Difference between revisions of "XPod Bridge - Strain gauge"

From OROS Wiki
Jump to navigation Jump to search
 
(45 intermediate revisions by 3 users not shown)
Line 1: Line 1:
===X-Pod Strain gauge Bridge===
[[category:NVGate]]
 
OROS analyzers feature optional XPod Wheatstone bridge conditioner.  
OROS analyzers feature optional Wheatstone bridge conditioner.  
[[File:xpod_strain.png|none]]
[[File:xpod_strain.png|none]]


Line 8: Line 7:




[[File:Xpod-plug.png|framed|none]]
[[File:Xpod-plug.png|600px|none]]
In the connected mode, when and XPod is present each corresponding input can be routed to:
In the connected mode, when and XPod is present each corresponding input can be routed to:


Line 23: Line 22:
[[File:xpod bridge.png|framed|none]]
[[File:xpod bridge.png|framed|none]]


* '''Bridge max current''': Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.[[File:xpod bridge2.png|400px]]
* '''Bridge max current''': Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.
* '''Bridge voltage''': Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.
* '''Bridge voltage''': Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.


===input settings===
===Input settings===
[[File:xpod bridge2.png|framed|right]]
[[File:xpod bridge2.png|framed|right]]
*'''Type:''' bridge signal conditioning for strain gage. <br>
For each [[NVGate_Front_End#Inputs|input]], you can select the following options dedidctaed for Xpod gauge:
 
*'''Type:''' bridge signal conditioning for strain gauge. <br>


*'''Gauge type''': Full, Half or Quarter bridge mount. The completion resistors are included in the Xpod.<br>
*'''Gauge type''': Full, Half or Quarter bridge mount. The completion resistors are included in the Xpod.<br>
Line 42: Line 43:
*'''Offset Comp.''': Bridge offset comp: Can be used for manual balance of the bridge.<br>
*'''Offset Comp.''': Bridge offset comp: Can be used for manual balance of the bridge.<br>


===tutorial: How to set up XPOD conditioners for strain gage===
==Connection==
The Xpods feature 8 connectors on the side of the analyzer. The connectors correspond to the front-end channels located at the Xpod level by set of 8.
[[File:correspondance_table.png|600px|none]]
 
===Type===
 
The bridge conditioner XPod is delivered with its own connector + a connection kit with the following characteristics:<br>
 
'''On the Xpod''' - Sub-D 9 pin - Female<br>
 
'''Connector kit''' - Sub-D 9 pin - Male - Screwed connector bloc - Removable metal cover<br>
 
===Cables===
We recommend using the following cable characteristics for wiring a strain gauge:<br>
 
- Twisted pairs<br>
 
- 6 wires<br>
 
- Independent ground shields for each pair<br>


====Hard Connection====
===Connectors Pinout===
[[File:Annotation 2020-04-16 184550.png|700px|none]]
 
===Bridge connection ===
====Quarter bridge====
 
[[File:XPOD_gauge_wiki_quarter.png|700px|none]]
 
====Half bridge====
[[Image:Strain_Gage_User_Guide_03.png|700px|none]]
 
====Full bridge====
[[File:full.png|700px|none]]
 
=== download ===
[https://orossas.sharepoint.com/:b:/g/support/EerOWioePhdAv8ioRwGCNAYBZe7TmmCnHF-oPxYOQz03Qw?e=KBG1I6 Download : OROS BRIDGE CONDITIONERS XPOD™ CONNECTIONS in .pdf]
 
==Tutorial: how to set up XPOD conditioners for strain gage==
 
===Hard Connection===
Let us see how to proceed a half bridge connection.
Let us see how to proceed a half bridge connection.


Use a male SUB D 9 pins connector and a "bullet proof" cable like Ethernet cable to link the strain gages to the connector.
Use a male SUB D 9 pins connector and a "bullet proof" cable like Ethernet cable to link the strain gauges to the connector.


[[Image:Strain_Gage_User_Guide_01.png|framed|none]]
[[Image:Strain_Gage_User_Guide_01.png|framed|none]]


[[Image:Strain_Gage_User_Guide_02.png|framed|none]]
[[Image:Strain_Gage_User_Guide_02.png|none]]


Then weld the connector pins according to the pin-out diagram below:
Then solder the connector pins according to the pin-out diagram below:


[[Image:Strain_Gage_User_Guide_03.png|framed|none]]
[[File:full.png]]
[[File:Annotation 2020-04-16 184550.png]]
[[File:Annotation 2020-04-16 184709.png]]


[[File:Annotation 2020-04-16 184550.png|700px|none]]
[[File:XPOD_gauge_wiki_quarter.png|700px|none]]


