Difference between revisions of "ORBIGate Manual User's guide"

From OROS Wiki
Jump to navigation Jump to search
 
(68 intermediate revisions by 3 users not shown)
Line 2: Line 2:
This chapter is an overview of the main features and set-up steps of the ORBIGate software. More details are provided in later chapters of the operating manual. The Set-Up wizard lets you quickly go through the steps to get started.
This chapter is an overview of the main features and set-up steps of the ORBIGate software. More details are provided in later chapters of the operating manual. The Set-Up wizard lets you quickly go through the steps to get started.


[[Image:ORBIGate_09.png|framed|none]]
[[Image:ORBIGate_09.png|200px]]


''The set up toolbar''
''The set up toolbar''
Line 10: Line 10:
- Choosing "Load Project" allows the user to load measurement settings.
- Choosing "Load Project" allows the user to load measurement settings.


- Choosing "Load Measurement", the acquired measurement data can be loaded.
- Choosing "Load Measurement", will load the acquired measurement data.


- Choosing "Load Signal" will allow the user to display a measurement based on a recorded signal
- Choosing "Load Signal" will allow the user to display a measurement based on a recorded signal.


To get started, choose "Create New project" and press GO.
To get started, choose "Create New project" and press GO.


[[Image:ORBIGate_10.png|framed|none]]
[[Image:ORBIGate_10.png|400px]]


''The Start <nowiki>’</nowiki>n Go window''
''The Start <nowiki>’</nowiki>n Go window''
Line 22: Line 22:
The next step is to configure the machine train.
The next step is to configure the machine train.


[[Image:ORBIGate_11.jpg|framed|none]]
[[Image:ORBIGate_11.jpg|400px]]


''The machine train configuration''
''The machine train configuration''
Line 28: Line 28:
The test point<nowiki>’</nowiki>s configuration screen is where the probe orientations are configured. It is also where the labels can be modified.
The test point<nowiki>’</nowiki>s configuration screen is where the probe orientations are configured. It is also where the labels can be modified.


[[Image:ORBIGate_12.jpg|framed|none]]
[[Image:ORBIGate_12.jpg|400px]]


''The test points configuration''
''The test points configuration''
Line 34: Line 34:
The next step is to configure the analyzer input properties (coupling, sensitivity etc…).
The next step is to configure the analyzer input properties (coupling, sensitivity etc…).


[[Image:ORBIGate_13.jpg|framed|none]]
[[Image:ORBIGate_13.jpg|400px]]


''The inputs configuration''
''The inputs configuration''
Line 40: Line 40:
The following step is to configure the speed properties.
The following step is to configure the speed properties.


[[Image:ORBIGate_14.jpg|framed|none]]
[[Image:OBG_7_tacho.png|400px]]


''The rotating speed configuration''
''The rotating speed configuration''
Line 46: Line 46:
The Analysis dialog box allows for the configuration of everything related to the analysis. There is a choice of 3 types of analysis: Sync analysis, ASync analysis, and signal recording.
The Analysis dialog box allows for the configuration of everything related to the analysis. There is a choice of 3 types of analysis: Sync analysis, ASync analysis, and signal recording.


[[Image:ORBIGate_15.jpg|framed|none]]
[[Image:OBG_7_analysis.png|400px]]


''The analysis window configuration''
''The analysis window configuration''


The next step, is the set-up of the alarms: alarm conditions and alarm actions. For the getting started stage, one doesn<nowiki>’</nowiki>t need to go through that step and one can just press next.
The next step, is the set-up of the alarms: alarm conditions and alarm actions. For the getting started stage, one doesn<nowiki>’</nowiki>t need to go through that step and can just press next.


[[Image:ORBIGate_16.png|framed|none]]
[[Image:ORBIGate_16.png|400px]]


The final step is to display the graphical windows. After choosing the graphical windows, click the "Save project" button to save the project.
The final step is to display the graphical windows. After choosing the graphical windows, click the "Save project" button to save the project.


[[Image:ORBIGate_17.jpg|framed|none]]
[[Image:ORBIGate_17.jpg|400px]]


''The Add/remove window configuration''
''The Add/remove window configuration''


The test is ready to be run.  Clicking on the monitoring icon   
The test is ready to be run.  Clicking on the monitoring icon   
[[Image:ORBIGate_18.gif|framed|none]]
[[Image:ORBIGate_18.gif]]
, as shown on the window below, will display the results.
as shown on the window below, will display the results.


[[Image:ORBIGate_19.jpg|framed|none]]
[[Image:ORBIGate_19.jpg|400px]]


''The acquisition working environment''
''The acquisition working environment''
Line 71: Line 71:
===Workspace===
===Workspace===


[[Image:ORBIGate_20.png|framed|none]]
[[Image:ORBIGate_20.png|500px]]


''The working environment''
''The working environment''
Line 98: Line 98:


===Start<nowiki>’</nowiki>n Go===
===Start<nowiki>’</nowiki>n Go===
The Start<nowiki>’</nowiki>n Go function helps the user to immediately start the test or set-up. If a set-up has already been defined, it can be loaded at this stage. Otherwise a new project can be created.
The Start<nowiki>’</nowiki>n Go function helps the user to immediately start the test or set-up. If a set-up has already been defined, it can be loaded at this stage. Otherwise, a new project can be created.


If the objective is to analyze a measurement, it can be loaded through the Start<nowiki>’</nowiki>n Go interface as shown below.
If the objective is to analyze a measurement, it can be loaded through the Start<nowiki>’</nowiki>n Go interface as shown below.


[[Image:ORBIGate_21.png|framed|none]]
[[Image:ORBIGate_21.png|400px]]


''The Start<nowiki>’</nowiki>n Go window''
''The Start<nowiki>’</nowiki>n Go window''


In order to avoid overwriting projects name, the "GO" button is protected: It will automatically be greyed if an existing name is entered.
In order to avoid overwriting project names, the "GO" button is protected: It will automatically be grayed if an existing name is entered.


If the purpose is to go through a post analysis, a signal should be loaded. The signal to be loaded should be located in the current NVGate project database. The signal should contain at least one tachometer channel recorded on the external trigger. This signal can be recorded from NVGate or ORBIGate.
If the purpose is to go through a post analysis, a signal should be loaded. The signal to be loaded should be located in the current NVGate project database. The signal should contain at least one tachometer channel recorded on the external trigger. This signal can be recorded from NVGate or ORBIGate.
Line 114: Line 114:
ORBIGate can be operated in three modes: the Acquisition mode, the Post Analysis and the Navigation mode. The user can switch from one to the other by using the operating mode icon toolbar as displayed below. The objective is to investigate measurement data: this is achieved by '''loading a''' '''Measurement File in the Navigation Mode'''. This measurement can be saved in the acquisition mode or in the post analysis mode.
ORBIGate can be operated in three modes: the Acquisition mode, the Post Analysis and the Navigation mode. The user can switch from one to the other by using the operating mode icon toolbar as displayed below. The objective is to investigate measurement data: this is achieved by '''loading a''' '''Measurement File in the Navigation Mode'''. This measurement can be saved in the acquisition mode or in the post analysis mode.


[[Image:ORBIGate_22.png|framed|none]]
[[Image:ORBIGate_22.png|1000px|none]]


''The 3 Modes''
''The 3 Modes''
Line 132: Line 132:
====Project management and operating modes====
====Project management and operating modes====


Projects are used to save a set-up. It concerns both the acquisition set-up and the display set-up.
Projects are used to save a set-up. It controls both the acquisition set-up and the display set-up.


Projects are saved in the directory defined at the installation and modifiable with the "environment" software. The default path is "C:\OROS\ORBIGate data\Projects". One directory is created for each saved project.
Projects are saved in the directory defined at the installation and modifiable with the "environment" software. The default path is "C:\OROS\ORBIGate data\Projects". One directory is created for each saved project.
Line 143: Line 143:


====Saving a project====
====Saving a project====
It is possible to save a project in different ways: either at the end of the setup wizard or through the "file menu". When switching from one mode another, it is recommended to save the project in each mode (for example, "test1 ACQ", "test1 PA", "test1 NAV") as the configuration is different. ORBIGate tracks any risks of overwriting projects when changing mode. A project was saved in one of the three modes (Acquisition, Post Analysis, and Navigation). Then, if after changing mode one attempt to overwrite the project, a message is issued as below. It proposes either to give it a new name (recommended) or gives the flexibility to overwrite the existing one (not recommended)
It is possible to save a project in different ways: either at the end of the setup wizard or through the "file menu". When switching from one mode another, it is recommended to save the project in each mode (for example, "test1 ACQ", "test1 PA", "test1 NAV") as the configuration is different. ORBIGate tracks any risks of overwriting projects when changing modes. A project was saved in one of the three modes (Acquisition, Post Analysis, and Navigation). Then, if after changing mode one attempt to overwrite the project, the below message is displayed. It proposes either provide a new name (recommended) or gives the flexibility to overwrite the existing one (not recommended).


[[Image:ORBIGate_25.png|framed|none]]
[[Image:ORBIGate_25.png|framed|none]]
Line 152: Line 152:
A project can be loaded through the menu and the command "Load…" (As shown in the picture below) or through the Start<nowiki>’</nowiki>n Go window.
A project can be loaded through the menu and the command "Load…" (As shown in the picture below) or through the Start<nowiki>’</nowiki>n Go window.


[[Image:ORBIGate_26.jpg|framed|none]]
[[Image:ORBIGate_26.jpg|400px]]


''Loading a project''
''Loading a project''
Line 171: Line 171:


====Loading a measurement====
====Loading a measurement====
A measurement is loaded through the command "File/Load Measurement". The following window is displayed. The measurements are browsable through the measurements database interface.
A measurement is loaded through the command "File/Load Measurement". The following window is displayed. The measurements can be browsed through the measurements database interface.


Files can be '''sorted''' based on their:
Files can be '''sorted''' based on their:
Line 182: Line 182:
One can also search for keywords coming from any of the columns. The interface looks as below.
One can also search for keywords coming from any of the columns. The interface looks as below.


[[Image:ORBIGate_28.png|framed|none]]
[[Image:ORBIGate_28.png|800px|none]]


''Measurement database interface''
''Measurement database interface''
Line 193: Line 193:
Once the sensors are selected, the set-up of the input is automatically done according to the predefined properties of the sensor.
Once the sensors are selected, the set-up of the input is automatically done according to the predefined properties of the sensor.


[[Image:ORBIGate_29.jpg|framed|none]]
[[Image:ORBIGate_29.jpg|400px]]


''Setting Up the inputs''
''Setting Up the inputs''
Line 205: Line 205:
The preferences allow storing the paths used for operating ORBIGate. The NVGate.exe position can be selected at this level.
The preferences allow storing the paths used for operating ORBIGate. The NVGate.exe position can be selected at this level.


