Wi-Fi operations
Hardware: OR3X TW Software NVGate : V10 - V11 - V2021 (incompatibility with NVgate V12)
All Teamwork instruments support Wi-Fi operation with NVGate V10.10. It is necessary to purchase the Wi-Fi dongle (ORNV-WIF-D) form OROS. Wi-Fi operation is not compatible with other Wi-Fi dongles type. The analyzer must be configured once prior to access it through Wi-Fi.
1. Connect your PC to the analyzer with the Ethernet cable 2. From the Connection dialog 3. Select DHCP server 4. Enable the Wi-Fi. 5. Enter the password and click on Ok 6. Plug the Wi-Fi dongle to the USB port on the back plane of the analyzer and restart the analyzer 7. Check your PC's Wi-Fi configuration application: The configured analyzer SSID should appear. 8. Connect to the Analyzer by Wi-Fi: use the configured pass-phrase. 9. Unplug the network cable BEFORE you launch NVGate 10. Launch NVGate 11. In the startup screen NVGate connection the analyzer is available for running NVGate through Wi-Fi
default password is : 12345678 (if not working try orosoros)