Difference between revisions of "NVGate X-Pod: Temperature probe conditioner"

From OROS Wiki
Jump to navigation Jump to search
Line 18: Line 18:


One XPod type selector is available for each XPod position.  
One XPod type selector is available for each XPod position.  
[[File:xpod bridge.png|framed|none]]
* '''Bridge max current''': Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.
* '''Bridge voltage''': Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.
[[File:Xpod_temp.png|Xpod_temp.png]]


* '''Cold Compensation''':''' '''Enable or disable the internal cold junction compensation for the corresponding active Temperature Xpod. This setting enables the compensation of the connectors metal junction thermocouple voltage at the Xpod side. It uses the Xpod internal temperature measurement (2 locations/Xpod) to correct the measured temperatures.


===Temperature===
===Temperature===

Revision as of 12:47, 8 May 2020

X-Pod: Temperature probe conditioner

Expander modules

The XPod is a device that can be fixed on OR3X TW analyzer. Each XPod is associated to a block of 8 inputs.


Xpod-plug.png

In the connected mode, when and XPod is present each corresponding input can be routed to:

· The BNC input (Standard)

· The XPod connector


General Xpod settings

  • Type: For acquisition setup preparation, the XPod presence can be simulated in the office mode:

One XPod type selector is available for each XPod position.

  • Cold Compensation: Enable or disable the internal cold junction compensation for the corresponding active Temperature Xpod. This setting enables the compensation of the connectors metal junction thermocouple voltage at the Xpod side. It uses the Xpod internal temperature measurement (2 locations/Xpod) to correct the measured temperatures.

Temperature

This Xpod handles thermocouples and RTD transducers.

  • PT100, PT1000 and J, K, T, N, E
  • Integrated linearization
  • Automatic cold junction compensation
  • Standard fl at pin connectors
Front end 08.png