==Set up NVGate==
===Set up NVGate===
<font color="#34A9DC">'''Sensitivity'''</font>
====Sensitivity====


In quarter bridge configuration, the gauge is usually glued toward the beam direction to measure the bending for example. In this case the right formula is
In quarter bridge configuration, the gauge is usually glued toward the beam direction to measure the bending for example. In this case the right formula is
Line 72: Line 110:
<font color="#333333">Where
<font color="#333333">Where


-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; S = Em/V is the sensitivity (*V / µdef)
-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; S = ΔEm/V is the sensitivity (*V / µdef)


-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; e , 2, 3 et 4 are the resistivity of the gage or generated by the hardware depending on the type of the
-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; e , 2, 3 et 4 are the resistivity of the gauge or generated by the hardware depending on the type of the


bridge.Here, 2, 3 and 4 are XPOD internal resistors
bridge.Here, 2, 3 and 4 are XPOD internal resistors
Line 84: Line 122:
-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; r, equal to R1/R4=R2/R3 often equal to 1.
-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; r, equal to R1/R4=R2/R3 often equal to 1.


-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Sg is the gage factor (often equal to 2 x 1E-06 volts per micro-def
-&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Sg is the gauge factor (often equal to 2 x 1E-06 volts per micro-def


{|border="0" cellspacing="2" width="100%"
{|border="0" cellspacing="2" width="100%"
|<nowiki>-</nowiki>  
|<nowiki>-</nowiki>  
|  n, the number of gages (For full bridge, it<nowiki>’</nowiki>s 4. For ¼ bridge it<nowiki>’</nowiki>s 1.)
|  n, the number of gauges (For full bridge, it<nowiki>’</nowiki>s 4. For ¼ bridge it<nowiki>’</nowiki>s 1.)


|}
|}
Line 94: Line 132:
</font>
</font>


You can notice that the sensitivity is not linear with the stress = L/L. For low strain, the formula is equivalent to:
You can notice that the sensitivity is not linear with the stress Δ = ΔL/L. For low strain, the formula is equivalent to:


[[Image:Strain_Gage_User_Guide_06.jpg|framed|none]]
[[Image:Strain_Gage_User_Guide_06.jpg|framed|none]]
Line 106: Line 144:
[[Image:Strain_Gage_User_Guide_08.jpg|framed|none]]
[[Image:Strain_Gage_User_Guide_08.jpg|framed|none]]


So the fact that 2 gages are used instead of only one gets the bridge sensitivity twice better.
So the fact that 2 gauges are used instead of only one gets the bridge sensitivity twice better.


<font color="#34A9DC">'''NVGate setup'''</font>
====NVGate setup====


It is advised to first add a "transducer" in NVGate database in order to register the bridge configuration.
It is advised to first add a "transducer" in NVGate database in order to register the bridge configuration.


Browse'' the Home menu Transducer tab Database''
Browse'' the Home menu > Transducer tab > Database''


[[Image:Strain_Gage_User_Guide_09.png|framed|none]]
[[Image:Strain_Gage_User_Guide_09.png|framed|none]]
Line 126: Line 164:
</font>
</font>


<font color="#34A9DC">'''Measurement'''</font>
====Measurement====


The XPOD contains 8 inputs. In case of XPOD connection, the 8 BNC inputs are disabled for using the XPOD inputs.
The XPOD contains 8 inputs. In case of XPOD connection, the 8 BNC inputs are disabled for using the XPOD inputs.
Line 142: Line 180:
Now, the measurement is ready to start.
Now, the measurement is ready to start.


[[category:Strain Gage User Guide]]
== Compute main stress with rosette ==
 
[[External_tools:_Rosette_computation|Read this page]]
=== download ===
[https://orossas.sharepoint.com/:b:/g/support/EerOWioePhdAv8ioRwGCNAYBZe7TmmCnHF-oPxYOQz03Qw?e=KBG1I6 Download : OROS BRIDGE CONDITIONERS XPOD™ CONNECTIONS]
 
=== See also ===
[[External_tools:_Rosette_computation|Compute Main stress with rosette]]

Latest revision as of 11:32, 13 March 2023

OROS analyzers feature optional XPod Wheatstone bridge conditioner.

Xpod strain.png

Expander modules

The XPod is a device that can be fixed on OR3X TW analyzer. Each XPod is associated to a block of 8 inputs.


Xpod-plug.png

In the connected mode, when and XPod is present each corresponding input can be routed to:

· The BNC input (Standard)

· The XPod connector


General Xpod settings

  • Type: For acquisition setup preparation, the XPod presence can be simulated in the office mode:

One XPod type selector is available for each XPod position.