[[Image:ORBIGate_30.gif|framed|none]]
[[Image:OBG_preferences.png]]


''Path preferences''
''Path preferences''


====Analysis preferences====
====Analysis preferences====
In the preferences, it is also possible to select the maximum sampling frequency allowed for the analyzer. Among others, it enables the user to limit the DSP load.
In the preferences, it is also possible to:<br>


It is possible to setup Post analysis signal playback speed in the user preferences: this allows controlling the post processing speed of the results. Increasing playback speed may introduce downsampling of the results collection in the measurement. It will cause results to be collected less often.
- '''Sampling frequency''': Select the maximum sampling frequency allowed for the analyzer. Among others, it enables the user to limit the DSP load. Placing the mouse on the “i” icon will give local contextual information.<br>


When the objective is to control the visualization speed of the results it is recommended to use the navigation mode which is designed for that purpose.
- '''Gap / Zero & low speed GAP''': Enable the acquisition in the measurement of the Gap at Zero and low speed from ASync analysis. For that:<br>


[[Image:ORBIGate_31.jpg|framed|none]]
* Corresponding ASync channels should be connected
* ASync analysis resolution is recommended to be set to high resolution to fit the low speed conditions (typically at least 0.625 Hz as set by default). <br>


''Analysis preferences''
- '''Post Analysis playback speed''': It is possible to setup Post analysis signal playback speed in the user preferences: this allows controlling the post processing speed of the results. Increasing playback speed may introduce down sampling of the results collection in the measurement. It will cause results to be collected less often. When the objective is to control the visualization speed of the results it is recommended to use the navigation mode which is designed for that purpose. <br>
 
[[Image:OBG_Pref_analysis.png|400px]]
 
''Analysis preferences''<br>


====Units preferences====
====Units preferences====
Users from different countries use different units: it is possible to select the desired "unit profile" from the preferences.
- '''Units''': Users from different countries use different units: it is possible to select the unit used for each physical quantity in the window below.<br>
 


[[Image:ORBIGate_32.jpg|framed|none]]
[[Image:OBG_Pref_unit.png|400px]]


''Units preferences''


* Trends can be displayed with the absolute time as an X-Scaling. This is especially useful when speed is not changing (steady state measurements) and when one wants to relate the vibration evolutions with other process phenomena (load, temperature or pressure changes). In order to get the absolute dating on the X axis, the corresponding preference should be chosen in the "preferences", as shown in the window below.
- '''Spectrum unit''' can be 0-Peak or Peak-Peak. It can be modified locally in the window as well.<br>
 
- '''Time format''': Trends can be displayed with the absolute time or relative time as an X-Scaling. Time scale is especially useful when speed is not changing (steady state measurements) and when one wants to relate the vibration evolutions with other process phenomena (load, temperature or pressure changes). In order to get the absolute dating on the X axis, the corresponding preference should be chosen in the “preferences”, as shown in the window below.
 


{|border="2" cellspacing="0" cellpadding="4" width="100%" align="center"
{|border="2" cellspacing="0" cellpadding="4" width="100%" align="center"
Line 238: Line 246:


|}<br clear="all">
|}<br clear="all">
- '''Phase''': The phase can selected to be displayed as 0°/360° or -180°/+180°.<br>


====Display preferences====
====Display preferences====


[[Image:ORBIGate_35.png|framed|none]]
[[Image:ORBIGate_35.png|700px]]


''Display and print preferences window''
''Display and print preferences window''
Line 249: Line 259:
* "Auto scale" corresponds to an "Auto scale up": the scale grows continuously adapting to the maximum spectrum value.
* "Auto scale" corresponds to an "Auto scale up": the scale grows continuously adapting to the maximum spectrum value.
* The "fixed" allows to setup the maximum value for a graph
* The "fixed" allows to setup the maximum value for a graph
These values are defined by physical units.
These values are defined by physical units (chosen in the “units” tab).


[[Image:ORBIGate_36.gif|framed|none]]
[[Image:OBG_pref_scale.png|400px]]


''"Auto scale" or "Fixed" Scaling''
''"Auto scale" or "Fixed" Scaling''


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|To activate the "fixed" settings one should press the toolbar icon to reset the scales:
|To apply the selection, one should press the corresponding toolbar icon to apply the scaling (second icon displayed). This is done for the current layout. The scaling preferences window can be accessed directly by using the “scaling preferences icon”.
|align = "center"|
|align = "center"|
[[Image:ORBIGate_37.png|framed|none]]
[[Image:OBG_scaling.png|framed|none]]
<br>''Reset scales button''
<br>''“Scaling preferences” and “apply scaling” toolbar icons''


|}
|}
Line 268: Line 278:
|<br>This preference aims at providing a better flexibility for the configuration of the probes positions in the graphics (Orbit, Shaft Centerline and Polar). This is meant to adapt to the different situation that may occur when connecting the systems to the buffered outputs on the online monitoring systems.<br><br>'''''Note:''' New XY preferences are taken into account for a new configuration.''<br>''For example: if a project was saved with older XY preferences, these older preferences will be taken into account.''
|<br>This preference aims at providing a better flexibility for the configuration of the probes positions in the graphics (Orbit, Shaft Centerline and Polar). This is meant to adapt to the different situation that may occur when connecting the systems to the buffered outputs on the online monitoring systems.<br><br>'''''Note:''' New XY preferences are taken into account for a new configuration.''<br>''For example: if a project was saved with older XY preferences, these older preferences will be taken into account.''
|align = "center"|
|align = "center"|
[[Image:ORBIGate_38.png|framed|none]]
[[Image:ORBIGate_38.png|400px]]
<br>''Probe settings preference''
<br>''Probe settings preference''


Line 274: Line 284:


=====Probes connections to the analyzer inputs=====
=====Probes connections to the analyzer inputs=====
Depending on the buffered outputs connections provided on the online monitoring systems, the user may have to choose between connecting the X first or the Y first. The choice is now available
Depending on the buffered outputs connections provided on the online monitoring systems, the user may have to choose between connecting the X first or the Y first. The choice is now available.


[[Image:ORBIGate_39.png|framed|none]]
[[Image:ORBIGate_39.png|400px]]


''Probe connections''
''Probe connections''
Line 283: Line 293:
The same preference above allows to choose also if one want to display orbits, shaft centerline and polar with the Y on the left or the Y on the right.
The same preference above allows to choose also if one want to display orbits, shaft centerline and polar with the Y on the left or the Y on the right.


[[Image:ORBIGate_40.png|framed|none]]
[[Image:ORBIGate_40.png|200px]]


''XY or YX orbit visualization''
''XY or YX orbit visualization''
Line 290: Line 300:
Consequently, there are 4 choices to allow the different configurations they are summarized in the graph below.
Consequently, there are 4 choices to allow the different configurations they are summarized in the graph below.


[[Image:ORBIGate_41.png|800px]]
[[Image:ORBIGate_41.png|600px]]


''4 choices for probes connections preferences''
''4 choices for probes connections preferences''
====Save preferences====
The save preferences options are shown below.
[[Image:OBG_7_preference.png|400px]]
''Save preferences dialog''
'''Save'''
The save option can be “With name confirmation” or “Without name confirmation”.
* If it is set to “With name confirmation”, when pressing stop, a dialog window is issued with fields available for editing the measurement name and comments. This setting is preferably used as a standard.
* In some cases, the choice “Without name confirmation” can be made: in particular when using an automatic stop based on alarms (cf the dedicated chapter for Alarms). The name used to save the measurement is the default one (based on date & time).
'''Export'''
Data can be exported either in Acquisition, Post Analysis or Navigation mode.
* To export in acquisition mode or post analysis, the option ”Automatic export on save” should be used.
* In navigation mode, one should use the function “export” from the file menu to trigger the export.
In both cases, the measurement is exported with the choosen preferences as described below.
'''Content'''
The export function uses MS Excel to export the data. The exported data are the data contained in the current loaded measurement (data contained in the grid columns).
* An option is also to export the data in separate CSV files.
* Checking the line “Repeat DATE/RPM/TIME on each sheet” will export this information on each spreadsheet of the Excel report
* “Export Graphs” will copy all the graphs displayed currently and paste them in a spreadsheet called “GRAPHS”.
'''Grouping'''
The grouping option allows exporting the data by grouping them either:
* By result (one spreadsheet for each type of results)
* By Sensor (one spreadsheet for each sensors)
Once the data have been exported, it is possible to reopen the Excel file and customize it within MS Excel. Then this file can be saved into the directory “C:\OROS\ORBIGate data\Templates” to become a new export template. This new report template can be used for data exported from the same project (same number of channels).
'''Path'''
The path of the exported document can be setup using this function. The default path is the directory of the project. It is where the file will be exported if the option “without name confirmation“ is selected.


==Set-up==
==Set-up==
Line 308: Line 358:
The interface allows the user to enter a configuration such as the one displayed on the figure below:
The interface allows the user to enter a configuration such as the one displayed on the figure below:


[[Image:ORBIGate_43.png|framed|none]]
[[Image:ORBIGate_43.png|700px|none]]


''A typical machine train''
''A typical machine train''
Line 320: Line 370:
This information will be used
This information will be used


- In the report edition
- In the report edition.


- It also helps when "loading a machine" as keywords.
- It also helps when "loading a machine" as keywords.


[[Image:ORBIGate_44.jpg|framed|none]]
[[Image:ORBIGate_44.jpg|400px]]


''The machine train configuration''
''The machine train configuration''
Line 338: Line 388:
{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
|
|
[[Image:ORBIGate_45.jpg|framed|none]]
[[Image:ORBIGate_45.jpg|50px]]


|Pair points are sensors such as proximity probes that go by set of pairs in order to display Orbits and Shaft centerline. Velocity probes and displacement probes can also be selected. In that case they are automatically integrated to get a displacement physical quantity.
|Pair points are sensors such as proximity probes that go by set of pairs in order to display Orbits and Shaft centerline. Velocity probes and displacement probes can also be selected. In that case they are automatically integrated to get a displacement physical quantity.
Line 344: Line 394:
|-
|-
|
|
[[Image:ORBIGate_46.jpg|framed|none]]
[[Image:ORBIGate_46.jpg|50px]]


|Mono points are extra sensors that come as standalone sensors. Accelerometers or velocimeters or axial proximity probes are typically used
|Mono points are extra sensors that come as standalone sensors. Accelerometers or velocimeters or axial proximity probes are typically used.


|}
|}


'''Machine train LOADING'''
=====Machine train LOADING=====


* The Machine train reload function allows loading a machine train that was saved in another project.
* The Machine train reload function allows loading a machine train that was saved in another project.
* Machine train configuration and parameters reload (inputs, configuration, gap reference etc…)
* Machine train configuration and parameters reload (inputs, configuration, gap reference etc…).
* Machine train search and sorting by key word: Machine train, Company, Plant, Project, Date
* Machine train search and sorting by key word: Machine train, Company, Plant, Project, Date.


