Difference between revisions of "NVGate X-Pod: Temperature probe conditioner"

From OROS Wiki
Jump to navigation Jump to search
Line 24: Line 24:
[[File:Xpod_temp.png|Xpod_temp.png]]
[[File:Xpod_temp.png|Xpod_temp.png]]


This Xpod handles thermocouples and RTD transducers.
 
===Temperature===
This Xpod handles [https://en.wikipedia.org/wiki/Thermocouple thermocouples] and [https://en.wikipedia.org/wiki/Resistance_thermometer RTD transducers].


* '''PT100, PT1000 and J, K, T, N, E'''
* '''PT100, PT1000 and J, K, T, N, E'''

Revision as of 12:46, 8 May 2020

X-Pod: Temperature probe conditioner

Expander modules

The XPod is a device that can be fixed on OR3X TW analyzer. Each XPod is associated to a block of 8 inputs.


Xpod-plug.png

In the connected mode, when and XPod is present each corresponding input can be routed to:

· The BNC input (Standard)

· The XPod connector


General Xpod settings

  • Type: For acquisition setup preparation, the XPod presence can be simulated in the office mode:

One XPod type selector is available for each XPod position.

Xpod bridge.png
  • Bridge max current: Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.
  • Bridge voltage: Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.

Xpod_temp.png


Temperature

This Xpod handles thermocouples and RTD transducers.

  • PT100, PT1000 and J, K, T, N, E
  • Integrated linearization
  • Automatic cold junction compensation
  • Standard fl at pin connectors
Front end 08.png