Xpod bridge.png
  • Bridge max current: Select the max excitation current for the corresponding active Bridge Xpod. Each bridge Xpod excitation current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.
  • Bridge voltage: Select the excitation voltage for the corresponding active Bridge Xpod. From 0 to 10 V.

Input settings

Xpod bridge2.png

For each input, you can select the following options dedidctaed for Xpod gauge:

  • Type: bridge signal conditioning for strain gauge.
  • Gauge type: Full, Half or Quarter bridge mount. The completion resistors are included in the Xpod.
  • Resistance: Bridge completion resistor: 120 or 350 Ohms for quarter bridge mount.
  • Gain: Bridge gain: 10 or 100 depending on the required precision and range.
  • Max current: the provided current can be limited to 30 mA up to 4 V and 12 mA up to 10 V.
  • Excitation voltage: Each Xpod provides an excitation voltage from 0 to 10 V.
  • Offset Comp.: Bridge offset comp: Can be used for manual balance of the bridge.

Connection

The Xpods feature 8 connectors on the side of the analyzer. The connectors correspond to the front-end channels located at the Xpod level by set of 8.

Correspondance table.png

Type

The bridge conditioner XPod is delivered with its own connector + a connection kit with the following characteristics:

On the Xpod - Sub-D 9 pin - Female

Connector kit - Sub-D 9 pin - Male - Screwed connector bloc - Removable metal cover

Cables

We recommend using the following cable characteristics for wiring a strain gauge:

- Twisted pairs

- 6 wires

- Independent ground shields for each pair

Connectors Pinout

Annotation 2020-04-16 184550.png

Bridge connection

Quarter bridge

XPOD gauge wiki quarter.png

Half bridge

Strain Gage User Guide 03.png

Full bridge

Full.png

download

Download : OROS BRIDGE CONDITIONERS XPOD™ CONNECTIONS in .pdf

Tutorial: how to set up XPOD conditioners for strain gage

Hard Connection

Let us see how to proceed a half bridge connection.

Use a male SUB D 9 pins connector and a "bullet proof" cable like Ethernet cable to link the strain gauges to the connector.

Strain Gage User Guide 01.png
Strain Gage User Guide 02.png

Then solder the connector pins according to the pin-out diagram below:


Annotation 2020-04-16 184550.png
XPOD gauge wiki quarter.png

Set up NVGate

Sensitivity

In quarter bridge configuration, the gauge is usually glued toward the beam direction to measure the bending for example. In this case the right formula is

Strain Gage User Guide 04.jpg

 

Strain Gage User Guide 05.jpg

Where

-       S = ΔEm/V is the sensitivity (*V / µdef)

-        e , 2, 3 et 4 are the resistivity of the gauge or generated by the hardware depending on the type of the

bridge.Here, 2, 3 and 4 are XPOD internal resistors

-       V, the excitation voltage (in NVGate, the settings is  "Expender module/Bridge voltage")

-        ΔEm, the strain measurement voltage

-         r, equal to R1/R4=R2/R3 often equal to 1.

-        Sg is the gauge factor (often equal to 2 x 1E-06 volts per micro-def

- n, the number of gauges (For full bridge, it’s 4. For ¼ bridge it’s 1.)

You can notice that the sensitivity is not linear with the stress Δ = ΔL/L. For low strain, the formula is equivalent to:

Strain Gage User Guide 06.jpg

For a half-bridge configuration, it is possible to glue 2 gauges toward the beam direction as shown in the figure below:

Strain Gage User Guide 07.jpg

In this case, the sensitivity is linear and is equal to:

Strain Gage User Guide 08.jpg

So the fact that 2 gauges are used instead of only one gets the bridge sensitivity twice better.

NVGate setup

It is advised to first add a "transducer" in NVGate database in order to register the bridge configuration.

Browse the Home menu > Transducer tab > Database

Strain Gage User Guide 09.png

Add a new transducer which is in fact a bridge configuration.

In the Conditioner area, fill the Gauge bridge type, the gauge resistance, the excitation voltage and so on.

The sensitivity to be filled is the bridge’s one (in red below). So, it is necessary to compute this sensitivity according to the above formula.

Strain Gage User Guide 10.jpg

Measurement

The XPOD contains 8 inputs. In case of XPOD connection, the 8 BNC inputs are disabled for using the XPOD inputs.

These XPOD inputs can be connected to different plug-ins like FFT or Time domain analyzer for example.

Then the bridge has to be balanced through the Bridge Auto-zero feature (this option has to be setup to ON in the Input settings).

Strain Gage User Guide 11.jpg

Press on « Bridge Auto-zero » from the "measurement" tab to start the bridge balancing. This step may last about 30 seconds.

Strain Gage User Guide 12.jpg

Now, the measurement is ready to start.

Compute main stress with rosette

Read this page