[[Image:ORBIGate_47.png|framed|none]]
[[Image:ORBIGate_47.png|400px]]


''Machine train loading''
''Machine train loading''
Line 364: Line 414:
{|border="0" cellspacing="2" width="100%"
{|border="0" cellspacing="2" width="100%"
|colspan = "4"|
|colspan = "4"|
[[Image:ORBIGate_48.png|framed|none]]
[[Image:ORBIGate_48.png|400px]]


|
|
[[Image:ORBIGate_49.png|framed|none]]
[[Image:ORBIGate_49.png|400px]]


|-
|-
Line 378: Line 428:
|}
|}


<font color="#3366FF">'''Tip:''' </font>This function is particularly useful when one switch from the acquisition mode where a signal has been saved to the post analysis mode. Once in the post-analysis mode, it is then easy to reload the machine train of the acquisition project.
<font color="#3366FF">'''Tip:''' </font>This function is particularly useful when a user switches from the acquisition mode where a signal has been saved to the post analysis mode. Once in the post-analysis mode, it is then easy to reload the machine train of the acquisition project.


====Test points configuration====
====Test points configuration====
Line 395: Line 445:
* The hardware input connection
* The hardware input connection


[[Image:ORBIGate_51.jpg|framed|none]]
[[Image:ORBIGate_51.jpg|400px]]


''The test points configuration''
''The test points configuration''


The default hardware input connection can be modified as in the figure below. It let the user for example inverse channel Y and X if those have been connected inversely.
The default hardware input connection can be modified as in the figure below. It allows the user for example to inverse channel Y and X if those have been connected inversely.


[[Image:ORBIGate_52.jpg|framed|none]]
[[Image:ORBIGate_52.jpg|400px]]


''Input connection choice''
''Input connection choice''
Line 407: Line 457:
* Rotation direction is reminded in the test points configuration window
* Rotation direction is reminded in the test points configuration window
* Cells can be copy/pasted (including the probe orientations).
* Cells can be copy/pasted (including the probe orientations).
====Inputs====
====Inputs====
The input properties are set-up through the input window accessed by clicking on the icon as shown on the figure below.
The input properties are set-up through the input window accessed by clicking on the icon as shown on the figure below.
Line 416: Line 467:
In the input window, the user can enter the setup of the input channels associated to the test points. Most of the input set-up can be collected from the sensor properties chosen from the sensor database.
In the input window, the user can enter the setup of the input channels associated to the test points. Most of the input set-up can be collected from the sensor properties chosen from the sensor database.


[[Image:ORBIGate_54.jpg|framed|none]]
[[Image:ORBIGate_54.jpg|400px]]


''The inputs configuration''
''The inputs configuration''
Line 439: Line 490:
The tachometer type can be set-up as: "External trigger". In this case, the speed is measured based on a tachometer sensor plugged to the external trigger channel.
The tachometer type can be set-up as: "External trigger". In this case, the speed is measured based on a tachometer sensor plugged to the external trigger channel.


[[Image:ORBIGate_57.jpg|framed|none]]
[[Image:OBG_7_tacho_2.png|400px]]


''External trigger configuration''
''External trigger configuration''
Line 445: Line 496:
If no speed can be measured, it is possible to use the "Manual" tachometer. This supposes that the speed is stable and is defined by the user in the analysis window. In that case, the user should not rely on the phase results.
If no speed can be measured, it is possible to use the "Manual" tachometer. This supposes that the speed is stable and is defined by the user in the analysis window. In that case, the user should not rely on the phase results.


[[Image:ORBIGate_58.jpg|framed|none]]
[[Image:ORBIGate_58.jpg|400px]]


''The manual tachometer type''
''The manual tachometer type''
Line 451: Line 502:
If a gear box is used in between two machines. It is possible to define a tachometer that depends on the ratio between each gear wheel. For that, the type "Fractional" should be selected.
If a gear box is used in between two machines. It is possible to define a tachometer that depends on the ratio between each gear wheel. For that, the type "Fractional" should be selected.


[[Image:ORBIGate_59.jpg|framed|none]]
[[Image:OBG_7_tacho_fraction.png|400px]]


''The fractional tachometer type''
''The fractional tachometer type''


In this case, it is not necessary to measure the second tachometer. It can be simply deduced from the first sensor and the gear ratio.
In this case, it is not necessary to measure the second tachometer. It can be simply calculated from the first sensor and the gear ratio.


===Analysis===
===Analysis===
Line 464: Line 515:
''Step 5, the analysis set up''
''Step 5, the analysis set up''


ORBIGate is able to proceed to several types of analysis simultaneously: Sync analysis (Order Tracking analysis for the so called NX data), ASync analysis (FFT Analysis) and Signal Recording for the raw time data recording.
ORBIGate is able to complete several types of analysis simultaneously: Sync analysis (Order Tracking analysis for the so called NX data), ASync analysis (FFT Analysis) and Signal Recording for the raw time data recording.


* Sync Analysis is referring to the Analysis Synchronous with the Rotating Speed (results coming from the SOA Analysis). There are two kinds of display X axis for the Sync spectrum: Frequency or Order. The Sync spectrum is available in the Navigation mode.
* Sync Analysis is referring to the Analysis Synchronous with the Rotating Speed (results coming from the SOA Analysis). There are two kinds of display X axis for the Sync spectrum: Frequency or Order. The Sync spectrum is available in the Navigation mode.
* ASync Analysis is referring to standard FFT analysis: It also gives a frequency spectrum. It is ASynchronous with the rotating speed: one advantage is that it can be monitored without any rotating speed.
* ASync Analysis is referring to standard FFT analysis: It also gives a frequency spectrum. It is ASynchronous with the rotating speed: one advantage is that it can be monitored without any rotating speed.


[[Image:ORBIGate_61.png|framed|none]]
[[Image:OBG_7_analysis.png|400px]]


''Step 5, the analysis set up''
''Step 5, the analysis set up''


When you press OK on this window, you might need to wait for a few seconds before it actually proceeds. It is the time allocated to connect the analysis plug-ins.
When you press OK on this window, you might need to wait for a few seconds before it actually proceeds. It is the time allocated to connect the analysis plug-ins.
====Signal recording====
The signal recording allows the user to capture the full time domain signal with no gap. It is also possible to capture the signal by portions: one signal file will be issued. This signal file can be post-processed using ORBIGate or NVGate post analysis module.
[[Image:OBG_7_analysis_recording.png|400px]]
''The signal recording set-up''
* The signal recording can be triggered using a manual start or can be synchronized with the measurement recording.
* The signal can be recorded in every case, including when no tachometer is detected.
[[Image:ORBIGate_71.gif|400px]]
''Select and Unselect all window''


====Sync analysis====
====Sync analysis====


* The "All In One" interactive settings panel has been designed to make the settings as easy as possible for the user: he has everything under his eyes to do his choices:
* The "All In One" interactive settings panel has been designed to make the settings as easy as possible for the user by displaying all available choices:
* Order synchronous analysis: Max order, order resolution and selected orders
* Order synchronous analysis: Max order, order resolution and selected orders
* Frequency synchronous analysis (Sync spectrum): Max frequency, Resolution
* Frequency synchronous analysis (Sync spectrum): Max frequency, Resolution
Line 509: Line 574:
An interesting use of ASync monitoring is that it can operate without a keyphasor signal.
An interesting use of ASync monitoring is that it can operate without a keyphasor signal.


[[Image:ORBIGate_64.png|framed|none]]
[[Image:OBG_7_analysis_async.png|400px]]


''Sync, ASync, Trigger, Signal recording''
''Signal recording, Sync, ASync, Trigger''


The Overall level between two cut off frequency can be determined as shown in the window above. This value will be noted as "OvB AS" (it is a RMS quantity).
The Overall level between two cut off frequencies can be determined as shown in the window above. This value will be noted as "OvB AS" (it is a RMS quantity).


====Trigger====
====Trigger====
Line 520: Line 585:
Four different types of triggers are possible:
Four different types of triggers are possible:


* Free Run: The data is saved in the array without any specific frequency specified
* Free Run: The data is saved in the array without any specific frequency specified.
* Delta RPM: The data is saved in the array every increase (or decrease) of speed with a defined step
* Delta RPM: The data is saved in the array every increase (or decrease) of speed with a defined step.
* Delta Time: The data is saved with a time period
* Delta Time: The data is saved with a time period.
* Delta RPM or Delta Time: The data is saved with a "or" condition. It is saved every DeltaRPM steps or DeltaTime steps depending on which comes first. This trigger type is especially interesting when machines go through a start-up with a sequence of Run-up sessions and stable speeds sessions.
* Delta RPM or Delta Time: The data is saved with a "or" condition. It is saved every DeltaRPM steps or DeltaTime steps depending on which comes first. This trigger type is especially interesting when machines go through a start-up with a sequence of Run-up sessions and stable speeds sessions.


[[Image:ORBIGate_65.png|framed|none]]
[[Image:OBG_7_analysis_trigger.png|400px]]


''nx Recording, free run sampling''
''nx Recording, free run sampling''
Line 531: Line 596:
'''Delta Time''' is used to collect data for a specific amount of time.
'''Delta Time''' is used to collect data for a specific amount of time.


[[Image:ORBIGate_66.jpg|framed|none]]
[[Image:ORBIGate_66.jpg|400px]]


''nX Recording, Delta Time sampling''
''nX Recording, Delta Time sampling''


'''Delta RPM''' is used to collect data every increase (or decrease) of the rotating speed
'''Delta RPM''' is used to collect data every increase (or decrease) of the rotating speed.


[[Image:ORBIGate_67.jpg|framed|none]]
[[Image:ORBIGate_67.jpg|400px]]


''nX Recording, Delta RPM sampling''
''nX Recording, Delta RPM sampling''


In some cases it is interesting to be able to mix the two above condition as shown in the window below.
In some cases it is interesting to be able to mix the two above conditions as shown in the window below.


[[Image:ORBIGate_68.jpg|framed|none]]
[[Image:ORBIGate_68.jpg|400px]]


''nX recording, the Delta RPM or Delta Time sampling''
''nX recording, the Delta RPM or Delta Time sampling''
Line 549: Line 614:
As a matter of fact, when the machine start-up is made of run ups and stable speed sequences as displayed on the figure below, it is interesting to alternate the delta RPM condition and delta time. In that case, the software detects which condition comes first. The amount of saved data is optimized: a high sampling during the run up periods (using delta RPM) and a lower sampling during the stabilization periods. These periods can be very long and often don<nowiki>’</nowiki>t have many changes in terms of vibrations.
As a matter of fact, when the machine start-up is made of run ups and stable speed sequences as displayed on the figure below, it is interesting to alternate the delta RPM condition and delta time. In that case, the software detects which condition comes first. The amount of saved data is optimized: a high sampling during the run up periods (using delta RPM) and a lower sampling during the stabilization periods. These periods can be very long and often don<nowiki>’</nowiki>t have many changes in terms of vibrations.


