Difference between revisions of "Notices and General safety instructions"
(2 intermediate revisions by one other user not shown) | |||
Line 194: | Line 194: | ||
|bgcolor = "#D9D9D9"|<font size = "1">'''WEEE'''</font> | |bgcolor = "#D9D9D9"|<font size = "1">'''WEEE'''</font> | ||
|bgcolor = "#D9D9D9"|<font size = "1">'''2012/19/EU'''</font> | |bgcolor = "#D9D9D9"|<font size = "1">'''2012/19/EU'''</font> | ||
|} | |||
---- | |||
====Environmental limits==== | |||
'''Never expose''' the instrument and its accessories '''over the environmental limits''' described hereafter. | |||
{|border="2" cellspacing="0" cellpadding="4" width="100%" | |||
|<font size = "1">'''Humidity'''</font> | |||
|bgcolor = "#D9D9D9"colspan = "2"|<font size = "1">Max''' 80 % RH''' at 40°C non-condensing</font> | |||
|- | |- | ||
|<font size = "1">''' | |rowspan = "1"|<font size = "1">'''Altitude'''</font> | ||
| | |<font size = "1">Operating, non tested above </font> | ||
|<font size = "1"> ≤ 2000 m (6562 feet) </font> | |||
|- | |- | ||
Line 217: | Line 227: | ||
|- | |- | ||
|rowspan = " | |rowspan = "6"|<font size = "1">'''Shocks/Bumps'''</font> | ||
|bgcolor = "#D9D9D9" colspan = "2"|<font size = "1">Complies with '''IEC 68-2-27'''</font> | |bgcolor = "#D9D9D9" colspan = "2"|<font size = "1">Complies with '''IEC 68-2-27 (all) and MIL-STD-810F (OR35TW)'''</font> | ||
|- | |- | ||
|<font size = "1">Operating</font> | |<font size = "1">Operating (OR36 & OR38)</font> | ||
|<font size = "1">''' | |<font size = "1">'''10 g '''(11 ms, ½ sine) and '''70 g''' (3 ms, ½ sine)</font> | ||
|- | |- | ||
|bgcolor = "#D9D9D9"|<font size = "1">Storage</font> | |bgcolor = "#D9D9D9"|<font size = "1">Storage (OR36 and OR38)</font> | ||
|bgcolor = "#D9D9D9"|<font size = "1">''' | |bgcolor = "#D9D9D9"|<font size = "1">'''40 g ('''11 ms, ½ sine)</font> | ||
|- | |||
|<font size = "1">Operating (OR35)</font> | |||
|<font size = "1">'''40 g '''(6 ms, ½ sine) and '''60 g''' (3 ms, ½ sine)</font><br> | |||
<font size = "1">MIL-STD-810F method 514.5</font> | |||
|- | |||
|bgcolor = "#D9D9D9"|<font size = "1">Storage (OR35)</font> | |||
|bgcolor = "#D9D9D9"|<font size = "1">'''60 g ('''11 ms, ½ sine)</font><br> | |||
<font size = "1">MIL-STD-810F method 516.5</font> | |||
|- | |- | ||
|<font size = "1">Absolute maximum rating<sup>ii</sup></font> | |<font size = "1">Absolute maximum rating<sup>ii</sup></font> | ||
|<font size = "1">''' | |<font size = "1">'''100 g''' (3 ms, ½ sine)</font> | ||
|- | |- | ||
|rowspan = "4"|<font size = "1">'''Vibrations'''</font> | |rowspan = "4"|<font size = "1">'''Vibrations'''</font> | ||
|bgcolor = "#D9D9D9" colspan = "2"|<font size = "1">Complies with '''IEC | |bgcolor = "#D9D9D9" colspan = "2"|<font size = "1">Complies with '''IEC 60068-2-6'''</font> | ||
|- | |- | ||
|<font size = "1">Operating</font> | |<font size = "1">Operating (OR36 & OR38)</font> | ||
|<font size = "1">''' | |<font size = "1">'''2 g, 5-500 Hz, 5mm'''</font> | ||
|- | |- | ||
|bgcolor = "#D9D9D9"|<font size = "1"> | |bgcolor = "#D9D9D9"|<font size = "1">Operating (OR35)</font> | ||
|bgcolor = "#D9D9D9"|<font size = "1">''' | |bgcolor = "#D9D9D9"|<font size = "1">'''2.5g, 5-500 Hz, 5mm'''</font> | ||
|- | |- |
Latest revision as of 15:05, 7 March 2022
This is the notice and general safety instruction for using OROS OR3X analyzer
Notices
The information contained in this document is subject to change without notice.
