and opens the corresponding setup.


Reports Tools Ribbons 370.png

Apply to: This button opens the filter dispatching dialog:

Filters are available only for tachometers, outputs, aux. Outputs, edge detection, FFT, and Sync. Order and only after being activated using the ’Inputs’, ’Outputs signals’, ’Tachometers’ or ’Event’ connections previously described.

All filters are available, and by adding any filters to any channel, the properties of filters and channel selected are displayed and filters are applied in the current measurement:

See Chapter 1 ASB - § Connection wizard for details

  • Filter properties: This button opens the corresponding filter properties dialog for setup.

The available filter types are:

Reports Tools Ribbons 373.png

High/Low pass: Filter the high or low (depends on setup) frequencies of a signal

Reports Tools Ribbons 374.png

Pass/stop band: Filter or reject (depends on setup) the specified bandwidth.

Reports Tools Ribbons 375.png

Integrator: Integrates or double integrates the signal. The high pass for avoiding divergences is included. Useful for acceleration to velocity/displacement conversion.

Reports Tools Ribbons 376.png

Differentiator: Derivate the signal. Useful for torsional velocity conversion in angular acceleration See Chapter 1 ASB - § Resources/Filter for details

Warning: Due to bandwidth compatibility, the filters use is exclusive of the plug-in. I.e: Once a filter is associated to a location (ex: a plug-in channel) it can be used on the same plug-in channels only. The front-end is considered as a plug-in.