[[Image:ORBIGate_69.jpg|framed|none]]
[[Image:ORBIGate_69.jpg|400px]]


''A typical machine start-up''
''A typical machine start-up''
====Signal recording====
The signal recording allows the user to capture the full time domain signal with no gap. It is also possible to capture the signal by portions: one signal file will be issued. This signal file can be post-processed using ORBIGate or NVGate post analysis module.
[[Image:ORBIGate_70.jpg|framed|none]]
''The signal recording set-up''
* The signal recording can be triggered using a manual start or can be synchronized with the measurement recording.
* The signal can be recorded in every case, including when no tachometer is detected.
[[Image:ORBIGate_71.gif|framed|none]]
''Select and Unselect all window''


===Alarm conditions and actions===
===Alarm conditions and actions===
Line 581: Line 632:
''Alarms list''
''Alarms list''


The first thing to do is to give a name to the alarm. For example "AlarmStart" as displayed in the window above. There is no chronology aspect in the list: each activated alarm condition is checked permanently.
The first thing to do is to give a name to the alarm. For example, "AlarmStart" as displayed in the window above. There is no chronology aspect in the list: each activated alarm condition is checked permanently.


The alarms are saved in the project. They should be configured for each project, as channels may be different from one machine to another.
The alarms are saved in the project. They should be configured for each project, as channels may be different from one machine to another.


=====Alarm conditions=====
=====Alarm conditions=====
An alarm is made of a condition and an action. If the condition is fullfilled, the action is started.
An alarm is made of a condition and an action. If the condition is fulfilled, the action is started.


The condition test is done for a number of:
The condition test is done for a number of:
Line 595: Line 646:
* A value
* A value


It should be specified if '''all tests have to be fullfilled''' or '''at least one of them should be fullfilled'''.
It should be specified if '''all tests have to be fulfilled''' or '''at least one of them should be fulfilled'''.


In the example shown in the window below one should read:
In the example shown in the window below one should read:
Line 601: Line 652:
If among "All displacement points", one of them has its 1X<nowiki>[</nowiki>RMS<nowiki>]</nowiki> component going above 60 microns, then the data record should be started.
If among "All displacement points", one of them has its 1X<nowiki>[</nowiki>RMS<nowiki>]</nowiki> component going above 60 microns, then the data record should be started.


[[Image:ORBIGate_74.png|framed|none]]
[[Image:ORBIGate_74.png|400px]]


''"Alarm start" configuration''
''"Alarm start" configuration''


=====Alarm actions=====
=====Alarm actions=====
If the alarm condition is fullfilled, then a number of actions can be achieved (like "Start data record" as shown in the window above).
If the alarm condition is fulfilled, then a number of actions can be achieved (like "Start data record" as shown in the window above).


====Alarm operation====
====Alarm operation====
When using the alarm, one needs to start the monitoring (
When using the alarm, one needs to start the monitoring  
[[Image:ORBIGate_75.png|framed|none]]
[[Image:ORBIGate_75.png|framed|none]]
) in order to start monitoring data and check conditions. From that point, the conditions are checked and actions started. Here is below an example of data record started on alarms and data record pause when the level goes down.
in order to start monitoring data and check conditions. From that point, the conditions are checked and actions started. Here is below an example of data record started on alarms and data record pause when the level goes down.


[[Image:ORBIGate_76.png|framed|none]]
[[Image:ORBIGate_76.png|framed|none]]
Line 629: Line 680:
The graphics to be displayed are chosen from the "Display type" section on the left side of the "Add/Remove" window. One type of display should be chosen. Then, the test points to be displayed should be chosen and finally the results type. When the selection is done the trace can be added (by clicking on the icon as indicated on the table below).
The graphics to be displayed are chosen from the "Display type" section on the left side of the "Add/Remove" window. One type of display should be chosen. Then, the test points to be displayed should be chosen and finally the results type. When the selection is done the trace can be added (by clicking on the icon as indicated on the table below).


[[Image:ORBIGate_78.jpg|framed|none]]
[[Image:ORBIGate_78.jpg|400px]]


''The add/remove window''
''The add/remove window''
Line 653: Line 704:
Use the layout feature to display more graphics on the same screen. Several graphic layouts can be configured. During the actual test, the different layouts can be swapped using the command "CTRL<nowiki>+</nowiki>SHIFT<nowiki>+</nowiki>SPACE".
Use the layout feature to display more graphics on the same screen. Several graphic layouts can be configured. During the actual test, the different layouts can be swapped using the command "CTRL<nowiki>+</nowiki>SHIFT<nowiki>+</nowiki>SPACE".


In the display toolbar, in the graphic zone, the layout can also be selected.
The layout can also be selected in the display toolbar or in the graphic zone.  


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
Line 663: Line 714:
|}
|}


Additional graphics are displayed during the online mode. The available graphics are described in the following table.
Additional graphics are displayed in online mode. The available graphics are described in the following table.


* Layouts can be renamed by right clicking in the window area
* Layouts can be renamed by right clicking in the window area
Line 821: Line 872:


''Functions table''
''Functions table''
====Infotrace====
The infotrace displays information about the graph:
* '''Graph mode:''' magnitude, real...
* '''List of traces with their states:''' available or not / displayed in all areas or not / not displayed. You can move into the infotrace to select a specified trace and to follow cursor<nowiki>’</nowiki>s value of this trace. Memorized traces are also displayed in the list (Memx)
* '''Overall levels:''' Displays the RMS level in 2D windows containing spectra or time signals.
* '''Cursor information:''' x, y, order, dx, dy if available
* '''Marker arrays:''' Using keyboard arrows, you can select the active trace in the infotrace.
[[File:infotrace orbigate.png|700px]]


====Scalar results====
====Scalar results====
Line 840: Line 904:


* SUB1X
* SUB1X
SUB1X is a value that gives the vibration energy contained in the sub harmonic region. It is there to evaluate instabilities such as oil whirl and oil whip.
SUB1X is a value that gives the vibration energy contained in the sub harmonic region. Its purpose is to evaluate instabilities such as oil whirl and oil whip.


[[Image:ORBIGate_103.jpg|framed|none]]
[[Image:ORBIGate_103.jpg|800px|none]]


''The SUB1X function''
''The SUB1X function''
Line 863: Line 927:
|}
|}


In the grid, Async results are noted with the initials AS after the result name
In the grid, Async results are noted with the initials AS after the result name.


====The orbit====
====The orbit====
Line 933: Line 997:


====The Bode diagram====
====The Bode diagram====
The Bode Diagram is one of the most useful graphics used for turbo machinery vibrations. It allows for the displaying of vibration levels as a function of the rotating speed or time.
The Bode Diagram is one of the most useful graphics used for turbo machinery vibrations. It allows for displaying vibration levels as a function of rotating speed or time.


[[Image:ORBIGate_116.png|framed|none]]
[[Image:ORBIGate_116.png|framed|none]]
Line 939: Line 1,003:
''The Bode diagram display''
''The Bode diagram display''


The properties can be accessed by a right mouse click on the graphic. At this point, it is possible for example to modify the cursor style used for the window.
The properties can be accessed by right clicking on the graphic. At this point, it is possible for example to modify the cursor style used for the window.


[[Image:ORBIGate_117.jpg|framed|none]]
[[Image:ORBIGate_117.jpg|framed|none]]
Line 999: Line 1,063:
3. Record the time domain data (for further post analysis).
3. Record the time domain data (for further post analysis).


4. Stop
4. Stop.


5. Update Scaling preference
5. Update Scaling preference.


[[Image:ORBIGate_123.png|framed|none]]
[[Image:ORBIGate_123.png|framed|none]]
Line 1,037: Line 1,101:
''Post Analysis signal selection''
''Post Analysis signal selection''


Once the signal is loaded, it will appear as in the window below
Once the signal is loaded, it will appear as in the window below:


[[Image:ORBIGate_127.jpg|framed|none]]
[[Image:ORBIGate_127.jpg|framed|none]]
Line 1,045: Line 1,109:
Loading a signal which has different XY preferences:
Loading a signal which has different XY preferences:


When one load a signal which has been recorded with a different preference, as shown on the figure below. One needs to inverse the tracks and the labels''.''
When a user loads a signal which has been recorded with a different preference, as shown on the figure below. The user needs to inverse the tracks and the labels''.''
[[Image:ORBIGate_128.png|framed|none]]
[[Image:ORBIGate_128.png|framed|none]]


Line 1,081: Line 1,145:
One can load several measurements in order to compare the results. In that case, one should tick at least two measurements from the list from the above window.
One can load several measurements in order to compare the results. In that case, one should tick at least two measurements from the list from the above window.


In the comparaison mode, each trace is labelled with an extension (suffix) that is added to the test point labels: that will be added to measurement name and traces names during the operation in the comparaison mode. This will make the comparaison easier when it is achieved on the same machine
In the comparison mode, each trace is labeled with an extension (suffix) that is added to the test point labels: that will be added to measurement name and traces names during the operation in the comparison mode. This will make the comparison easier when it is achieved on the same machine.


[[Image:ORBIGate_131.gif|framed|none]]
[[Image:ORBIGate_131.gif|framed|none]]
''Renaming the measurement for an easier comparaison''
''Renaming the measurement for an easier comparison''


[[Image:ORBIGate_132.png|framed|none]]
[[Image:ORBIGate_132.png|framed|none]]
Line 1,090: Line 1,154:
''Steps when comparing measurement data''
''Steps when comparing measurement data''


* If one chooses to load several measurements, one enters the comparaison mode. The navigation functionalities are restricted in that case
* If a user chooses to load several measurements and enters comparison mode. The navigation functionalities will be restricted in that case.
===Navigation main cursor===
===Navigation main cursor===
Once a measurement has been saved, the next step is to be able to use it for analysis and reporting. The navigation mode of ORBIGate allows looking through the data using the navigation cursor.
Once a measurement has been saved, the next step is to be able to use it for analysis and reporting. The navigation mode of ORBIGate allows looking through the data using the navigation cursor.
Line 1,098: Line 1,162:
''The navigation toolbar and cursor''
''The navigation toolbar and cursor''


Using the navigation cursor, it is possible to go through data that have been saved for different RPMs and different date
Using the navigation cursor, it is possible to go through data that have been saved for different RPMs and different dates.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
Line 1,162: Line 1,226:
|}
|}


It is also possible to sweep forward and backward the history using the keyboard arrows
It is also possible to sweep forward and backward the history using the keyboard arrows.


Using the navigation cursor, all the data are updated in the graphic zone.
Using the navigation cursor, all the data is updated in the graphic zone.


[[Image:ORBIGate_144.png|framed|none]]
[[Image:ORBIGate_144.png|1500px|none]]


''The working environment in the navigation mode''
''The working environment in the navigation mode''
Line 1,177: Line 1,241:
''Range selection''
''Range selection''


During that stage the number of data available in the measurement can also be resampled in order to reduce the number of data. The "Nb grid selected" value gives the amount of array exported to the report.
During that stage the number of data available in the measurement can also be resampled in order to reduce the amount of data. The "Nb grid selected" value gives the amount of array exported to the report.


The selected range can be saved as a new measurement using the file menu "save measurement as …"
The selected range can be saved as a new measurement using the file menu "save measurement as …"
Line 1,236: Line 1,300:
|}
|}


The reference values can be accessed in two ways: In the test point<nowiki>’</nowiki>s configuration or through the toolbar icon
The reference values can be accessed in two ways: In the test point<nowiki>’</nowiki>s configuration or through the toolbar icon.