Responsibilities
OROS is not liable for any errors contained in this document.
OROS makes no warranties of any kind with regards to this document, whether express or implied. OROS disclaims the implied warranties of merchantability and fitness for a particular purpose.
OROS shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory, in connection with the furnishing of this document or the use of the information in this document.
Warranty Information
A copy of the warranty terms applicable to your OROS product and replacement parts can be obtained from your local distributor. A copy of the extended warranty (service contracts) terms applicable to your OROS product can be obtained from your local distributor.
Calibration/Verification
OROS instruments needs to be calibrated regularly. OROS recommend to verify and calibrate your instrument yearly.
Note: The calibration does not guarantee the measurements accuracy for the coming year. It guarantees that your previous measurements do not show erroneous results.
Support
Support (hotline, trainings, coaching, renting, customization, update) are managed locally by country with the support of OROS customer care team. For any support please contact your local distributor first.
Service/repairs
OROS offers local service and repairs through its maintenance centers available within a worldwide coverage (China, Europe, India, Japan, South Korea, USA). Local technicians are certified on a regular basis by OROS specialists, enabling them to repair, calibrate and upgrade all OROS systems reducing the maintenance downtime.
Find your local distributor details.
General Safety Instructions
Read the manual before use. It provides important safety information.
Your safety is our primary concern. Please carefully read the following information to be safe with your OROS instrument.
Safety symbols in the manual
Warning
Calls attention to a procedure, practice, or condition that could cause body injury or death.
Caution
Calls attention to a procedure, practice, or condition that could possibly cause damage to equipment or permanent loss of data.
Warnings
The following general safety precautions must be observed during all operations with OROS instruments.
Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the product. OROS assumes no liability for the customer’s failure to comply with these requirements.
Notice
Some accessories shown in this document are available as option and will not be shipped as standard parts.
Environmental Considerations
The following guidelines must be observed to avoid environmental pollution while using and recycling the OROS products.
Product End-of-Life Handling
Observe the following guidelines when recycling an OROS instrument:
System and Components Recycling: Production of these components required the extraction and use of natural resources. The substances contained in the system could be harmful to your health and to the environment if the system is improperly handled at its end of life.
Please recycle this product in an appropriate way to avoid unnecessary pollution of the environment and to keep natural resources.
This symbol indicates that this system complies with the European Union’s requirements according to Directive 2012/19/EU on waste electrical and electronic equipment (WEEE) According to article 7 of directive 2012/19/EU, costs for collection, treatment, recovery and environmentally sound disposal of waste electrical and electronic equipment (WEEE) shall be provided by the users. French users, please consult our local policy @ www.oros.fr.
Restriction of Hazardous Substances
OROS products comply with the 2011/65/EU et 2015/863 RoHs directives. Our products are lead free. Pollution degree 2.
General safety and hazard warnings for OROS Instruments
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified. Only qualified personnel should perform service procedures. To avoid fire and personal injury: Use proper power cord. Use only the power cord specified in this product and certified for the country of use.
Connect OROS Teamwork instruments to the ground. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding connector must be connected to earth ground.
To avoid electric hazards during battery operation, connect OROS Teamwork instrument to the ground .
Respect all terminal ratings. To avoid fire or shock hazard, observe all ratings and marking on the product. Do not apply voltages over +/-60 V to any terminals and covers of the instruments.
Do not operate without covers. Do not operate this product with covers or panels removed.
Avoid exposed circuitry. Do not touch exposed connections and components when AC outlet is connected.
Do not operate with suspected failures. If you suspect there is any damage to this product, get it inspected by qualified service personnel.
Only use original and approved devices, components and accessories. Apply them in accordance with their intended use.
Maintenance of the instruments and accessories is operated in OROS certified maintenance center only.