{|border="2" cellspacing="0" cellpadding="4" width="100%"
{|border="2" cellspacing="0" cellpadding="4" width="100%"
Line 1,257: Line 1,321:
When accessing the "set gap reference" values through the toolbar icon. The values measured (shown in the grid) appear in the table when the button "new" is checked. If the button "current" is checked, the current values are shown.
When accessing the "set gap reference" values through the toolbar icon. The values measured (shown in the grid) appear in the table when the button "new" is checked. If the button "current" is checked, the current values are shown.


At this stage, one should choose the corresponding position of the shaft and the preference for the 0 position (0 at center or 0 at bottom): So 4 choices are possible.
At this stage, one should choose the corresponding position of the shaft and the preference for the 0 position (0 at center or 0 at bottom): Four choices are possible:


{|border="2" cellspacing="0" cellpadding="4" width="77%"
{|border="2" cellspacing="0" cellpadding="4" width="77%"
Line 1,296: Line 1,360:
Full shaft motion (Orbit<nowiki>+</nowiki> Shaft centerline <nowiki>+</nowiki> Clearance circle) can be displayed simultaneously on the same plot for having a full display of the shaft motion.
Full shaft motion (Orbit<nowiki>+</nowiki> Shaft centerline <nowiki>+</nowiki> Clearance circle) can be displayed simultaneously on the same plot for having a full display of the shaft motion.


The shaft centerline can be displayed taking the offset reference voltage into account or not. This choice is mage through the 2 new icons implemented in version 3: "Apply Gap Reference Voltage" or "Disable Gap Reference Voltage". In order to see the Shaft Centerline displayed together with the Clearance Circle, one should use "Apply Gap Reference Voltage".
The shaft centerline can be displayed taking the offset reference voltage into account or not. This choice is made through the 2 new icons implemented in version 3: "Apply Gap Reference Voltage" or "Disable Gap Reference Voltage". In order to see the Shaft Centerline displayed together with the Clearance Circle, one should use "Apply Gap Reference Voltage".


[[Image:ORBIGate_159.png|framed|none]]
[[Image:ORBIGate_159.png|framed|none]]
Line 1,303: Line 1,367:


==Exporting==
==Exporting==
In the offline mode, it is possible to start the report tool. The report tool uses MS Excel to export the data. The exported data are the data contained in the current loaded measurement (data contained in the array columns).
In offline mode, it is possible to start the report tool. The report tool uses MS Excel to export the data. The exported data are the data contained in the current loaded measurement (data contained in the array columns).


- An option is also to export the data in separate CSV files.
- An option is also to export the data in separate CSV files.


- Checking the line "Repeat DATE/RPM/TIME on each sheet" will export this information on each spreadsheet of the Excel report
- Checking the line "Repeat DATE/RPM/TIME on each sheet" will export this information on each spreadsheet of the Excel report.


- "Export Graphs" will copy all the graphs displayed currently and paste them in a spreadsheet called "GRAPHS".
- "Export Graphs" will copy all the graphs displayed currently and paste them in a spreadsheet called "GRAPHS".
Line 1,320: Line 1,384:
''The export set up window''
''The export set up window''


Once the data have been exported, it is possible to reopen the Excel file and customize it within MS Excel. Then this file can be saved into the directory "C:\OROS\ORBIGate data\Templates" to become a new template. This new report template can be used for data exported from the same project (same number of channels).
Once the data has been exported, it is possible to reopen the Excel file and customize it within MS Excel. Then this file can be saved into the directory "C:\OROS\ORBIGate data\Templates" to become a new template. This new report template can be used for data exported from the same project (same number of channels).


==Reporting==
==Reporting==
Line 1,328: Line 1,392:
There are two ways to make a report:
There are two ways to make a report:


<font color="#17365D">'''FULL REPORT:''' </font>The template decides what is going to be printed: windows to be printed in the report don<nowiki>’</nowiki>t need to be displayed by the user. The Full report is a batch report, accessible in the navigation mode
<font color="#17365D">'''FULL REPORT:''' </font>The template decides what is going to be printed: windows to be printed in the report don<nowiki>’</nowiki>t need to be displayed by the user. The Full report is a batch report, accessible in the navigation mode.


'''INSTANT REPORT:''' Print all windows displayed in the current layout (on screen). The online report is available anytime during a test.
'''INSTANT REPORT:''' Print all windows displayed in the current layout (on screen). The online report is available anytime during a test.

Latest revision as of 17:02, 12 November 2024

Getting started

This chapter is an overview of the main features and set-up steps of the ORBIGate software. More details are provided in later chapters of the operating manual. The Set-Up wizard lets you quickly go through the steps to get started.

ORBIGate 09.png

The set up toolbar

The Start’n Go window appears when ORBIGate software is opened. The user has 3 choices to get started from this window: Create New Project, Load Project and Load Measurement:

- Choosing "Load Project" allows the user to load measurement settings.

- Choosing "Load Measurement", will load the acquired measurement data.

- Choosing "Load Signal" will allow the user to display a measurement based on a recorded signal.

To get started, choose "Create New project" and press GO.

ORBIGate 10.png

The Start ’n Go window

The next step is to configure the machine train.

ORBIGate 11.jpg

The machine train configuration

The test point’s configuration screen is where the probe orientations are configured. It is also where the labels can be modified.

ORBIGate 12.jpg

The test points configuration

The next step is to configure the analyzer input properties (coupling, sensitivity etc…).

ORBIGate 13.jpg

The inputs configuration

The following step is to configure the speed properties.

OBG 7 tacho.png

The rotating speed configuration

The Analysis dialog box allows for the configuration of everything related to the analysis. There is a choice of 3 types of analysis: Sync analysis, ASync analysis, and signal recording.

OBG 7 analysis.png

The analysis window configuration

The next step, is the set-up of the alarms: alarm conditions and alarm actions. For the getting started stage, one doesn’t need to go through that step and can just press next.

ORBIGate 16.png

The final step is to display the graphical windows. After choosing the graphical windows, click the "Save project" button to save the project.

ORBIGate 17.jpg

The Add/remove window configuration

The test is ready to be run. Clicking on the monitoring icon ORBIGate 18.gif as shown on the window below, will display the results.

ORBIGate 19.jpg

The acquisition working environment

Working environment

Workspace

ORBIGate 20.png

The working environment

The ORBIGate solution interface is made of a number of toolbars and displays organized as in the figure above. The reference indexes from the figure are:

1. Analysis Mode Toolbar

2. Set-Up Toolbar

3. Operation Toolbar

4. Run-Out Toolbar

5. ORBIGate status zone

6. Overview grid

7. Display zone

8. Display toolbar

9. Hardware status zone

10. Clearance circle and reference voltage toolbar

Start’n Go

The Start’n Go function helps the user to immediately start the test or set-up. If a set-up has already been defined, it can be loaded at this stage. Otherwise, a new project can be created.

If the objective is to analyze a measurement, it can be loaded through the Start’n Go interface as shown below.

ORBIGate 21.png

The Start’n Go window

In order to avoid overwriting project names, the "GO" button is protected: It will automatically be grayed if an existing name is entered.

If the purpose is to go through a post analysis, a signal should be loaded. The signal to be loaded should be located in the current NVGate project database. The signal should contain at least one tachometer channel recorded on the external trigger. This signal can be recorded from NVGate or ORBIGate.

Acquisition, Post Analysis, Navigation: 3 Modes to operate ORBIGate

ORBIGate can be operated in three modes: the Acquisition mode, the Post Analysis and the Navigation mode. The user can switch from one to the other by using the operating mode icon toolbar as displayed below. The objective is to investigate measurement data: this is achieved by loading a Measurement File in the Navigation Mode. This measurement can be saved in the acquisition mode or in the post analysis mode.

ORBIGate 22.png

The 3 Modes

  • The acquisition mode is used to acquire measurement data. A Measurement file and/or a Signal file can be acquired. The acquisition mode is chosen by pressing the first icon.
  • The post analysis mode is used to create a Measurement file data based on a recorded Signal. The post analysis is selected by pressing the second icon.
  • The navigation mode is used to load, study and report a Measurement file that was acquired in the Acquisition mode or in the Post Analysis mode. The navigation mode is selected by pressing the third icon.

ORBIGate 23.jpg


The operating mode toolbar

Project and Measurement Management

Project management and operating modes

Projects are used to save a set-up. It controls both the acquisition set-up and the display set-up.

Projects are saved in the directory defined at the installation and modifiable with the "environment" software. The default path is "C:\OROS\ORBIGate data\Projects". One directory is created for each saved project.

When switching from one mode to the other, it is important to remember that the set-up is not the same. For this reason, the set-up is reset when the user switches modes. It is recommended to save the project when going from one mode to another. Consequently, when going from one mode to another, the following message appears in order to avoid losing the current set-up by mistake. In the figure below, as an example, the message appearing when switching to the Post Analysis mode is displayed.

ORBIGate 24.jpg

Switching to the offline mode

Saving a project

It is possible to save a project in different ways: either at the end of the setup wizard or through the "file menu". When switching from one mode another, it is recommended to save the project in each mode (for example, "test1 ACQ", "test1 PA", "test1 NAV") as the configuration is different. ORBIGate tracks any risks of overwriting projects when changing modes. A project was saved in one of the three modes (Acquisition, Post Analysis, and Navigation). Then, if after changing mode one attempt to overwrite the project, the below message is displayed. It proposes either provide a new name (recommended) or gives the flexibility to overwrite the existing one (not recommended).

ORBIGate 25.png

Save overwriting check

Loading a project

A project can be loaded through the menu and the command "Load…" (As shown in the picture below) or through the Start’n Go window.

ORBIGate 26.jpg

Loading a project

Removing a project

It is possible to remove projects through the project windows by selecting the project and pressing "Delete" on the keyboard.

Measurement management

A measurement allows storing data during a test. The data are saved when pressing the record icon. A measurement name is given at the end of the test. The list of data saved is shown in the function table in the "Displays" chapter.

ORBIGate 27.gif

The record measurement icon

Measurement data is saved in a file with the extension "OBG". It is possible to share this file with other users (it is for example possible to send this file through e-mail). In order to be usable, this file should be placed directly in the "ORBIGate data" directory or in one of the projects directory.

By default, the measurement file is saved in the project directory used for the measurement.

Loading a measurement

A measurement is loaded through the command "File/Load Measurement". The following window is displayed. The measurements can be browsed through the measurements database interface.

Files can be sorted based on their:

  • Measurement name
  • Machine train, Company, Plant
  • Project
  • Date
  • Comments

One can also search for keywords coming from any of the columns. The interface looks as below.

ORBIGate 28.png

Measurement database interface

Sensor database

The sensor database is a great tool to store the sensors used frequently. When using it, the user can just choose the sensor from the database without redefining them each time. The sensor database should be completed in NVGate before starting ORBIGate.

When using ORBIGate, the sensor can be selected by clicking in the transducer column of the selected input. The sensor database is automatically displayed. Only the appropriate sensors are shown. For the pair points, the usable sensors are accelerometers, velocity probes, and displacement probes. For the mono points all sensors from the sensor database can be used.

Once the sensors are selected, the set-up of the input is automatically done according to the predefined properties of the sensor.

ORBIGate 29.jpg

Setting Up the inputs

User preferences

User preferences allow storing set-ups that will be kept from one project to the other.

Path preferences

The location of the ORBIGate data can also be specified through this dialog window.

The preferences allow storing the paths used for operating ORBIGate. The NVGate.exe position can be selected at this level.

OBG preferences.png

Path preferences

Analysis preferences

In the preferences, it is also possible to:

- Sampling frequency: Select the maximum sampling frequency allowed for the analyzer. Among others, it enables the user to limit the DSP load. Placing the mouse on the “i” icon will give local contextual information.

- Gap / Zero & low speed GAP: Enable the acquisition in the measurement of the Gap at Zero and low speed from ASync analysis. For that:

  • Corresponding ASync channels should be connected
  • ASync analysis resolution is recommended to be set to high resolution to fit the low speed conditions (typically at least 0.625 Hz as set by default).

- Post Analysis playback speed: It is possible to setup Post analysis signal playback speed in the user preferences: this allows controlling the post processing speed of the results. Increasing playback speed may introduce down sampling of the results collection in the measurement. It will cause results to be collected less often. When the objective is to control the visualization speed of the results it is recommended to use the navigation mode which is designed for that purpose.

OBG Pref analysis.png

Analysis preferences

Units preferences

- Units: Users from different countries use different units: it is possible to select the unit used for each physical quantity in the window below.


OBG Pref unit.png


- Spectrum unit can be 0-Peak or Peak-Peak. It can be modified locally in the window as well.

- Time format: Trends can be displayed with the absolute time or relative time as an X-Scaling. Time scale is especially useful when speed is not changing (steady state measurements) and when one wants to relate the vibration evolutions with other process phenomena (load, temperature or pressure changes). In order to get the absolute dating on the X axis, the corresponding preference should be chosen in the “preferences”, as shown in the window below.


ORBIGate 33.png


Absolute time scaling for trend plots

ORBIGate 34.gif


Absolute time scaling for trend plots


- Phase: The phase can selected to be displayed as 0°/360° or -180°/+180°.

Display preferences

ORBIGate 35.png

Display and print preferences window

They are two options for the Y scaling settings: Auto scale or Fixed

  • "Auto scale" corresponds to an "Auto scale up": the scale grows continuously adapting to the maximum spectrum value.
  • The "fixed" allows to setup the maximum value for a graph

These values are defined by physical units (chosen in the “units” tab).

OBG pref scale.png

"Auto scale" or "Fixed" Scaling

To apply the selection, one should press the corresponding toolbar icon to apply the scaling (second icon displayed). This is done for the current layout. The scaling preferences window can be accessed directly by using the “scaling preferences icon”.
OBG scaling.png


“Scaling preferences” and “apply scaling” toolbar icons

XY preferences

This preference aims at providing a better flexibility for the configuration of the probes positions in the graphics (Orbit, Shaft Centerline and Polar). This is meant to adapt to the different situation that may occur when connecting the systems to the buffered outputs on the online monitoring systems.

Note: New XY preferences are taken into account for a new configuration.
For example: if a project was saved with older XY preferences, these older preferences will be taken into account.

ORBIGate 38.png
Probe settings preference

Probes connections to the analyzer inputs

Depending on the buffered outputs connections provided on the online monitoring systems, the user may have to choose between connecting the X first or the Y first. The choice is now available.

ORBIGate 39.png

Probe connections

XY or YX display

The same preference above allows to choose also if one want to display orbits, shaft centerline and polar with the Y on the left or the Y on the right.

ORBIGate 40.png

XY or YX orbit visualization

Preferences choices

Consequently, there are 4 choices to allow the different configurations they are summarized in the graph below.

ORBIGate 41.png

4 choices for probes connections preferences

Save preferences

The save preferences options are shown below.

OBG 7 preference.png

Save preferences dialog


Save

The save option can be “With name confirmation” or “Without name confirmation”.

  • If it is set to “With name confirmation”, when pressing stop, a dialog window is issued with fields available for editing the measurement name and comments. This setting is preferably used as a standard.
  • In some cases, the choice “Without name confirmation” can be made: in particular when using an automatic stop based on alarms (cf the dedicated chapter for Alarms). The name used to save the measurement is the default one (based on date & time).

Export

Data can be exported either in Acquisition, Post Analysis or Navigation mode.

  • To export in acquisition mode or post analysis, the option ”Automatic export on save” should be used.
  • In navigation mode, one should use the function “export” from the file menu to trigger the export.

In both cases, the measurement is exported with the choosen preferences as described below.

Content

The export function uses MS Excel to export the data. The exported data are the data contained in the current loaded measurement (data contained in the grid columns).

  • An option is also to export the data in separate CSV files.
  • Checking the line “Repeat DATE/RPM/TIME on each sheet” will export this information on each spreadsheet of the Excel report
  • “Export Graphs” will copy all the graphs displayed currently and paste them in a spreadsheet called “GRAPHS”.

Grouping

The grouping option allows exporting the data by grouping them either:

  • By result (one spreadsheet for each type of results)
  • By Sensor (one spreadsheet for each sensors)

Once the data have been exported, it is possible to reopen the Excel file and customize it within MS Excel. Then this file can be saved into the directory “C:\OROS\ORBIGate data\Templates” to become a new export template. This new report template can be used for data exported from the same project (same number of channels).

Path

The path of the exported document can be setup using this function. The default path is the directory of the project. It is where the file will be exported if the option “without name confirmation“ is selected.

Set-up

The set-up concerns the acquisition parameters and the display parameters.

Acquisition

Machine train

The initial step is to type in the machine train configuration. From that point, a set-up of the inputs is automatically displayed.

ORBIGate 42.png

First step, the machine train configuration

The interface allows the user to enter a configuration such as the one displayed on the figure below:

ORBIGate 43.png

A typical machine train

The first step is to enter the general information:

  • The company name,
  • The plant name
  • The machine train name.

This information will be used

- In the report edition.

- It also helps when "loading a machine" as keywords.

ORBIGate 44.jpg

The machine train configuration

Then, it is time to enter the machine train details. It is possible to enter machines by pressing "add" or "insert". The machine name is automatically entered and can be modified by simply clicking in the machine cell.

A rotating speed can be selected. Up to 2 speeds can be chosen for each machine train.

The rotation direction can be selected: CW or CCW

Then, a number of pair points and mono points should be selected:

ORBIGate 45.jpg

Pair points are sensors such as proximity probes that go by set of pairs in order to display Orbits and Shaft centerline. Velocity probes and displacement probes can also be selected. In that case they are automatically integrated to get a displacement physical quantity.

ORBIGate 46.jpg

Mono points are extra sensors that come as standalone sensors. Accelerometers or velocimeters or axial proximity probes are typically used.
Machine train LOADING
  • The Machine train reload function allows loading a machine train that was saved in another project.
  • Machine train configuration and parameters reload (inputs, configuration, gap reference etc…).
  • Machine train search and sorting by key word: Machine train, Company, Plant, Project, Date.

ORBIGate 47.png

Machine train loading

It allows reloading the information contained into the two first windows (below).

ORBIGate 48.png

ORBIGate 49.png

    Machine train configuration   Test points configuration

Tip: This function is particularly useful when a user switches from the acquisition mode where a signal has been saved to the post analysis mode. Once in the post-analysis mode, it is then easy to reload the machine train of the acquisition project.

Test points configuration

The test point configuration allows the set-up of the properties of the test points: channel label name and probes orientation.

ORBIGate 50.png

Step 2, the test points configuration

As shown on the figure below, the test point’s configuration allows entering:

  • The labels of the pair points
  • The Clearance circle radius
  • Proximity probes angular positioning: it is possible to choose the orientation of the probes by clicking on the probe orientation icon. The dialog box appears as displayed. The probes orientation can be set by steps of 1°. The angle between the probes is always 90°.
  • The label of the measurement point
  • The hardware input connection

ORBIGate 51.jpg

The test points configuration

The default hardware input connection can be modified as in the figure below. It allows the user for example to inverse channel Y and X if those have been connected inversely.

ORBIGate 52.jpg

Input connection choice

  • Rotation direction is reminded in the test points configuration window
  • Cells can be copy/pasted (including the probe orientations).

Inputs

The input properties are set-up through the input window accessed by clicking on the icon as shown on the figure below.

ORBIGate 53.png

Step 3, The Inputs configuration

In the input window, the user can enter the setup of the input channels associated to the test points. Most of the input set-up can be collected from the sensor properties chosen from the sensor database.

ORBIGate 54.jpg

The inputs configuration

Integration

  • Pair points are always displacement values (because they are meant to display orbits). So Accelerometer or velocity probes are integrated once or twice.
  • Monopoints can be integrated or derivate as desired during acquisition.
ORBIGate 55.gif

Monopoints integration

Tachometer

ORBIGate 56.png

Step 4, The tachometer configuration

The properties of the speed measurement can be entered in the "Speeds" window: Three different modes are enabled.

The tachometer type can be set-up as: "External trigger". In this case, the speed is measured based on a tachometer sensor plugged to the external trigger channel.

OBG 7 tacho 2.png

External trigger configuration

If no speed can be measured, it is possible to use the "Manual" tachometer. This supposes that the speed is stable and is defined by the user in the analysis window. In that case, the user should not rely on the phase results.

ORBIGate 58.jpg

The manual tachometer type

If a gear box is used in between two machines. It is possible to define a tachometer that depends on the ratio between each gear wheel. For that, the type "Fractional" should be selected.

OBG 7 tacho fraction.png

The fractional tachometer type

In this case, it is not necessary to measure the second tachometer. It can be simply calculated from the first sensor and the gear ratio.

Analysis

The analysis set-up window can be accessed through the icon shown below.

ORBIGate 60.png

Step 5, the analysis set up

ORBIGate is able to complete several types of analysis simultaneously: Sync analysis (Order Tracking analysis for the so called NX data), ASync analysis (FFT Analysis) and Signal Recording for the raw time data recording.

  • Sync Analysis is referring to the Analysis Synchronous with the Rotating Speed (results coming from the SOA Analysis). There are two kinds of display X axis for the Sync spectrum: Frequency or Order. The Sync spectrum is available in the Navigation mode.
  • ASync Analysis is referring to standard FFT analysis: It also gives a frequency spectrum. It is ASynchronous with the rotating speed: one advantage is that it can be monitored without any rotating speed.

OBG 7 analysis.png

Step 5, the analysis set up

When you press OK on this window, you might need to wait for a few seconds before it actually proceeds. It is the time allocated to connect the analysis plug-ins.

Signal recording

The signal recording allows the user to capture the full time domain signal with no gap. It is also possible to capture the signal by portions: one signal file will be issued. This signal file can be post-processed using ORBIGate or NVGate post analysis module.

OBG 7 analysis recording.png

The signal recording set-up

  • The signal recording can be triggered using a manual start or can be synchronized with the measurement recording.
  • The signal can be recorded in every case, including when no tachometer is detected.