Adverse environmental conditions : Take specific protections when operating the instrument exposed to the following conditions
- Do not operate in wet/damp conditions for outdoor operations.
indoor, dry location use only.
- Do not operate in explosive atmosphere (Dust, flammable gases/vapors, fumes or dissolver).
- Keep product surfaces clean and dry. Disconnect and power off the instrument before cleaning external surfaces with a damp cloth. Power on and reconnect when surfaces are dry.
- Provide proper ventilation. See the "Fan" chapter.
- Do not operate the instrument out of the temperature, shock and vibration absolute maximum ratings (see compliance with standards).
- Do not operate in environments where pollutants may be present.
Electrical and radio frequencies
Electrical and radio environments
OROS instruments are designed to operate in office, laboratory, on-board and industrial environments regarding electrical and radio disturbances.
Radio frequencies sensibility
Input measured with 50 Ω terminator | |
Radiated RF: 80-1000 MHz, 80% AM 1 kHz, 10 V/m | < 20 µV |
Conducted RF: 0.15-80 MHz, 80% AM 1 kHz, 10 V | < 100 µV |
Magnetic field: 30 A/m, 50 Hz | < 2 µV |
Compliance with standards
CE\CB\FCC | Indicates compliance with EMC Directive 89/336/EEC, 2014/30/EU and Low Voltage Directive 73/23/EEC,2014/35/EU | |
Safety | IEC 61010-1 | Safety requirements for electrical equipment for measurement, control and laboratory use. |
IEC 61010-2-30 | Particular requirements for testing and measuring circuits. | |
Over-voltage Cat. | II (Local level mains, appliance, and portable equipment) | |
EMC Emission | IEC 61000-6-3 | Emission standard for residential, commercial and light-industrial environments. |
IEC 61000-6-4 | Emission standard for industrial environments. | |
IEC 61326-1 | Electrical equipment for measurement control and laboratory use EMC requirements. | |
CISPR 11 | Radio disturbance characteristics of industrial and scientific equipment.: Class B limit. | |
FCC Rules | Complies with the limits for a Class B digital device. | |
EMC Immunity | IEC 61000-6-1 | Immunity standard for residential, commercial and light-industrial environments. |
IEC 61000-6-2 | Immunity standard for industrial environments | |
IEC 61326-1 | Electrical equipment for measurement control and laboratory use EMC requirements. | |
EN 50082-2 | Generic immunity standard: Industrial environment. | |
EMF | Recommendations 199/519/CE EN 62311 | Evaluation of person exposure to electromagnetic fields |
Materials | ROHS | 2011/65/EU and 2015/863 |
WEEE | 2012/19/EU |
Environmental limits
Never expose the instrument and its accessories over the environmental limits described hereafter.
Humidity | Max 80 % RH at 40°C non-condensing | |
Altitude | Operating, non tested above | ≤ 2000 m (6562 feet) |
Temperature | OR34, OR35, OR36 Operating | -20°C to 50°C (32°F to 122°F) |
OR38 Operating | -20°C4 to 45°C (32°F to 113°F) | |
Storage | -20°C to 65°C (-4°F to 149°F) | |
Absolute maximum ratingii | -35°C to 70°C (-31°F to 158°F) | |
Shocks/Bumps | Complies with IEC 68-2-27 (all) and MIL-STD-810F (OR35TW) | |
Operating (OR36 & OR38) | 10 g (11 ms, ½ sine) and 70 g (3 ms, ½ sine) | |
Storage (OR36 and OR38) | 40 g (11 ms, ½ sine) | |
Operating (OR35) | 40 g (6 ms, ½ sine) and 60 g (3 ms, ½ sine) MIL-STD-810F method 514.5
| |
Storage (OR35) | 60 g (11 ms, ½ sine) MIL-STD-810F method 516.5 | |
Absolute maximum ratingii | 100 g (3 ms, ½ sine) | |
Vibrations | Complies with IEC 60068-2-6 | |
Operating (OR36 & OR38) | 2 g, 5-500 Hz, 5mm | |
Operating (OR35) | 2.5g, 5-500 Hz, 5mm | |
Absolute maximum ratingii | 30 m/s², 5-500 Hz, 5mm | |
Enclosure | OR35 | IP 40 |
OR34, OR36, OR38 | IP 42 |