ORBIGate 71.gif

Select and Unselect all window

Sync analysis

  • The "All In One" interactive settings panel has been designed to make the settings as easy as possible for the user by displaying all available choices:
  • Order synchronous analysis: Max order, order resolution and selected orders
  • Frequency synchronous analysis (Sync spectrum): Max frequency, Resolution
  • Rotating speed: Max and Min speed (Ratio between Max and Min speed is 256). When the speed is "manual", its value is chosen here as well.
  • Number of displayed revolutions and number of points per rev
ORBIGate 62.png
ORBIGate 63.png

Sync analysis settings

Settings can be entered by hand or modified graphically on the image.

NX analysis allows to set-up what orders will be calculated and displayed as:

  • NX orbits
  • Grid

It is possible to calculate up to 3 default orders: 1X (Order 1), 2X (Order 2), 3X (Order 3) as well as 3 custom orders.

The user can specify how many rotations will be displayed for the raw orbit as well as how many points per rotation.

ASync analysis

In addition to the NX recording, it can be interesting to acquire an FFT analysis. In that case, the FFT parameters used can be defined in the following window.

An interesting use of ASync monitoring is that it can operate without a keyphasor signal.

OBG 7 analysis async.png

Signal recording, Sync, ASync, Trigger

The Overall level between two cut off frequencies can be determined as shown in the window above. This value will be noted as "OvB AS" (it is a RMS quantity).

Trigger

Finally, it allows selecting the sampling frequency used for saving the data in the array.

Four different types of triggers are possible:

  • Free Run: The data is saved in the array without any specific frequency specified.
  • Delta RPM: The data is saved in the array every increase (or decrease) of speed with a defined step.
  • Delta Time: The data is saved with a time period.
  • Delta RPM or Delta Time: The data is saved with a "or" condition. It is saved every DeltaRPM steps or DeltaTime steps depending on which comes first. This trigger type is especially interesting when machines go through a start-up with a sequence of Run-up sessions and stable speeds sessions.

OBG 7 analysis trigger.png

nx Recording, free run sampling

Delta Time is used to collect data for a specific amount of time.

ORBIGate 66.jpg

nX Recording, Delta Time sampling

Delta RPM is used to collect data every increase (or decrease) of the rotating speed.

ORBIGate 67.jpg

nX Recording, Delta RPM sampling

In some cases it is interesting to be able to mix the two above conditions as shown in the window below.

ORBIGate 68.jpg

nX recording, the Delta RPM or Delta Time sampling

As a matter of fact, when the machine start-up is made of run ups and stable speed sequences as displayed on the figure below, it is interesting to alternate the delta RPM condition and delta time. In that case, the software detects which condition comes first. The amount of saved data is optimized: a high sampling during the run up periods (using delta RPM) and a lower sampling during the stabilization periods. These periods can be very long and often don’t have many changes in terms of vibrations.

ORBIGate 69.jpg

A typical machine start-up

Alarm conditions and actions

Alarm conditions and actions are set-up through icon 6 of the set-up toolbar.

ORBIGate 72.png

Step 6, the alarm conditions and actions

Configurating Alarms

A list of alarms can be created for one particular project. The alarm can be activated or not.

ORBIGate 73.png

Alarms list

The first thing to do is to give a name to the alarm. For example, "AlarmStart" as displayed in the window above. There is no chronology aspect in the list: each activated alarm condition is checked permanently.

The alarms are saved in the project. They should be configured for each project, as channels may be different from one machine to another.

Alarm conditions

An alarm is made of a condition and an action. If the condition is fulfilled, the action is started.

The condition test is done for a number of:

  • Measurement points
  • Parameters (1X, Overall etc …)
  • Below / Above
  • A value

It should be specified if all tests have to be fulfilled or at least one of them should be fulfilled.

In the example shown in the window below one should read:

If among "All displacement points", one of them has its 1X[RMS] component going above 60 microns, then the data record should be started.

ORBIGate 74.png

"Alarm start" configuration

Alarm actions

If the alarm condition is fulfilled, then a number of actions can be achieved (like "Start data record" as shown in the window above).

Alarm operation

When using the alarm, one needs to start the monitoring

ORBIGate 75.png

in order to start monitoring data and check conditions. From that point, the conditions are checked and actions started. Here is below an example of data record started on alarms and data record pause when the level goes down.

ORBIGate 76.png

Example of operations between data record and data monitoring

Displays

A large number of displays can be selected depending on the needs of the user.

Adding and removing a window

Graphic windows can be shown or removed using the "add/remove window" icon as shown in the toolbar below.

ORBIGate 77.png

Step 7, the add/remove window

The graphics to be displayed are chosen from the "Display type" section on the left side of the "Add/Remove" window. One type of display should be chosen. Then, the test points to be displayed should be chosen and finally the results type. When the selection is done the trace can be added (by clicking on the icon as indicated on the table below).

ORBIGate 78.jpg

The add/remove window

It is possible to have multiple traces on the same graph or several graphs per window.

ORBIGate 79.png
Allows adding a trace to a graph. If there is no graph, a new one is automatically created.
ORBIGate 80.png
Add a new graph.

Delete the selected graph.

Layouts:

Use the layout feature to display more graphics on the same screen. Several graphic layouts can be configured. During the actual test, the different layouts can be swapped using the command "CTRL+SHIFT+SPACE".

The layout can also be selected in the display toolbar or in the graphic zone.

ORBIGate 81.png
Add a new layout

Delete the selected a layout

Additional graphics are displayed in online mode. The available graphics are described in the following table.

  • Layouts can be renamed by right clicking in the window area
  • Windows can be renamed by right clicking on their label
  • Windows and traces can be deleted by right clicking on them

Display Acquisition
& Post Analysis
Saved and displayed in Navigation mode
ORBIGate 82.jpg
ORBIGate 83.jpg
RPM vu-meter and profile YES YES
ORBIGate 84.png
Full Shaft Motion YES YES
ORBIGate 85.jpg
Raw Orbit YES YES
ORBIGate 86.jpg
nX Orbit YES YES
ORBIGate 87.jpg
Shaft Centerline with Time/RPM ticks YES YES
ORBIGate 88.jpg
Profile vs. Time/RPM YES YES
ORBIGate 89.jpg
Bode Diagram vs. Time/RPM YES YES
ORBIGate 90.jpg
Polar plot with time/RPM ticks YES YES
ORBIGate 91.jpg
Signal vs. angle YES YES
ORBIGate 92.jpg
ORBIGate 93.jpg
Order spectrum YES YES
ORBIGate 94.jpg
ORBIGate 95.jpg
Frequency spectrum YES YES
ORBIGate 96.jpg
Sync Vumeters YES YES
ORBIGate 97.jpg
Recorded Signal YES NO
ORBIGate 98.jpg
Monitoring Shaft View YES NO
ORBIGate 99.jpg
Signal vs. time YES NO
ORBIGate 100.jpg
ASync Monitoring YES YES
ORBIGate 101.jpg
Async Vumeters YES YES


Functions table

Infotrace

The infotrace displays information about the graph:

  • Graph mode: magnitude, real...
  • List of traces with their states: available or not / displayed in all areas or not / not displayed. You can move into the infotrace to select a specified trace and to follow cursor’s value of this trace. Memorized traces are also displayed in the list (Memx)
  • Overall levels: Displays the RMS level in 2D windows containing spectra or time signals.
  • Cursor information: x, y, order, dx, dy if available
  • Marker arrays: Using keyboard arrows, you can select the active trace in the infotrace.

Infotrace orbigate.png

Scalar results

Graphic windows should be added and removed through ORBIGate "add/remove window" function.

  • Overall values are available simultaneously in the overview grid as: RMS, pk-pk and 0-pk values. They are calculated on the synchronous signal blocks available for the orbit.
  • nX orders are also available as RMS values in the grid and as trends vs. Time or RPM

The Sync grid allows visualizing the main required parameters during a measurement: the grid columns can be setup by double clicking on one of the columns or by a right click. The "Sync grid settings" appears as shown beside.

  • Ov RO- : It is the overall value corrected by the Run Out whenever a run out correction is applied.
  • nX Vectors can be shown as 0-Pk or Pk-Pk (depending on the preference) or RMS
  • Smax

SMax is defined as the maximum value of the radius of the orbit, as displayed on the graph below.

ORBIGate 102.jpg

SMax definition display

  • SUB1X

SUB1X is a value that gives the vibration energy contained in the sub harmonic region. Its purpose is to evaluate instabilities such as oil whirl and oil whip.

ORBIGate 103.jpg

The SUB1X function

This value is displayed in the array and can also be displayed as a RPM profile or time profile.

The grid

Results from the Synchronous and ASynchronous analysis are available in the grid. The grid can be configured by a right click on it.

There are two tabs available in the grid: one for setting up the Sync analysis and one for the ASync analysis.

ORBIGate 104.png



Grid Sync results settings

ORBIGate 105.png



Grid ASync results settings

In the grid, Async results are noted with the initials AS after the result name.

The orbit

The Orbit can be displayed in a multi-trace mode or in a multi-graph mode.

ORBIGate 106.png
ORBIGate 107.png

Multi trace orbit

  • The Orbit display shows the important parameters of the test:
  • Rotation direction
  • Blank/Dot indication
  • X and Y probes position.
  • The cursors are synchronized and can help to describe the deflection of the shaft as shown in figure 47. For example, each cursor 1 displayed is synchronized from one graph to another.
ORBIGate 108.png

The Orbit display

The orbit cursors let the user measure a number of dimensions of the orbit. It is accessed by dragging to the left the info trace section (on the right side of the orbit window).

ORBIGate 109.png

The Orbit cursors

The options of the Orbit window can also be modified by using a right click and selecting properties.

ORBIGate 110.jpg

The orbit display properties

The shaft centerline

  • The Shaft center line (also called GAP) display shows the important parameters of the test:
  • Rotation direction,
  • X and Y probes position
ORBIGate 111.png

The Shaft Centerline display

The shaft center line cursors let the user measure a number of dimensions of the shaft center line position. It is accessed by dragging to the left the info trace section (on the right side of the shaft center line window).

ORBIGate 112.png

The Shaft Centerline cursors

The options of the Shaft Center line window can also be modified by using a right click and selecting properties.

ORBIGate 113.jpg

The Shaft centerline display properties

The cursor in the tags section can be moved to select the number of tags in the shaft center line display.

The full shaft motion

The Full Shaft Motion is displayed using the add/remove window in two steps. One should select first the Shaft Centerline display and then select the orbit. The orbit can be a raw orbit or an nX orbit.

An information section has been added in the add/remove window to describe specific information for the selected plot. The way to display the Orbit+Shaft Centerline + Clearance circle is described in the information section of the shaft centerline window.

ORBIGate 114.png

The Full shaft motion in the add/remove window

ORBIGate 115.jpg

The Full Shaft Motion: Orbit & Shaft Centerline

The Bode diagram

The Bode Diagram is one of the most useful graphics used for turbo machinery vibrations. It allows for displaying vibration levels as a function of rotating speed or time.

ORBIGate 116.png

The Bode diagram display

The properties can be accessed by right clicking on the graphic. At this point, it is possible for example to modify the cursor style used for the window.

ORBIGate 117.jpg

The Bode diagram properties

1X Amplitude, and Phase and the Direct can be displayed together on the same plot. This is achieved by choosing first to display a Bode Diagram with the 1X result, then select a profile with the "Direct" selection and adding it in the Bode window. This graph is particularly useful to see to check the evolutions of other components than the 1X data (such as sub synchronous components for example).

ORBIGate 118.jpg

1X and Direct

The polar diagram

  • For turbo machinery measurements the expected polar diagram should have the following features:
ORBIGate 119.png

Typical polar graph display

  • 0 Phase positioned as the measured probe
  • Phase lag positive counting in opposite direction as machine rotation
  • Machine Rotation & Keyphasor position indication
  • (0,0) at center
Tachometer (KeyphaSor) position
  • The Tachometer position is indicated in the "Speeds" dialog box.
ORBIGate 120.png

Keyphasor position setting

Polar graph scaling
  • Using a fixed scale, the polar graph is displayed so that the (0, 0) origin is at the graph center so that results coming from different sensors can be easily compared.
ORBIGate 121.png

Polar graph scaling

Polar graph tags

Tags are available for the polar graphs: it can be "time tags" or "RPM" tags. This choice is available in the add/remove window.
ORBIGate 122.png



Polar graphs with RPM tags

Data acquisition

Once the set-up is done, the user can proceed to data acquisition.

Operation control

At this stage, three possibilities exist:

1. Monitor the data, without saving.

2. Record the data (displayed in the array).

3. Record the time domain data (for further post analysis).

4. Stop.

5. Update Scaling preference.

ORBIGate 123.png

The operating toolbar

GAP Monitoring

The GAP monitoring function allows the user to check the GAP and GAP Voltage even when the machine is at rest or running at very slow speed. It is useful also to determine the DC Offset.

ORBIGate 124.jpg

GAP monitoring toolbar

Saving data

When an array record or a time domain record has been started, saving the file is automatically recommended as shown in the dialog box.

By checking "Optional save as NVGate results", the displayed results are also saved within the NVGate project manager as NVGate results. It allows comparing traces from the NVGate project manager.

ORBIGate 125.png

The Save measurement dialog box

It is important to be able to correlate the measurements with the conditions under which it has been recorded. For that purpose, it is possible to enter a comment for each measurement. This is achieved when saving a measurement as shown on the screen nearby. It is possible to do it for a new measurement or for an existing measurement using the function "save measurement as...". The comment can be printed in the reports. It is also displayed during the measurement load in the database interface as shown in the display above.

Post analysis

Loading a signal

A new measurement can be created based on a time domain signal previously recorded. In that case, the first step is to load a signal with the oxf format. Then one should go through the analysis set-up as previously described in chapter 4. It follows the same principles as a real time acquisition. One can save a measurement as in the acquisition mode.

The signal to be loaded should be located in the current NVGate project database. The signal should contain at least one tachometer channel recorded on the external trigger. This signal can be recorded from NVGate or ORBIGate.

ORBIGate 126.jpg

Post Analysis signal selection

Once the signal is loaded, it will appear as in the window below:

ORBIGate 127.jpg

Post analysis working environment

Loading a signal which has different XY preferences:

When a user loads a signal which has been recorded with a different preference, as shown on the figure below. The user needs to inverse the tracks and the labels.

ORBIGate 128.png

Loading a signal with different XY preferences

Hard drive Management

  • The Hard drive management is accessible through ORBIGate.
  • In addition, the recorded signal can be post-processed directly from the hardware. There is no need to download it to the PC. One can also stay in the connected mode and consequently it is not required to exit the software.
ORBIGate 129.jpg

Hard drive management

Navigation mode

Loading one measurement

A measurement or several measurements is loaded through the command "File/Load Measurements". The following window is displayed. The measurements are browsable through the measurements database interface.

Files can be sorted based on their:

  • Measurement name
  • Machine train, Company, Plant
  • Project
  • Date
  • Comments

One can also search for keywords coming from any of the columns. The interface looks as below.

ORBIGate 130.png

Measurement database interface

Once the measurement is loaded, one can look at it in the navigation mode.

Loading measurements

One can load several measurements in order to compare the results. In that case, one should tick at least two measurements from the list from the above window.

In the comparison mode, each trace is labeled with an extension (suffix) that is added to the test point labels: that will be added to measurement name and traces names during the operation in the comparison mode. This will make the comparison easier when it is achieved on the same machine.

ORBIGate 131.gif

Renaming the measurement for an easier comparison

ORBIGate 132.png

Steps when comparing measurement data

  • If a user chooses to load several measurements and enters comparison mode. The navigation functionalities will be restricted in that case.

Navigation main cursor

Once a measurement has been saved, the next step is to be able to use it for analysis and reporting. The navigation mode of ORBIGate allows looking through the data using the navigation cursor.

ORBIGate 133.png

The navigation toolbar and cursor

Using the navigation cursor, it is possible to go through data that have been saved for different RPMs and different dates.

ORBIGate 134.png
By pressing play, the navigator sweep the time history
ORBIGate 135.png
The pause button allows to pause the sweep
ORBIGate 136.png
The stop button stops the sweep
ORBIGate 137.png
The Backward button allows to go one step back
ORBIGate 138.png
The Forward button allows to go one step forward
ORBIGate 139.png
The decrease button allows to reduce the sweep speed
ORBIGate 140.png
The reset button allows to reset the sweep speed to the default value
ORBIGate 141.png
The increase button allows to increase the sweep speed
ORBIGate 142.png
The navigation cursor can be grabbed with the mouse and moved
ORBIGate 143.png
The Range selector interface allows to select a range of the measurement

It is also possible to sweep forward and backward the history using the keyboard arrows.

Using the navigation cursor, all the data is updated in the graphic zone.

ORBIGate 144.png

The working environment in the navigation mode

Range Selection

The Range Selector allows the user in the Navigation mode to select one part of the measurement range. The range selection allows choosing the START DATE/RPM value and the END DATE/RPM value. For doing that, the user should grab the Start/End of the blue line and drag it to the expected value.

ORBIGate 145.png

Range selection

During that stage the number of data available in the measurement can also be resampled in order to reduce the amount of data. The "Nb grid selected" value gives the amount of array exported to the report.

The selected range can be saved as a new measurement using the file menu "save measurement as …"

Run Out Correction

The Run Out error can be corrected on the displayed results. This can be done in the navigation mode but also in real time or during post analysis using the toolbar shown below.

ORBIGate 146.jpg

The Run Out toolbar

For that, the first step is to define a Slow Roll during the test. The selection of the Slow Roll is done by pressing the following icon.

ORBIGate 147.gif

Slow Roll selection icon

Once this is done, one can switch from corrected results to uncorrected results using the two icons below:

ORBIGate 148.jpg

1- Uncorrected
2- Corrected

Switching from corrected to uncorrected results

On the graph below, one can see the effect of the correction at the polar diagram of order 1. Such a graph can be created by saving the polar diagram in the NVGate project manager using the function "optional save as NVGate result".

ORBIGate 149.jpg

A 1X polar diagram with and without run out correction

Reference voltage determination

GAP reference offset determination

The GAP Reference Offset can be determined by pressing the "Reference Offset" icon. It is used to position the shaft centerline referred to the (0, 0) of the plot.

Two choices are possible: the reference offset can be determined when the shaft is at bottom or when it is at center. In most cases it is carried out at bottom when the machine is at rest. In this case, the position is well known. When the machine is already running and one wants to start a shutdown, the only possibility is to choose the calibration "at center".

ORBIGate 150.png

Gap reference voltage determination

  • One action to be carried out before achieving a Full Shaft Motion display is the gap voltage reference determination. Actually, the shaft centerline has to be positioned in a coordinate system referred to a (0, 0) origin. To achieve that, the measured GAP voltage (DC component of the signal) should be associated to a known (or at least assumed) position of the shaft.
  • Typically, at rest, the shaft is assumed to be laying on the bottom of the bearing:
ORBIGate 151.png
   

The reference values can be accessed in two ways: In the test point’s configuration or through the toolbar icon.

ORBIGate 152.png


ORBIGate 153.png

  • Reference in the "test points" configuration window

  • Reference values setup through the toolbar

In this window the current values can be visualized or entered by hand.

When accessing the "set gap reference" values through the toolbar icon. The values measured (shown in the grid) appear in the table when the button "new" is checked. If the button "current" is checked, the current values are shown.

At this stage, one should choose the corresponding position of the shaft and the preference for the 0 position (0 at center or 0 at bottom): Four choices are possible:

  0 at bottom 0 at center
  • Shaft assumed at bottom:
ORBIGate 154.gif
ORBIGate 155.gif
  • Shaft assumed at center:
ORBIGate 156.gif
ORBIGate 157.png

The gap reference is saved in the measurement and in the project as well as in the machine train.

Bearing clearance circle

The bearing clearance radius can be entered and so the GAP clearance circle can be displayed. The value is entered in Microns or Mils depending on the unit defined in the preferences. If the choice is made to be "Show Clearance Circle", the circle will be shown automatically when a shaft centerline is displayed. If a value is not entered the circle won’t be shown and the shaft centerline will be displayed as standard.

ORBIGate 158.png

Bearing clearance radius set-up

Acquiring the Gap reference voltage

Full shaft motion (Orbit+ Shaft centerline + Clearance circle) can be displayed simultaneously on the same plot for having a full display of the shaft motion.

The shaft centerline can be displayed taking the offset reference voltage into account or not. This choice is made through the 2 new icons implemented in version 3: "Apply Gap Reference Voltage" or "Disable Gap Reference Voltage". In order to see the Shaft Centerline displayed together with the Clearance Circle, one should use "Apply Gap Reference Voltage".

ORBIGate 159.png

Apply gap reference voltage for shaft centerline representation

Exporting

In offline mode, it is possible to start the report tool. The report tool uses MS Excel to export the data. The exported data are the data contained in the current loaded measurement (data contained in the array columns).

- An option is also to export the data in separate CSV files.

- Checking the line "Repeat DATE/RPM/TIME on each sheet" will export this information on each spreadsheet of the Excel report.

- "Export Graphs" will copy all the graphs displayed currently and paste them in a spreadsheet called "GRAPHS".

The grouping option allows exporting the data by grouping them either:

  • By result (one spreadsheet for each type of results)
  • By Sensor (one spreadsheet for each sensors)
ORBIGate 160.png

The export set up window

Once the data has been exported, it is possible to reopen the Excel file and customize it within MS Excel. Then this file can be saved into the directory "C:\OROS\ORBIGate data\Templates" to become a new template. This new report template can be used for data exported from the same project (same number of channels).

Reporting

ORBIGate 01.png

There are two ways to make a report:

FULL REPORT: The template decides what is going to be printed: windows to be printed in the report don’t need to be displayed by the user. The Full report is a batch report, accessible in the navigation mode.

INSTANT REPORT: Print all windows displayed in the current layout (on screen). The online report is available anytime during a test.

One of the great advantages is that the report template doesn’t need to be changed according to the number of test points.

The print report setup window